Published in: Radiation Oncology 1/2014

Open Access 01-12-2014 | Research

PTV margin definition in hypofractionated IGRT of localized prostate cancer using cone beam CT and orthogonal image pairs with fiducial markers

Authors: Christoph Oehler, Stephanie Lang, Peter Dimmerling, Christian Bolesch, Stephan Kloeck, Alessandra Tini, Christoph Glanzmann, Yousef Najafi, Gabriela Studer, Daniel R Zwahlen

Published in: Radiation Oncology | Issue 1/2014

Login to get access

Abstract

Purpose

To evaluate PTV margins for hypofractionated IGRT of prostate comparing kV/kV imaging or CBCT.

Patients and methods

Between 2009 and 2012, 20 patients with low- (LR), intermediate- (IR) and high-risk (HR) prostate cancer were treated with VMAT in supine position with fiducial markers (FM), endorectal balloon (ERB) and full bladder. CBCT’s and kV/kV imaging were performed before and additional CBCT’s after treatment assessing intra-fraction motion. CTVP for 5 patients with LR and CTVPSV for 5 patients with IR/HR prostate cancer were contoured independently by 3 radiation oncologists using MRI. The van Hark formula (PTV margin =2.5Σ +0.7σ) was applied to calculate PTV margins of prostate/seminal vesicles (P/PSV) using CBCT or FM.

Results

172 and 52 CBCTs before and after RT and 507 kV/kV images before RT were analysed. Differences between FM in CBCT or in planar kV image pairs were below 1 mm. Accounting for both random and systematic uncertainties anisotropic PTV margins were 5-8 mm for P (LR) and 6-11 mm for PSV (IR/HR). Random uncertainties like intra-fraction and inter-fraction (setup) uncertainties were of similar magnitude (0.9-1.4 mm). Largest uncertainty was introduced by CTV delineation (LR: 1-2 mm, IR/HR: 1.6-3.5 mm). Patient positioning using bone matching or ERB-matching resulted in larger PTV margins.

Conclusions

For IGRT CBCT or kV/kV-image pairs with FM are interchangeable in respect of accuracy. Especially for hypofractionated RT, PTV margins can be kept in the range of 5 mm or below if stringent daily IGRT, ideally including prostate tracking, is applied. MR-based CTV delineation optimization is recommended.
Appendix
Available only for authorised users
Literature
1.
go back to reference Zelefsky MJ, Kollmeier M, Cox B, Fidaleo A, Sperling D, Pei X, Carver B, Coleman J, Lovelock M, Hunt M: Improved clinical outcomes with high-dose image guided radiotherapy compared with non-IGRT for the treatment of clinically localized prostate cancer. Int J Radiat Oncol Biol Phys. 2012, 84: 125-129. 10.1016/j.ijrobp.2011.11.047.CrossRefPubMed Zelefsky MJ, Kollmeier M, Cox B, Fidaleo A, Sperling D, Pei X, Carver B, Coleman J, Lovelock M, Hunt M: Improved clinical outcomes with high-dose image guided radiotherapy compared with non-IGRT for the treatment of clinically localized prostate cancer. Int J Radiat Oncol Biol Phys. 2012, 84: 125-129. 10.1016/j.ijrobp.2011.11.047.CrossRefPubMed
2.
go back to reference Dearnaley DP, Sydes MR, Graham JD, Aird EG, Bottomley D, Cowan RA, Huddart RA, Jose CC, Matthews JH, Millar J, Moore AR, Morgan RC, Russell JM, Scrase CD, Stephens RJ, Syndikus I, Parmar MK, RT01 collaborators: Escalated-dose versus standard-dose conformal radiotherapy in prostate cancer: first results from the MRC RT01 randomised controlled trial. Lancet Oncol. 2007, 8: 475-487. 10.1016/S1470-2045(07)70143-2.CrossRefPubMed Dearnaley DP, Sydes MR, Graham JD, Aird EG, Bottomley D, Cowan RA, Huddart RA, Jose CC, Matthews JH, Millar J, Moore AR, Morgan RC, Russell JM, Scrase CD, Stephens RJ, Syndikus I, Parmar MK, RT01 collaborators: Escalated-dose versus standard-dose conformal radiotherapy in prostate cancer: first results from the MRC RT01 randomised controlled trial. Lancet Oncol. 2007, 8: 475-487. 10.1016/S1470-2045(07)70143-2.CrossRefPubMed
3.
go back to reference Crehange G, Mirjolet C, Gauthier M, Martin E, Truc G, Peignaux-Casasnovas K, Azelie C, Bonnetain F, Naudy S, Maingon P: Clinical impact of margin reduction on late toxicity and short-term biochemical control for patients treated with daily on-line image guided IMRT for prostate cancer. Radiotherapy and oncology: journal of the European Society for Therapeutic Radiology and Oncology. 2012, 103: 244-246. 10.1016/j.radonc.2011.10.025.CrossRef Crehange G, Mirjolet C, Gauthier M, Martin E, Truc G, Peignaux-Casasnovas K, Azelie C, Bonnetain F, Naudy S, Maingon P: Clinical impact of margin reduction on late toxicity and short-term biochemical control for patients treated with daily on-line image guided IMRT for prostate cancer. Radiotherapy and oncology: journal of the European Society for Therapeutic Radiology and Oncology. 2012, 103: 244-246. 10.1016/j.radonc.2011.10.025.CrossRef
4.
go back to reference Langsenlehner T, Doller C, Winkler P, Galle G, Kapp KS: Impact of inter- and intrafraction deviations and residual set-up errors on PTV margins. Different alignment techniques in 3D conformal prostate cancer radiotherapy. Strahlentherapie und Onkologie. Organ der Deutschen Rontgengesellschaft [et al]. 2013, 189: 321-328. 10.1007/s00066-012-0303-0.CrossRef Langsenlehner T, Doller C, Winkler P, Galle G, Kapp KS: Impact of inter- and intrafraction deviations and residual set-up errors on PTV margins. Different alignment techniques in 3D conformal prostate cancer radiotherapy. Strahlentherapie und Onkologie. Organ der Deutschen Rontgengesellschaft [et al]. 2013, 189: 321-328. 10.1007/s00066-012-0303-0.CrossRef
5.
go back to reference Barney BM, Lee RJ, Handrahan D, Welsh KT, Cook JT, Sause WT: Image-guided radiotherapy (IGRT) for prostate cancer comparing kV imaging of fiducial markers with cone beam computed tomography (CBCT). Int J Radiat Oncol Biol Phys. 2011, 80: 301-305. 10.1016/j.ijrobp.2010.06.007.CrossRefPubMed Barney BM, Lee RJ, Handrahan D, Welsh KT, Cook JT, Sause WT: Image-guided radiotherapy (IGRT) for prostate cancer comparing kV imaging of fiducial markers with cone beam computed tomography (CBCT). Int J Radiat Oncol Biol Phys. 2011, 80: 301-305. 10.1016/j.ijrobp.2010.06.007.CrossRefPubMed
6.
go back to reference Rasch C, Steenbakkers R, van Herk M: Target definition in prostate, head, and neck. Semin Radiat Oncol. 2005, 15: 136-145. 10.1016/j.semradonc.2005.01.005.CrossRefPubMed Rasch C, Steenbakkers R, van Herk M: Target definition in prostate, head, and neck. Semin Radiat Oncol. 2005, 15: 136-145. 10.1016/j.semradonc.2005.01.005.CrossRefPubMed
7.
go back to reference van Herk M, Remeijer P, Rasch C, Lebesque JV: The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys. 2000, 47: 1121-1135. 10.1016/S0360-3016(00)00518-6.CrossRefPubMed van Herk M, Remeijer P, Rasch C, Lebesque JV: The probability of correct target dosage: dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys. 2000, 47: 1121-1135. 10.1016/S0360-3016(00)00518-6.CrossRefPubMed
8.
go back to reference Dearnaley D: CHHiP: Conventional or hypofractionated high dose intensity modulated radiotherapy for prostate cancer, protocol version 8.0. Sutton, Surrey, UK: Institute of Cancer Research. Clinical Trial & Statistics Unit. 2009, 52. Dearnaley D: CHHiP: Conventional or hypofractionated high dose intensity modulated radiotherapy for prostate cancer, protocol version 8.0. Sutton, Surrey, UK: Institute of Cancer Research. Clinical Trial & Statistics Unit. 2009, 52.
9.
go back to reference Zwahlen DR, Lang S, Hrbacek J, Glanzmann C, Kloeck S, Najafi Y, Streller T, Studer G, Zaugg K, Luetolf UM: The use of photon beams of a flattening filter-free linear accelerator for hypofractionated volumetric modulated arc therapy in localized prostate cancer. Int J Radiat Oncol Biol Phys. 2012, 83: 1655-1660. 10.1016/j.ijrobp.2011.10.019.CrossRefPubMed Zwahlen DR, Lang S, Hrbacek J, Glanzmann C, Kloeck S, Najafi Y, Streller T, Studer G, Zaugg K, Luetolf UM: The use of photon beams of a flattening filter-free linear accelerator for hypofractionated volumetric modulated arc therapy in localized prostate cancer. Int J Radiat Oncol Biol Phys. 2012, 83: 1655-1660. 10.1016/j.ijrobp.2011.10.019.CrossRefPubMed
10.
go back to reference Skarsgard D, Cadman P, El-Gayed A, Pearcey R, Tai P, Pervez N, Wu J: Planning target volume margins for prostate radiotherapy using daily electronic portal imaging and implanted fiducial markers. Radiat Oncol. 2010, 5: 52-10.1186/1748-717X-5-52.PubMedCentralCrossRefPubMed Skarsgard D, Cadman P, El-Gayed A, Pearcey R, Tai P, Pervez N, Wu J: Planning target volume margins for prostate radiotherapy using daily electronic portal imaging and implanted fiducial markers. Radiat Oncol. 2010, 5: 52-10.1186/1748-717X-5-52.PubMedCentralCrossRefPubMed
11.
go back to reference van Herk M: Errors and margins in radiotherapy. Semin Radiat Oncol. 2004, 14: 52-64. 10.1053/j.semradonc.2003.10.003.CrossRefPubMed van Herk M: Errors and margins in radiotherapy. Semin Radiat Oncol. 2004, 14: 52-64. 10.1053/j.semradonc.2003.10.003.CrossRefPubMed
12.
go back to reference Nairz O, Merz F, Deutschmann H, Kopp P, Schöller H, Zehentmayr F, Wurstbauer K, Kametriser G, Sedlmayer F: A strategy for the use of image-guided radiotherapy (IGRT) on linear accelerators and its impact on treatment margins for prostate cancer patients. Strahlentherapie und Onkologie. Organ der Deutschen Rontgengesellschaft [et al]. 2008, 184: 663-667. 10.1007/s00066-008-1874-7.CrossRef Nairz O, Merz F, Deutschmann H, Kopp P, Schöller H, Zehentmayr F, Wurstbauer K, Kametriser G, Sedlmayer F: A strategy for the use of image-guided radiotherapy (IGRT) on linear accelerators and its impact on treatment margins for prostate cancer patients. Strahlentherapie und Onkologie. Organ der Deutschen Rontgengesellschaft [et al]. 2008, 184: 663-667. 10.1007/s00066-008-1874-7.CrossRef
13.
go back to reference Adamson J, Wu Q, Yan D: Dosimetric effect of intrafraction motion and residual setup error for hypofractionated prostate intensity-modulated radiotherapy with online cone beam computed tomography image guidance. Int J Radiat Oncol Biol Phys. 2011, 80: 453-461. 10.1016/j.ijrobp.2010.02.033.PubMedCentralCrossRefPubMed Adamson J, Wu Q, Yan D: Dosimetric effect of intrafraction motion and residual setup error for hypofractionated prostate intensity-modulated radiotherapy with online cone beam computed tomography image guidance. Int J Radiat Oncol Biol Phys. 2011, 80: 453-461. 10.1016/j.ijrobp.2010.02.033.PubMedCentralCrossRefPubMed
14.
go back to reference Polat B, Guenther I, Wilbert J, Goebel J, Sweeney RA, Flentje M, Guckenberger M: Intra-fractional uncertainties in image-guided intensity-modulated radiotherapy (IMRT) of prostate cancer. Strahlentherapie und Onkologie. Organ der Deutschen Rontgengesellschaft [et al]. 2008, 184: 668-673. 10.1007/s00066-008-1875-6.CrossRef Polat B, Guenther I, Wilbert J, Goebel J, Sweeney RA, Flentje M, Guckenberger M: Intra-fractional uncertainties in image-guided intensity-modulated radiotherapy (IMRT) of prostate cancer. Strahlentherapie und Onkologie. Organ der Deutschen Rontgengesellschaft [et al]. 2008, 184: 668-673. 10.1007/s00066-008-1875-6.CrossRef
15.
go back to reference Mak D, Gill S, Paul R, Stillie A, Haworth A, Kron T, Cramb J, Knight K, Thomas J, Duchesne G, Foroudi F: Seminal vesicle interfraction displacement and margins in image guided radiotherapy for prostate cancer. Radiat Oncol. 2012, 7: 139-10.1186/1748-717X-7-139.PubMedCentralCrossRefPubMed Mak D, Gill S, Paul R, Stillie A, Haworth A, Kron T, Cramb J, Knight K, Thomas J, Duchesne G, Foroudi F: Seminal vesicle interfraction displacement and margins in image guided radiotherapy for prostate cancer. Radiat Oncol. 2012, 7: 139-10.1186/1748-717X-7-139.PubMedCentralCrossRefPubMed
16.
go back to reference Remeijer P, Rasch C, Lebesque JV, van Herk M: A general methodology for three-dimensional analysis of variation in target volume delineation. Med Phys. 1999, 26: 931-940. 10.1118/1.598485.CrossRefPubMed Remeijer P, Rasch C, Lebesque JV, van Herk M: A general methodology for three-dimensional analysis of variation in target volume delineation. Med Phys. 1999, 26: 931-940. 10.1118/1.598485.CrossRefPubMed
17.
go back to reference Khoo EL, Schick K, Plank AW, Poulsen M, Wong WW, Middleton M, Martin JM: Prostate contouring variation: can it be fixed? International Journal of Radiation Oncology, Biology. Physics. 2012, 82: 1923-1929. Khoo EL, Schick K, Plank AW, Poulsen M, Wong WW, Middleton M, Martin JM: Prostate contouring variation: can it be fixed? International Journal of Radiation Oncology, Biology. Physics. 2012, 82: 1923-1929.
18.
go back to reference Lukka H: A randomized pphase II trial of hypofractionated radiotherapy for favorable-risk prostate cancer - RTOG CCOP study. 2013 Lukka H: A randomized pphase II trial of hypofractionated radiotherapy for favorable-risk prostate cancer - RTOG CCOP study. 2013
19.
go back to reference Deutschmann H, Kametriser G, Steininger P, Scherer P, Schöller H, Gaisberger C, Mooslechner M, Mitterlechner B, Weichenberger H, Fastner G, Wurstbauer K, Jeschke S, Forstner R, Sedlmayer F: First clinical release of an online, adaptive, aperture-based image-guided radiotherapy strategy in intensity-modulated radiotherapy to correct for inter- and intrafractional rotations of the prostate. Int J Radiat Oncol Biol Phys. 2012, 83: 1624-1632. 10.1016/j.ijrobp.2011.10.009.CrossRefPubMed Deutschmann H, Kametriser G, Steininger P, Scherer P, Schöller H, Gaisberger C, Mooslechner M, Mitterlechner B, Weichenberger H, Fastner G, Wurstbauer K, Jeschke S, Forstner R, Sedlmayer F: First clinical release of an online, adaptive, aperture-based image-guided radiotherapy strategy in intensity-modulated radiotherapy to correct for inter- and intrafractional rotations of the prostate. Int J Radiat Oncol Biol Phys. 2012, 83: 1624-1632. 10.1016/j.ijrobp.2011.10.009.CrossRefPubMed
20.
go back to reference Gates LL, Gladstone DJ, Kasibhatla MS, Marshall JF, Seigne JD, Hug E, Hartford AC: Stability of serrated gold coil markers in prostate localization. Journal of Applied Clinical Medical Physics/American College of Medical Physics. 2011, 12: 3453-PubMed Gates LL, Gladstone DJ, Kasibhatla MS, Marshall JF, Seigne JD, Hug E, Hartford AC: Stability of serrated gold coil markers in prostate localization. Journal of Applied Clinical Medical Physics/American College of Medical Physics. 2011, 12: 3453-PubMed
21.
go back to reference Piziorska M, Kukolowicz P, Zawadzka A, Pilichowska M, Peczkowski P: Adaptive off-line protocol for prostate external radiotherapy with cone beam computer tomography. Strahlentherapie und Onkologie. Organ der Deutschen Rontgengesellschaft [et al]. 2012, 188: 1003-1009. 10.1007/s00066-012-0226-9.CrossRef Piziorska M, Kukolowicz P, Zawadzka A, Pilichowska M, Peczkowski P: Adaptive off-line protocol for prostate external radiotherapy with cone beam computer tomography. Strahlentherapie und Onkologie. Organ der Deutschen Rontgengesellschaft [et al]. 2012, 188: 1003-1009. 10.1007/s00066-012-0226-9.CrossRef
Metadata
Title
PTV margin definition in hypofractionated IGRT of localized prostate cancer using cone beam CT and orthogonal image pairs with fiducial markers
Authors
Christoph Oehler
Stephanie Lang
Peter Dimmerling
Christian Bolesch
Stephan Kloeck
Alessandra Tini
Christoph Glanzmann
Yousef Najafi
Gabriela Studer
Daniel R Zwahlen
Publication date
01-12-2014
Publisher
BioMed Central
Published in
Radiation Oncology / Issue 1/2014
Electronic ISSN: 1748-717X
DOI
https://doi.org/10.1186/s13014-014-0229-z

Other articles of this Issue 1/2014

Radiation Oncology 1/2014 Go to the issue