Published in: BMC Ophthalmology 1/2016

Open Access 01-12-2016 | Research article

Comparing DTL microfiber and Neuroline skin electrode in the Mini Ganzfeld ERG

Authors: Anastasia Lapkovska, Anja M. Palmowski-Wolfe, Margarita G. Todorova

Published in: BMC Ophthalmology | Issue 1/2016

Login to get access

Abstract

Background

In infant ERG recordings skin electrodes frequently result in a better compliance. In order to assess the quality of such recordings, we compared the recording characteristics of DTL microfiber and Neuroline surface electrodes using a modified ISCEV protocol in the Mini Ganzfeld ERG.

Methods

A prospective cohort study on healthy adult subjects was conducted at the Department of Ophthalmology, University of Basel, Switzerland. Thirty healthy volunteers were tested. The microfiber electrode (DTL Plus Electrode) was placed across the cornea, above the lower eyelid. The Neuroline skin electrode was placed on the surface of the lower lid on the opposite eye. The eye on which each electrode type was placed was randomised.
Amplitudes of the rod, standard combined, standard flash cone, light-adapted 3.0 Hz flicker and red cone responses were analysed, as well as their respective implicit times.

Results

Both electrode recordings showed the same waveform characteristics. Responses with the Neuroline electrode were significantly weaker than those from the DTL electrode. Amplitudes of the rod, standard combined, standard flash cone, light-adapted 3.0 Hz flicker and red cone responses were up to four times larger when recorded with the DTL electrode (p < 0.005, ANOVA). Implicit times of the red cone ERGs were slightly faster for the Neuroline skin electrode recordings (p ≤ 0.039).

Conclusions

Comparison of full-field ERG recordings with microfiber DTL and Neuroline skin electrodes showed that DTL electrodes produce larger ERGs. Hence, we provide evidence that both electrode types allow successful full-field ERG recording, although separate normative data for both electrodes are necessary.
Literature
1.
go back to reference Marmor MF, Arden G, Nilsson SE, Zrenner E. Standard for clinical electroretinography. International Standardization Committee. Arch Ophthalmol. 1989;107:816–9.CrossRef Marmor MF, Arden G, Nilsson SE, Zrenner E. Standard for clinical electroretinography. International Standardization Committee. Arch Ophthalmol. 1989;107:816–9.CrossRef
2.
go back to reference Marmor MF, Fulton A, Holder GE, Miyake Y, Brigell M, Bach M, et al. ISCEV Standard for full-field clinical electroretinography (2008 update). Doc Ophthalmol. 2009;118:69–77.CrossRefPubMed Marmor MF, Fulton A, Holder GE, Miyake Y, Brigell M, Bach M, et al. ISCEV Standard for full-field clinical electroretinography (2008 update). Doc Ophthalmol. 2009;118:69–77.CrossRefPubMed
3.
go back to reference McCulloch DL, Marmor MF, Brigell MG, Hamilton R, Holder GE, Tzekov R, et al. ISCEV Standard for full-field clinical electroretinography (2015 update). Doc Ophthalmol. 2015;130:1–12.CrossRefPubMed McCulloch DL, Marmor MF, Brigell MG, Hamilton R, Holder GE, Tzekov R, et al. ISCEV Standard for full-field clinical electroretinography (2015 update). Doc Ophthalmol. 2015;130:1–12.CrossRefPubMed
4.
go back to reference Brodie S. Tips and tricks for successful electroretinography in children. Curr Opin Ophthalmol. 2014;25:366–73.CrossRefPubMed Brodie S. Tips and tricks for successful electroretinography in children. Curr Opin Ophthalmol. 2014;25:366–73.CrossRefPubMed
5.
go back to reference Beeler P, Barthelmes D, Sutter FK, Helbig H, Fleischhauer JC. Comparison of performance and patient satisfaction of two types of ERG electrodes. Klin Monatsbl Augenheilkd. 2007;224:265–8.CrossRefPubMed Beeler P, Barthelmes D, Sutter FK, Helbig H, Fleischhauer JC. Comparison of performance and patient satisfaction of two types of ERG electrodes. Klin Monatsbl Augenheilkd. 2007;224:265–8.CrossRefPubMed
6.
go back to reference ISCEV Committee for Pediatric Clinical Electrophysiology Guidelines, Fulton A, Brecelj J, Lorenz B, Moskowitz A, Thompson D, et al. Pediatric clinical visual electrophysiology: a survey of actual practice. Doc Ophthalmol. 2006;113:193–204.CrossRef ISCEV Committee for Pediatric Clinical Electrophysiology Guidelines, Fulton A, Brecelj J, Lorenz B, Moskowitz A, Thompson D, et al. Pediatric clinical visual electrophysiology: a survey of actual practice. Doc Ophthalmol. 2006;113:193–204.CrossRef
7.
go back to reference Parness-Yossifon R, Mets MB. The electroretinogram in children. Curr Opin Ophthalmol. 2008;19:398–402.CrossRefPubMed Parness-Yossifon R, Mets MB. The electroretinogram in children. Curr Opin Ophthalmol. 2008;19:398–402.CrossRefPubMed
8.
go back to reference Wongpichedchai S, Hansen R, Koka B, Gudas VM, Fulton AB. Effects of halothane on children's electroretinograms. Ophthalmology. 1992;99:1309–12.CrossRefPubMed Wongpichedchai S, Hansen R, Koka B, Gudas VM, Fulton AB. Effects of halothane on children's electroretinograms. Ophthalmology. 1992;99:1309–12.CrossRefPubMed
9.
go back to reference Iohom G, Gardiner C, Whyte A, O'Connor G, Shorten G. Abnormalities of contrast sensitivity and electroretinogram following sevoflurane anaesthesia. Eur J Anaesthesiol. 2004;21:646–52.CrossRefPubMed Iohom G, Gardiner C, Whyte A, O'Connor G, Shorten G. Abnormalities of contrast sensitivity and electroretinogram following sevoflurane anaesthesia. Eur J Anaesthesiol. 2004;21:646–52.CrossRefPubMed
10.
go back to reference Tremblay F, Parkinson J. Alteration of electroretinographic recordings when performed under sedation or halogenate anesthesia in a pediatric population. Doc Ophthalmol. 2003;107:271–9.CrossRefPubMed Tremblay F, Parkinson J. Alteration of electroretinographic recordings when performed under sedation or halogenate anesthesia in a pediatric population. Doc Ophthalmol. 2003;107:271–9.CrossRefPubMed
11.
go back to reference Kriss A. Skin ERGs: their effectiveness in paediatric visual assessment, confounding factors, and comparison with ERGs recorded using various types of corneal electrode. Int J Psychophysiol. 1994;16:137–46.CrossRefPubMed Kriss A. Skin ERGs: their effectiveness in paediatric visual assessment, confounding factors, and comparison with ERGs recorded using various types of corneal electrode. Int J Psychophysiol. 1994;16:137–46.CrossRefPubMed
12.
go back to reference Papathanasiou ES, Papacostas SS. Flash electroretinography: normative values with surface skin electrodes and no pupil dilation using a standard stimulation protocol. Doc Ophthalmol. 2008;116:61–73.CrossRefPubMed Papathanasiou ES, Papacostas SS. Flash electroretinography: normative values with surface skin electrodes and no pupil dilation using a standard stimulation protocol. Doc Ophthalmol. 2008;116:61–73.CrossRefPubMed
13.
go back to reference Coupland SG, Janaky M. ERG electrode in pediatric patients: comparison of DTL fiber, PVA-gel, and non-corneal skin electrodes. Doc Ophthalmol. 1989;71:427–33.CrossRefPubMed Coupland SG, Janaky M. ERG electrode in pediatric patients: comparison of DTL fiber, PVA-gel, and non-corneal skin electrodes. Doc Ophthalmol. 1989;71:427–33.CrossRefPubMed
14.
go back to reference Zhang G, Nao-i N, Honda Y. Topographical distribution of electroretinographic potentials monitored by skin-electrodes. J Pediatr Ophthalmol Strabismus. 1983;20:153–7.PubMed Zhang G, Nao-i N, Honda Y. Topographical distribution of electroretinographic potentials monitored by skin-electrodes. J Pediatr Ophthalmol Strabismus. 1983;20:153–7.PubMed
15.
go back to reference Jandeck C, Kellner U, Kraus H, Foerster MH. Electrophysiologic studies according to ISCEV (International Society for Clinical Electrophysiology of Vision) standards in children under 10 years of age. Ophthalmologe. 1997;94:796–800.CrossRefPubMed Jandeck C, Kellner U, Kraus H, Foerster MH. Electrophysiologic studies according to ISCEV (International Society for Clinical Electrophysiology of Vision) standards in children under 10 years of age. Ophthalmologe. 1997;94:796–800.CrossRefPubMed
16.
go back to reference Bradshaw K, Hansen R, Fulton A. Comparison of ERGs recorded with skin and corneal-contact electrodes in normal children and adults. Doc Ophthalmol. 2004;109:43–55.CrossRefPubMed Bradshaw K, Hansen R, Fulton A. Comparison of ERGs recorded with skin and corneal-contact electrodes in normal children and adults. Doc Ophthalmol. 2004;109:43–55.CrossRefPubMed
17.
go back to reference Van Quill KR, Dioguardi PK, Tong CT, Gilbert JA, Aaberg Jr TM, Grossniklaus HE, et al. Subconjunctival carboplatin in fibrin sealant in the treatment of transgenic murine retinoblastoma. Ophthalmology. 2005;112:1151–8.CrossRefPubMed Van Quill KR, Dioguardi PK, Tong CT, Gilbert JA, Aaberg Jr TM, Grossniklaus HE, et al. Subconjunctival carboplatin in fibrin sealant in the treatment of transgenic murine retinoblastoma. Ophthalmology. 2005;112:1151–8.CrossRefPubMed
18.
go back to reference Church ML, Norman JC. Electroretinogram responses of the normal thoroughbred horse sedated with detomidine hydrochloride. Vet Ophthalmol. 2012;2:77–83.CrossRef Church ML, Norman JC. Electroretinogram responses of the normal thoroughbred horse sedated with detomidine hydrochloride. Vet Ophthalmol. 2012;2:77–83.CrossRef
19.
go back to reference Sachidanandam R, Krishnakumar S, Gopal L, O'Brien JM, Khetan V, Sen P. Full-field electroretinography under general anesthesia in retinoblastoma. Doc Ophthalmol. 2013;126:149–57.CrossRefPubMed Sachidanandam R, Krishnakumar S, Gopal L, O'Brien JM, Khetan V, Sen P. Full-field electroretinography under general anesthesia in retinoblastoma. Doc Ophthalmol. 2013;126:149–57.CrossRefPubMed
20.
go back to reference Brodie SE, Gobin Y, Dunkel IJ, Kim JW, Abramson DH. Persistence of retinal function after selective ophthalmic artery chemotherapy infusion for retinoblastoma. Doc Ophthalmol. 2009;119:13–22.CrossRefPubMed Brodie SE, Gobin Y, Dunkel IJ, Kim JW, Abramson DH. Persistence of retinal function after selective ophthalmic artery chemotherapy infusion for retinoblastoma. Doc Ophthalmol. 2009;119:13–22.CrossRefPubMed
21.
go back to reference Brodie SE, Paulus YM, Patel M, Gobin YP, Dunkel IJ, Marr BP, et al. ERG monitoring of retinal function during systemic chemotherapy for retinoblastoma. Br J Ophthalmol. 2012;96:877–80.CrossRefPubMed Brodie SE, Paulus YM, Patel M, Gobin YP, Dunkel IJ, Marr BP, et al. ERG monitoring of retinal function during systemic chemotherapy for retinoblastoma. Br J Ophthalmol. 2012;96:877–80.CrossRefPubMed
22.
go back to reference Marmor MF, Cabael L, Shukla S, Hwang JC, Marcus M. Clinical S-cone ERG recording with a commercial hand-held full-field stimulator. Doc Ophthalmol. 2004;109:101–7.CrossRefPubMed Marmor MF, Cabael L, Shukla S, Hwang JC, Marcus M. Clinical S-cone ERG recording with a commercial hand-held full-field stimulator. Doc Ophthalmol. 2004;109:101–7.CrossRefPubMed
23.
go back to reference Sachidanandam R, Khetan V, Sen P. Comparison between fullfield electroretinography obtained from handheld and tabletop devices in normal subjects. Can J Ophthalmol. 2015;50:166–71.CrossRefPubMed Sachidanandam R, Khetan V, Sen P. Comparison between fullfield electroretinography obtained from handheld and tabletop devices in normal subjects. Can J Ophthalmol. 2015;50:166–71.CrossRefPubMed
24.
go back to reference Trick GL, Nesher R, Cooper DG, Shields SM. The human pattern ERG: alteration of response properties with aging. Optom Vis Sci. 1992;69:122–8.CrossRefPubMed Trick GL, Nesher R, Cooper DG, Shields SM. The human pattern ERG: alteration of response properties with aging. Optom Vis Sci. 1992;69:122–8.CrossRefPubMed
25.
go back to reference Wright CE, Williams D, Drasdo N, Harding GF. The influence of age on the electroretinogram and visual evoked potential. Doc Ophthalmol. 1985;59:365–84.CrossRefPubMed Wright CE, Williams D, Drasdo N, Harding GF. The influence of age on the electroretinogram and visual evoked potential. Doc Ophthalmol. 1985;59:365–84.CrossRefPubMed
26.
go back to reference Hébert M, Vaegan, Lachapelle P. Reproducibility of ERG responses obtained with the DTL electrode. Vision Res. 1999;39:1069–70.CrossRefPubMed Hébert M, Vaegan, Lachapelle P. Reproducibility of ERG responses obtained with the DTL electrode. Vision Res. 1999;39:1069–70.CrossRefPubMed
27.
go back to reference Kuze M, Uji Y. Comparison between Dawson, Trick, and Litzkow electrode and contact lens electrodes used in clinical electroretinography. Jpn J Ophthalmol. 2000;44:374–80.CrossRefPubMed Kuze M, Uji Y. Comparison between Dawson, Trick, and Litzkow electrode and contact lens electrodes used in clinical electroretinography. Jpn J Ophthalmol. 2000;44:374–80.CrossRefPubMed
28.
go back to reference Yin H, Pardue M. Performance of the DTL electrode compared to the jet contact lens electrode in clinical testing. Doc Ophthalmol. 2004;108:77–86.CrossRefPubMed Yin H, Pardue M. Performance of the DTL electrode compared to the jet contact lens electrode in clinical testing. Doc Ophthalmol. 2004;108:77–86.CrossRefPubMed
29.
go back to reference Mortlock KE, Binns AM, Aldebasi YH, North RV. Inter-subject, inter-ocular and inter-session repeatability of the photopic negative response of the electroretinogram recorded using DTL and skin electrodes. Doc Ophthalmol. 2010;121:123–34.CrossRefPubMed Mortlock KE, Binns AM, Aldebasi YH, North RV. Inter-subject, inter-ocular and inter-session repeatability of the photopic negative response of the electroretinogram recorded using DTL and skin electrodes. Doc Ophthalmol. 2010;121:123–34.CrossRefPubMed
30.
go back to reference Hawlina M, Konec B. New corneal HK*-loop electrode for clinical electroretinography. Doc Ophthalmol. 1992;81:253–9.CrossRefPubMed Hawlina M, Konec B. New corneal HK*-loop electrode for clinical electroretinography. Doc Ophthalmol. 1992;81:253–9.CrossRefPubMed
31.
go back to reference Esakowitz L, Kriss A, Shawkat F. A comparison of flash electroretinograms recorded from Burian Allen, JET, C-glide, gold foil, DTL and skin electrodes. Eye (Lond). 1993;7:169–71.CrossRef Esakowitz L, Kriss A, Shawkat F. A comparison of flash electroretinograms recorded from Burian Allen, JET, C-glide, gold foil, DTL and skin electrodes. Eye (Lond). 1993;7:169–71.CrossRef
32.
go back to reference McCulloch DL, Van Boemel GB, Borchert MS. Comparisons of contact lens, foil, fiber and skin electrodes for patterns electroretinograms. Doc Ophthalmol. 1998;94:327–40.CrossRef McCulloch DL, Van Boemel GB, Borchert MS. Comparisons of contact lens, foil, fiber and skin electrodes for patterns electroretinograms. Doc Ophthalmol. 1998;94:327–40.CrossRef
Metadata
Title
Comparing DTL microfiber and Neuroline skin electrode in the Mini Ganzfeld ERG
Authors
Anastasia Lapkovska
Anja M. Palmowski-Wolfe
Margarita G. Todorova
Publication date
01-12-2016
Publisher
BioMed Central
Published in
BMC Ophthalmology / Issue 1/2016
Electronic ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-016-0311-4

Other articles of this Issue 1/2016

BMC Ophthalmology 1/2016 Go to the issue