Improved lag screw positioning in the treatment of proximal femur fractures using a novel computer assisted surgery method: a cadaveric study

Authors: Matthias Regling, Arno Blau, Robert A Probe, James W Maxey, Brian D Solberg

Published in: BMC Musculoskeletal Disorders | Issue 1/2014

Login to get access

Abstract

Background

The importance of the tip-apex distance (TAD) to predict the cut-out risk of fixed angle hip implants has been widely discussed in the scientific literature. Intra-operative determination of TAD is difficult and can be hampered by image quality, body habitus, and image projection. The purpose of this paper is to evaluate, through a cadaveric study, a novel computer assisted surgery system (ADAPT), which is intended for intraoperative optimisation of lag screw positioning during antegrade femoral nailing. A 3D measure for optimal lag screw position, the tip-to-head-surface distance (TSD), is introduced.

Methods

45 intra-medullary hip screw procedures were performed by experienced and less experienced surgeons in a cadaveric test series: in 23 surgeries the ADAPT system was used, and in 22 it was not used. The position of the lag screw within the femoral head and neck was evaluated using post-operative CT scans. TAD, TSD, fluoroscopy as well as procedure time and variability were assessed.

Results

The use of the ADAPT system increased accuracy in TSD values (i.e. smaller variability around the target value) for both groups of surgeons (interquartile range (IQR) of experienced surgeons: 4.10 mm (Conventional) vs. 1.35 mm (ADAPT) (p = 0.004)/IQR of less experienced surgeons: 3.60 mm (Conventional) vs. 0.85 mm (ADAPT) (p = 0.002)). The accuracy gain in TAD values did not prove to be significant in the grouped analysis (p = 0.269 for experienced surgeons; p = 0.066 for less experienced surgeons); however, the overall analysis showed a significant increase in accuracy (IQR: 4.50 mm (Conventional) vs. 2.00 mm (ADAPT) (p = 0.042)). The fluoroscopy time was significantly decreased by the use of the ADAPT system with a median value of 29.00 seconds (Conventional) vs. 17.00 seconds (ADAPT) for the less experienced surgeons (p = 0.046). There was no statistically significant impact on the procedure time (p = 0.739).

Conclusions

The ADAPT system improved the position of the lag screw within the femoral head, regardless of the surgeon’s level of clinical experience, and at the same time decreased overall fluoroscopy usage. These positive effects are achieved without increasing procedure time.
Appendix
Available only for authorised users
Literature
1.
go back to reference Kukla C, Heinz T, Gaebler C, Heinze G, Vecsei V: The standard Gamma nail: a critical analysis of 1,000 cases. J Trauma. 2001, 51 (1): 77-83. 10.1097/00005373-200107000-00012.CrossRefPubMed Kukla C, Heinz T, Gaebler C, Heinze G, Vecsei V: The standard Gamma nail: a critical analysis of 1,000 cases. J Trauma. 2001, 51 (1): 77-83. 10.1097/00005373-200107000-00012.CrossRefPubMed
2.
go back to reference Ahrengart L, Tornkvist H, Fornander P, Thorngren KG, Pasanen L, Wahlstrom P, Honkonen S, Lindgren U: A randomized study of the compression hip screw and Gamma nail in 426 fractures. Clin Orthop. 2002, 401: 209-222.CrossRefPubMed Ahrengart L, Tornkvist H, Fornander P, Thorngren KG, Pasanen L, Wahlstrom P, Honkonen S, Lindgren U: A randomized study of the compression hip screw and Gamma nail in 426 fractures. Clin Orthop. 2002, 401: 209-222.CrossRefPubMed
3.
go back to reference Hesse B, Gachter A: Complications following the treatment of trochanteric fractures with the gamma nail. Arch Orthop Trauma Surg. 2004, 124: 692-698. 10.1007/s00402-004-0744-8.CrossRefPubMed Hesse B, Gachter A: Complications following the treatment of trochanteric fractures with the gamma nail. Arch Orthop Trauma Surg. 2004, 124: 692-698. 10.1007/s00402-004-0744-8.CrossRefPubMed
4.
go back to reference Schipper IB, Steyerberg EW, Castelein RM, van der Heijden FH, den Hoed PT, Kerver AJ, van Vugt AB: Treatment of unstable trochanteric fractures. Randomised comparison of the gamma nail and the proximal femoral nail. J Bone Joint Surg (Br). 2004, 86 (1): 86-94. Schipper IB, Steyerberg EW, Castelein RM, van der Heijden FH, den Hoed PT, Kerver AJ, van Vugt AB: Treatment of unstable trochanteric fractures. Randomised comparison of the gamma nail and the proximal femoral nail. J Bone Joint Surg (Br). 2004, 86 (1): 86-94.
5.
go back to reference Chirodian N, Arch B, Parker MJ: Sliding hip screw fixation of trochanteric hip fractures: outcome of 1024 procedures. Injury. 2005, 36: 793-800. 10.1016/j.injury.2005.01.017.CrossRefPubMed Chirodian N, Arch B, Parker MJ: Sliding hip screw fixation of trochanteric hip fractures: outcome of 1024 procedures. Injury. 2005, 36: 793-800. 10.1016/j.injury.2005.01.017.CrossRefPubMed
6.
go back to reference Utrilla AL, Reig JS, Munoz FM, Tufanisco CB: Trochanteric gamma nail and compression hip screw for trochanteric fractures: a randomized, prospective, comparative study in 210 elderly patients with a new design of the gamma nail. J Orthop Trauma. 2005, 19 (4): 229-233. 10.1097/01.bot.0000151819.95075.ad.CrossRefPubMed Utrilla AL, Reig JS, Munoz FM, Tufanisco CB: Trochanteric gamma nail and compression hip screw for trochanteric fractures: a randomized, prospective, comparative study in 210 elderly patients with a new design of the gamma nail. J Orthop Trauma. 2005, 19 (4): 229-233. 10.1097/01.bot.0000151819.95075.ad.CrossRefPubMed
7.
go back to reference Lenich A, Mayr E, Ruter A, Mockl C, Fuchtmeier B: First results with the trochanter fixation nail (TFN): a report on 120 cases. Arch Orthop Trauma Surg. 2006, 126: 706-712. 10.1007/s00402-006-0117-6.CrossRefPubMed Lenich A, Mayr E, Ruter A, Mockl C, Fuchtmeier B: First results with the trochanter fixation nail (TFN): a report on 120 cases. Arch Orthop Trauma Surg. 2006, 126: 706-712. 10.1007/s00402-006-0117-6.CrossRefPubMed
8.
go back to reference Simmermacher RK, Ljungqvist J, Bail H, Hockertz T, Vochteloo AJ, Ochs U, Werken CV: The new proximal femoral nail antirotation (PFNA) in daily practice: results of a multicentre clinical study. Injury. 2008, 39 (8): 932-939. 10.1016/j.injury.2008.02.005.CrossRefPubMed Simmermacher RK, Ljungqvist J, Bail H, Hockertz T, Vochteloo AJ, Ochs U, Werken CV: The new proximal femoral nail antirotation (PFNA) in daily practice: results of a multicentre clinical study. Injury. 2008, 39 (8): 932-939. 10.1016/j.injury.2008.02.005.CrossRefPubMed
9.
go back to reference Walton M, Barnett A, Jackson M: Tip-apex distance as a predictor of failure following cephalo-medullary fixation for unstable fractures of the proximal femur. Eur J Trauma Emerg S. 2008, 34: 273-276. 10.1007/s00068-008-7075-3.CrossRef Walton M, Barnett A, Jackson M: Tip-apex distance as a predictor of failure following cephalo-medullary fixation for unstable fractures of the proximal femur. Eur J Trauma Emerg S. 2008, 34: 273-276. 10.1007/s00068-008-7075-3.CrossRef
10.
go back to reference Geller JA, Saifi C, Morrison TA, Macaulay W: Tip-apex distance of intramedullary devices as a predictor of cut-out failure in the treatment of peritrochanteric elderly hip fractures. Int Orthop. 2010, 34: 719-722. 10.1007/s00264-009-0837-7.CrossRefPubMed Geller JA, Saifi C, Morrison TA, Macaulay W: Tip-apex distance of intramedullary devices as a predictor of cut-out failure in the treatment of peritrochanteric elderly hip fractures. Int Orthop. 2010, 34: 719-722. 10.1007/s00264-009-0837-7.CrossRefPubMed
11.
go back to reference Mereddy P, Kamath S, Ramakrishnan M, Malik H, Donnachie N: The AO/ASIF proximal femoral nail antirotation (PFNA): a new design for the treatment of unstable proximal femoral fractures. Injury. 2009, 40 (4): 428-432. 10.1016/j.injury.2008.10.014.CrossRefPubMed Mereddy P, Kamath S, Ramakrishnan M, Malik H, Donnachie N: The AO/ASIF proximal femoral nail antirotation (PFNA): a new design for the treatment of unstable proximal femoral fractures. Injury. 2009, 40 (4): 428-432. 10.1016/j.injury.2008.10.014.CrossRefPubMed
12.
go back to reference Barton TM, Gleeson R, Topliss C, Greenwood R, Harries WJ, Chesser TJ: A comparison of the long gamma nail with the sliding hip screw for the treatment of AO/OTA 31-A2 fractures of the proximal part of the femur: a prospective randomized trial. J Bone Joint Surg Am. 2010, 92: 792-798.CrossRefPubMed Barton TM, Gleeson R, Topliss C, Greenwood R, Harries WJ, Chesser TJ: A comparison of the long gamma nail with the sliding hip screw for the treatment of AO/OTA 31-A2 fractures of the proximal part of the femur: a prospective randomized trial. J Bone Joint Surg Am. 2010, 92: 792-798.CrossRefPubMed
13.
go back to reference Bojan AJ, Beimel C, Speitling A, Taglang G, Ekholm C, Jönsson A: 3066 consecutive Gamma Nails. 12 years experience at a single centre. BMC Musculoskelet Disord. 2010, 26 (11): 133-CrossRef Bojan AJ, Beimel C, Speitling A, Taglang G, Ekholm C, Jönsson A: 3066 consecutive Gamma Nails. 12 years experience at a single centre. BMC Musculoskelet Disord. 2010, 26 (11): 133-CrossRef
14.
go back to reference Hsueh KK, Fang CK, Chen CM, Su YP, Wu HF, Chiu FY: Risk factors in cutout of sliding hip screw in intertrochanteric fractures: an evaluation of 937 patients. Int Orthop. 2010, 34: 1273-1276. 10.1007/s00264-009-0866-2.CrossRefPubMed Hsueh KK, Fang CK, Chen CM, Su YP, Wu HF, Chiu FY: Risk factors in cutout of sliding hip screw in intertrochanteric fractures: an evaluation of 937 patients. Int Orthop. 2010, 34: 1273-1276. 10.1007/s00264-009-0866-2.CrossRefPubMed
15.
go back to reference Lobo-Escolar A, Joven E, Iglesias D, Herrera A: Predictive factors for cutting-out in femoral intramedullary nailing. Injury. 2010, 41: 1312-1316. 10.1016/j.injury.2010.08.009.CrossRefPubMed Lobo-Escolar A, Joven E, Iglesias D, Herrera A: Predictive factors for cutting-out in femoral intramedullary nailing. Injury. 2010, 41: 1312-1316. 10.1016/j.injury.2010.08.009.CrossRefPubMed
16.
go back to reference Stern R, Lübbeke A, Suva D: Prospective randomized study comparing screw versus helical blade in the treatment of low-energy trochanteric fractures. Int Orthop. 2011, 35: 1855-1861. 10.1007/s00264-011-1232-8.CrossRefPubMedPubMedCentral Stern R, Lübbeke A, Suva D: Prospective randomized study comparing screw versus helical blade in the treatment of low-energy trochanteric fractures. Int Orthop. 2011, 35: 1855-1861. 10.1007/s00264-011-1232-8.CrossRefPubMedPubMedCentral
17.
go back to reference Herman A, Landau Y, Gutman G, Ougortsin V, Chechick A, Shazar N: Radiological evaluation of intertrochanteric fracture fixation by the proximal femoral nail. Injury. 2011, 43: 856-863.CrossRefPubMed Herman A, Landau Y, Gutman G, Ougortsin V, Chechick A, Shazar N: Radiological evaluation of intertrochanteric fracture fixation by the proximal femoral nail. Injury. 2011, 43: 856-863.CrossRefPubMed
18.
go back to reference Davis TRC, Sher JL, Horsman A, Simpson M, Porter BB, Checketts RG: Intertrochanteric femoral fractures: mechanical failure after internal fixation. J Bone Joint Surg (Br). 1990, 72-B: 26-31. Davis TRC, Sher JL, Horsman A, Simpson M, Porter BB, Checketts RG: Intertrochanteric femoral fractures: mechanical failure after internal fixation. J Bone Joint Surg (Br). 1990, 72-B: 26-31.
19.
go back to reference Bannister GC, Gibson AG, Ackroyd CE, Newman JH: The fixation and prognosis of trochanteric fractures. A randomized prospective controlled trial. Clin Orthop Relat Res. 1990, 254: 242-246.PubMed Bannister GC, Gibson AG, Ackroyd CE, Newman JH: The fixation and prognosis of trochanteric fractures. A randomized prospective controlled trial. Clin Orthop Relat Res. 1990, 254: 242-246.PubMed
20.
go back to reference Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77: 1058-1064.PubMed Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM: The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995, 77: 1058-1064.PubMed
21.
go back to reference Pervez H, Parker MJ, Vowler S: Prediction of fixation failure after sliding hip screw fixation. Injury. 2004, 35: 994-998. 10.1016/j.injury.2003.10.028.CrossRefPubMed Pervez H, Parker MJ, Vowler S: Prediction of fixation failure after sliding hip screw fixation. Injury. 2004, 35: 994-998. 10.1016/j.injury.2003.10.028.CrossRefPubMed
22.
go back to reference Kraus M, Krischak G, Wiedmann K, Riepl C, Gebhard F, Jöckel JA, Scola A: Clinical evaluation of PFNA and relationship between the tip-apex distance and mechanical failure. Unfallchirurg. 2011, 114 (6): 470-478. 10.1007/s00113-011-1975-0.CrossRefPubMed Kraus M, Krischak G, Wiedmann K, Riepl C, Gebhard F, Jöckel JA, Scola A: Clinical evaluation of PFNA and relationship between the tip-apex distance and mechanical failure. Unfallchirurg. 2011, 114 (6): 470-478. 10.1007/s00113-011-1975-0.CrossRefPubMed
23.
go back to reference Parker MJ: Cutting-out of the dynamic hip screw related to its position. J Bone Joint Surg (Br). 1992, 74-B: 625- Parker MJ: Cutting-out of the dynamic hip screw related to its position. J Bone Joint Surg (Br). 1992, 74-B: 625-
24.
go back to reference Den Hartog BD, Bartal E, Cooke F: Treatment of the unstable intertrochanteric fracture. Effect of the placement of the screw, its angle of insertion, and osteotomy. J Bone Joint Surg Am. 1991, 73 (5): 726-733.PubMed Den Hartog BD, Bartal E, Cooke F: Treatment of the unstable intertrochanteric fracture. Effect of the placement of the screw, its angle of insertion, and osteotomy. J Bone Joint Surg Am. 1991, 73 (5): 726-733.PubMed
25.
go back to reference Galanakis IA, Steriopoulos KA, Dretakis EK: Correct placement of the screw or nail in trochanteric fractures: Effect of the initial placement in the migration. Clin Orthop. 1995, 313: 206-213.PubMed Galanakis IA, Steriopoulos KA, Dretakis EK: Correct placement of the screw or nail in trochanteric fractures: Effect of the initial placement in the migration. Clin Orthop. 1995, 313: 206-213.PubMed
26.
go back to reference Kawaguchi S, Sawada K, Nabata Y: Cutting-out of the lag screw after internal fixation with the Asiatic gamma nail. Injury. 1998, 29 (1): 47-53. 10.1016/S0020-1383(97)00158-7.CrossRefPubMed Kawaguchi S, Sawada K, Nabata Y: Cutting-out of the lag screw after internal fixation with the Asiatic gamma nail. Injury. 1998, 29 (1): 47-53. 10.1016/S0020-1383(97)00158-7.CrossRefPubMed
27.
go back to reference Mainds CC, Newman RJ: Implant failures in patients with proximal fractures of the femur treated with a sliding screw device. Injury. 1989, 20 (2): 98-100. 10.1016/0020-1383(89)90151-4.CrossRefPubMed Mainds CC, Newman RJ: Implant failures in patients with proximal fractures of the femur treated with a sliding screw device. Injury. 1989, 20 (2): 98-100. 10.1016/0020-1383(89)90151-4.CrossRefPubMed
28.
go back to reference Larsson S, Friberg S, Hansson LI: Trochanteric fractures: influence of reduction and implant position on impaction and complications. Clin Orthop. 1990, 259: 130-139.PubMed Larsson S, Friberg S, Hansson LI: Trochanteric fractures: influence of reduction and implant position on impaction and complications. Clin Orthop. 1990, 259: 130-139.PubMed
29.
go back to reference Leung KS, So WS, Shen WY, Hui PW: Gamma nails and dynamic hip screws for peritrochanteric fractures. A randomized prospective study in elderly patients. J Bone Joint Surg (Br). 1992, 74 (3): 345-351. Leung KS, So WS, Shen WY, Hui PW: Gamma nails and dynamic hip screws for peritrochanteric fractures. A randomized prospective study in elderly patients. J Bone Joint Surg (Br). 1992, 74 (3): 345-351.
30.
go back to reference Wu CC, Shih CH, Lee MY, Tai CL: Biomechanical analysis of location of lag screw of a dynamic hip screw in treatment of unstable intertrochanteric fracture. J Trauma. 1996, 41 (4): 699-702. 10.1097/00005373-199610000-00017.CrossRefPubMed Wu CC, Shih CH, Lee MY, Tai CL: Biomechanical analysis of location of lag screw of a dynamic hip screw in treatment of unstable intertrochanteric fracture. J Trauma. 1996, 41 (4): 699-702. 10.1097/00005373-199610000-00017.CrossRefPubMed
31.
go back to reference Güven M, Yavuz U, Kadioglu B, Akman B, Kilincoglu V, Ünay K, Altintas F: Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw. Orthop Traumatol Surg Res. 2010, 96: 21-27. 10.1016/j.otsr.2009.10.008.CrossRefPubMed Güven M, Yavuz U, Kadioglu B, Akman B, Kilincoglu V, Ünay K, Altintas F: Importance of screw position in intertrochanteric femoral fractures treated by dynamic hip screw. Orthop Traumatol Surg Res. 2010, 96: 21-27. 10.1016/j.otsr.2009.10.008.CrossRefPubMed
32.
go back to reference De Bruijn K, den Hartog D, Tuinebreijer W, Roukema G: Reliability of predictors for screw cutout in intertrochanteric hip fractures. J Bone Joint Surg Am. 2012, 94: 1266-1272. 10.2106/JBJS.K.00357.CrossRefPubMed De Bruijn K, den Hartog D, Tuinebreijer W, Roukema G: Reliability of predictors for screw cutout in intertrochanteric hip fractures. J Bone Joint Surg Am. 2012, 94: 1266-1272. 10.2106/JBJS.K.00357.CrossRefPubMed
33.
go back to reference Kuzyk PRT, Zdero R, Shah S, Olsen M, Waddell JP, Schemitsch EH: Femoral head lag screw position for Cephalomedullary nails: a biomechanical analysis. J Orthop Trauma. 2012, 26 (7): 414-421. 10.1097/BOT.0b013e318229acca.CrossRefPubMed Kuzyk PRT, Zdero R, Shah S, Olsen M, Waddell JP, Schemitsch EH: Femoral head lag screw position for Cephalomedullary nails: a biomechanical analysis. J Orthop Trauma. 2012, 26 (7): 414-421. 10.1097/BOT.0b013e318229acca.CrossRefPubMed
34.
go back to reference Davies J, Mokawem M, Guy S: Tip Apex Distance: a useful method to assess extra-capsular neck of femur fixation?. J Trauma Treatment. 2012, 1: 119- Davies J, Mokawem M, Guy S: Tip Apex Distance: a useful method to assess extra-capsular neck of femur fixation?. J Trauma Treatment. 2012, 1: 119-
35.
go back to reference Liebergall M, Ben-David D, Weil Y, Peyser A, Mosheiff R: Computerized navigation for the internal fixation of femoral neck fractures. J Bone Joint Surg Am. 2006, 88: 1748-1754. 10.2106/JBJS.E.00137.CrossRefPubMed Liebergall M, Ben-David D, Weil Y, Peyser A, Mosheiff R: Computerized navigation for the internal fixation of femoral neck fractures. J Bone Joint Surg Am. 2006, 88: 1748-1754. 10.2106/JBJS.E.00137.CrossRefPubMed
36.
go back to reference Chong KW, Wong MK, Rikhraj IS, Howe TS: The use of computer navigation in performing minimally invasive surgery for intertrochanteric hip fractures - the experience in Singapore. Injury. 2006, 37: 755-762. 10.1016/j.injury.2006.03.002.CrossRefPubMed Chong KW, Wong MK, Rikhraj IS, Howe TS: The use of computer navigation in performing minimally invasive surgery for intertrochanteric hip fractures - the experience in Singapore. Injury. 2006, 37: 755-762. 10.1016/j.injury.2006.03.002.CrossRefPubMed
37.
go back to reference Atesok K, Schemitsch EH: Computer-assisted trauma surgery. J Am Acad Orthop Surg. 2010, 18 (5): 247-258.CrossRefPubMed Atesok K, Schemitsch EH: Computer-assisted trauma surgery. J Am Acad Orthop Surg. 2010, 18 (5): 247-258.CrossRefPubMed
38.
go back to reference Johnson LJ, Cope MR, Shahrokhi S, Tamblyn P: Measuring tip-apex distance using a picture archiving and communication system (PACS). Injury. 2008, 39: 786-790. 10.1016/j.injury.2007.12.019.CrossRefPubMed Johnson LJ, Cope MR, Shahrokhi S, Tamblyn P: Measuring tip-apex distance using a picture archiving and communication system (PACS). Injury. 2008, 39: 786-790. 10.1016/j.injury.2007.12.019.CrossRefPubMed
39.
go back to reference Bojan A, Beimel C, Taglang G, Collin D, Ekholm C, Jönsson A: Critical factors in cut-out complication after gamma nail treatment of proximal femoral fractures. BMC Musculoskelet Disord. 2013, 14: 1-10.1186/1471-2474-14-1.CrossRefPubMedPubMedCentral Bojan A, Beimel C, Taglang G, Collin D, Ekholm C, Jönsson A: Critical factors in cut-out complication after gamma nail treatment of proximal femoral fractures. BMC Musculoskelet Disord. 2013, 14: 1-10.1186/1471-2474-14-1.CrossRefPubMedPubMedCentral
40.
go back to reference Baumgaertner MR, Solberg BD: Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg (Br). 1997, 79: 969-971. 10.1302/0301-620X.79B6.7949.CrossRef Baumgaertner MR, Solberg BD: Awareness of tip-apex distance reduces failure of fixation of trochanteric fractures of the hip. J Bone Joint Surg (Br). 1997, 79: 969-971. 10.1302/0301-620X.79B6.7949.CrossRef
Metadata
Title
Improved lag screw positioning in the treatment of proximal femur fractures using a novel computer assisted surgery method: a cadaveric study
Authors
Matthias Regling
Arno Blau
Robert A Probe
James W Maxey
Brian D Solberg
Publication date
01-12-2014
Publisher
BioMed Central
Published in
BMC Musculoskeletal Disorders / Issue 1/2014
Electronic ISSN: 1471-2474
DOI
https://doi.org/10.1186/1471-2474-15-189

Other articles of this Issue 1/2014

BMC Musculoskeletal Disorders 1/2014 Go to the issue