Intrasession repeatability of ocular anatomical measurements obtained with a multidiagnostic device in healthy eyes

Authors: David P. Piñero, Inmaculada Cabezos, Alberto López-Navarro, Dolores de Fez, María T. Caballero, Vicente J. Camps

Published in: BMC Ophthalmology | Issue 1/2017

Login to get access

Abstract

Background

To evaluate the intrasession repeatability of anterior chamber depth (ACD), central (CCT) and peripheral corneal thickness (PCT), white-to-white diameter (WTW), and irido-corneal angle (IA) measurements obtained with a multidiagnostic device in healthy eyes.

Methods

A total of 107 eyes of 107 patients ranging in age from 23 to 65 years were examined with the VX120 system (Visionix-Luneau Technologies). Three consecutive measurements were obtained with this device to assess the intrasession repeatability of ACD, CCT, PCT at different nasal and temporal locations, WTW, and nasal and temporal IA. Data analysis included the calculation of within-subject standard deviation (Sw), intrasubject precision (1.96xSw), coefficient of variation (CV) and intraclass correlation coefficient (ICC).

Results

The Sw and CV for ACD was 0.03 mm and 1.16%, respectively, with an ICC of 0.992. The Sw values for central and peripheral pachymetric measurements were below 9 μm, with CV of less than 1.6% and ICC of 0.976 or higher. For IA measurements, Sw values of 0.84 or lower were found, with a CV between 1 and 2%, and an ICC of more than 0.970. The Sw for WTW was 0.24 mm and the CV was 1.95%. No statistically significant correlations were found between any anatomical parameter evaluated and their Sw and CV values associated (−0.220 ≤ r ≤ 0.204, p ≥ 0.125).

Conclusions

The VX120 system is able to provide repeatable measurements of anatomical parameters in healthy eyes. Inter-observer repeatability should be evaluated in future studies.
Literature
1.
go back to reference Piñero DP. Technologies for anatomical and geometric characterization of the corneal structure and anterior segment: a review. Semin Ophthalmol. 2015;30:161–70.CrossRefPubMed Piñero DP. Technologies for anatomical and geometric characterization of the corneal structure and anterior segment: a review. Semin Ophthalmol. 2015;30:161–70.CrossRefPubMed
2.
go back to reference Kurian M, Negalur N, Das S, Puttaiah NK, Haria D, TS J, Thakkar MM. Biometry with a new swept-source optical coherence tomography biometer: repeatability and agreement agreement with an optical low-coherence reflectometry device. J Cataract Refract Surg. 2016;42:577–81.CrossRefPubMed Kurian M, Negalur N, Das S, Puttaiah NK, Haria D, TS J, Thakkar MM. Biometry with a new swept-source optical coherence tomography biometer: repeatability and agreement agreement with an optical low-coherence reflectometry device. J Cataract Refract Surg. 2016;42:577–81.CrossRefPubMed
3.
go back to reference Shajari M, Lehmann UC, Kohnen T. Comparison of corneal diameter and anterior chamber depth measurements using 4 different devices. Cornea. 2016;35:838–42.CrossRefPubMed Shajari M, Lehmann UC, Kohnen T. Comparison of corneal diameter and anterior chamber depth measurements using 4 different devices. Cornea. 2016;35:838–42.CrossRefPubMed
4.
go back to reference Kunert KS, Peter M, Blum M, Haigis W, Sekundo W, Schütze J, Büehren T. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg. 2016;42:76–83.CrossRefPubMed Kunert KS, Peter M, Blum M, Haigis W, Sekundo W, Schütze J, Büehren T. Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry. J Cataract Refract Surg. 2016;42:76–83.CrossRefPubMed
5.
go back to reference Shin MC, Chung SY, Hwang HS, Han KE. Comparison of two optical biometers. Optom Vis Sci. 2016;93:259–65.CrossRefPubMed Shin MC, Chung SY, Hwang HS, Han KE. Comparison of two optical biometers. Optom Vis Sci. 2016;93:259–65.CrossRefPubMed
6.
go back to reference Prakash G, Srivastava D. Single session, intrauser repeatability of anterior chamber biometric and corneal pachy-volumetric parameters using a new Scheimpflug+Placido device. J Optom. 2016;9:85–92.CrossRefPubMed Prakash G, Srivastava D. Single session, intrauser repeatability of anterior chamber biometric and corneal pachy-volumetric parameters using a new Scheimpflug+Placido device. J Optom. 2016;9:85–92.CrossRefPubMed
7.
go back to reference Cerviño A, Dominguez-Vicent A, Ferrer-Blasco T, García-Lázaro S, Albarrán-Diego C. Intrasubject repeatability of corneal power, thickness, and wavefront aberrations with a new version of a dual rotating Scheimpflug-Placido system. J Cataract Refract Surg. 2015;41:186–92.CrossRefPubMed Cerviño A, Dominguez-Vicent A, Ferrer-Blasco T, García-Lázaro S, Albarrán-Diego C. Intrasubject repeatability of corneal power, thickness, and wavefront aberrations with a new version of a dual rotating Scheimpflug-Placido system. J Cataract Refract Surg. 2015;41:186–92.CrossRefPubMed
8.
go back to reference Huang J, Savini G, Wu F, Yu X, Yang J, Yu A, Yu Y, Wang Q. Repeatability and reproducibility of ocular biometry using a new noncontact optical low-coherence interferometer. J Cataract Refract Surg. 2015;41:2233–41.CrossRefPubMed Huang J, Savini G, Wu F, Yu X, Yang J, Yu A, Yu Y, Wang Q. Repeatability and reproducibility of ocular biometry using a new noncontact optical low-coherence interferometer. J Cataract Refract Surg. 2015;41:2233–41.CrossRefPubMed
9.
go back to reference Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg. 2015;41:2224–32.CrossRefPubMed Srivannaboon S, Chirapapaisan C, Chonpimai P, Loket S. Clinical comparison of a new swept-source optical coherence tomography-based optical biometer and a time-domain optical coherence tomography-based optical biometer. J Cataract Refract Surg. 2015;41:2224–32.CrossRefPubMed
10.
go back to reference Masoud M, Livny E, Bahar I. Repeatability and intrasession reproducibility obtained by the Sirius anterior segment analysis system. Eye Contact Lens. 2015;41:107–10.CrossRefPubMed Masoud M, Livny E, Bahar I. Repeatability and intrasession reproducibility obtained by the Sirius anterior segment analysis system. Eye Contact Lens. 2015;41:107–10.CrossRefPubMed
11.
go back to reference Wang Q, Ding X, Savini G, Chen H, Feng Y, Pan C, Hua Y, Huang J. Anterior chamber depth measurements using Scheimpflug imaging and optical coherence tomography:repeatability, reproducibility, and agreement. J Cataract Refract Surg. 2015;41:178–85.CrossRefPubMed Wang Q, Ding X, Savini G, Chen H, Feng Y, Pan C, Hua Y, Huang J. Anterior chamber depth measurements using Scheimpflug imaging and optical coherence tomography:repeatability, reproducibility, and agreement. J Cataract Refract Surg. 2015;41:178–85.CrossRefPubMed
12.
go back to reference Hernández-Camarena JC, Chirinos-Saldaña P, Navas A, Ramirez-Miranda A, de la Mota A, Jimenez-Corona A, Graue-Hernindez EO. Repeatability, reproducibility, and agreement between three different Scheimpflug systems in measuring corneal and anterior segment biometry. J Refract Surg. 2014;30:616–21.CrossRefPubMed Hernández-Camarena JC, Chirinos-Saldaña P, Navas A, Ramirez-Miranda A, de la Mota A, Jimenez-Corona A, Graue-Hernindez EO. Repeatability, reproducibility, and agreement between three different Scheimpflug systems in measuring corneal and anterior segment biometry. J Refract Surg. 2014;30:616–21.CrossRefPubMed
13.
go back to reference Shetty R, Arora V, Jayadev C, Nuijts RM, Kumar M, Puttaiah NK, Kummelil MK. Repeatability and agreement of three Scheimpflug-based imaging systems for measuring anterior segment parameters in keratoconus. Invest Ophthalmol Vis Sci. 2014;55:5263–8.CrossRefPubMed Shetty R, Arora V, Jayadev C, Nuijts RM, Kumar M, Puttaiah NK, Kummelil MK. Repeatability and agreement of three Scheimpflug-based imaging systems for measuring anterior segment parameters in keratoconus. Invest Ophthalmol Vis Sci. 2014;55:5263–8.CrossRefPubMed
14.
go back to reference Montalbán R, Alió JL, Javaloy J, Piñero DP. Intrasubject repeatability in keratoconus-eye measurements obtained with a new Scheimpflug photography-based system. J Cataract Refract Surg. 2013;39:211–8.CrossRefPubMed Montalbán R, Alió JL, Javaloy J, Piñero DP. Intrasubject repeatability in keratoconus-eye measurements obtained with a new Scheimpflug photography-based system. J Cataract Refract Surg. 2013;39:211–8.CrossRefPubMed
15.
go back to reference Zhao J, Chen Z, Zhou Z, Ding L, Zhou X. Evaluation of the repeatability of the Lenstar and comparison with two other non-contact biometric devices in myopes. Clin Exp Optom. 2013;96:92–9.CrossRefPubMed Zhao J, Chen Z, Zhou Z, Ding L, Zhou X. Evaluation of the repeatability of the Lenstar and comparison with two other non-contact biometric devices in myopes. Clin Exp Optom. 2013;96:92–9.CrossRefPubMed
16.
go back to reference Nasser CK, Singer R, Barkana Y, Zadok D, Avni I, Goldich Y. Repeatability of the Sirius imaging system and agreement with the Pentacam HR. J Refract Surg. 2012;28:493–7.CrossRefPubMed Nasser CK, Singer R, Barkana Y, Zadok D, Avni I, Goldich Y. Repeatability of the Sirius imaging system and agreement with the Pentacam HR. J Refract Surg. 2012;28:493–7.CrossRefPubMed
17.
go back to reference Chen W, McAlinden C, Pesudovs K, Wang Q, Lu F, Feng Y, Chen J, Huang J. Scheimpflug-Placido topographer and optical low-coherence reflectometry biometer: repeatability and agreement. J Cataract Refract Surg. 2012;38:1626–32.CrossRefPubMed Chen W, McAlinden C, Pesudovs K, Wang Q, Lu F, Feng Y, Chen J, Huang J. Scheimpflug-Placido topographer and optical low-coherence reflectometry biometer: repeatability and agreement. J Cataract Refract Surg. 2012;38:1626–32.CrossRefPubMed
18.
go back to reference Montalbán R, Piñero DP, Javaloy J, Alió JL. Intrasubject repeatability of corneal morphology measurements obtained with a new Scheimpflug photography-based system. J Cataract Refract Surg. 2012;38:971–7.CrossRefPubMed Montalbán R, Piñero DP, Javaloy J, Alió JL. Intrasubject repeatability of corneal morphology measurements obtained with a new Scheimpflug photography-based system. J Cataract Refract Surg. 2012;38:971–7.CrossRefPubMed
19.
go back to reference Savini G, Carbonelli M, Barboni P, Hoffer KJ. Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes. J Cataract Refract Surg. 2011;37:302–9.CrossRefPubMed Savini G, Carbonelli M, Barboni P, Hoffer KJ. Repeatability of automatic measurements performed by a dual Scheimpflug analyzer in unoperated and post-refractive surgery eyes. J Cataract Refract Surg. 2011;37:302–9.CrossRefPubMed
20.
go back to reference Milla M, Piñero DP, Amparo F, Alió JL. Pachymetric measurements with a new Scheimpflug photography-based system. Intraobserver repeatability and agreement with optical coherence tomography pachymetry. J Cataract Refract Surg. 2011;37:310–6.CrossRefPubMed Milla M, Piñero DP, Amparo F, Alió JL. Pachymetric measurements with a new Scheimpflug photography-based system. Intraobserver repeatability and agreement with optical coherence tomography pachymetry. J Cataract Refract Surg. 2011;37:310–6.CrossRefPubMed
21.
go back to reference Asgari S, Hashemi H, Jafarzadehpur E, Mohamadi A, Rezvan F, Fotouhi A. OPD-Scan III: a repeatability and inter-device agreement study of a multifunctional device in emmetropia, ametropia, and keratoconus. Int Ophthalmol. 2016; [Epub ahead of print] Asgari S, Hashemi H, Jafarzadehpur E, Mohamadi A, Rezvan F, Fotouhi A. OPD-Scan III: a repeatability and inter-device agreement study of a multifunctional device in emmetropia, ametropia, and keratoconus. Int Ophthalmol. 2016; [Epub ahead of print]
22.
go back to reference Piñero DP, Juan JT, Alió JL. Intrasubject repeatability of internal aberrometry obtained with a new integrated aberrometer. J Refract Surg. 2011;27:509–17.CrossRefPubMed Piñero DP, Juan JT, Alió JL. Intrasubject repeatability of internal aberrometry obtained with a new integrated aberrometer. J Refract Surg. 2011;27:509–17.CrossRefPubMed
23.
go back to reference López-Miguel A, Martínez-Almeida L, González-García MJ, Coco-Martín MB, Sobrado-Calvo P, Maldonado MJ. Precision of higher-order aberration measurements with a new Placido-disk topographer and Hartmann-Shack wavefront sensor. J Cataract Refract Surg. 2013;39:242–9.CrossRefPubMed López-Miguel A, Martínez-Almeida L, González-García MJ, Coco-Martín MB, Sobrado-Calvo P, Maldonado MJ. Precision of higher-order aberration measurements with a new Placido-disk topographer and Hartmann-Shack wavefront sensor. J Cataract Refract Surg. 2013;39:242–9.CrossRefPubMed
Metadata
Title
Intrasession repeatability of ocular anatomical measurements obtained with a multidiagnostic device in healthy eyes
Authors
David P. Piñero
Inmaculada Cabezos
Alberto López-Navarro
Dolores de Fez
María T. Caballero
Vicente J. Camps
Publication date
01-12-2017
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2017
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-017-0589-x

Other articles of this Issue 1/2017

BMC Ophthalmology 1/2017 Go to the issue