Open Access 01-12-2019 | Urinary Tract Infection | Research article

Clinical usefulness of iQ200/iChem Velocity workstation for screening of urine culture

Authors: Jong-Mi Lee, Doo-Jin Baek, Kang Gyun Park, Eunhee Han, Yeon-Joon Park

Published in: BMC Infectious Diseases | Issue 1/2019

Login to get access

Abstract

Background

Clinical microbiology laboratories are asked to process large numbers of urine specimens for culture, but only 20–40% of them are positive. Therefore, a rapid, reliable screening method is necessary to speed up the reporting of a negative result. In this study, we evaluated the iQ200/iChem workstation, which is a combination of digital imaging software and a strip reader to predict negative urine culture.

Method

A total of 1942 urine specimens were processed through both culture and iQ200/ iChem workstation. We analyzed the performance using two definition of positive urine culture; one or two potential uropathogens at a concentration of ≥105 CFU/ml and ≥ 104 CFU/ml. We assessed combinations of parameters (ASP; all small particles, WBC; leukocyte, BACT; bcteria, LE; leukocyte esterase) applying various cut-offs which can achieve the negative predictive value (NPV) ≥97% and culture reduction rate ≥ 50%.

Results

The culture positive rate was 12.8 and 18.4% applying the criteria of ≥105 CFU/ml and ≥ 104 CFU/ml, respectively. The area under the curve (AUC) of each parameter for ≥105 CFU/ml / ≥104 CFU/ml bacteriuria was 795 /0.719 for WBC, 0.722 / 0.701 for ASP and 0.740 /0.704 for bacteria. Therefore, we investigated the combination of the parameters. With the fixed parameter of BACT≥1/HPF and positive LE, the combinations of WBC ≥ 4/HPF and ASP ≥8500/μl or WBC ≥ 6/HPF and ASP≥5500/μl showed good performance for detecting ≥105 CFU/ml uropathogen. The ranges of sensitivity, specificity, negative predictive value and culture reduction rate were 91.5–92.3%, 49.8–52.6%, 97.7–97.9% and 50.4–53.0%, respectively. However, none of the combined setting yielded acceptable range of NPV for detecting ≥104 CFU/ml uropathogen (NPV 92.9–94.9%). Enterococcus spp. was the most common uropathogen causing the false negative results (55.7%), and also the main pathogen among the positive culture of 104-5 CFU/ml bacteriuria (45%).

Conclusions

iQ200/iChem workstation was excellent in detection of ≥105 CFU/ml uropathogen, but unsatisfactory in detection of 104–5 CFU/ml uropathogen and Enterococcus spp. It can be useful for screening of urine specimens to reduce bacterial culture. However, notice from clinician will be necessary for specimens from the patients with high risk for UTI, such as pregnant woman, infant, elderly or immune compromised patients.
Appendix
Available only for authorised users
Literature
1.
go back to reference Peleg AY, Hooper DC. Hospital-acquired infections due to gram-negative bacteria. N Engl J Med. 2010;362(19):1804–13.CrossRef Peleg AY, Hooper DC. Hospital-acquired infections due to gram-negative bacteria. N Engl J Med. 2010;362(19):1804–13.CrossRef
3.
go back to reference Shang YJ, Wang QQ, Zhang JR, Xu YL, Zhang WW, Chen Y, Gu ML, Hu ZD, Deng AM. Systematic review and meta-analysis of flow cytometry in urinary tract infection screening. Clin Chim Acta. 2013;424:90–5.CrossRef Shang YJ, Wang QQ, Zhang JR, Xu YL, Zhang WW, Chen Y, Gu ML, Hu ZD, Deng AM. Systematic review and meta-analysis of flow cytometry in urinary tract infection screening. Clin Chim Acta. 2013;424:90–5.CrossRef
4.
go back to reference Winquist AG, Orrico MA, Peterson LR. Evaluation of the cytocentrifuge gram stain as a screening test for bacteriuria in specimens from specific patient populations. Am J Clin Pathol. 1997;108(5):515–24.CrossRef Winquist AG, Orrico MA, Peterson LR. Evaluation of the cytocentrifuge gram stain as a screening test for bacteriuria in specimens from specific patient populations. Am J Clin Pathol. 1997;108(5):515–24.CrossRef
5.
go back to reference Semeniuk H, Church D. Evaluation of the leukocyte esterase and nitrite urine dipstick screening tests for detection of bacteriuria in women with suspected uncomplicated urinary tract infections. J Clin Microbiol. 1999;37(9):3051–2.PubMedPubMedCentral Semeniuk H, Church D. Evaluation of the leukocyte esterase and nitrite urine dipstick screening tests for detection of bacteriuria in women with suspected uncomplicated urinary tract infections. J Clin Microbiol. 1999;37(9):3051–2.PubMedPubMedCentral
6.
go back to reference Stürenburg E, Kramer J, Schön G, Cachovan G, Sobottka I. Detection of significant bacteriuria by use of the iQ200 automated urine microscope. J Clin Microbiol. 2014;52(8):2855–60.CrossRef Stürenburg E, Kramer J, Schön G, Cachovan G, Sobottka I. Detection of significant bacteriuria by use of the iQ200 automated urine microscope. J Clin Microbiol. 2014;52(8):2855–60.CrossRef
7.
go back to reference Parta M, Hudson BY, Le TP, Ittmann M, Musher DM, Stager C. IRIS iQ200 workstation as a screen for performing urine culture. Diagn Microbiol Infect Dis. 2013;75(1):5–8.CrossRef Parta M, Hudson BY, Le TP, Ittmann M, Musher DM, Stager C. IRIS iQ200 workstation as a screen for performing urine culture. Diagn Microbiol Infect Dis. 2013;75(1):5–8.CrossRef
8.
go back to reference Ronald A. The etiology of urinary tract infection: traditional and emerging pathogens. Am J Med. 2002;113(1):14–9.CrossRef Ronald A. The etiology of urinary tract infection: traditional and emerging pathogens. Am J Med. 2002;113(1):14–9.CrossRef
9.
go back to reference Leber AL. Clinical microbiology procedures handbook, fourth edition, vol. 2, 4 edn: American society of. Microbiology. 2016. Leber AL. Clinical microbiology procedures handbook, fourth edition, vol. 2, 4 edn: American society of. Microbiology. 2016.
10.
go back to reference Lamchiagdhase P, Preechaborisutkul K, Lomsomboon P, Srisuchart P, Tantiniti P, Khan-u-Ra N, Preechaborisutkul B. Urine sediment examination: a comparison between the manual method and the iQ200 automated urine microscopy analyzer. Clin Chim Acta. 2005;358(1–2):167–74.CrossRef Lamchiagdhase P, Preechaborisutkul K, Lomsomboon P, Srisuchart P, Tantiniti P, Khan-u-Ra N, Preechaborisutkul B. Urine sediment examination: a comparison between the manual method and the iQ200 automated urine microscopy analyzer. Clin Chim Acta. 2005;358(1–2):167–74.CrossRef
11.
go back to reference Linko S, Kouri TT, Toivonen E, Ranta PH, Chapoulaud E, Lalla M. Analytical performance of the Iris iQ200 automated urine microscopy analyzer. Clin Chim Acta. 2006;372(1–2):54–64.CrossRef Linko S, Kouri TT, Toivonen E, Ranta PH, Chapoulaud E, Lalla M. Analytical performance of the Iris iQ200 automated urine microscopy analyzer. Clin Chim Acta. 2006;372(1–2):54–64.CrossRef
12.
go back to reference Russcher A, Kusters E, Wolterbeek R, Kuijper EJ, Cobbaert CM, van der Beek MT. Interlaboratory collaboration for optimized screening for urinary tract infection. J Clin Microbiol. 2016;54(1):93–8.CrossRef Russcher A, Kusters E, Wolterbeek R, Kuijper EJ, Cobbaert CM, van der Beek MT. Interlaboratory collaboration for optimized screening for urinary tract infection. J Clin Microbiol. 2016;54(1):93–8.CrossRef
13.
go back to reference Luciano R, Piga S, Federico L, Argentieri M, Fina F, Cuttini M, Misirocchi E, Emma F, Muraca M. Development of a score based on urinalysis to improve the management of urinary tract infection in children. Clin Chim Acta. 2012;413(3):478–82.CrossRef Luciano R, Piga S, Federico L, Argentieri M, Fina F, Cuttini M, Misirocchi E, Emma F, Muraca M. Development of a score based on urinalysis to improve the management of urinary tract infection in children. Clin Chim Acta. 2012;413(3):478–82.CrossRef
14.
go back to reference Alves L, Ballester F, Camps J, Joven J. Preliminary evaluation of the Iris IQ™ 200 automated urine analyser. Clinical Chemical Laboratory Medicine. 2005;43(9):967–70.CrossRef Alves L, Ballester F, Camps J, Joven J. Preliminary evaluation of the Iris IQ™ 200 automated urine analyser. Clinical Chemical Laboratory Medicine. 2005;43(9):967–70.CrossRef
15.
go back to reference Brilha S, Proenca H, Cristino JM, Hänscheid T. Use of flow cytometry (Sysmex® UF-100) to screen for positive urine cultures: in search for the ideal cut-off. Clin Chem Lab Med. 2010;48(2):289–92.CrossRef Brilha S, Proenca H, Cristino JM, Hänscheid T. Use of flow cytometry (Sysmex® UF-100) to screen for positive urine cultures: in search for the ideal cut-off. Clin Chem Lab Med. 2010;48(2):289–92.CrossRef
16.
go back to reference Yusuf E, Van Herendael B, van Schaeren J. Performance of urinalysis tests and their ability in predicting results of urine cultures: a comparison between automated test strip analyser and flow cytometry in various subpopulations and types of samples. J Clin Pathol. 2016:jclinpath–2016-204108. Yusuf E, Van Herendael B, van Schaeren J. Performance of urinalysis tests and their ability in predicting results of urine cultures: a comparison between automated test strip analyser and flow cytometry in various subpopulations and types of samples. J Clin Pathol. 2016:jclinpath–2016-204108.
17.
go back to reference Foxman B. Epidemiology of urinary tract infections: incidence, morbidity, and economic costs. Am J Med. 2002;113(1):5–13.CrossRef Foxman B. Epidemiology of urinary tract infections: incidence, morbidity, and economic costs. Am J Med. 2002;113(1):5–13.CrossRef
Metadata
Title
Clinical usefulness of iQ200/iChem Velocity workstation for screening of urine culture
Authors
Jong-Mi Lee
Doo-Jin Baek
Kang Gyun Park
Eunhee Han
Yeon-Joon Park
Publication date
01-12-2019
Publisher
BioMed Central
Published in
BMC Infectious Diseases / Issue 1/2019
Electronic ISSN: 1471-2334
DOI
https://doi.org/10.1186/s12879-019-4141-x

Other articles of this Issue 1/2019

BMC Infectious Diseases 1/2019 Go to the issue