Published in: BMC Cardiovascular Disorders 1/2014

Open Access 01-12-2014 | Research article

Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function

Authors: Jonathan R Linder, Harald M Stauss, Holly Gindes, Gary L Pierce, Nicholas H Von Bergen, William G Haynes, Jess G Fiedorowicz

Published in: BMC Cardiovascular Disorders | Issue 1/2014

Login to get access

Abstract

Background

We sought to determine whether heart rate variability (HRV), blood pressure (BP) variability, and baroreceptor-heart rate reflex sensitivity can be reliably assessed using finger volume pulse waveforms obtained from the commercially available EndoPAT device.

Methods

Non-invasive BP (Finometer Pro as a non-invasive standard) and finger volume (EndoPAT) waveforms were recorded in 65 adults (37 ± 14 years; 60% female) and systolic BP and heart rate (HR) time series were derived after calibrating the EndoPAT signal based on systolic and diastolic BP values obtained by a sphygomomanometer. Transfer function analyses were performed to test for coherence between systolic BP and HR time series derived from the Finometer and EndoPAT devices. Time-domain HRV parameters, frequency domain HR and systolic BP variability parameters, and baroreflex sensitivity (sequence technique) were computed from Finometer- and EndoPAT-derived time series and intraclass correlation coefficients (ICC) were calculated.

Results

Squared coherence between systolic BP time series derived from the Finometer and EndoPAT devices was low, suggesting poor correlation. In contrast, squared coherence between HR time series derived from the two devices was excellent [High Frequency (HF) = 0.80, Low Frequency (LF) = 0.81], with gain values close to 1.0. ICC values for time- and frequency-domain HRV parameters were excellent (>0.9 except for relative HF HRV, which was 0.77), while ICC values for frequency-domain BP variability parameters and baroreceptor-HR reflex sensitivity were low.

Conclusions

Finger volume pulse waveforms can be used to reliably assess both time-domain and frequency-domain HR variability. However, frequency domain BP variability parameters cannot be reliably assessed from finger volume pulse waveforms using the simple calibration technique used in this study.
Appendix
Available only for authorised users
Literature
1.
go back to reference Malik M, Bigger JT, Camm AJ, Kleiger RE, Malliani A, Moss AJ, Schwartz PJ: Heart rate variability: Standards of measurement, physiological interpretation, and clinical use. Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Eur Heart J. 1996, 17 (3): 354-381. 10.1093/oxfordjournals.eurheartj.a014868.CrossRef Malik M, Bigger JT, Camm AJ, Kleiger RE, Malliani A, Moss AJ, Schwartz PJ: Heart rate variability: Standards of measurement, physiological interpretation, and clinical use. Task Force of the European Society of Cardiology and the North American Society of Pacing and Electrophysiology. Eur Heart J. 1996, 17 (3): 354-381. 10.1093/oxfordjournals.eurheartj.a014868.CrossRef
2.
go back to reference Cavalcanti S, Severi S, Chiari L, Avanzolini G, Enzmann G, Bianco G, Panzetta G: Autonomic nervous function during haemodialysis assessed by spectral analysis of heart-rate variability. Clin Sci (Lond). 1997, 92 (4): 351-359.CrossRef Cavalcanti S, Severi S, Chiari L, Avanzolini G, Enzmann G, Bianco G, Panzetta G: Autonomic nervous function during haemodialysis assessed by spectral analysis of heart-rate variability. Clin Sci (Lond). 1997, 92 (4): 351-359.CrossRef
3.
go back to reference Ewing DJ, Borsey DQ, Bellavere F, Clarke BF: Cardiac autonomic neuropathy in diabetes: comparison of measures of R-R interval variation. Diabetologia. 1981, 21 (1): 18-24.CrossRefPubMed Ewing DJ, Borsey DQ, Bellavere F, Clarke BF: Cardiac autonomic neuropathy in diabetes: comparison of measures of R-R interval variation. Diabetologia. 1981, 21 (1): 18-24.CrossRefPubMed
4.
go back to reference Stauss HM: Identification of blood pressure control mechanisms by power spectral analysis. Clin Exp Pharmacol Physiol. 2007, 34 (4): 362-368. 10.1111/j.1440-1681.2007.04588.x.CrossRefPubMed Stauss HM: Identification of blood pressure control mechanisms by power spectral analysis. Clin Exp Pharmacol Physiol. 2007, 34 (4): 362-368. 10.1111/j.1440-1681.2007.04588.x.CrossRefPubMed
5.
go back to reference Pagani M, Lombardi F, Guzzetti S, Rimoldi O, Furlan R, Pizzinelli P, Sandrone G, Malfatto G, Dell'Orto S, Piccaluga E, Turiel M, Baselli G, Cerutti S, Malliani A: Power spectral analysis of heart rate and arterial pressure variabilities as a marker of sympatho-vagal interaction in man and conscious dog. Circ Res. 1986, 59 (2): 178-193. 10.1161/01.RES.59.2.178.CrossRefPubMed Pagani M, Lombardi F, Guzzetti S, Rimoldi O, Furlan R, Pizzinelli P, Sandrone G, Malfatto G, Dell'Orto S, Piccaluga E, Turiel M, Baselli G, Cerutti S, Malliani A: Power spectral analysis of heart rate and arterial pressure variabilities as a marker of sympatho-vagal interaction in man and conscious dog. Circ Res. 1986, 59 (2): 178-193. 10.1161/01.RES.59.2.178.CrossRefPubMed
6.
go back to reference Sandercock GR, Bromley P, Brodie DA: Reliability of three commercially available heart rate variability instruments using short-term (5-min) recordings. Clin Physiol Funct Imaging. 2004, 24 (6): 359-367. 10.1111/j.1475-097X.2004.00584.x.CrossRefPubMed Sandercock GR, Bromley P, Brodie DA: Reliability of three commercially available heart rate variability instruments using short-term (5-min) recordings. Clin Physiol Funct Imaging. 2004, 24 (6): 359-367. 10.1111/j.1475-097X.2004.00584.x.CrossRefPubMed
7.
go back to reference Latalova K, Prasko J, Diveky T, Grambal A, Kamaradova D, Velartova H, Salinger J, Opavsky J: Autonomic nervous system in euthymic patients with bipolar affective disorder. Neuro Endocrinol Lett. 2010, 31 (6): 829-836.PubMed Latalova K, Prasko J, Diveky T, Grambal A, Kamaradova D, Velartova H, Salinger J, Opavsky J: Autonomic nervous system in euthymic patients with bipolar affective disorder. Neuro Endocrinol Lett. 2010, 31 (6): 829-836.PubMed
8.
go back to reference Parati G, Saul JP, Di Rienzo M, Mancia G: Spectral analysis of blood pressure and heart rate variability in evaluating cardiovascular regulation. A critical appraisal. Hypertens. 1995, 25 (6): 1276-1286. 10.1161/01.HYP.25.6.1276.CrossRef Parati G, Saul JP, Di Rienzo M, Mancia G: Spectral analysis of blood pressure and heart rate variability in evaluating cardiovascular regulation. A critical appraisal. Hypertens. 1995, 25 (6): 1276-1286. 10.1161/01.HYP.25.6.1276.CrossRef
9.
go back to reference Parati G: Arterial baroreflex control of heart rate: determining factors and methods to assess its spontaneous modulation. J Physiol. 2005, 565 (Pt 3): 706-707.CrossRefPubMedPubMedCentral Parati G: Arterial baroreflex control of heart rate: determining factors and methods to assess its spontaneous modulation. J Physiol. 2005, 565 (Pt 3): 706-707.CrossRefPubMedPubMedCentral
10.
go back to reference Parati G, Di Rienzo M, Mancia G: How to measure baroreflex sensitivity: from the cardiovascular laboratory to daily life. J Hypertens. 2000, 18 (1): 7-19. 10.1097/00004872-200018010-00003.CrossRefPubMed Parati G, Di Rienzo M, Mancia G: How to measure baroreflex sensitivity: from the cardiovascular laboratory to daily life. J Hypertens. 2000, 18 (1): 7-19. 10.1097/00004872-200018010-00003.CrossRefPubMed
11.
go back to reference Ponchon P, Elghozi JL: Contribution of the renin-angiotensin and kallikrein-kinin systems to short-term variability of blood pressure in two-kidney, one-clip hypertensive rats. Eur J Pharmacol. 1996, 297 (1–2): 61-70.CrossRefPubMed Ponchon P, Elghozi JL: Contribution of the renin-angiotensin and kallikrein-kinin systems to short-term variability of blood pressure in two-kidney, one-clip hypertensive rats. Eur J Pharmacol. 1996, 297 (1–2): 61-70.CrossRefPubMed
12.
go back to reference Hamburg NM, Keyes MJ, Larson MG, Vasan RS, Schnabel R, Pryde MM, Mitchell GF, Sheffy J, Vita JA, Benjamin EJ: Cross-sectional relations of digital vascular function to cardiovascular risk factors in the Framingham Heart Study. Circulation. 2008, 117 (19): 2467-2474. 10.1161/CIRCULATIONAHA.107.748574.CrossRefPubMedPubMedCentral Hamburg NM, Keyes MJ, Larson MG, Vasan RS, Schnabel R, Pryde MM, Mitchell GF, Sheffy J, Vita JA, Benjamin EJ: Cross-sectional relations of digital vascular function to cardiovascular risk factors in the Framingham Heart Study. Circulation. 2008, 117 (19): 2467-2474. 10.1161/CIRCULATIONAHA.107.748574.CrossRefPubMedPubMedCentral
13.
go back to reference Chin KY, Panerai RB: Comparative study of Finapres devices. Blood Press Monit. 2012, 17 (4): 171-178. 10.1097/MBP.0b013e328356e1b3.CrossRefPubMed Chin KY, Panerai RB: Comparative study of Finapres devices. Blood Press Monit. 2012, 17 (4): 171-178. 10.1097/MBP.0b013e328356e1b3.CrossRefPubMed
14.
go back to reference Pinna GD, Maestri R: New criteria for estimating baroreflex sensitivity using the transfer function method. Med Biol Eng Comput. 2002, 40 (1): 79-84. 10.1007/BF02347699.CrossRefPubMed Pinna GD, Maestri R: New criteria for estimating baroreflex sensitivity using the transfer function method. Med Biol Eng Comput. 2002, 40 (1): 79-84. 10.1007/BF02347699.CrossRefPubMed
15.
go back to reference Guelen I, Westerhof BE, Van Der Sar GL, Van Montfrans GA, Kiemeneij F, Wesseling KH, Bos WJ: Finometer, finger pressure measurements with the possibility to reconstruct brachial pressure. Blood Press Monit. 2003, 8 (1): 27-30. 10.1097/00126097-200302000-00006.CrossRefPubMed Guelen I, Westerhof BE, Van Der Sar GL, Van Montfrans GA, Kiemeneij F, Wesseling KH, Bos WJ: Finometer, finger pressure measurements with the possibility to reconstruct brachial pressure. Blood Press Monit. 2003, 8 (1): 27-30. 10.1097/00126097-200302000-00006.CrossRefPubMed
16.
go back to reference Guelen I, Westerhof BE, van der Sar GL, van Montfrans GA, Kiemeneij F, Wesseling KH, Bos WJ: Validation of brachial artery pressure reconstruction from finger arterial pressure. J Hypertens. 2008, 26 (7): 1321-1327. 10.1097/HJH.0b013e3282fe1d28.CrossRefPubMed Guelen I, Westerhof BE, van der Sar GL, van Montfrans GA, Kiemeneij F, Wesseling KH, Bos WJ: Validation of brachial artery pressure reconstruction from finger arterial pressure. J Hypertens. 2008, 26 (7): 1321-1327. 10.1097/HJH.0b013e3282fe1d28.CrossRefPubMed
17.
go back to reference Schutte AE, Huisman HW, van Rooyen JM, Malan NT, Schutte R: Validation of the Finometer device for measurement of blood pressure in black women. J Hum Hypertens. 2004, 18 (2): 79-84. 10.1038/sj.jhh.1001639.CrossRefPubMed Schutte AE, Huisman HW, van Rooyen JM, Malan NT, Schutte R: Validation of the Finometer device for measurement of blood pressure in black women. J Hum Hypertens. 2004, 18 (2): 79-84. 10.1038/sj.jhh.1001639.CrossRefPubMed
18.
go back to reference Bush K, Kivlahan DR, McDonell MB, Fihn SD, Bradley KA: The AUDIT alcohol consumption questions (AUDIT-C): an effective brief screening test for problem drinking. Ambulatory Care Quality Improvement Project (ACQUIP). Alcohol Use Disorders Identification Test. Arch Intern Med. 1998, 158 (16): 1789-1795. 10.1001/archinte.158.16.1789.CrossRefPubMed Bush K, Kivlahan DR, McDonell MB, Fihn SD, Bradley KA: The AUDIT alcohol consumption questions (AUDIT-C): an effective brief screening test for problem drinking. Ambulatory Care Quality Improvement Project (ACQUIP). Alcohol Use Disorders Identification Test. Arch Intern Med. 1998, 158 (16): 1789-1795. 10.1001/archinte.158.16.1789.CrossRefPubMed
19.
go back to reference La Fountaine MF, Wecht JM, Spungen AM, Bauman WA: Intra-inter visit reproducibility of short-term linear and nonlinear measurement of heart rate variability in tetraplegia and neurologically intact controls. Physiol Meas. 2010, 31 (3): 363-374. 10.1088/0967-3334/31/3/006.CrossRefPubMed La Fountaine MF, Wecht JM, Spungen AM, Bauman WA: Intra-inter visit reproducibility of short-term linear and nonlinear measurement of heart rate variability in tetraplegia and neurologically intact controls. Physiol Meas. 2010, 31 (3): 363-374. 10.1088/0967-3334/31/3/006.CrossRefPubMed
20.
go back to reference Rothwell PM, Howard SC, Dolan E, O'Brien E, Dobson JE, Dahlof B, Sever PS, Poulter NR: Prognostic significance of visit-to-visit variability, maximum systolic blood pressure, and episodic hypertension. Lancet. 2010, 375 (9718): 895-905. 10.1016/S0140-6736(10)60308-X.CrossRefPubMed Rothwell PM, Howard SC, Dolan E, O'Brien E, Dobson JE, Dahlof B, Sever PS, Poulter NR: Prognostic significance of visit-to-visit variability, maximum systolic blood pressure, and episodic hypertension. Lancet. 2010, 375 (9718): 895-905. 10.1016/S0140-6736(10)60308-X.CrossRefPubMed
Metadata
Title
Finger volume pulse waveforms facilitate reliable assessment of heart rate variability, but not blood pressure variability or baroreflex function
Authors
Jonathan R Linder
Harald M Stauss
Holly Gindes
Gary L Pierce
Nicholas H Von Bergen
William G Haynes
Jess G Fiedorowicz
Publication date
01-12-2014
Publisher
BioMed Central
Published in
BMC Cardiovascular Disorders / Issue 1/2014
Electronic ISSN: 1471-2261
DOI
https://doi.org/10.1186/1471-2261-14-180

Other articles of this Issue 1/2014

BMC Cardiovascular Disorders 1/2014 Go to the issue