New Imaging Protocols for New Single Photon Emission CT Technologies

Authors: Piotr J. Slomka, Daniel S. Berman, Guido Germano

Published in: Current Cardiovascular Imaging Reports | Issue 3/2010

Login to get access

Abstract

Nuclear cardiology practitioners have several new technologies available with which to perform myocardial perfusion single photon emission CT (MPS). These include dedicated small-footprint cardiac scanners, new stationary or semi-stationary three-dimensional detectors, and advanced software algorithms for optimal image reconstruction. These new technologies have been employed to reduce imaging time and radiation exposure. They require less technologist and camera time and offer improved patient comfort. They have potential for the overall cost reduction of MPS and at the same time for improved accuracy by increased resolution, or accurate attenuation correction. Furthermore, these new technologies offer potential for new protocols such as simultaneous dual isotope, new combinations of isotopes, stress only MPS, or dynamic first-pass imaging. In addition, new imaging technologies in coronary CT angiography (CCTA) allow novel hybrid stress only MPS/CCTA protocols with reduced radiation burden. Additional developments further improving efficiency and diagnostic accuracy of MPS are on the horizon.
Literature
1.
go back to reference Dilsizian V, Narula J: Atlas of Nuclear Cardiology, edn 3. Springer; Philadelphia, PA: 2009. Dilsizian V, Narula J: Atlas of Nuclear Cardiology, edn 3. Springer; Philadelphia, PA: 2009.
2.
go back to reference • Slomka PJ, Patton JA, Berman DS, Germano G: Advances in technical aspects of myocardial perfusion SPECT imaging. J Nucl Cardiol 2009, 16:255–276. This is a comprehensive review of the technical aspects of the new systems and software.CrossRefPubMed • Slomka PJ, Patton JA, Berman DS, Germano G: Advances in technical aspects of myocardial perfusion SPECT imaging. J Nucl Cardiol 2009, 16:255–276. This is a comprehensive review of the technical aspects of the new systems and software.CrossRefPubMed
3.
go back to reference Lewin HC, Hyun MC: A clinical comparison of an upright triple-head digital detector system to a standard supine dual-head gamma camera [abstract]. J Nucl Cardiol 2005, 12:113.CrossRef Lewin HC, Hyun MC: A clinical comparison of an upright triple-head digital detector system to a standard supine dual-head gamma camera [abstract]. J Nucl Cardiol 2005, 12:113.CrossRef
5.
go back to reference Jaszczak RJ, Li J, Wang H, et al.: Pinhole collimation for ultra-high-resolution, small-field-of-view SPECT. Phys Med Biol 1994, 39:425–437.CrossRefPubMed Jaszczak RJ, Li J, Wang H, et al.: Pinhole collimation for ultra-high-resolution, small-field-of-view SPECT. Phys Med Biol 1994, 39:425–437.CrossRefPubMed
6.
go back to reference Schramm NU, Ebel G, Engeland U, et al.: High-resolution SPECT using multipinhole collimation. IEEE Transactions on Nuclear Science 2003, 50:315–320.CrossRef Schramm NU, Ebel G, Engeland U, et al.: High-resolution SPECT using multipinhole collimation. IEEE Transactions on Nuclear Science 2003, 50:315–320.CrossRef
7.
go back to reference Beekman FJ, Vastenhouw B: Design and simulation of a high-resolution stationary SPECT system for small animals. Phys Med Biol 2004, 49:4579–4592.CrossRefPubMed Beekman FJ, Vastenhouw B: Design and simulation of a high-resolution stationary SPECT system for small animals. Phys Med Biol 2004, 49:4579–4592.CrossRefPubMed
8.
go back to reference Volokh L, Hugg J, Blevis I, et al.: Effect of detector energy response on image quality of myocardial perfusion SPECT. Paper presented at IEEE Nuclear Science Symposium and Medical Imaging Conference. Dresden, Germany; October 19–26, 2008. Volokh L, Hugg J, Blevis I, et al.: Effect of detector energy response on image quality of myocardial perfusion SPECT. Paper presented at IEEE Nuclear Science Symposium and Medical Imaging Conference. Dresden, Germany; October 19–26, 2008.
9.
go back to reference Vija H, Chapman J, Ray M: IQ•SPECT technology White Paper. Siemens Medical Solutions USA Molecular Imaging; 2008:1–7. Vija H, Chapman J, Ray M: IQ•SPECT technology White Paper. Siemens Medical Solutions USA Molecular Imaging; 2008:1–7.
10.
go back to reference Hudson HM, Larkin RS: Accelerated image reconstruction using ordered subsets of projection data. IEEE Transactions on Medical Imaging 1994, 13:601–609.CrossRefPubMed Hudson HM, Larkin RS: Accelerated image reconstruction using ordered subsets of projection data. IEEE Transactions on Medical Imaging 1994, 13:601–609.CrossRefPubMed
12.
go back to reference Borges-Neto S, Pagnanelli RA, Shaw LK, et al.: Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies. J Nucl Cardiol 2007, 14:555–565.CrossRefPubMed Borges-Neto S, Pagnanelli RA, Shaw LK, et al.: Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies. J Nucl Cardiol 2007, 14:555–565.CrossRefPubMed
13.
go back to reference Herzog BA, Buechel RR, Katz R, et al.: Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: optimized protocol for scan time reduction. J Nucl Med 2010, 51:46–51.CrossRefPubMed Herzog BA, Buechel RR, Katz R, et al.: Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: optimized protocol for scan time reduction. J Nucl Med 2010, 51:46–51.CrossRefPubMed
14.
go back to reference Buechel RR, Herzog BA, Husmann L, et al.: Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation. Eur J Nucl Med Mol Imaging 2009, In press. Buechel RR, Herzog BA, Husmann L, et al.: Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: first clinical validation. Eur J Nucl Med Mol Imaging 2009, In press.
15.
go back to reference • Esteves FP, Raggi P, Folks RD, et al.: Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector cameras. J Nucl Cardiol 2009, 16:927–934. This is the first clinical US study of the new multi-pinhole detector system.CrossRefPubMed • Esteves FP, Raggi P, Folks RD, et al.: Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector cameras. J Nucl Cardiol 2009, 16:927–934. This is the first clinical US study of the new multi-pinhole detector system.CrossRefPubMed
16.
go back to reference Sharir T, Ben-Haim S, Merzon K, et al.: High-speed myocardial perfusion imaging: initial clinical comparison with conventional dual detector anger camera imaging. JACC Cardiovasc Imaging 2008, 1:156–163.CrossRefPubMed Sharir T, Ben-Haim S, Merzon K, et al.: High-speed myocardial perfusion imaging: initial clinical comparison with conventional dual detector anger camera imaging. JACC Cardiovasc Imaging 2008, 1:156–163.CrossRefPubMed
17.
go back to reference • Berman DS, Kang X, Tamarappoo B, et al.: Stress thallium-201/rest technetium-99m sequential dual isotope high-speed myocardial perfusion imaging. JACC Cardiovasc Imaging 2009, 2:273–282. This is a novel Tl/Tc protocol for dose-efficient fast stress and rest MPS scanning.CrossRefPubMed • Berman DS, Kang X, Tamarappoo B, et al.: Stress thallium-201/rest technetium-99m sequential dual isotope high-speed myocardial perfusion imaging. JACC Cardiovasc Imaging 2009, 2:273–282. This is a novel Tl/Tc protocol for dose-efficient fast stress and rest MPS scanning.CrossRefPubMed
18.
go back to reference Nishina H, Slomka PJ, Abidov A, et al.: Combined supine and prone quantitative myocardial perfusion SPECT: method development and clinical validation in patients with no known coronary artery disease. J Nucl Med 2006, 47:51–58.PubMed Nishina H, Slomka PJ, Abidov A, et al.: Combined supine and prone quantitative myocardial perfusion SPECT: method development and clinical validation in patients with no known coronary artery disease. J Nucl Med 2006, 47:51–58.PubMed
19.
go back to reference •• Sharir T, Slomka P, Hayes S, et al.: Multicenter trial of high-speed vs. conventional SPECT imaging: quantitative results of myocardial perfusion and left ventricular function. J Am Coll Cardiol 2010, In press. This article includes a comparison of fully automated perfusion quantification for conventional and fast SPECT. •• Sharir T, Slomka P, Hayes S, et al.: Multicenter trial of high-speed vs. conventional SPECT imaging: quantitative results of myocardial perfusion and left ventricular function. J Am Coll Cardiol 2010, In press. This article includes a comparison of fully automated perfusion quantification for conventional and fast SPECT.
20.
go back to reference Slomka PJ, Nishina H, Berman DS, et al.: Automated quantification of myocardial perfusion SPECT using simplified normal limits. J Nucl Cardiol 2005, 12:66–77.CrossRefPubMed Slomka PJ, Nishina H, Berman DS, et al.: Automated quantification of myocardial perfusion SPECT using simplified normal limits. J Nucl Cardiol 2005, 12:66–77.CrossRefPubMed
21.
go back to reference Germano G, Kiat H, Kavanagh PB, et al.: Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med 1995, 36:2138–2147.PubMed Germano G, Kiat H, Kavanagh PB, et al.: Automatic quantification of ejection fraction from gated myocardial perfusion SPECT. J Nucl Med 1995, 36:2138–2147.PubMed
22.
go back to reference Maddahi J, Mendez R, Mahmarian JJ, et al.: Prospective multicenter evaluation of rapid, gated SPECT myocardial perfusion upright imaging. J Nucl Cardiol 2009, 16:351–357.CrossRefPubMed Maddahi J, Mendez R, Mahmarian JJ, et al.: Prospective multicenter evaluation of rapid, gated SPECT myocardial perfusion upright imaging. J Nucl Cardiol 2009, 16:351–357.CrossRefPubMed
23.
go back to reference DePuey EG, Bommireddipalli S, Clark J, et al.: Wide beam reconstruction “quarter-time” gated myocardial perfusion SPECT functional imaging: a comparison to “full-time” ordered subset expectation maximum. J Nucl Cardiol 2009, 16:736–752.CrossRefPubMed DePuey EG, Bommireddipalli S, Clark J, et al.: Wide beam reconstruction “quarter-time” gated myocardial perfusion SPECT functional imaging: a comparison to “full-time” ordered subset expectation maximum. J Nucl Cardiol 2009, 16:736–752.CrossRefPubMed
24.
go back to reference Cahill JM, Ficaro EP, Corbett JR: 4.11: Value of same-session coronary artery calcium (CAC) scoring in the setting of stress myocardial perfusion imaging (MPI) using CT-based attenuation correction (CTAC). J Nucl Cardiol 2008, 15:12.CrossRef Cahill JM, Ficaro EP, Corbett JR: 4.11: Value of same-session coronary artery calcium (CAC) scoring in the setting of stress myocardial perfusion imaging (MPI) using CT-based attenuation correction (CTAC). J Nucl Cardiol 2008, 15:12.CrossRef
25.
go back to reference Koss J, Kirch D, Steele P: Fast acquisition of myocardial perfusion and LV contractile indices with multi-pinhole SPECT [abstract]. J Nucl Med 2009, 50(Suppl 2):1144. Koss J, Kirch D, Steele P: Fast acquisition of myocardial perfusion and LV contractile indices with multi-pinhole SPECT [abstract]. J Nucl Med 2009, 50(Suppl 2):1144.
26.
go back to reference Heller GV, Bateman TM, Johnson LL, et al.: Clinical value of attenuation correction in stress-only Tc-99m sestamibi SPECT imaging. J Nucl Cardiol 2004, 11:273–281.CrossRefPubMed Heller GV, Bateman TM, Johnson LL, et al.: Clinical value of attenuation correction in stress-only Tc-99m sestamibi SPECT imaging. J Nucl Cardiol 2004, 11:273–281.CrossRefPubMed
27.
go back to reference Iskandrian AE: Stress-only myocardial perfusion imaging: a new paradigm. J Am Coll Cardiol 2009, In press. Iskandrian AE: Stress-only myocardial perfusion imaging: a new paradigm. J Am Coll Cardiol 2009, In press.
28.
go back to reference Venero CV, Heller GV, Bateman TM, et al.: A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol 2009, 16:714–725.CrossRefPubMed Venero CV, Heller GV, Bateman TM, et al.: A multicenter evaluation of a new post-processing method with depth-dependent collimator resolution applied to full-time and half-time acquisitions without and with simultaneously acquired attenuation correction. J Nucl Cardiol 2009, 16:714–725.CrossRefPubMed
29.
go back to reference Ali I, Ruddy TD, Almgrahi A, et al.: Half-time SPECT myocardial perfusion imaging with attenuation correction. J Nucl Med 2009, 50:554.CrossRefPubMed Ali I, Ruddy TD, Almgrahi A, et al.: Half-time SPECT myocardial perfusion imaging with attenuation correction. J Nucl Med 2009, 50:554.CrossRefPubMed
30.
go back to reference • Bateman TM, Heller GV, McGhie AI, et al.: Multicenter investigation comparing a highly efficient half-time stress-only attenuation correction approach against standard rest-stress Tc-99m SPECT imaging. J Nucl Cardiol 2009, 16:726–735. This article demonstrates the feasibility of stress-only fast MPS protocols.CrossRefPubMed • Bateman TM, Heller GV, McGhie AI, et al.: Multicenter investigation comparing a highly efficient half-time stress-only attenuation correction approach against standard rest-stress Tc-99m SPECT imaging. J Nucl Cardiol 2009, 16:726–735. This article demonstrates the feasibility of stress-only fast MPS protocols.CrossRefPubMed
31.
go back to reference Gambhir SS, Berman DS, Ziffer J, et al.: A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera. J Nucl Med 2009, 50:635.CrossRefPubMed Gambhir SS, Berman DS, Ziffer J, et al.: A novel high-sensitivity rapid-acquisition single-photon cardiac imaging camera. J Nucl Med 2009, 50:635.CrossRefPubMed
32.
go back to reference Kacperski K, Erlandsson K, Ben-Haim S, et al.: Iterative deconvolution of simultaneous dual radionuclide projections for CdZnTe based cardiac SPECT. IEEE Nuclear Science Symposium Conference Record; 2008:5260–5263. Kacperski K, Erlandsson K, Ben-Haim S, et al.: Iterative deconvolution of simultaneous dual radionuclide projections for CdZnTe based cardiac SPECT. IEEE Nuclear Science Symposium Conference Record; 2008:5260–5263.
33.
go back to reference Kiat H, Germano G, Friedman J, et al.: Comparative feasibility of separate or simultaneous rest thallium-201/stress technetium-99m-sestamibi dual-isotope myocardial perfusion SPECT. J Nucl Med 1994, 35:542.PubMed Kiat H, Germano G, Friedman J, et al.: Comparative feasibility of separate or simultaneous rest thallium-201/stress technetium-99m-sestamibi dual-isotope myocardial perfusion SPECT. J Nucl Med 1994, 35:542.PubMed
34.
go back to reference Steele PP, Kirch DL, Koss JE: Comparison of simultaneous dual-isotope multipinhole SPECT with rotational SPECT in a group of patients with coronary artery disease. J Nucl Med 2008, 49:1080. Steele PP, Kirch DL, Koss JE: Comparison of simultaneous dual-isotope multipinhole SPECT with rotational SPECT in a group of patients with coronary artery disease. J Nucl Med 2008, 49:1080.
35.
go back to reference Ben-Haim S, Kacperski K, Hain S, et al.: Simultaneous dual-radionuclide myocardial perfusion imaging with a solid-state dedicated cardiac camera. Eur J Nucl Med Mol Imaging 2010, In press. Ben-Haim S, Kacperski K, Hain S, et al.: Simultaneous dual-radionuclide myocardial perfusion imaging with a solid-state dedicated cardiac camera. Eur J Nucl Med Mol Imaging 2010, In press.
36.
go back to reference Beller GA: Will cardiac positron emission tomography ultimately replace SPECT for myocardial perfusion imaging? J Nucl Cardiol 2009, 16:841–843.CrossRefPubMed Beller GA: Will cardiac positron emission tomography ultimately replace SPECT for myocardial perfusion imaging? J Nucl Cardiol 2009, 16:841–843.CrossRefPubMed
37.
go back to reference Funk T, Kirch DL, Koss JE, et al.: A novel approach to multipinhole SPECT for myocardial perfusion imaging. J Nucl Med 2006, 47:595–602.PubMed Funk T, Kirch DL, Koss JE, et al.: A novel approach to multipinhole SPECT for myocardial perfusion imaging. J Nucl Med 2006, 47:595–602.PubMed
38.
go back to reference Huang Q, Boutchko R, Reutter B, Gullberg G: Dynamic imaging with a novel dedicated cardiac SPECT system. J Nucl Med 2009, 50(Suppl 2):525P. Huang Q, Boutchko R, Reutter B, Gullberg G: Dynamic imaging with a novel dedicated cardiac SPECT system. J Nucl Med 2009, 50(Suppl 2):525P.
39.
go back to reference Breault C, Roth N, Slomka P, et al.: Quantification of coronary vasodilator reserve using dynamic SPECT imaging: a feasibility study [abstract]. J Nucl Cardiol 2010, (in press). Breault C, Roth N, Slomka P, et al.: Quantification of coronary vasodilator reserve using dynamic SPECT imaging: a feasibility study [abstract]. J Nucl Cardiol 2010, (in press).
40.
go back to reference Slomka P, Kavanagh PB, Javadi M, et al.: Evaluation of a new automatic algorithm for quantification of ECG-gated 82Rb cardiac PET. J Nucl Med 2009, 50(Suppl 2):217P. Slomka P, Kavanagh PB, Javadi M, et al.: Evaluation of a new automatic algorithm for quantification of ECG-gated 82Rb cardiac PET. J Nucl Med 2009, 50(Suppl 2):217P.
41.
go back to reference Slomka PJ, Berman DS, Germano G: Applications and software techniques for integrated cardiac multimodality imaging. Exp Rev Cardiovasc Ther 2008, 6:27–41.CrossRef Slomka PJ, Berman DS, Germano G: Applications and software techniques for integrated cardiac multimodality imaging. Exp Rev Cardiovasc Ther 2008, 6:27–41.CrossRef
42.
go back to reference Gutstein A, Wolak A, Lee C, et al.: Predicting success of prospective and retrospective gating with dual-source coronary computed tomography angiography: development of selection criteria and initial experience. J Cardiovasc Comput Tomogr 2008, 2:81-90.CrossRefPubMed Gutstein A, Wolak A, Lee C, et al.: Predicting success of prospective and retrospective gating with dual-source coronary computed tomography angiography: development of selection criteria and initial experience. J Cardiovasc Comput Tomogr 2008, 2:81-90.CrossRefPubMed
43.
go back to reference • Husmann L, Herzog BA, Gaemperli O, et al.: Diagnostic accuracy of computed tomography coronary angiography and evaluation of stress-only single-photon emission computed tomography/computed tomography hybrid imaging: comparison of prospective electrocardiogram-triggering vs. retrospective gating. Eur Heart J 2009, 30:600. Novel protocols are proposed in this article, which combine MPS and CCTA.CrossRefPubMed • Husmann L, Herzog BA, Gaemperli O, et al.: Diagnostic accuracy of computed tomography coronary angiography and evaluation of stress-only single-photon emission computed tomography/computed tomography hybrid imaging: comparison of prospective electrocardiogram-triggering vs. retrospective gating. Eur Heart J 2009, 30:600. Novel protocols are proposed in this article, which combine MPS and CCTA.CrossRefPubMed
44.
go back to reference DePuey EG: New software methods to cope with reduced counting statistics: shorter SPECT acquisitions and many more possibilities. J Nucl Cardiol 2009, 16:335–338.CrossRefPubMed DePuey EG: New software methods to cope with reduced counting statistics: shorter SPECT acquisitions and many more possibilities. J Nucl Cardiol 2009, 16:335–338.CrossRefPubMed
45.
go back to reference Achenbach S, Marwan M, Ropers D, et al.: Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J 2010, 31:340–346.CrossRefPubMed Achenbach S, Marwan M, Ropers D, et al.: Coronary computed tomography angiography with a consistent dose below 1 mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition. Eur Heart J 2010, 31:340–346.CrossRefPubMed
46.
go back to reference Santana CA, Garcia EV, Faber TL, et al.: Diagnostic performance of fusion of myocardial perfusion imaging (MPI) and computed tomography coronary angiography. J Nucl Cardiol 2009, 16:201–211.CrossRefPubMed Santana CA, Garcia EV, Faber TL, et al.: Diagnostic performance of fusion of myocardial perfusion imaging (MPI) and computed tomography coronary angiography. J Nucl Cardiol 2009, 16:201–211.CrossRefPubMed
47.
go back to reference Slomka PJ, Cheng VY, Dey D, et al.: Quantitative analysis of myocardial perfusion SPECT anatomically guided by coregistered 64-slice coronary CT angiography. J Nucl Med 2009, 50:1621–1630.CrossRefPubMed Slomka PJ, Cheng VY, Dey D, et al.: Quantitative analysis of myocardial perfusion SPECT anatomically guided by coregistered 64-slice coronary CT angiography. J Nucl Med 2009, 50:1621–1630.CrossRefPubMed
48.
49.
go back to reference Dey J: Conference record for IEEE NSS-MIC. 2009, In press. Dey J: Conference record for IEEE NSS-MIC. 2009, In press.
Metadata
Title
New Imaging Protocols for New Single Photon Emission CT Technologies
Authors
Piotr J. Slomka
Daniel S. Berman
Guido Germano
Publication date
01-06-2010
Publisher
Current Science Inc.
Published in
Current Cardiovascular Imaging Reports / Issue 3/2010
Print ISSN: 1941-9066
Electronic ISSN: 1941-9074
DOI
https://doi.org/10.1007/s12410-010-9021-0

Other articles of this Issue 3/2010

Current Cardiovascular Imaging Reports 3/2010 Go to the issue