Published in: Journal of Clinical Monitoring and Computing 5/2019

01-10-2019 | Original Research

Anesthesia depth monitoring using alternative placement of entropy sensors: a prospective study

Authors: David I. Vachnadze, Boris A. Akselrod, Denis A. Guskov, Alevtina V. Goncharova

Published in: Journal of Clinical Monitoring and Computing | Issue 5/2019

Login to get access

Abstract

Spectral entropy is based on analysis of variations in electroencephalography and frontal electromyography, and is a safe and reliable method for anesthesia depth monitoring. However, standard frontal positioning of entropy electrodes in patients undergoing cardiac surgery is sometimes challenging. The present study aimed to compare standard entropy sensor placement with an alternative (infraorbital) site. This prospective study included 20 patients who underwent cardiovascular surgery at the authors’ center. Monitoring was performed with standard and alternative entropy electrode positions from patient admission to surgery to transfer to the intensive care unit. Data were recorded every 15 s; all data were analyzed and compared using Bland–Altman, scatter plot with Pearson correlation coefficient, and sensitivity/specificity analyses. Overall, 20,784 pairs of response entropy (RE) and state entropy (SE) indexes were collected. Bland–Altman analysis revealed a mean difference in RE of 0.37 (95% LOA − 7.09, 7.88) and SE 0.69 (95% LOA − 5.95, 7.31); with 3.46% (720/20,784) RE and 3.40% (706/10,790) SE values lying outside of the limits of agreement. Correlation analysis revealed strong positive correlation in both cases: RE, r = 0.983, p < 0.05; SE, r = 0.984; p < 0.05. Sensitivity/specificity analysis revealed 98.1% sensitivity, 93.3% specificity and 97.1% test efficiency for RE, and 99.2%, 95.1% and 98.5% for SE, respectively. Infraorbital entropy sensor placement in patients undergoing cardiovascular surgery is reliable and effective. The strong positive correlation between the two methods of registration enables alternative entropy measurement when frontal placement is not possible.
Literature
1.
go back to reference Arutyunyan OM, Yavorovsky AG, Seleznev MN, Guleshov VA, Bunyatyan AA. Entropy monitoring of anesthetic depth at cardiac surgery. Vestnik Anestesiologii I Reanimatologii. 2011;6:17–23. Arutyunyan OM, Yavorovsky AG, Seleznev MN, Guleshov VA, Bunyatyan AA. Entropy monitoring of anesthetic depth at cardiac surgery. Vestnik Anestesiologii I Reanimatologii. 2011;6:17–23.
2.
go back to reference Chhabra A, Subramaniam R, Srivastava A, Prabhakar H, Kalaivani M, Paranjape S. Spectral entropy monitoring for adults and children undergoing general anaesthesia. Cochrane Database Syst Rev. 2016;3:CD010135.PubMed Chhabra A, Subramaniam R, Srivastava A, Prabhakar H, Kalaivani M, Paranjape S. Spectral entropy monitoring for adults and children undergoing general anaesthesia. Cochrane Database Syst Rev. 2016;3:CD010135.PubMed
3.
go back to reference Dahaba AA, Xue JX, Zhao GG, et al. BIS-vista occipital montage in patients undergoing neurosurgical procedures during propofol remifentanil anesthesia. Anesthesiology. 2010;112:645–51.CrossRefPubMed Dahaba AA, Xue JX, Zhao GG, et al. BIS-vista occipital montage in patients undergoing neurosurgical procedures during propofol remifentanil anesthesia. Anesthesiology. 2010;112:645–51.CrossRefPubMed
4.
5.
go back to reference Nelson P, Nelson JA, Chen AJ, Kofke WA. An alternative position for the BIS-Vista montage in frontal approach neurosurgical cases. J Neurosurg Anesthesiol. 2013;25:135–42.CrossRefPubMed Nelson P, Nelson JA, Chen AJ, Kofke WA. An alternative position for the BIS-Vista montage in frontal approach neurosurgical cases. J Neurosurg Anesthesiol. 2013;25:135–42.CrossRefPubMed
6.
go back to reference Viertiö-Oja H, Maja V, Särkelä M, Talja P, Tenkanen N, Tolvanen-Laakso H, et al. Description of the entropy algorithm as applied in the Datex-Ohmeda S/5 entropy module. Acta Anaesthesiol Scand. 2004;48:154–61.CrossRefPubMed Viertiö-Oja H, Maja V, Särkelä M, Talja P, Tenkanen N, Tolvanen-Laakso H, et al. Description of the entropy algorithm as applied in the Datex-Ohmeda S/5 entropy module. Acta Anaesthesiol Scand. 2004;48:154–61.CrossRefPubMed
7.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRefPubMed
8.
go back to reference Jiahai M. Spectral entropy monitoring reduces anesthetic dosage for patients undergoing off-pump coronary artery bypass graft surgery. J Cardiothorac Vasc Anesth. 2012;26:818–21.CrossRefPubMed Jiahai M. Spectral entropy monitoring reduces anesthetic dosage for patients undergoing off-pump coronary artery bypass graft surgery. J Cardiothorac Vasc Anesth. 2012;26:818–21.CrossRefPubMed
9.
go back to reference Shepherd J, Jones J, Frampton G, Bryant J, Baxter L, Cooper K. Clinical effectiveness and cost-effectiveness of depth of anaesthesia monitoring (E-Entropy, Bispectral Index and Narcotrend): a systematic review and economic evaluation. Health Technol Assess. 2013;17(34):1–264.CrossRefPubMedPubMedCentral Shepherd J, Jones J, Frampton G, Bryant J, Baxter L, Cooper K. Clinical effectiveness and cost-effectiveness of depth of anaesthesia monitoring (E-Entropy, Bispectral Index and Narcotrend): a systematic review and economic evaluation. Health Technol Assess. 2013;17(34):1–264.CrossRefPubMedPubMedCentral
10.
go back to reference Leon S. Awareness in cardiac anesthesia. Curr Opin Anesthesiol. 2010;23:103–8.CrossRef Leon S. Awareness in cardiac anesthesia. Curr Opin Anesthesiol. 2010;23:103–8.CrossRef
11.
go back to reference Yu H, Wu D. Effects of different methods of general anesthesia on intraoperative awareness in surgical patients. Medicine (Baltimore). 2017;96(42):e6428.CrossRef Yu H, Wu D. Effects of different methods of general anesthesia on intraoperative awareness in surgical patients. Medicine (Baltimore). 2017;96(42):e6428.CrossRef
12.
go back to reference Messina AG, Wang M, Ward MJ, Wilker CC, Eggers M, Miller DB, et al. The effectiveness of anaesthetic interventions for prevention of wakefulness and awareness during and after surgery. Cochrane Database Syst Rev. 2008;3:803–8. Messina AG, Wang M, Ward MJ, Wilker CC, Eggers M, Miller DB, et al. The effectiveness of anaesthetic interventions for prevention of wakefulness and awareness during and after surgery. Cochrane Database Syst Rev. 2008;3:803–8.
13.
go back to reference Shiraishi T, Uchino H, Sagara T, et al. A comparison of frontal and occipital bispectral index values obtained during neurosurgical procedures. Anesth Analg. 2004;98:1773–5.CrossRefPubMed Shiraishi T, Uchino H, Sagara T, et al. A comparison of frontal and occipital bispectral index values obtained during neurosurgical procedures. Anesth Analg. 2004;98:1773–5.CrossRefPubMed
14.
go back to reference Singh S, Bansal S, Kumar G, Gupta I, Thakur JR. Entropy as an indicator to measure depth of anaesthesia for laryngeal mask airway (LMA) insertion during sevoflurane and propofol anaesthesia. J Clin Diagn Res. 2017;11(7):1–3. Singh S, Bansal S, Kumar G, Gupta I, Thakur JR. Entropy as an indicator to measure depth of anaesthesia for laryngeal mask airway (LMA) insertion during sevoflurane and propofol anaesthesia. J Clin Diagn Res. 2017;11(7):1–3.
15.
go back to reference Messina AG, Wang M, Ward MJ, Wilker CC, Smith BB, Vezina DP, Pace NL. Anaesthetic interventions for prevention of awareness during surgery. Cochrane Database Syst Rev. 2016;10. Messina AG, Wang M, Ward MJ, Wilker CC, Smith BB, Vezina DP, Pace NL. Anaesthetic interventions for prevention of awareness during surgery. Cochrane Database Syst Rev. 2016;10.
16.
go back to reference Akavipat P, Hungsawanich N, Jansin R. Alternative placement of bispectral index electrode for monitoring depth of anesthesia during neurosurgery. Acta Med Okayama. 2014;68:151–5.PubMed Akavipat P, Hungsawanich N, Jansin R. Alternative placement of bispectral index electrode for monitoring depth of anesthesia during neurosurgery. Acta Med Okayama. 2014;68:151–5.PubMed
17.
go back to reference Lee SY, Kim YS, Lim BG, Kim H, Kong MH, Lee IO. Comparison of bispectral index scores from the standard frontal sensor position with those from an alternative mandibular position. Korean J Anesthesiol. 2014;66:267–73.CrossRefPubMedPubMedCentral Lee SY, Kim YS, Lim BG, Kim H, Kong MH, Lee IO. Comparison of bispectral index scores from the standard frontal sensor position with those from an alternative mandibular position. Korean J Anesthesiol. 2014;66:267–73.CrossRefPubMedPubMedCentral
Metadata
Title
Anesthesia depth monitoring using alternative placement of entropy sensors: a prospective study
Authors
David I. Vachnadze
Boris A. Akselrod
Denis A. Guskov
Alevtina V. Goncharova
Publication date
01-10-2019
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 5/2019
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-018-00238-z

Other articles of this Issue 5/2019

Journal of Clinical Monitoring and Computing 5/2019 Go to the issue