Published in: Sleep and Breathing 3/2014

01-09-2014 | Original Article

A pilot validation study for the NOX T3TM portable monitor for the detection of OSA

Authors: Alyssa Cairns, Emerson Wickwire, Edward Schaefer, David Nyanjom

Published in: Sleep and Breathing | Issue 3/2014

Login to get access

Abstract

Purpose

The aim of the current pilot study is to compare the diagnostic accuracy of the NOX T3TM (T3) portable sleep monitor (PM) to that of simultaneously recorded in-lab polysomnogram (PSG).

Methods

A total of 40 participants were recruited following face-to-face evaluation at a sleep disorders clinic. Each participant wore both PSG and PM equipment simultaneously during their in-lab PSG. PSG records were manually scored using the American Academy of Sleep Medicine (AASM) criteria, and PM records were double-scored using the device’s autoscore algorithm as well as manual scoring.

Results

The final sample consisted of 32 participants (56 % male, 50 % black) with a mean ESS, BMI, and apnea–hypopnea index (AHI) of 10.4, 32.8, and 16.3, respectively. Three participants (7.5 %) were excluded for poor PM signal quality. Mean AHI derived from the T3’s autoscore algorithm was similar to that from manual scoring (19.6 ± 18.9 vs. 18.6 ± 19.1, respectively). Autoscore-derived T3 AHI and PSG-derived AHI were strongly related (r = .93). The T3 (autoscored AHI) demonstrated a high degree of sensitivity for the presence of obstructive sleep apnea syndrome (OSA; 100 %) and acceptable specificity for the exclusion of OSA using an AHI cutoff of ≥5 events/h (70 %). The unit (autoscored) had a high degree of both sensitivity (92 %) and specificity (85 %) when the presence of OSA was defined more conservatively (AHI > 15 events/h). For OSA defined as an AHI of ≥5, the T3 (autoscored) correctly identified 88 % of positive cases and 100 % of negative cases.

Conclusions

In this small, clinic-based sample, the T3 demonstrated very good measurement agreement compared to PSG and a high degree of sensitivity for detecting even mild OSA. False positives appeared to be due to respiratory effort-related arousals (RERAs) being autoscored as obstructive apneas and may be due to inherent discrepancy in flow measurement sensitivity between PSG and portable monitors.
Literature
1.
go back to reference Young T, Palta M, Dempsey J, Peppard PE, Nieto FJ, Hla KM (2009) Burden of sleep apnea: rationale, design, and major findings of the Wisconsin Sleep Cohort study. WMJ 108(5):246–249PubMedCentralPubMed Young T, Palta M, Dempsey J, Peppard PE, Nieto FJ, Hla KM (2009) Burden of sleep apnea: rationale, design, and major findings of the Wisconsin Sleep Cohort study. WMJ 108(5):246–249PubMedCentralPubMed
2.
go back to reference Young T, Peppard PE, Gottlieb DJ (2002) Epidemiology of obstructive sleep apnea: a population health perspective. Am J Respir Crit Care Med 165(9):1217–1239PubMedCrossRef Young T, Peppard PE, Gottlieb DJ (2002) Epidemiology of obstructive sleep apnea: a population health perspective. Am J Respir Crit Care Med 165(9):1217–1239PubMedCrossRef
3.
go back to reference United Healthcare (2012) Polysomnography and portable monitoring for evaluation of sleep related breathing disorders: medical policy. United Healthcare, Minnetonka United Healthcare (2012) Polysomnography and portable monitoring for evaluation of sleep related breathing disorders: medical policy. United Healthcare, Minnetonka
4.
go back to reference Aetna (2013) Clinical policy bulletin: obstructive sleep apnea in adults. Aetna, Hartford Aetna (2013) Clinical policy bulletin: obstructive sleep apnea in adults. Aetna, Hartford
5.
go back to reference Collop NA, Anderson WM, Boehlecke B, Claman D, Goldberg R, Gottlieb DJ, Hudgel D, Sateia M, Schwab R (2007) Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med 3(7):737–747PubMed Collop NA, Anderson WM, Boehlecke B, Claman D, Goldberg R, Gottlieb DJ, Hudgel D, Sateia M, Schwab R (2007) Clinical guidelines for the use of unattended portable monitors in the diagnosis of obstructive sleep apnea in adult patients. Portable Monitoring Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med 3(7):737–747PubMed
6.
go back to reference Driver HS, Pereira EJ, Bjerring K, Toop F, Stewart SC, Munt PW, Fitzpatrick MF (2011) Validation of the MediByte(R) type 3 portable monitor compared with polysomnography for screening of obstructive sleep apnea. Can Respir J 18(3):137–143PubMedCentralPubMed Driver HS, Pereira EJ, Bjerring K, Toop F, Stewart SC, Munt PW, Fitzpatrick MF (2011) Validation of the MediByte(R) type 3 portable monitor compared with polysomnography for screening of obstructive sleep apnea. Can Respir J 18(3):137–143PubMedCentralPubMed
7.
go back to reference Santos-Silva R, Sartori DE, Truksinas V, Truksinas E, Alonso FF, Tufik S, Bittencourt LR (2009) Validation of a portable monitoring system for the diagnosis of obstructive sleep apnea syndrome. Sleep 32(5):629–636PubMedCentralPubMed Santos-Silva R, Sartori DE, Truksinas V, Truksinas E, Alonso FF, Tufik S, Bittencourt LR (2009) Validation of a portable monitoring system for the diagnosis of obstructive sleep apnea syndrome. Sleep 32(5):629–636PubMedCentralPubMed
8.
go back to reference Flemons WW, Littner MR, Rowley JA, Gay P, Anderson WM, Hudgel DW, McEvoy RD, Loube DI (2003) Home diagnosis of sleep apnea: a systematic review of the literature. An evidence review cosponsored by the American Academy of Sleep Medicine, the American College of Chest Physicians, and the American Thoracic Society. Chest 124(4):1543–1579PubMedCrossRef Flemons WW, Littner MR, Rowley JA, Gay P, Anderson WM, Hudgel DW, McEvoy RD, Loube DI (2003) Home diagnosis of sleep apnea: a systematic review of the literature. An evidence review cosponsored by the American Academy of Sleep Medicine, the American College of Chest Physicians, and the American Thoracic Society. Chest 124(4):1543–1579PubMedCrossRef
9.
go back to reference Collop NA, Tracy SL, Kapur V, Mehra R, Kuhlmann D, Fleishman SA, Ojile JM (2011) Obstructive sleep apnea devices for out-of-center (OOC) testing: technology evaluation. J Clin Sleep Med 7(5):531–548PubMedCentralPubMed Collop NA, Tracy SL, Kapur V, Mehra R, Kuhlmann D, Fleishman SA, Ojile JM (2011) Obstructive sleep apnea devices for out-of-center (OOC) testing: technology evaluation. J Clin Sleep Med 7(5):531–548PubMedCentralPubMed
10.
go back to reference Berry RB, Budhiraja R, Gottlieb DJ, Gozal D, Iber C, Kapur VK, Marcus CL, Mehra R, Parthasarathy S, Quan SF, Redline S, Strohl KP, Davidson Ward SL, Tangredi MM (2012) Rules for scoring respiratory events in sleep: update of the 2007 AASM Manual for the Scoring of Sleep and Associated Events. Deliberations of the Sleep Apnea Definitions Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med 8(5):597–619PubMedCentralPubMed Berry RB, Budhiraja R, Gottlieb DJ, Gozal D, Iber C, Kapur VK, Marcus CL, Mehra R, Parthasarathy S, Quan SF, Redline S, Strohl KP, Davidson Ward SL, Tangredi MM (2012) Rules for scoring respiratory events in sleep: update of the 2007 AASM Manual for the Scoring of Sleep and Associated Events. Deliberations of the Sleep Apnea Definitions Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med 8(5):597–619PubMedCentralPubMed
11.
go back to reference Epstein LJ, Kristo D, Strollo PJ Jr, Friedman N, Malhotra A, Patil SP, Ramar K, Rogers R, Schwab RJ, Weaver EM, Weinstein MD (2009) Clinical guideline for the evaluation, management and long-term care of obstructive sleep apnea in adults. J Clin Sleep Med 5(3):263–276PubMed Epstein LJ, Kristo D, Strollo PJ Jr, Friedman N, Malhotra A, Patil SP, Ramar K, Rogers R, Schwab RJ, Weaver EM, Weinstein MD (2009) Clinical guideline for the evaluation, management and long-term care of obstructive sleep apnea in adults. J Clin Sleep Med 5(3):263–276PubMed
12.
go back to reference Kushida CA, Littner MR, Morgenthaler T, Alessi CA, Bailey D, Coleman J Jr, Friedman L, Hirshkowitz M, Kapen S, Kramer M, Lee-Chiong T, Loube DL, Owens J, Pancer JP, Wise M (2005) Practice parameters for the indications for polysomnography and related procedures: an update for 2005. Sleep 28(4):499–521PubMed Kushida CA, Littner MR, Morgenthaler T, Alessi CA, Bailey D, Coleman J Jr, Friedman L, Hirshkowitz M, Kapen S, Kramer M, Lee-Chiong T, Loube DL, Owens J, Pancer JP, Wise M (2005) Practice parameters for the indications for polysomnography and related procedures: an update for 2005. Sleep 28(4):499–521PubMed
13.
go back to reference Iber C, Ancoli-Israel S, Chesson AL, Quan SF (2007) The AASM manual for the scoring of sleep and associated events: rules, terminology and technical specifications. The American Academy of Sleep Medicine, Westchester Iber C, Ancoli-Israel S, Chesson AL, Quan SF (2007) The AASM manual for the scoring of sleep and associated events: rules, terminology and technical specifications. The American Academy of Sleep Medicine, Westchester
14.
go back to reference Yin M, Miyazaki S, Ishikawa K (2006) Evaluation of type 3 portable monitoring in unattended home setting for suspected sleep apnea: factors that may affect its accuracy. Otolaryngol Head Neck Surg 134(2):204–209PubMedCrossRef Yin M, Miyazaki S, Ishikawa K (2006) Evaluation of type 3 portable monitoring in unattended home setting for suspected sleep apnea: factors that may affect its accuracy. Otolaryngol Head Neck Surg 134(2):204–209PubMedCrossRef
Metadata
Title
A pilot validation study for the NOX T3TM portable monitor for the detection of OSA
Authors
Alyssa Cairns
Emerson Wickwire
Edward Schaefer
David Nyanjom
Publication date
01-09-2014
Publisher
Springer Berlin Heidelberg
Published in
Sleep and Breathing / Issue 3/2014
Print ISSN: 1520-9512
Electronic ISSN: 1522-1709
DOI
https://doi.org/10.1007/s11325-013-0924-2

Other articles of this Issue 3/2014

Sleep and Breathing 3/2014 Go to the issue
Live Webinar | 27-06-2024 | 18:00 (CEST)

Keynote webinar | Spotlight on medication adherence

Live: Thursday 27th June 2024, 18:00-19:30 (CEST)

WHO estimates that half of all patients worldwide are non-adherent to their prescribed medication. The consequences of poor adherence can be catastrophic, on both the individual and population level.

Join our expert panel to discover why you need to understand the drivers of non-adherence in your patients, and how you can optimize medication adherence in your clinics to drastically improve patient outcomes.

Prof. Kevin Dolgin
Prof. Florian Limbourg
Prof. Anoop Chauhan
Developed by: Springer Medicine
Obesity Clinical Trial Summary

At a glance: The STEP trials

A round-up of the STEP phase 3 clinical trials evaluating semaglutide for weight loss in people with overweight or obesity.

Developed by: Springer Medicine

Highlights from the ACC 2024 Congress

Year in Review: Pediatric cardiology

Watch Dr. Anne Marie Valente present the last year's highlights in pediatric and congenital heart disease in the official ACC.24 Year in Review session.

Year in Review: Pulmonary vascular disease

The last year's highlights in pulmonary vascular disease are presented by Dr. Jane Leopold in this official video from ACC.24.

Year in Review: Valvular heart disease

Watch Prof. William Zoghbi present the last year's highlights in valvular heart disease from the official ACC.24 Year in Review session.

Year in Review: Heart failure and cardiomyopathies

Watch this official video from ACC.24. Dr. Biykem Bozkurt discusses last year's major advances in heart failure and cardiomyopathies.