Published in: BMC Medical Imaging 1/2016

Open Access 01-12-2016 | Research article

Development of a morphology-based modeling technique for tracking solid-body displacements: examining the reliability of a potential MRI-only approach for joint kinematics assessment

Authors: Niladri K. Mahato, Stephane Montuelle, John Cotton, Susan Williams, James Thomas, Brian Clark

Published in: BMC Medical Imaging | Issue 1/2016

Login to get access

Abstract

Background

Single or biplanar video radiography and Roentgen stereophotogrammetry (RSA) techniques used for the assessment of in-vivo joint kinematics involves application of ionizing radiation, which is a limitation for clinical research involving human subjects. To overcome this limitation, our long-term goal is to develop a magnetic resonance imaging (MRI)-only, three dimensional (3-D) modeling technique that permits dynamic imaging of joint motion in humans. Here, we present our initial findings, as well as reliability data, for an MRI-only protocol and modeling technique.

Methods

We developed a morphology-based motion-analysis technique that uses MRI of custom-built solid-body objects to animate and quantify experimental displacements between them. The technique involved four major steps. First, the imaging volume was calibrated using a custom-built grid. Second, 3-D models were segmented from axial scans of two custom-built solid-body cubes. Third, these cubes were positioned at pre-determined relative displacements (translation and rotation) in the magnetic resonance coil and scanned with a T1 and a fast contrast-enhanced pulse sequences. The digital imaging and communications in medicine (DICOM) images were then processed for animation. The fourth step involved importing these processed images into an animation software, where they were displayed as background scenes. In the same step, 3-D models of the cubes were imported into the animation software, where the user manipulated the models to match their outlines in the scene (rotoscoping) and registered the models into an anatomical joint system. Measurements of displacements obtained from two different rotoscoping sessions were tested for reliability using coefficient of variations (CV), intraclass correlation coefficients (ICC), Bland-Altman plots, and Limits of Agreement analyses.

Results

Between-session reliability was high for both the T1 and the contrast-enhanced sequences. Specifically, the average CVs for translation were 4.31 % and 5.26 % for the two pulse sequences, respectively, while the ICCs were 0.99 for both. For rotation measures, the CVs were 3.19 % and 2.44 % for the two pulse sequences with the ICCs being 0.98 and 0.97, respectively. A novel biplanar imaging approach also yielded high reliability with mean CVs of 2.66 % and 3.39 % for translation in the x- and z-planes, respectively, and ICCs of 0.97 in both planes.

Conclusions

This work provides basic proof-of-concept for a reliable marker-less non-ionizing-radiation-based quasi-dynamic motion quantification technique that can potentially be developed into a tool for real-time joint kinematics analysis.
Literature
1.
go back to reference Cappozzo A et al. Human movement analysis using stereophotogrammetry. Part 1: theoretical background. Gait Posture. 2005;21(2):186–96.PubMed Cappozzo A et al. Human movement analysis using stereophotogrammetry. Part 1: theoretical background. Gait Posture. 2005;21(2):186–96.PubMed
2.
go back to reference Della Croce U et al. Human movement analysis using stereophotogrammetry. Part 4: assessment of anatomical landmark misplacement and its effects on joint kinematics. Gait Posture. 2005;21(2):226–37.CrossRefPubMed Della Croce U et al. Human movement analysis using stereophotogrammetry. Part 4: assessment of anatomical landmark misplacement and its effects on joint kinematics. Gait Posture. 2005;21(2):226–37.CrossRefPubMed
3.
go back to reference Reinschmidt C et al. Effect of skin movement on the analysis of skeletal knee joint motion during running. J Biomech. 1997;30(7):729–32.CrossRefPubMed Reinschmidt C et al. Effect of skin movement on the analysis of skeletal knee joint motion during running. J Biomech. 1997;30(7):729–32.CrossRefPubMed
4.
go back to reference Cereatti A, Della Croce U, Cappozzo A. Reconstruction of skeletal movement using skin markers: comparative assessment of bone pose estimators. J Neuroeng Rehabil. 2006;3:7.CrossRefPubMedPubMedCentral Cereatti A, Della Croce U, Cappozzo A. Reconstruction of skeletal movement using skin markers: comparative assessment of bone pose estimators. J Neuroeng Rehabil. 2006;3:7.CrossRefPubMedPubMedCentral
5.
go back to reference Chiari L et al. Human movement analysis using stereophotogrammetry. Part 2: instrumental errors. Gait Posture. 2005;21(2):197–211.CrossRefPubMed Chiari L et al. Human movement analysis using stereophotogrammetry. Part 2: instrumental errors. Gait Posture. 2005;21(2):197–211.CrossRefPubMed
6.
go back to reference Brainerd EL et al. X-ray reconstruction of moving morphology (XROMM): precision, accuracy and applications in comparative biomechanics research. J Exp Zool A Ecol Genet Physiol. 2010;313(5):262–79.PubMed Brainerd EL et al. X-ray reconstruction of moving morphology (XROMM): precision, accuracy and applications in comparative biomechanics research. J Exp Zool A Ecol Genet Physiol. 2010;313(5):262–79.PubMed
7.
go back to reference Yamazaki T et al. Improvement of depth position in 2-D/3-D registration of knee implants using single-plane fluoroscopy. IEEE Trans Med Imaging. 2004;23(5):602–12.CrossRefPubMed Yamazaki T et al. Improvement of depth position in 2-D/3-D registration of knee implants using single-plane fluoroscopy. IEEE Trans Med Imaging. 2004;23(5):602–12.CrossRefPubMed
8.
go back to reference Axelsson P, Karlsson BS. Intervertebral mobility in the progressive degenerative process. A radiostereometric analysis. Eur Spine J. 2004;13(6):567–72.CrossRefPubMedPubMedCentral Axelsson P, Karlsson BS. Intervertebral mobility in the progressive degenerative process. A radiostereometric analysis. Eur Spine J. 2004;13(6):567–72.CrossRefPubMedPubMedCentral
10.
go back to reference Bojan AJ et al. Three-dimensional bone-implant movements in trochanteric hip fractures: Precision and accuracy of radiostereometric analysis in a phantom model. J Orthop Res. 2015;33(5):705–11.CrossRefPubMed Bojan AJ et al. Three-dimensional bone-implant movements in trochanteric hip fractures: Precision and accuracy of radiostereometric analysis in a phantom model. J Orthop Res. 2015;33(5):705–11.CrossRefPubMed
11.
go back to reference Bey MJ et al. Accuracy of biplane x-ray imaging combined with model-based tracking for measuring in-vivo patellofemoral joint motion. J Orthop Surg Res. 2008;3:38.CrossRefPubMedPubMedCentral Bey MJ et al. Accuracy of biplane x-ray imaging combined with model-based tracking for measuring in-vivo patellofemoral joint motion. J Orthop Surg Res. 2008;3:38.CrossRefPubMedPubMedCentral
12.
go back to reference Martin DE et al. Model-based tracking of the hip: implications for novel analyses of hip pathology. J Arthroplasty. 2011;26(1):88–97.CrossRefPubMed Martin DE et al. Model-based tracking of the hip: implications for novel analyses of hip pathology. J Arthroplasty. 2011;26(1):88–97.CrossRefPubMed
13.
go back to reference Lai JY et al. A new registration method for three-dimensional knee nearthrosis model using two X-ray images. Comput Methods Biomech Biomed Engin. 2010;13(2):265–78.CrossRefPubMed Lai JY et al. A new registration method for three-dimensional knee nearthrosis model using two X-ray images. Comput Methods Biomech Biomed Engin. 2010;13(2):265–78.CrossRefPubMed
14.
go back to reference Hurschler C et al. Comparison of the model-based and marker-based roentgen stereophotogrammetry methods in a typical clinical setting. J Arthroplasty. 2009;24(4):594–606.CrossRefPubMed Hurschler C et al. Comparison of the model-based and marker-based roentgen stereophotogrammetry methods in a typical clinical setting. J Arthroplasty. 2009;24(4):594–606.CrossRefPubMed
15.
go back to reference Bey MJ et al. Validation of a new model-based tracking technique for measuring three-dimensional, in vivo glenohumeral joint kinematics. J Biomech Eng. 2006;128(4):604–9.CrossRefPubMedPubMedCentral Bey MJ et al. Validation of a new model-based tracking technique for measuring three-dimensional, in vivo glenohumeral joint kinematics. J Biomech Eng. 2006;128(4):604–9.CrossRefPubMedPubMedCentral
16.
go back to reference Dial KP, Goslow GE, Jenkins FA. The Functional-Anatomy of the Shoulder in the European Starling (Sturnus-Vulgaris). J Morphol. 1991;207(3):327–44.CrossRef Dial KP, Goslow GE, Jenkins FA. The Functional-Anatomy of the Shoulder in the European Starling (Sturnus-Vulgaris). J Morphol. 1991;207(3):327–44.CrossRef
17.
go back to reference Jenkins Jr FA, Dial KP, Goslow Jr GE. A cineradiographic analysis of bird flight: the wishbone in starlings is a spring. Science. 1988;241(4872):1495–8.CrossRefPubMed Jenkins Jr FA, Dial KP, Goslow Jr GE. A cineradiographic analysis of bird flight: the wishbone in starlings is a spring. Science. 1988;241(4872):1495–8.CrossRefPubMed
18.
go back to reference Zhu Z et al. The accuracy and repeatability of an automatic 2D–3D fluoroscopic image-model registration technique for determining shoulder joint kinematics. Med Eng Phys. 2012;34(9):1303–9.CrossRefPubMed Zhu Z et al. The accuracy and repeatability of an automatic 2D–3D fluoroscopic image-model registration technique for determining shoulder joint kinematics. Med Eng Phys. 2012;34(9):1303–9.CrossRefPubMed
19.
go back to reference Baier DB, Gatesy SM. Three-dimensional skeletal kinematics of the shoulder girdle and forelimb in walking Alligator. J Anat. 2013;223(5):462–73.PubMedPubMedCentral Baier DB, Gatesy SM. Three-dimensional skeletal kinematics of the shoulder girdle and forelimb in walking Alligator. J Anat. 2013;223(5):462–73.PubMedPubMedCentral
20.
go back to reference Gatesy SM, David BB. The Origin of the Avian Flight Stroke: A Kinematic and Kinetic Perspective. Paleobiology. 2005;31(3):382–399.CrossRef Gatesy SM, David BB. The Origin of the Avian Flight Stroke: A Kinematic and Kinetic Perspective. Paleobiology. 2005;31(3):382–399.CrossRef
21.
go back to reference Gatesy SM et al. Scientific rotoscoping: a morphology-based method of 3-D motion analysis and visualization. J Exp Zool A Ecol Genet Physiol. 2010;313(5):244–61.PubMed Gatesy SM et al. Scientific rotoscoping: a morphology-based method of 3-D motion analysis and visualization. J Exp Zool A Ecol Genet Physiol. 2010;313(5):244–61.PubMed
22.
go back to reference Anderst WJ, Vaidya R, Tashman S. A technique to measure three-dimensional in vivo rotation of fused and adjacent lumbar vertebrae. Spine J. 2008;8(6):991–7.CrossRefPubMed Anderst WJ, Vaidya R, Tashman S. A technique to measure three-dimensional in vivo rotation of fused and adjacent lumbar vertebrae. Spine J. 2008;8(6):991–7.CrossRefPubMed
23.
go back to reference Tashman S et al. Dynamic function of the ACL-reconstructed knee during running. Clin Orthop Relat Res. 2007;454:66–73.CrossRefPubMed Tashman S et al. Dynamic function of the ACL-reconstructed knee during running. Clin Orthop Relat Res. 2007;454:66–73.CrossRefPubMed
24.
go back to reference Schueler BA. The AAPM/RSNA physics tutorial for residents: general overview of fluoroscopic imaging. Radiographics. 2000;20(4):1115–26.CrossRefPubMed Schueler BA. The AAPM/RSNA physics tutorial for residents: general overview of fluoroscopic imaging. Radiographics. 2000;20(4):1115–26.CrossRefPubMed
25.
go back to reference Latini G et al. Reproductive effects of low-to-moderate medical radiation exposure. Curr Med Chem. 2012;19(36):6171–7.CrossRefPubMed Latini G et al. Reproductive effects of low-to-moderate medical radiation exposure. Curr Med Chem. 2012;19(36):6171–7.CrossRefPubMed
26.
go back to reference Kesavachandran CN, Haamann F, Nienhaus A. Radiation exposure of eyes, thyroid gland and hands in orthopaedic staff: a systematic review. Eur J Med Res. 2012;17:28.CrossRefPubMedPubMedCentral Kesavachandran CN, Haamann F, Nienhaus A. Radiation exposure of eyes, thyroid gland and hands in orthopaedic staff: a systematic review. Eur J Med Res. 2012;17:28.CrossRefPubMedPubMedCentral
27.
go back to reference Hricak H et al. Managing radiation use in medical imaging: a multifaceted challenge. Radiology. 2011;258(3):889–905.CrossRefPubMed Hricak H et al. Managing radiation use in medical imaging: a multifaceted challenge. Radiology. 2011;258(3):889–905.CrossRefPubMed
28.
go back to reference Mellor FE, Thomas P, Breen A. Moving back: The radiation dose received from lumbar spine quantitative fluoroscopy compared to lumbar spine radiographs with suggestions for dose reduction. Radiography. 2014;20(3):251–7.CrossRefPubMedPubMedCentral Mellor FE, Thomas P, Breen A. Moving back: The radiation dose received from lumbar spine quantitative fluoroscopy compared to lumbar spine radiographs with suggestions for dose reduction. Radiography. 2014;20(3):251–7.CrossRefPubMedPubMedCentral
29.
go back to reference Boice Jr JD, Mandel JS, Doody MM. Breast cancer among radiologic technologists. JAMA. 1995;274(5):394–401.CrossRefPubMed Boice Jr JD, Mandel JS, Doody MM. Breast cancer among radiologic technologists. JAMA. 1995;274(5):394–401.CrossRefPubMed
30.
go back to reference Licciardone JC, Gatchel R, Dagenais S. Assessment and management of back pain. JAMA Intern Med. 2014;174(3):478–9.CrossRefPubMed Licciardone JC, Gatchel R, Dagenais S. Assessment and management of back pain. JAMA Intern Med. 2014;174(3):478–9.CrossRefPubMed
31.
go back to reference Freddolini M, Strike S, Lee RY. Stiffness properties of the trunk in people with low back pain. Hum Mov Sci. 2014;36:70–9.CrossRefPubMed Freddolini M, Strike S, Lee RY. Stiffness properties of the trunk in people with low back pain. Hum Mov Sci. 2014;36:70–9.CrossRefPubMed
32.
33.
go back to reference Pearcy MJ, Bogduk N. Instantaneous axes of rotation of the lumbar intervertebral joints. Spine (Phila Pa 1976). 1988;13(9):1033–41.CrossRef Pearcy MJ, Bogduk N. Instantaneous axes of rotation of the lumbar intervertebral joints. Spine (Phila Pa 1976). 1988;13(9):1033–41.CrossRef
34.
go back to reference Gertzbein SD et al. Centrode characteristics of the lumbar spine as a function of segmental instability. Clin Orthop Relat Res. 1986;208:48–51.PubMed Gertzbein SD et al. Centrode characteristics of the lumbar spine as a function of segmental instability. Clin Orthop Relat Res. 1986;208:48–51.PubMed
35.
go back to reference Panjabi MM. Clinical spinal instability and low back pain. J Electromyogr Kinesiol. 2003;13(4):371–9.CrossRefPubMed Panjabi MM. Clinical spinal instability and low back pain. J Electromyogr Kinesiol. 2003;13(4):371–9.CrossRefPubMed
36.
go back to reference Dvorak J et al. Clinical validation of functional flexion-extension roentgenograms of the lumbar spine. Spine (Phila Pa 1976). 1991;16(8):943–50.CrossRef Dvorak J et al. Clinical validation of functional flexion-extension roentgenograms of the lumbar spine. Spine (Phila Pa 1976). 1991;16(8):943–50.CrossRef
37.
go back to reference McGill SM, Cholewicki J. Biomechanical basis for stability: an explanation to enhance clinical utility. J Orthop Sports Phys Ther. 2001;31(2):96–100.CrossRefPubMed McGill SM, Cholewicki J. Biomechanical basis for stability: an explanation to enhance clinical utility. J Orthop Sports Phys Ther. 2001;31(2):96–100.CrossRefPubMed
38.
go back to reference Passias PG et al. Segmental lumbar rotation in patients with discogenic low back pain during functional weight-bearing activities. J Bone Joint Surg Am. 2011;93(1):29–37.CrossRefPubMedPubMedCentral Passias PG et al. Segmental lumbar rotation in patients with discogenic low back pain during functional weight-bearing activities. J Bone Joint Surg Am. 2011;93(1):29–37.CrossRefPubMedPubMedCentral
39.
go back to reference Weiler PJ, King GJ, Gertzbein SD. Analysis of sagittal plane instability of the lumbar spine in vivo. Spine (Phila Pa 1976). 1990;15(12):1300–6.CrossRef Weiler PJ, King GJ, Gertzbein SD. Analysis of sagittal plane instability of the lumbar spine in vivo. Spine (Phila Pa 1976). 1990;15(12):1300–6.CrossRef
40.
go back to reference Yeager MS, Cook DJ, Cheng BC. Reliability of computer-assisted lumbar intervertebral measurements using a novel vertebral motion analysis system. Spine J. 2014;14(2):274–81.CrossRefPubMed Yeager MS, Cook DJ, Cheng BC. Reliability of computer-assisted lumbar intervertebral measurements using a novel vertebral motion analysis system. Spine J. 2014;14(2):274–81.CrossRefPubMed
41.
42.
go back to reference Tan Y et al. Kinetic magnetic resonance imaging analysis of lumbar segmental mobility in patients without significant spondylosis. Eur Spine J. 2012;21(12):2673–9.CrossRefPubMedPubMedCentral Tan Y et al. Kinetic magnetic resonance imaging analysis of lumbar segmental mobility in patients without significant spondylosis. Eur Spine J. 2012;21(12):2673–9.CrossRefPubMedPubMedCentral
43.
go back to reference Yao Q et al. Motion characteristics of the lumbar spinous processes with degenerative disc disease and degenerative spondylolisthesis. Eur Spine J. 2013;22(12):2702–9.CrossRefPubMedPubMedCentral Yao Q et al. Motion characteristics of the lumbar spinous processes with degenerative disc disease and degenerative spondylolisthesis. Eur Spine J. 2013;22(12):2702–9.CrossRefPubMedPubMedCentral
44.
45.
go back to reference Pearcy MJ, Hindle RJ. New method for the non-invasive three-dimensional measurement of human back movement. Clin Biomech (Bristol, Avon). 1989;4(2):73–9.CrossRef Pearcy MJ, Hindle RJ. New method for the non-invasive three-dimensional measurement of human back movement. Clin Biomech (Bristol, Avon). 1989;4(2):73–9.CrossRef
46.
go back to reference Schneider G, Pearcy MJ, Bogduk N. Abnormal motion in spondylolytic spondylolisthesis. Spine (Phila Pa 1976). 2005;30(10):1159–64.CrossRef Schneider G, Pearcy MJ, Bogduk N. Abnormal motion in spondylolytic spondylolisthesis. Spine (Phila Pa 1976). 2005;30(10):1159–64.CrossRef
47.
go back to reference Yang JS et al. Dynamic Radiographic Results of Different Semi-Rigid Fusion Devices for Degenerative Lumbar Spondylolisthesis: "Dynamic Rod" vs. "Dynamic Screw Head". Turk Neurosurg. 2016;26(2):268–73.PubMed Yang JS et al. Dynamic Radiographic Results of Different Semi-Rigid Fusion Devices for Degenerative Lumbar Spondylolisthesis: "Dynamic Rod" vs. "Dynamic Screw Head". Turk Neurosurg. 2016;26(2):268–73.PubMed
48.
go back to reference Keorochana G et al. Effect of sagittal alignment on kinematic changes and degree of disc degeneration in the lumbar spine: an analysis using positional MRI. Spine (Phila Pa 1976). 2011;36(11):893–8.CrossRef Keorochana G et al. Effect of sagittal alignment on kinematic changes and degree of disc degeneration in the lumbar spine: an analysis using positional MRI. Spine (Phila Pa 1976). 2011;36(11):893–8.CrossRef
49.
go back to reference Lee SH, Daffner SD, Wang JC, Davis BC, Alanay A, Kim JS. The change of whole lumbar segmental motion according to the mobility of degenerated disc in the lower lumbar spine: a kinetic MRI study. Eur Spine J. 2014. doi:10.1007/s00586-014-3277-z. Lee SH, Daffner SD, Wang JC, Davis BC, Alanay A, Kim JS. The change of whole lumbar segmental motion according to the mobility of degenerated disc in the lower lumbar spine: a kinetic MRI study. Eur Spine J. 2014. doi:10.​1007/​s00586-014-3277-z.
50.
go back to reference Chung SS et al. Effect of low back posture on the morphology of the spinal canal. Skeletal Radiol. 2000;29(4):217–23.CrossRefPubMed Chung SS et al. Effect of low back posture on the morphology of the spinal canal. Skeletal Radiol. 2000;29(4):217–23.CrossRefPubMed
51.
go back to reference Rebmann AJ, Sheehan FT. Precise 3D skeletal kinematics using fast phase contrast magnetic resonance imaging. J Magn Reson Imaging. 2003;17(2):206–13.CrossRefPubMed Rebmann AJ, Sheehan FT. Precise 3D skeletal kinematics using fast phase contrast magnetic resonance imaging. J Magn Reson Imaging. 2003;17(2):206–13.CrossRefPubMed
52.
go back to reference Behnam AJ, Herzka DA, Sheehan FT. Assessing the accuracy and precision of musculoskeletal motion tracking using cine-PC MRI on a 3.0 T platform. J Biomech. 2011;44(1):193–7.CrossRefPubMedPubMedCentral Behnam AJ, Herzka DA, Sheehan FT. Assessing the accuracy and precision of musculoskeletal motion tracking using cine-PC MRI on a 3.0 T platform. J Biomech. 2011;44(1):193–7.CrossRefPubMedPubMedCentral
53.
go back to reference Ward SR et al. Assessment of patellofemoral relationships using kinematic MRI: Comparison between qualitative and quantitative methods. J Magn Reson Imaging. 2002;16(1):69–74.CrossRefPubMed Ward SR et al. Assessment of patellofemoral relationships using kinematic MRI: Comparison between qualitative and quantitative methods. J Magn Reson Imaging. 2002;16(1):69–74.CrossRefPubMed
54.
go back to reference Quick HH et al. Real-time MRI of joint movement with TrueFISP. J Magn Reson Imaging. 2002;15(6):710–5.CrossRefPubMed Quick HH et al. Real-time MRI of joint movement with TrueFISP. J Magn Reson Imaging. 2002;15(6):710–5.CrossRefPubMed
55.
go back to reference Shellock FG et al. Kinematic magnetic resonance imaging of the joints: techniques and clinical applications. Magn Reson Q. 1991;7(2):104–35.PubMed Shellock FG et al. Kinematic magnetic resonance imaging of the joints: techniques and clinical applications. Magn Reson Q. 1991;7(2):104–35.PubMed
56.
go back to reference Clarke EC et al. A non-invasive, 3D, dynamic MRI method for measuring muscle moment arms in vivo: demonstration in the human ankle joint and Achilles tendon. Med Eng Phys. 2015;37(1):93–9.CrossRefPubMed Clarke EC et al. A non-invasive, 3D, dynamic MRI method for measuring muscle moment arms in vivo: demonstration in the human ankle joint and Achilles tendon. Med Eng Phys. 2015;37(1):93–9.CrossRefPubMed
57.
go back to reference Fellows RA et al. Repeatability of a novel technique for in vivo measurement of three-dimensional patellar tracking using magnetic resonance imaging. J Magn Reson Imaging. 2005;22(1):145–53.CrossRefPubMed Fellows RA et al. Repeatability of a novel technique for in vivo measurement of three-dimensional patellar tracking using magnetic resonance imaging. J Magn Reson Imaging. 2005;22(1):145–53.CrossRefPubMed
58.
go back to reference Fellows RA et al. Magnetic resonance imaging for in vivo assessment of three-dimensional patellar tracking. J Biomech. 2005;38(8):1643–52.CrossRefPubMed Fellows RA et al. Magnetic resonance imaging for in vivo assessment of three-dimensional patellar tracking. J Biomech. 2005;38(8):1643–52.CrossRefPubMed
59.
go back to reference d'Entremont AG et al. Do dynamic-based MR knee kinematics methods produce the same results as static methods? Magn Reson Med. 2013;69(6):1634–44.CrossRefPubMed d'Entremont AG et al. Do dynamic-based MR knee kinematics methods produce the same results as static methods? Magn Reson Med. 2013;69(6):1634–44.CrossRefPubMed
60.
go back to reference Whitehurst DG, et al. Implementing Stratified Primary care Management for low Back Pain: Cost Utility Analysis alongside a Prospective, Population-based, Sequential Comparison Study. Spine (Phila Pa 1976). 2015. Whitehurst DG, et al. Implementing Stratified Primary care Management for low Back Pain: Cost Utility Analysis alongside a Prospective, Population-based, Sequential Comparison Study. Spine (Phila Pa 1976). 2015.
61.
go back to reference Manchikanti L et al. Epidemiology of low back pain in adults. Neuromodulation. 2014;17 Suppl 2:3–10.CrossRefPubMed Manchikanti L et al. Epidemiology of low back pain in adults. Neuromodulation. 2014;17 Suppl 2:3–10.CrossRefPubMed
62.
go back to reference Schaefer C et al. Pain severity and the economic burden of neuropathic pain in the United States: BEAT Neuropathic Pain Observational Study. Clinicoecon Outcomes Res. 2014;6:483–96.PubMedPubMedCentral Schaefer C et al. Pain severity and the economic burden of neuropathic pain in the United States: BEAT Neuropathic Pain Observational Study. Clinicoecon Outcomes Res. 2014;6:483–96.PubMedPubMedCentral
63.
go back to reference North RB et al. A review of economic factors related to the delivery of health care for chronic low back pain. Neuromodulation. 2014;17 Suppl 2:69–76.CrossRefPubMed North RB et al. A review of economic factors related to the delivery of health care for chronic low back pain. Neuromodulation. 2014;17 Suppl 2:69–76.CrossRefPubMed
64.
go back to reference March L et al. Burden of disability due to musculoskeletal (MSK) disorders. Best Pract Res Clin Rheumatol. 2014;28(3):353–66.CrossRefPubMed March L et al. Burden of disability due to musculoskeletal (MSK) disorders. Best Pract Res Clin Rheumatol. 2014;28(3):353–66.CrossRefPubMed
65.
go back to reference Wu G et al. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion--part I: ankle, hip, and spine. International Society of Biomechanics. J Biomech. 2002;35(4):543–8.CrossRefPubMed Wu G et al. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human joint motion--part I: ankle, hip, and spine. International Society of Biomechanics. J Biomech. 2002;35(4):543–8.CrossRefPubMed
66.
go back to reference Cheung JP, Shigematsu H, Cheung KM. Verification of measurements of lumbar spinal dimensions in T1- and T2-weighted magnetic resonance imaging sequences. Spine J. 2014;14(8):1476–83.CrossRefPubMed Cheung JP, Shigematsu H, Cheung KM. Verification of measurements of lumbar spinal dimensions in T1- and T2-weighted magnetic resonance imaging sequences. Spine J. 2014;14(8):1476–83.CrossRefPubMed
67.
go back to reference Ostrofsky KR, Churchill SE. Sex determination by discriminant function analysis of lumbar vertebrae. J Forensic Sci. 2015;60(1):21–8.CrossRefPubMed Ostrofsky KR, Churchill SE. Sex determination by discriminant function analysis of lumbar vertebrae. J Forensic Sci. 2015;60(1):21–8.CrossRefPubMed
68.
go back to reference Rhoad RC et al. A new in vivo technique for three-dimensional shoulder kinematics analysis. Skeletal Radiol. 1998;27(2):92–7.CrossRefPubMed Rhoad RC et al. A new in vivo technique for three-dimensional shoulder kinematics analysis. Skeletal Radiol. 1998;27(2):92–7.CrossRefPubMed
69.
70.
go back to reference Nyakatura JA, Fischer MS. Three-dimensional kinematic analysis of the pectoral girdle during upside-down locomotion of two-toed sloths (Choloepus didactylus, Linne 1758). Front Zool. 2010;7:21.CrossRefPubMedPubMedCentral Nyakatura JA, Fischer MS. Three-dimensional kinematic analysis of the pectoral girdle during upside-down locomotion of two-toed sloths (Choloepus didactylus, Linne 1758). Front Zool. 2010;7:21.CrossRefPubMedPubMedCentral
71.
go back to reference Clark BC, Cook SB, Ploutz-Snyder LL. Reliability of techniques to assess human neuromuscular function in vivo. J Electromyogr Kinesiol. 2007;17(1):90–101.CrossRefPubMed Clark BC, Cook SB, Ploutz-Snyder LL. Reliability of techniques to assess human neuromuscular function in vivo. J Electromyogr Kinesiol. 2007;17(1):90–101.CrossRefPubMed
72.
go back to reference Atkinson G, Nevill AM. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med. 1998;26(4):217–38.CrossRefPubMed Atkinson G, Nevill AM. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med. 1998;26(4):217–38.CrossRefPubMed
73.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.CrossRefPubMed
74.
Metadata
Title
Development of a morphology-based modeling technique for tracking solid-body displacements: examining the reliability of a potential MRI-only approach for joint kinematics assessment
Authors
Niladri K. Mahato
Stephane Montuelle
John Cotton
Susan Williams
James Thomas
Brian Clark
Publication date
01-12-2016
Publisher
BioMed Central
Published in
BMC Medical Imaging / Issue 1/2016
Electronic ISSN: 1471-2342
DOI
https://doi.org/10.1186/s12880-016-0140-1

Other articles of this Issue 1/2016

BMC Medical Imaging 1/2016 Go to the issue