Open Access 01-08-2006 | Research

Ventilator-associated pneumonia using a heated humidifier or a heat and moisture exchanger: a randomized controlled trial [ISRCTN88724583]

Authors: Leonardo Lorente, María Lecuona, Alejandro Jiménez, María L Mora, Antonio Sierra

Published in: Critical Care | Issue 4/2006

Login to get access

Abstract

Introduction

Some guidelines to prevent ventilator-associated pneumonia (VAP) do not establish a recommendation for the preferential use of either heat and moisture exchangers (HMEs) or heated humidifiers (HHs), while other guidelines clearly advocate the use of HMEs. The aim of this study was to determine the incidence of VAP associated with HHs or HMEs.

Methods

A randomized study was conducted in the intensive care unit of a university hospital involving patients expected to require mechanical ventilation for >5 days. Patients were assigned to two groups; one group received HH and the other group received HME. Tracheal aspirate samples were obtained on endotracheal intubation, then twice a week, and finally on extubation, in order to diagnose VAP. Throat swabs were taken on admission to the intensive care unit, then twice a week, and finally at discharge from the intensive care unit in order to classify VAP as primary endogenous, secondary endogenous, or exogenous.

Results

A total of 120 patients were assigned to HMEs (60 patients) and HHs (60 patients); 16 patients received mechanical ventilation for less than five days and were excluded from the analysis. Data analysis of the remaining 104 patients (53 HMEs and 51 HHs) showed no significant differences between groups regarding sex, age, Acute Physiology and Chronic Health Evaluation II score, pre-VAP use of antibiotics, days on mechanical ventilation, and diagnosis group. VAP was found in eight of 51 (15.69%) patients in the HH group and in 21 of 53 (39.62%) patients in the HME group (P = 0.006). The median time free of VAP was 20 days (95% confidence interval, 13.34–26.66) for the HH group and was 42 days (95% confidence interval, 35.62–48.37) for the HME group (P < 0.001). Cox regression analysis showed the HME as a risk factor for VAP (hazard rate, 16.2; 95% confidence interval, 4.54–58.04; P < 0.001).

Conclusion

The patients mechanically ventilated for more than five days developed a lower incidence of VAP with a HH than with a HME.
Appendix
Available only for authorised users
Literature
1.
go back to reference Williams R, Rankin N, Smith T, Galler D, Seakins P: Relationship between the humidity and temperature of inspired gas and the function of the airway mucosa. Crit Care Med 1996, 24: 1920-1929. 10.1097/00003246-199611000-00025CrossRefPubMed Williams R, Rankin N, Smith T, Galler D, Seakins P: Relationship between the humidity and temperature of inspired gas and the function of the airway mucosa. Crit Care Med 1996, 24: 1920-1929. 10.1097/00003246-199611000-00025CrossRefPubMed
2.
go back to reference Irlbeck D: Normal mechanisms of heat and moisture exchange in the respiratory tract. Respir Care Clin N Am 1998, 4: 189-198.PubMed Irlbeck D: Normal mechanisms of heat and moisture exchange in the respiratory tract. Respir Care Clin N Am 1998, 4: 189-198.PubMed
3.
go back to reference McFadden ER Jr, Pichurko BM, Bowman HF, Ingenito E, Burns S, Dowling N, Solway J: Thermal mapping of the airways in humans. J Appl Physiol 1985, 58: 564-570. 10.1063/1.335663CrossRefPubMed McFadden ER Jr, Pichurko BM, Bowman HF, Ingenito E, Burns S, Dowling N, Solway J: Thermal mapping of the airways in humans. J Appl Physiol 1985, 58: 564-570. 10.1063/1.335663CrossRefPubMed
4.
go back to reference Chatburn R, Primiano F: A rational basis for humidity therapy. Respir Care 1987, 32: 249-254. Chatburn R, Primiano F: A rational basis for humidity therapy. Respir Care 1987, 32: 249-254.
5.
go back to reference Branson RD: Humidification for patients with artificial airways. Respir Care 1999, 44: 639-642. Branson RD: Humidification for patients with artificial airways. Respir Care 1999, 44: 639-642.
6.
go back to reference Jackson C, Webb AR: An evaluation of the heat and moisture exchange performance of four ventilator circuit filters. Intensive Care Med 1992, 18: 264-268. 10.1007/BF01706470CrossRefPubMed Jackson C, Webb AR: An evaluation of the heat and moisture exchange performance of four ventilator circuit filters. Intensive Care Med 1992, 18: 264-268. 10.1007/BF01706470CrossRefPubMed
7.
go back to reference Fritz KW, Osterhaus A, Mottner J: The humidification of anesthetic gases during anesthesia using heat and moisture exchangers. Anaesthesiol Reanim 1992, 17: 133-144.PubMed Fritz KW, Osterhaus A, Mottner J: The humidification of anesthetic gases during anesthesia using heat and moisture exchangers. Anaesthesiol Reanim 1992, 17: 133-144.PubMed
8.
go back to reference Kugimiya T, Phuc TG, Numata K: Laboratory evaluation of heat-and-moisture exchangers. J Anesth 1989, 3: 80-85. 10.1007/s0054090030080CrossRefPubMed Kugimiya T, Phuc TG, Numata K: Laboratory evaluation of heat-and-moisture exchangers. J Anesth 1989, 3: 80-85. 10.1007/s0054090030080CrossRefPubMed
9.
go back to reference Mebius C: A comparative evaluation of disposable humidifiers. Acta Anaesthesiol Scand 1983, 27: 403-409.CrossRefPubMed Mebius C: A comparative evaluation of disposable humidifiers. Acta Anaesthesiol Scand 1983, 27: 403-409.CrossRefPubMed
10.
go back to reference Sottiaux T, Mignolet G, Damas P, Lamy M: Comparative evaluation of three heat and moisture exchangers during short-term postoperative mechanical ventilation. Chest 1993, 104: 220-224.CrossRefPubMed Sottiaux T, Mignolet G, Damas P, Lamy M: Comparative evaluation of three heat and moisture exchangers during short-term postoperative mechanical ventilation. Chest 1993, 104: 220-224.CrossRefPubMed
11.
go back to reference Rankin N: What is optimum humidity? Respir Care Clin N Am 1998, 4: 321-328.PubMed Rankin N: What is optimum humidity? Respir Care Clin N Am 1998, 4: 321-328.PubMed
12.
go back to reference Luchetti M, Stuani A, Castelli G, Marraro G: Comparison of three different humidification systems during prolonged mechanical ventilation. Minerva Anestesiol 1998, 64: 75-81.PubMed Luchetti M, Stuani A, Castelli G, Marraro G: Comparison of three different humidification systems during prolonged mechanical ventilation. Minerva Anestesiol 1998, 64: 75-81.PubMed
13.
go back to reference Lawrence JC: Humidification practices in the Adult Intensive Care Unit, Prince of Wales Hospital. Respir Care Clin N Am 1998, 4: 301-304.PubMed Lawrence JC: Humidification practices in the Adult Intensive Care Unit, Prince of Wales Hospital. Respir Care Clin N Am 1998, 4: 301-304.PubMed
14.
go back to reference Ryan SN, Rankin N, Meyer E, Williams R: Energy balance in the intubated human airway is an indicator of optimal gas conditioning. Crit Care Med 2002, 30: 355-361. 10.1097/00003246-200202000-00015CrossRefPubMed Ryan SN, Rankin N, Meyer E, Williams R: Energy balance in the intubated human airway is an indicator of optimal gas conditioning. Crit Care Med 2002, 30: 355-361. 10.1097/00003246-200202000-00015CrossRefPubMed
15.
go back to reference Kirton OC, De Haven B, Morgan J, Morejon O, Civetta J: A prospective, randomized comparison of an in-line heat moisture exchange filter and heated wire humidifiers. Rates of ventilator-associated early-onset (community-acquired) or late-onset (hospital-acquired) pneumonia and incidence of endotracheal tube occlusion. Chest 1997, 112: 1055-1059.CrossRefPubMed Kirton OC, De Haven B, Morgan J, Morejon O, Civetta J: A prospective, randomized comparison of an in-line heat moisture exchange filter and heated wire humidifiers. Rates of ventilator-associated early-onset (community-acquired) or late-onset (hospital-acquired) pneumonia and incidence of endotracheal tube occlusion. Chest 1997, 112: 1055-1059.CrossRefPubMed
16.
go back to reference Martin C, Perrin G, Gevaudan MJ, Saux P, Gouin F: Heat and moisture exchangers and vaporizing humidifiers in the intensive care unit. Chest 1990, 97: 144-149.CrossRefPubMed Martin C, Perrin G, Gevaudan MJ, Saux P, Gouin F: Heat and moisture exchangers and vaporizing humidifiers in the intensive care unit. Chest 1990, 97: 144-149.CrossRefPubMed
17.
go back to reference Roustan JP, Kienlen J, Aubas S, Du Cailar J: Comparison of hydrophobic heat and moisture exchangers with heated humidifier during prolonged mechanical ventilation. Intensive Care Med 1992, 18: 97-100. 10.1007/BF01705040CrossRefPubMed Roustan JP, Kienlen J, Aubas S, Du Cailar J: Comparison of hydrophobic heat and moisture exchangers with heated humidifier during prolonged mechanical ventilation. Intensive Care Med 1992, 18: 97-100. 10.1007/BF01705040CrossRefPubMed
18.
go back to reference Misset B, Escudier B, Rivara D, Leclercq B, Nitenberg G: Heat and moisture exchanger vs heated humidifier during long-term mechanical ventilation. A prospective randomized study. Chest 1991, 100: 160-163.CrossRefPubMed Misset B, Escudier B, Rivara D, Leclercq B, Nitenberg G: Heat and moisture exchanger vs heated humidifier during long-term mechanical ventilation. A prospective randomized study. Chest 1991, 100: 160-163.CrossRefPubMed
19.
go back to reference Dreyfuss D, Djedaïni K, Gros I, Mier L, Le Bourdellés G, Cohen Y, Estagnasié P, Coste F, Boussougant : Mechanical ventilation with heated humidifiers or heat and moisture exchangers: effects on patient colonization and incidence of nosocomial pneumonia. Am J Respir Crit Care Med 1995, 151: 986-992.PubMed Dreyfuss D, Djedaïni K, Gros I, Mier L, Le Bourdellés G, Cohen Y, Estagnasié P, Coste F, Boussougant : Mechanical ventilation with heated humidifiers or heat and moisture exchangers: effects on patient colonization and incidence of nosocomial pneumonia. Am J Respir Crit Care Med 1995, 151: 986-992.PubMed
20.
go back to reference Kollef MH, Shapiro SD, Boyd V, Silver P, Von Harz B, Trovillion E, Prentice D: A randomized clinical trial comparing an extended-use hygroscopic condenser humidifier with heated-water humidification in mechanically ventilated patients. Chest 1998, 113: 759-767.CrossRefPubMed Kollef MH, Shapiro SD, Boyd V, Silver P, Von Harz B, Trovillion E, Prentice D: A randomized clinical trial comparing an extended-use hygroscopic condenser humidifier with heated-water humidification in mechanically ventilated patients. Chest 1998, 113: 759-767.CrossRefPubMed
21.
go back to reference Branson R, Davis K, Campbell RS, Johnson DJ, Porembka DT: Humidification in the intensive care unit. Prospective study of a new protocol utilizing heated humidification and a hygroscopic condenser humidifier. Chest 1993, 104: 1800-1805.CrossRefPubMed Branson R, Davis K, Campbell RS, Johnson DJ, Porembka DT: Humidification in the intensive care unit. Prospective study of a new protocol utilizing heated humidification and a hygroscopic condenser humidifier. Chest 1993, 104: 1800-1805.CrossRefPubMed
22.
go back to reference Boots RJ, Howe S, George N, Harris FM, Faoagali J: Clinical utility of hygroscopic heat and moisture exchangers in intensive care patients. Crit Care Med 1997, 25: 1707-1712. 10.1097/00003246-199710000-00021CrossRefPubMed Boots RJ, Howe S, George N, Harris FM, Faoagali J: Clinical utility of hygroscopic heat and moisture exchangers in intensive care patients. Crit Care Med 1997, 25: 1707-1712. 10.1097/00003246-199710000-00021CrossRefPubMed
23.
go back to reference Memish ZA, Oni GA, Djazmati W, Cunningham G, Mah MW: A randomized trial to compare the effects of a heat and moisture exchanger with a heated humidifying system on the occurrence rate of ventilator-associated pneumonia. Am J Infect Control 2001, 29: 301-305. 10.1067/mic.2001.115404CrossRefPubMed Memish ZA, Oni GA, Djazmati W, Cunningham G, Mah MW: A randomized trial to compare the effects of a heat and moisture exchanger with a heated humidifying system on the occurrence rate of ventilator-associated pneumonia. Am J Infect Control 2001, 29: 301-305. 10.1067/mic.2001.115404CrossRefPubMed
24.
go back to reference Lacherade JC, Auburtin M, Cerf C, Van de Louw A, Soufir L, Rebufat Y, Rezaiguia S, Ricard JD, Lellouche F, Brun-Buisson C, Brochard L: Impact of humidification systems on ventilator-associated pneumonia: a randomized multicenter trial. Am J Respir Crit Care Med 2005, 172: 1276-1282. 10.1164/rccm.200408-1028OCCrossRefPubMed Lacherade JC, Auburtin M, Cerf C, Van de Louw A, Soufir L, Rebufat Y, Rezaiguia S, Ricard JD, Lellouche F, Brun-Buisson C, Brochard L: Impact of humidification systems on ventilator-associated pneumonia: a randomized multicenter trial. Am J Respir Crit Care Med 2005, 172: 1276-1282. 10.1164/rccm.200408-1028OCCrossRefPubMed
25.
go back to reference Kranabetter R, Leier M, Kammermeier D, Just HM, Heuser D: The effects of active and passive humidification on ventilation-associated nosocomial pneumonia. Anaesthesist 2004, 53: 29-35. 10.1007/s00101-003-0607-7CrossRefPubMed Kranabetter R, Leier M, Kammermeier D, Just HM, Heuser D: The effects of active and passive humidification on ventilation-associated nosocomial pneumonia. Anaesthesist 2004, 53: 29-35. 10.1007/s00101-003-0607-7CrossRefPubMed
26.
go back to reference Cohen IL, Weinberg PF, Alan I, Rowinsky GS: Endotracheal tube occlusion associated with the use of heat and moisture exchangers in the intensive care unit. Crit Care Med 1988, 16: 277-279.CrossRefPubMed Cohen IL, Weinberg PF, Alan I, Rowinsky GS: Endotracheal tube occlusion associated with the use of heat and moisture exchangers in the intensive care unit. Crit Care Med 1988, 16: 277-279.CrossRefPubMed
27.
go back to reference Blin F, Fraïsse F, Chauveau P, Lahilaire P, Boulet E, Bouhaddi R, Sadourny F, Schlatter J, Trouillet G, Hermitte M, Kleinknecht D, the members of the Nosocomial Infections Research Group: Incidence of nosocomial pneumonias among 1788 ventilated patients in 6 ICU according to the type of humidification used. Intensive Care Med 1996,22(Suppl 3):S324. [abstract]CrossRef Blin F, Fraïsse F, Chauveau P, Lahilaire P, Boulet E, Bouhaddi R, Sadourny F, Schlatter J, Trouillet G, Hermitte M, Kleinknecht D, the members of the Nosocomial Infections Research Group: Incidence of nosocomial pneumonias among 1788 ventilated patients in 6 ICU according to the type of humidification used. Intensive Care Med 1996,22(Suppl 3):S324. [abstract]CrossRef
28.
go back to reference Torres A, Carlet J, the European Task Force on Ventilator Associated Pneumonia: Ventilator associated pneumonia. Eur Respir J 2001, 17: 1034-1045. 10.1183/09031936.01.17510340CrossRefPubMed Torres A, Carlet J, the European Task Force on Ventilator Associated Pneumonia: Ventilator associated pneumonia. Eur Respir J 2001, 17: 1034-1045. 10.1183/09031936.01.17510340CrossRefPubMed
29.
go back to reference American Thoracic Society and Infectious Diseases Society of America: Guidelines for the management of adults with hospital-acquired, ventilator associated, and healthcare associated pneumonia. Am J Respir Crit Care Med 2005, 171: 388-416. 10.1164/rccm.200405-644STCrossRef American Thoracic Society and Infectious Diseases Society of America: Guidelines for the management of adults with hospital-acquired, ventilator associated, and healthcare associated pneumonia. Am J Respir Crit Care Med 2005, 171: 388-416. 10.1164/rccm.200405-644STCrossRef
30.
go back to reference Tablan OC, Anderson LJ, Besser R, Bridges C, Hajjeh R, CDC and Healthcare Infection Control Practices Advisory Committee: Guidelines for preventing health-care – associated pneumonia, 2003: recommendations of CDC and the Healthcare Infection Control Practices Advisory Committee. MMWR Recomm Rep 2004, 53: 1-36.PubMed Tablan OC, Anderson LJ, Besser R, Bridges C, Hajjeh R, CDC and Healthcare Infection Control Practices Advisory Committee: Guidelines for preventing health-care – associated pneumonia, 2003: recommendations of CDC and the Healthcare Infection Control Practices Advisory Committee. MMWR Recomm Rep 2004, 53: 1-36.PubMed
31.
go back to reference Dodek P, Keenan S, Cook D, Heyland D, Jacka M, Hand L, Muscedere J, Foster D, Mehta N, Hall R, Brun-Buisson C, the Canadian Critical Care Trials Group, Canadian Critical Care Society: Evidence based clinical practice guideline for the prevention of ventilator associated pneumonia. Ann Intern Med 2004, 141: 305-313.CrossRefPubMed Dodek P, Keenan S, Cook D, Heyland D, Jacka M, Hand L, Muscedere J, Foster D, Mehta N, Hall R, Brun-Buisson C, the Canadian Critical Care Trials Group, Canadian Critical Care Society: Evidence based clinical practice guideline for the prevention of ventilator associated pneumonia. Ann Intern Med 2004, 141: 305-313.CrossRefPubMed
32.
go back to reference Hess DR, Kallstrom TJ, Mottram CD, Myers TR, Sorenson HM, Vines DL: American Association for Respiratory Care. Care of the ventilator circuit and its relation to Guidelines for ventilator associated pneumonia. Respir Care 2003, 48: 869-879.PubMed Hess DR, Kallstrom TJ, Mottram CD, Myers TR, Sorenson HM, Vines DL: American Association for Respiratory Care. Care of the ventilator circuit and its relation to Guidelines for ventilator associated pneumonia. Respir Care 2003, 48: 869-879.PubMed
33.
go back to reference Lorente L, Lecuona M, Jiménez A, Mora ML, Sierra A: Tracheal suction by closed system without daily change versus open system. Intensive Care Med 2006, 32: 538-544. 10.1007/s00134-005-0057-6CrossRefPubMed Lorente L, Lecuona M, Jiménez A, Mora ML, Sierra A: Tracheal suction by closed system without daily change versus open system. Intensive Care Med 2006, 32: 538-544. 10.1007/s00134-005-0057-6CrossRefPubMed
34.
go back to reference Cook D, De Jonghe B, Brochard L, Brun-Buisson C: Influence of airway management on ventilator-associated pneumonia: evidence from randomized trials. JAMA 1998, 279: 781-787. 10.1001/jama.279.10.781CrossRefPubMed Cook D, De Jonghe B, Brochard L, Brun-Buisson C: Influence of airway management on ventilator-associated pneumonia: evidence from randomized trials. JAMA 1998, 279: 781-787. 10.1001/jama.279.10.781CrossRefPubMed
35.
go back to reference Kola A, Eckmanns T, Gastmeier P: Efficacy of heat and moisture exchangers in preventing ventilator-associated pneumonia: meta-analysis of randomized controlled trials. Intensive Care Med 2005, 31: 5-11. 10.1007/s00134-004-2431-1CrossRefPubMed Kola A, Eckmanns T, Gastmeier P: Efficacy of heat and moisture exchangers in preventing ventilator-associated pneumonia: meta-analysis of randomized controlled trials. Intensive Care Med 2005, 31: 5-11. 10.1007/s00134-004-2431-1CrossRefPubMed
36.
go back to reference Garibaldi RA, Britt MR, Webster C, Pace NL: Failure of bacterial filters to reduce the incidence of pneumonia after inhalation anesthesia. Anesthesiology 1981, 54: 364-368.CrossRefPubMed Garibaldi RA, Britt MR, Webster C, Pace NL: Failure of bacterial filters to reduce the incidence of pneumonia after inhalation anesthesia. Anesthesiology 1981, 54: 364-368.CrossRefPubMed
37.
go back to reference Feeley TW, Hamilton WK, Xavier B, Moyers J, Eger EI: Sterile anesthesia breathing circuits do not prevent postoperative pulmonary infection. Anesthesiology 1981, 54: 369-372.CrossRefPubMed Feeley TW, Hamilton WK, Xavier B, Moyers J, Eger EI: Sterile anesthesia breathing circuits do not prevent postoperative pulmonary infection. Anesthesiology 1981, 54: 369-372.CrossRefPubMed
38.
go back to reference Lorente L, Lecuona M, Málaga J, Revert C, Mora ML, Sierra A: Bacterial filters in respiratory circuits: an unnecessary cost? Crit Care Med 2003, 31: 2126-2130. 10.1097/01.CCM.0000069733.24843.07CrossRefPubMed Lorente L, Lecuona M, Málaga J, Revert C, Mora ML, Sierra A: Bacterial filters in respiratory circuits: an unnecessary cost? Crit Care Med 2003, 31: 2126-2130. 10.1097/01.CCM.0000069733.24843.07CrossRefPubMed
39.
go back to reference Hess DR: And now for the rest of the store. Respir Care 2002, 47: 696-699. Hess DR: And now for the rest of the store. Respir Care 2002, 47: 696-699.
40.
go back to reference Luchetti M, Stuani A, Castelli G, Marraro G: Comparison of three different humidification systems during prolonged mechanical ventilation. Minerva Anestesiol 1998, 64: 75-81.PubMed Luchetti M, Stuani A, Castelli G, Marraro G: Comparison of three different humidification systems during prolonged mechanical ventilation. Minerva Anestesiol 1998, 64: 75-81.PubMed
41.
go back to reference Hurni JM, Feihl F, Lazor R, Leuenberger P, Perret C: Safety of combined heat and moisture exchanger filters in long-term mechanical ventilation. Chest 1997, 111: 686-691.CrossRefPubMed Hurni JM, Feihl F, Lazor R, Leuenberger P, Perret C: Safety of combined heat and moisture exchanger filters in long-term mechanical ventilation. Chest 1997, 111: 686-691.CrossRefPubMed
Metadata
Title
Ventilator-associated pneumonia using a heated humidifier or a heat and moisture exchanger: a randomized controlled trial [ISRCTN88724583]
Authors
Leonardo Lorente
María Lecuona
Alejandro Jiménez
María L Mora
Antonio Sierra
Publication date
01-08-2006
Publisher
BioMed Central
Published in
Critical Care / Issue 4/2006
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc5009

Other articles of this Issue 4/2006

Critical Care 4/2006 Go to the issue