Open Access 01-12-2019 | Hip-TEP | Research article

Fourth-generation ceramic-on-ceramic THA results in improvements in midterm outcomes compared to third-generation THA but does not resolve noise problems: a cohort study of a single-hip system

Authors: Seung-Chan Kim, Young-Wook Lim, Woo-Lam Jo, Hyun-Woo Park, Sung-Bin Han, Soon-Yong Kwon, Yong-Sik Kim

Published in: BMC Musculoskeletal Disorders | Issue 1/2019

Login to get access

Abstract

Background

Using data from the Korean Hip Registry, we aimed to investigate mid-term clinical and radiographic outcomes, including the prevalence of periprosthetic joint infection (PJI), osteolysis, and component loosening or dislocation, and to analyze the incidence of bearing-related complications following modern ceramic-on-ceramic (COC) total hip arthroplasty (THA) using a single cementless hip system.

Methods

Four hundred eighty-two patients (602 hips) who underwent Forte or Delta COC THAs with a single hip system and had a minimum 5-year follow-up were identified. The sample included 243 (50.4%) women and 239 (49.6%) men with a mean age of 50.6 years (range: 18–83 years). The Forte group comprised 310 hips, and the Delta group comprised 292 hips. The mean follow-up was 6.1 years (range: 5–10.2 years).

Results

Cup orientation did not differ between groups. No hip had a PJI or osteolysis in either group. All acetabular components and all but two femoral components (in the Delta group) were well fixed. Dislocations occurred in six (1.9%) hips in the Forte group and one (0.3%) hip in the Delta group (p = 0.124). A total of nine (1.5%) revisions were performed. The 5-year survival rates for all-cause revisions were 98.4 and 98.6%, respectively. One (0.3%) ceramic head fracture occurred in the Forte group. Sixteen (5%) hips exhibited clicking and 6 (2%) hips had squeaking in the Forte group; 16 (6%) hips exhibited clicking and 5 (2%) hips had squeaking in the Delta group. Multiple regression analysis revealed that noise generation was unassociated with any factor.

Conclusions

From the Korean Hip Registry data, THA with modern ceramic bearings showed encouraging results, with lower risks of PJI, osteolysis, and component loosening. In particular, Delta COC THA resulted in no PJI or ceramic fracture and had a reduced dislocation risk. However, associated noise remains a concern.
Literature
1.
go back to reference Hamilton WG, McAuley JP, Blumenfeld TJ, Lesko JP, Himden SE, Dennis DA. Midterm results of Delta ceramic-on-ceramic Total hip arthroplasty. J Arthroplast. 2015;30:110–5.CrossRef Hamilton WG, McAuley JP, Blumenfeld TJ, Lesko JP, Himden SE, Dennis DA. Midterm results of Delta ceramic-on-ceramic Total hip arthroplasty. J Arthroplast. 2015;30:110–5.CrossRef
2.
go back to reference The SoFCOT Study Group, Migaud H, Putman S, Kern G, Isida R, Girard J, et al. Do the reasons for ceramic-on-ceramic revisions differ from other bearings in total hip arthroplasty? Clin Orthop Relat Res. 2016;474:2190–9.CrossRef The SoFCOT Study Group, Migaud H, Putman S, Kern G, Isida R, Girard J, et al. Do the reasons for ceramic-on-ceramic revisions differ from other bearings in total hip arthroplasty? Clin Orthop Relat Res. 2016;474:2190–9.CrossRef
3.
go back to reference Hernigou P, Homma Y, Pidet O, Guissou I, Hernigou J. Ceramic-on-ceramic bearing decreases the cumulative long-term risk of dislocation. Clin Orthop Relat Res. 2013;471:3875–82.CrossRef Hernigou P, Homma Y, Pidet O, Guissou I, Hernigou J. Ceramic-on-ceramic bearing decreases the cumulative long-term risk of dislocation. Clin Orthop Relat Res. 2013;471:3875–82.CrossRef
4.
go back to reference Hernigou P, Roussignol X, Delambre J, Poignard A, Flouzat-Lachaniette C-H. Ceramic-on-ceramic THA associated with fewer dislocations and less muscle degeneration by preserving muscle progenitors. Clin Orthop. 2015;473:3762–9.CrossRef Hernigou P, Roussignol X, Delambre J, Poignard A, Flouzat-Lachaniette C-H. Ceramic-on-ceramic THA associated with fewer dislocations and less muscle degeneration by preserving muscle progenitors. Clin Orthop. 2015;473:3762–9.CrossRef
5.
go back to reference Pitto RP, Sedel L. Periprosthetic joint infection in hip arthroplasty: is there an association between infection and bearing surface type? Clin Orthop Relat Res. 2016;474:2213–8.CrossRef Pitto RP, Sedel L. Periprosthetic joint infection in hip arthroplasty: is there an association between infection and bearing surface type? Clin Orthop Relat Res. 2016;474:2213–8.CrossRef
6.
go back to reference Kurtz SM, Lau E, Baykal D, Springer BD. Outcomes of ceramic bearings after primary total hip arthroplasty in the Medicare population. J Arthroplast. 2017;32:743–9.CrossRef Kurtz SM, Lau E, Baykal D, Springer BD. Outcomes of ceramic bearings after primary total hip arthroplasty in the Medicare population. J Arthroplast. 2017;32:743–9.CrossRef
7.
go back to reference Lee G-C, Kim RH. Incidence of modern alumina ceramic and alumina matrix composite femoral head failures in nearly 6 million hip implants. J Arthroplast. 2017;32:546–51.CrossRef Lee G-C, Kim RH. Incidence of modern alumina ceramic and alumina matrix composite femoral head failures in nearly 6 million hip implants. J Arthroplast. 2017;32:546–51.CrossRef
8.
go back to reference Hamilton WG, McAuley JP, Dennis DA, Murphy JA, Blumenfeld TJ, Politi J. THA with Delta ceramic on ceramic: results of a multicenter investigational device exemption trial. Clin Orthop Relat Res. 2010;468:358–66.CrossRef Hamilton WG, McAuley JP, Dennis DA, Murphy JA, Blumenfeld TJ, Politi J. THA with Delta ceramic on ceramic: results of a multicenter investigational device exemption trial. Clin Orthop Relat Res. 2010;468:358–66.CrossRef
9.
go back to reference Yoon PW, Yoo JJ, Kim Y, Yoo S, Lee S, Kim HJ. The epidemiology and National Trends of bearing surface usage in primary Total hip arthroplasty in Korea. Clin Orthop Surg. 2016;8:29.CrossRef Yoon PW, Yoo JJ, Kim Y, Yoo S, Lee S, Kim HJ. The epidemiology and National Trends of bearing surface usage in primary Total hip arthroplasty in Korea. Clin Orthop Surg. 2016;8:29.CrossRef
10.
go back to reference Baek S-H, Kim WK, Kim JY, Kim S-Y. Do alumina matrix composite bearings decrease hip noises and bearing fractures at a minimum of 5 years after THA? Clin Orthop Relat Res. 2015;473:3796–802.CrossRef Baek S-H, Kim WK, Kim JY, Kim S-Y. Do alumina matrix composite bearings decrease hip noises and bearing fractures at a minimum of 5 years after THA? Clin Orthop Relat Res. 2015;473:3796–802.CrossRef
11.
go back to reference Schroder D, Bornstein L, Bostrom MPG, Nestor BJ, Padgett DE, Westrich GH. Ceramic-on-ceramic total hip arthroplasty: incidence of instability and noise. Clin Orthop Relat Res. 2011;469:437–42.CrossRef Schroder D, Bornstein L, Bostrom MPG, Nestor BJ, Padgett DE, Westrich GH. Ceramic-on-ceramic total hip arthroplasty: incidence of instability and noise. Clin Orthop Relat Res. 2011;469:437–42.CrossRef
12.
go back to reference Stanat SJC, Capozzi JD. Squeaking in third- and fourth-generation ceramic-on-ceramic total hip arthroplasty. J Arthroplast. 2012;27:445–53.CrossRef Stanat SJC, Capozzi JD. Squeaking in third- and fourth-generation ceramic-on-ceramic total hip arthroplasty. J Arthroplast. 2012;27:445–53.CrossRef
13.
go back to reference Jarrett CA. The squeaking hip: a phenomenon of ceramic-on-ceramic Total hip arthroplasty. J Bone Jt Surg Am. 2009;91:1344.CrossRef Jarrett CA. The squeaking hip: a phenomenon of ceramic-on-ceramic Total hip arthroplasty. J Bone Jt Surg Am. 2009;91:1344.CrossRef
14.
go back to reference Restrepo C, Matar WY, Parvizi J, Rothman RH, Hozack WJ. Natural history of squeaking after total hip arthroplasty. Clin Orthop Relat Res. 2010;468:2340–5.CrossRef Restrepo C, Matar WY, Parvizi J, Rothman RH, Hozack WJ. Natural history of squeaking after total hip arthroplasty. Clin Orthop Relat Res. 2010;468:2340–5.CrossRef
15.
go back to reference Nam D, Barrack T, Johnson SR, Nunley RM, Barrack RL. Hard-on-hard bearings are associated with increased noise generation in young patients undergoing hip arthroplasty. Clin Orthop Relat Res. 2016;474:2115–22.CrossRef Nam D, Barrack T, Johnson SR, Nunley RM, Barrack RL. Hard-on-hard bearings are associated with increased noise generation in young patients undergoing hip arthroplasty. Clin Orthop Relat Res. 2016;474:2115–22.CrossRef
16.
go back to reference Langdown AJ, Pickard RJ, Hobbs CM, Clarke HJ, Dalton DJN, Grover ML. Incomplete seating of the liner with the trident acetabular system: a cause for concern? J Bone Joint Surg Br. 2007;89:291–5.CrossRef Langdown AJ, Pickard RJ, Hobbs CM, Clarke HJ, Dalton DJN, Grover ML. Incomplete seating of the liner with the trident acetabular system: a cause for concern? J Bone Joint Surg Br. 2007;89:291–5.CrossRef
17.
go back to reference Kim S-C, Lim Y-W, Kwon S-Y, Lee J-K, Park I-K, Kim Y-S. Comparative Analysis of Radiographic Hip Joint Geometry Using Measurement Tools on Picture Archiving and Communication System: A Prospective Study of 100 Pelvic Radiographs of Koreans. J Arthroplasty. 2016;31:2597–602.CrossRef Kim S-C, Lim Y-W, Kwon S-Y, Lee J-K, Park I-K, Kim Y-S. Comparative Analysis of Radiographic Hip Joint Geometry Using Measurement Tools on Picture Archiving and Communication System: A Prospective Study of 100 Pelvic Radiographs of Koreans. J Arthroplasty. 2016;31:2597–602.CrossRef
18.
go back to reference Kim S-C, Lim Y-W, Kwon S-Y, Jo W-L, Ju S-H, Park C-J, et al. Level of surgical experience is associated with change in hip center of rotation following cementless total hip arthroplasty: a radiographic assessment. PLoS One. 2017;12:e0178300 Nazarian A, editor.CrossRef Kim S-C, Lim Y-W, Kwon S-Y, Jo W-L, Ju S-H, Park C-J, et al. Level of surgical experience is associated with change in hip center of rotation following cementless total hip arthroplasty: a radiographic assessment. PLoS One. 2017;12:e0178300 Nazarian A, editor.CrossRef
19.
go back to reference Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Jt Surg. 1978;60:217–20.CrossRef Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip-replacement arthroplasties. J Bone Jt Surg. 1978;60:217–20.CrossRef
20.
go back to reference Lu M, Zhou Y-X, Du H, Zhang J, Liu J. Reliability and validity of measuring acetabular component orientation by plain anteroposterior radiographs. Clin Orthop Relat Res. 2013;471:2987–94.CrossRef Lu M, Zhou Y-X, Du H, Zhang J, Liu J. Reliability and validity of measuring acetabular component orientation by plain anteroposterior radiographs. Clin Orthop Relat Res. 2013;471:2987–94.CrossRef
21.
go back to reference Woolson ST, Hartford JM, Sawyer A. Results of a method of leg-length equalization for patients undergoing primary total hip replacement. J Arthroplast. 1999;14:159–64.CrossRef Woolson ST, Hartford JM, Sawyer A. Results of a method of leg-length equalization for patients undergoing primary total hip replacement. J Arthroplast. 1999;14:159–64.CrossRef
22.
go back to reference Latimer HA, Lachiewicz PF. Porous-coated acetabular components with screw fixation. Five to ten-year results*. J Bone Jt Surg Am. 1996;78:975–81.CrossRef Latimer HA, Lachiewicz PF. Porous-coated acetabular components with screw fixation. Five to ten-year results*. J Bone Jt Surg Am. 1996;78:975–81.CrossRef
23.
go back to reference Engh CA, Massin P, Suthers KE. Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin. Orthop. 1990:107–28. Engh CA, Massin P, Suthers KE. Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin. Orthop. 1990:107–28.
24.
go back to reference Engh CA, Hooten JP, Zettl-Schaffer KF, Ghaffarpour M, McGovern TF, Macalino GE, et al. Porous-coated total hip replacement. Clin Orthop. 1994;298:89–96. Engh CA, Hooten JP, Zettl-Schaffer KF, Ghaffarpour M, McGovern TF, Macalino GE, et al. Porous-coated total hip replacement. Clin Orthop. 1994;298:89–96.
25.
go back to reference DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop. 1976;121:20–32. DeLee JG, Charnley J. Radiological demarcation of cemented sockets in total hip replacement. Clin Orthop. 1976;121:20–32.
26.
go back to reference Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop. 1979;141:17–27. Gruen TA, McNeice GM, Amstutz HC. “Modes of failure” of cemented stem-type femoral components: a radiographic analysis of loosening. Clin Orthop. 1979;141:17–27.
27.
go back to reference Engh CA, Bobyn JD. The influence of stem size and extent of porous coating on femoral bone resorption after primary cementless hip arthroplasty. Clin Orthop. 1988;231:7–28. Engh CA, Bobyn JD. The influence of stem size and extent of porous coating on femoral bone resorption after primary cementless hip arthroplasty. Clin Orthop. 1988;231:7–28.
28.
go back to reference Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–28.CrossRef Livermore J, Ilstrup D, Morrey B. Effect of femoral head size on wear of the polyethylene acetabular component. J Bone Joint Surg Am. 1990;72:518–28.CrossRef
29.
go back to reference Khalily C, Lester DK. Results of a tapered cementless femoral stem implanted in varus. J Arthroplast. 2002;17:463–6.CrossRef Khalily C, Lester DK. Results of a tapered cementless femoral stem implanted in varus. J Arthroplast. 2002;17:463–6.CrossRef
30.
go back to reference Sedel L. Evolution of alumina-on-alumina implants: a review. Clin Orthop. 2000;379:48–54.CrossRef Sedel L. Evolution of alumina-on-alumina implants: a review. Clin Orthop. 2000;379:48–54.CrossRef
31.
go back to reference D’Antonio JA, Capello WN, Naughton M. Ceramic bearings for Total hip arthroplasty have high survivorship at 10 years. Clin Orthop Relat Res. 2012;470:373–81.CrossRef D’Antonio JA, Capello WN, Naughton M. Ceramic bearings for Total hip arthroplasty have high survivorship at 10 years. Clin Orthop Relat Res. 2012;470:373–81.CrossRef
32.
go back to reference Kim Y-H, Park J-W, Kim J-S. Alumina Delta-on-Alumina Delta Bearing in Cementless Total Hip Arthroplasty in Patients Aged <50 Years. J Arthroplasty. 2017;32:1048–53.CrossRef Kim Y-H, Park J-W, Kim J-S. Alumina Delta-on-Alumina Delta Bearing in Cementless Total Hip Arthroplasty in Patients Aged <50 Years. J Arthroplasty. 2017;32:1048–53.CrossRef
33.
go back to reference Kim Y-H, Park J-W, Kim J-S. Long-term results of third-generation ceramic-on-ceramic bearing cementless total hip arthroplasty in Young patients. J Arthroplast. 2016;31:2520–4.CrossRef Kim Y-H, Park J-W, Kim J-S. Long-term results of third-generation ceramic-on-ceramic bearing cementless total hip arthroplasty in Young patients. J Arthroplast. 2016;31:2520–4.CrossRef
34.
go back to reference Lee Y-K, Ha Y-C, Yoo JJ, Koo K-H, Yoon KS, Kim HJ. Alumina-on-Alumina Total Hip Arthroplasty: A Concise Follow-up, at a Minimum of Ten Years, of a Previous Report*. J Bone Jt Surg-Am Vol. 2010;92:1715–9.CrossRef Lee Y-K, Ha Y-C, Yoo JJ, Koo K-H, Yoon KS, Kim HJ. Alumina-on-Alumina Total Hip Arthroplasty: A Concise Follow-up, at a Minimum of Ten Years, of a Previous Report*. J Bone Jt Surg-Am Vol. 2010;92:1715–9.CrossRef
35.
go back to reference Murphy SB, Ecker TM, Tannast M. Two- to 9-year clinical results of alumina ceramic-on-ceramic THA. Clin Orthop. 2006;453:97–102.CrossRef Murphy SB, Ecker TM, Tannast M. Two- to 9-year clinical results of alumina ceramic-on-ceramic THA. Clin Orthop. 2006;453:97–102.CrossRef
36.
go back to reference Parvizi J, Gehrke T, Chen AF. Proceedings of the international consensus on periprosthetic joint infection. Bone Jt J. 2013;95–B:1450–2.CrossRef Parvizi J, Gehrke T, Chen AF. Proceedings of the international consensus on periprosthetic joint infection. Bone Jt J. 2013;95–B:1450–2.CrossRef
37.
go back to reference Bozic KJ, Lau EC, Ong KL, Vail TP, Rubash HE, Berry DJ. Comparative effectiveness of metal-on-metal and metal-on-polyethylene bearings in Medicare Total hip arthroplasty patients. J Arthroplast. 2012;27:37–40.CrossRef Bozic KJ, Lau EC, Ong KL, Vail TP, Rubash HE, Berry DJ. Comparative effectiveness of metal-on-metal and metal-on-polyethylene bearings in Medicare Total hip arthroplasty patients. J Arthroplast. 2012;27:37–40.CrossRef
38.
go back to reference Hosman AH, van der Mei HC, Bulstra SK, Busscher HJ, Neut D. Effects of metal-on-metal wear on the host immune system and infection in hip arthroplasty. Acta Orthop. 2010;81:526–34.CrossRef Hosman AH, van der Mei HC, Bulstra SK, Busscher HJ, Neut D. Effects of metal-on-metal wear on the host immune system and infection in hip arthroplasty. Acta Orthop. 2010;81:526–34.CrossRef
39.
go back to reference Boutin P, Christel P, Dorlot JM, Meunier A, de Roquancourt A, Blanquaert D, et al. The use of dense alumina-alumina ceramic combination in total hip replacement. J Biomed Mater Res. 1988;22:1203–32.CrossRef Boutin P, Christel P, Dorlot JM, Meunier A, de Roquancourt A, Blanquaert D, et al. The use of dense alumina-alumina ceramic combination in total hip replacement. J Biomed Mater Res. 1988;22:1203–32.CrossRef
40.
go back to reference Lerouge S, Huk O, Yahia L, Witvoet J, Sedel L. Ceramic-ceramic and metal-polyethylene total hip replacements: comparison of pseudomembranes after loosening. J Bone Joint Surg Br. 1997;79:135–9.CrossRef Lerouge S, Huk O, Yahia L, Witvoet J, Sedel L. Ceramic-ceramic and metal-polyethylene total hip replacements: comparison of pseudomembranes after loosening. J Bone Joint Surg Br. 1997;79:135–9.CrossRef
41.
go back to reference Lee SJ, Yoo JJ, Kim HJ. Cementless Total hip arthroplasty involving trochanteric osteotomy without subtrochanteric shortening for high hip dislocation. Clin Orthop Surg. 2017;9:19.CrossRef Lee SJ, Yoo JJ, Kim HJ. Cementless Total hip arthroplasty involving trochanteric osteotomy without subtrochanteric shortening for high hip dislocation. Clin Orthop Surg. 2017;9:19.CrossRef
42.
go back to reference Lee S-H, Lee G-W, Seol Y-J, Park K-S, Yoon T-R. Comparison of outcomes of Total hip arthroplasty between patients with ankylosing spondylitis and avascular necrosis of the femoral head. Clin Orthop Surg. 2017;9:263.CrossRef Lee S-H, Lee G-W, Seol Y-J, Park K-S, Yoon T-R. Comparison of outcomes of Total hip arthroplasty between patients with ankylosing spondylitis and avascular necrosis of the femoral head. Clin Orthop Surg. 2017;9:263.CrossRef
43.
go back to reference Esposito C, Maclean F, Campbell P, Walter WL, Walter WK, Bonar SF. Periprosthetic tissues from third generation alumina-on-alumina total hip arthroplasties. J Arthroplast. 2013;28:860–6.CrossRef Esposito C, Maclean F, Campbell P, Walter WL, Walter WK, Bonar SF. Periprosthetic tissues from third generation alumina-on-alumina total hip arthroplasties. J Arthroplast. 2013;28:860–6.CrossRef
44.
go back to reference Lee Y-K, Ha Y-C, Jo W-L, Kim T-Y, Jung W-H, Koo K-H. Could larger diameter of 4th generation ceramic bearing decrease the rate of dislocation after THA? J Orthop Sci. 2016;21:327–31.CrossRef Lee Y-K, Ha Y-C, Jo W-L, Kim T-Y, Jung W-H, Koo K-H. Could larger diameter of 4th generation ceramic bearing decrease the rate of dislocation after THA? J Orthop Sci. 2016;21:327–31.CrossRef
45.
go back to reference McDonnell SM, Boyce G, Baré J, Young D, Shimmin AJ. The incidence of noise generation arising from the large-diameter Delta Motion ceramic total hip bearing. Bone Jt J. 2013;95–B:160–5.CrossRef McDonnell SM, Boyce G, Baré J, Young D, Shimmin AJ. The incidence of noise generation arising from the large-diameter Delta Motion ceramic total hip bearing. Bone Jt J. 2013;95–B:160–5.CrossRef
46.
go back to reference Massin P, Lopes R, Masson B, Mainard D. Does Biolox® Delta ceramic reduce the rate of component fractures in total hip replacement? Orthop Traumatol Surg Res. 2014;100:S317–21.CrossRef Massin P, Lopes R, Masson B, Mainard D. Does Biolox® Delta ceramic reduce the rate of component fractures in total hip replacement? Orthop Traumatol Surg Res. 2014;100:S317–21.CrossRef
47.
go back to reference Koo K-H, Ha Y-C, Kim S-Y, Yoon K-S, Min B-W, Kim S-R. Revision of ceramic head fracture after third generation ceramic-on-ceramic total hip arthroplasty. J Arthroplast. 2014;29:214–8.CrossRef Koo K-H, Ha Y-C, Kim S-Y, Yoon K-S, Min B-W, Kim S-R. Revision of ceramic head fracture after third generation ceramic-on-ceramic total hip arthroplasty. J Arthroplast. 2014;29:214–8.CrossRef
48.
go back to reference Abdel MP, Heyse TJ, Elpers ME, Mayman DJ, Su EP, Pellicci PM, et al. Ceramic liner fractures presenting as squeaking after primary Total hip arthroplasty. J Bone Jt Surg Am. 2014;96:27.CrossRef Abdel MP, Heyse TJ, Elpers ME, Mayman DJ, Su EP, Pellicci PM, et al. Ceramic liner fractures presenting as squeaking after primary Total hip arthroplasty. J Bone Jt Surg Am. 2014;96:27.CrossRef
49.
go back to reference Walter WL, Waters TS, Gillies M, Donohoo S, Kurtz SM, Ranawat AS, et al. Squeaking hips. J Bone Joint Surg Am. 2008;90(Suppl 4):102–11.CrossRef Walter WL, Waters TS, Gillies M, Donohoo S, Kurtz SM, Ranawat AS, et al. Squeaking hips. J Bone Joint Surg Am. 2008;90(Suppl 4):102–11.CrossRef
Metadata
Title
Fourth-generation ceramic-on-ceramic THA results in improvements in midterm outcomes compared to third-generation THA but does not resolve noise problems: a cohort study of a single-hip system
Authors
Seung-Chan Kim
Young-Wook Lim
Woo-Lam Jo
Hyun-Woo Park
Sung-Bin Han
Soon-Yong Kwon
Yong-Sik Kim
Publication date
01-12-2019
Publisher
BioMed Central
Keywords
Hip-TEP
Hip-TEP
Published in
BMC Musculoskeletal Disorders / Issue 1/2019
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/s12891-019-2641-x

Other articles of this Issue 1/2019

BMC Musculoskeletal Disorders 1/2019 Go to the issue