Published in: Thrombosis Journal 1/2021

Open Access 01-12-2021 | Care | Research

Evaluation of the analytical performance of the PC100 platelet counter

Authors: Magdolna Nagy, Sepanta Fazaeli, René van Oerle, Hugo ten Cate, Marcel Schemmann, John Sherry, Gillian Kelleher, Henri M. H. Spronk

Published in: Thrombosis Journal | Issue 1/2021

Login to get access

Abstract

Introduction

Platelet count can be altered in various diseases and treatments and measuring it may provide better insight into the expected outcome. So far, quantification of platelet count is done within laboratory conditions by using established hematology analyzers, whereas a point-of-care device could be used for this purpose outside of the clinical laboratories.

Aim

Our aim was to assess the closeness of agreement between a newly developed point-of-care PC100 platelet counter and two reference methods (Sysmex® XP-300, Sysmex® XN-9000) in measuring platelet counts in whole blood and platelet-rich-plasma (PRP).

Method

Whole blood was obtained from 119 individuals, of which 74 were used to prepare PRP samples. Whole blood platelet count was measured by the two reference methods and the PC100 platelet counter. PRP was prepared from the whole blood and platelet count was adjusted to the range of 250–3600 × 103/μl and measured with the PC100 platelet counter and Sysmex® XP-300.

Results

A median difference of − 1.35% and − 2.98% occurred in whole blood platelet count between the PC100 platelet counter and the Sysmex® XP-300 and Sysmex® XN-9000, respectively. A strong linear correlation (r ≥ 0.98) was seen in both cases and regression equations indicated neither a constant nor a proportional bias between the methods. Direct comparison of the two reference methods revealed a median difference of − 1.15% and a strongly linear relationship (r = 0.99). Platelet count in PRP resulted in a median difference of 1.42% between the PC100 platelet counter and the reference method, Sysmex® XP-300. While the difference between two methods increased with concentration of platelets in PRP, a strong linear relationship remained throughout the whole measuring interval indicated by the high correlation coefficient (r = 0.99). Assessment of the predicted bias at predefined platelet counts showed that the bias in platelet counts falls within the acceptance criterion for both whole blood and PRP measurements.

Conclusions

Our results show that the PC100 platelet counter can be used interchangeably with the reference methods for determining platelet counts.
Literature
1.
go back to reference Vinholt PJ, Hvas AM, Frederiksen H, Bathum L, Jørgensen MK, Nybo M. Platelet count is associated with cardiovascular disease, cancer and mortality: a population-based cohort study. Thromb Res. 2016;148:136–42.CrossRef Vinholt PJ, Hvas AM, Frederiksen H, Bathum L, Jørgensen MK, Nybo M. Platelet count is associated with cardiovascular disease, cancer and mortality: a population-based cohort study. Thromb Res. 2016;148:136–42.CrossRef
2.
go back to reference Oh SE, Seo JE, An JY, Lee JH, Sohn TS, Bae JM, et al. Prognostic impact of increased perioperative platelet count in gastric cancer patients. J Surg Res. 2019;242:296–303.CrossRef Oh SE, Seo JE, An JY, Lee JH, Sohn TS, Bae JM, et al. Prognostic impact of increased perioperative platelet count in gastric cancer patients. J Surg Res. 2019;242:296–303.CrossRef
3.
go back to reference Rumi E, Cazzola M. Essential Thrombocythemia. New Engl. J Med. 2020;382(12):e21. Rumi E, Cazzola M. Essential Thrombocythemia. New Engl. J Med. 2020;382(12):e21.
4.
go back to reference Briggs C, Harrison P, Machin SJ. Continuing developments with the automated platelet count. Int J Lab Hematol. 2007;29:77–91.CrossRef Briggs C, Harrison P, Machin SJ. Continuing developments with the automated platelet count. Int J Lab Hematol. 2007;29:77–91.CrossRef
5.
go back to reference Green R, Wachsmann-Hogiu S. Development, history, and future of automated cell counters. Clin Lab Med. 2015;35:1–10.CrossRef Green R, Wachsmann-Hogiu S. Development, history, and future of automated cell counters. Clin Lab Med. 2015;35:1–10.CrossRef
6.
go back to reference Abbasi U, Chowdhury P, Subramaniam S, Jain P, Muthe N, Sheikh F, et al. A cartridge based point-of-care device for complete blood count. Sci Rep. 2019;9:18583.CrossRef Abbasi U, Chowdhury P, Subramaniam S, Jain P, Muthe N, Sheikh F, et al. A cartridge based point-of-care device for complete blood count. Sci Rep. 2019;9:18583.CrossRef
7.
go back to reference Rao LV, Ekberg BA, Connor D, Jakubiak F, Vallaro GM, Snyder M. Evaluation of a new point of care automated complete blood count (CBC) analyzer in various clinical settings. Clin Chim Acta. 2008;389:120–5.CrossRef Rao LV, Ekberg BA, Connor D, Jakubiak F, Vallaro GM, Snyder M. Evaluation of a new point of care automated complete blood count (CBC) analyzer in various clinical settings. Clin Chim Acta. 2008;389:120–5.CrossRef
8.
go back to reference Larsson A, Smekal D, Lipcsey M. Rapid testing of red blood cells, white blood cells and platelets in intensive care patients using the HemoScreen™ point-of-care analyzer. Platelets. 2019;30:1013–6.CrossRef Larsson A, Smekal D, Lipcsey M. Rapid testing of red blood cells, white blood cells and platelets in intensive care patients using the HemoScreen™ point-of-care analyzer. Platelets. 2019;30:1013–6.CrossRef
9.
go back to reference Giusti I, D’Ascenzo S, Macchiarelli G, Dolo V. In vitro evidence supporting applications of platelet derivatives in regenerative medicine. Blood Transfus. 2020;18(2):117–29.PubMedPubMedCentral Giusti I, D’Ascenzo S, Macchiarelli G, Dolo V. In vitro evidence supporting applications of platelet derivatives in regenerative medicine. Blood Transfus. 2020;18(2):117–29.PubMedPubMedCentral
10.
go back to reference Etulain J. Platelets in wound healing and regenerative medicine. Platelets. 2018;29:556–68.CrossRef Etulain J. Platelets in wound healing and regenerative medicine. Platelets. 2018;29:556–68.CrossRef
11.
go back to reference Harrison P. Subcommittee on platelet physiology. The use of platelets in regenerative medicine and proposal for a new classification system: guidance from the SSC of the ISTH. J Thromb Haemost. 2018;16:1895–900.CrossRef Harrison P. Subcommittee on platelet physiology. The use of platelets in regenerative medicine and proposal for a new classification system: guidance from the SSC of the ISTH. J Thromb Haemost. 2018;16:1895–900.CrossRef
13.
go back to reference Hardy M, Lessire S, Kasikci S, Baudar J, Guldenpfennig M, Collard A, et al. Effects of time-interval since blood draw and of anticoagulation on platelet testing (count, indices and impedance aggregometry): a systematic study with blood from healthy volunteers. J Clin Med Res. 2020;9. https://doi.org/10.3390/jcm9082515. Hardy M, Lessire S, Kasikci S, Baudar J, Guldenpfennig M, Collard A, et al. Effects of time-interval since blood draw and of anticoagulation on platelet testing (count, indices and impedance aggregometry): a systematic study with blood from healthy volunteers. J Clin Med Res. 2020;9. https://​doi.​org/​10.​3390/​jcm9082515.
14.
go back to reference Schapkaitz E, Raburabu S. Performance evaluation of the new measurement channels on the automated Sysmex XN-9000 hematology analyzer. Clin Biochem. 2018;53:132–8.CrossRef Schapkaitz E, Raburabu S. Performance evaluation of the new measurement channels on the automated Sysmex XN-9000 hematology analyzer. Clin Biochem. 2018;53:132–8.CrossRef
15.
go back to reference van Dievoet MA, Louagie H, Ghys T. Performance evaluation of the Sysmex(®) XP-300 in an oncology setting: evaluation and comparison of hematological parameters with the Sysmex(®) XN-3000. Int J Lab Hematol. 2016;38:490–6.CrossRef van Dievoet MA, Louagie H, Ghys T. Performance evaluation of the Sysmex(®) XP-300 in an oncology setting: evaluation and comparison of hematological parameters with the Sysmex(®) XN-3000. Int J Lab Hematol. 2016;38:490–6.CrossRef
17.
go back to reference Obrien T, Dijkstra H, Schemmann MF. Optical platelet counter method. US Patent. 2017. Patent number: US9690974B2. Obrien T, Dijkstra H, Schemmann MF. Optical platelet counter method. US Patent. 2017. Patent number: US9690974B2.
18.
go back to reference CLSI. In: Wayne PA, editor. Measurement Procedure Comparison and bias estimation using patient samples; Approved Guideline-Third Edition: Clinical and Laboratory Standards Institute; 2013. CLSI. In: Wayne PA, editor. Measurement Procedure Comparison and bias estimation using patient samples; Approved Guideline-Third Edition: Clinical and Laboratory Standards Institute; 2013.
20.
go back to reference Larsson A, Greig-Pylypczuk R, Huisman A. The state of point-of-care testing: a European perspective. Ups J Med Sci. 2015;120:1–10.CrossRef Larsson A, Greig-Pylypczuk R, Huisman A. The state of point-of-care testing: a European perspective. Ups J Med Sci. 2015;120:1–10.CrossRef
21.
go back to reference Bland JM, Altman DG. Correlation in restricted ranges of data. BMJ. 2011;342:d556.CrossRef Bland JM, Altman DG. Correlation in restricted ranges of data. BMJ. 2011;342:d556.CrossRef
22.
go back to reference Ludbrook J. Linear regression analysis for comparing two measurers or methods of measurement: but which regression? Clin Exp Pharmacol Physiol. 2010;37:692–9.CrossRef Ludbrook J. Linear regression analysis for comparing two measurers or methods of measurement: but which regression? Clin Exp Pharmacol Physiol. 2010;37:692–9.CrossRef
Metadata
Title
Evaluation of the analytical performance of the PC100 platelet counter
Authors
Magdolna Nagy
Sepanta Fazaeli
René van Oerle
Hugo ten Cate
Marcel Schemmann
John Sherry
Gillian Kelleher
Henri M. H. Spronk
Publication date
01-12-2021
Publisher
BioMed Central
Keywords
Care
Hematology
Published in
Thrombosis Journal / Issue 1/2021
Electronic ISSN: 1477-9560
DOI
https://doi.org/10.1186/s12959-021-00283-w