Published in: Journal of Clinical Monitoring and Computing 5/2019

01-10-2019 | Original Research

Mindray 3-directional NMT Module (a new generation “Tri-axial” neuromuscular monitor) versus the Relaxometer mechanomyograph and versus the TOF-Watch SX acceleromyograph

Authors: Ashraf A. Dahaba, Ismet Suljevic, Zhao Yang Xiao, Kun Wang

Published in: Journal of Clinical Monitoring and Computing | Issue 5/2019

Login to get access

Abstract

Recently introduced Mindray “3-directional” neuromuscular transmission transducer (NMT, Shenzhen, China) acceleromyograph) claim to monitor thumb movement in 3 different directions. We compared NMT with the gold standard Relaxometer® mechanomyograph (MMG, Groningen University, Netherlands) in Study-1 and with TOF-Watch SX™ (WTCH) acceleromyograph from which it was developed in Study-2. We used first twitch (T1%) and train-of-four (TOF) ratio rocuronium 0.6 mg kg−1 neuromuscular block to evaluate NMT diagnostic accuracy in indicating 3 clinically relevant time points namely; MMG T1 5% (95% twitch depression) for tracheal intubation, MMG T1 25% for repeat neuromuscular blocking agents (NMBAs) administration, and MMG 0.9 TOF ratio full neuromuscular block recovery. We compared onset time (time from beginning of rocuronium administration until maximal depression), Dur25 (time until T1 25% recovery) and Dur0.9 (time until 0.9 TOF ratio recovery). In Study-1, NMT showed low sensitivity in indicating MMG time for tracheal intubation, repeat NMBAs administration and full neuromuscular block recovery (6.25%, 38.9% and 38.9% respectively). NMT onset time, Dur25 and Dur0.9 (2:51 ± 00:57, 36:50 ± 24:25, 70:08 ± 25:27 min:s) were significantly longer than MMG (1:56 ± 00:46, 30:26 ± 20:24, 62:03 ± 20:01). In Study-2, NMT onset time, Dur25 and Dur0.9 (02:37 ± 00:53, 35:38 ± 11:54, 53:40 ± 13:49) were not significantly different than WTCH (02:23 ± 00:45, 33:27 ± 12:51, 53:57 ± 12:47). NMT could not efficaciously detect MMG time for tracheal intubation; NMBAs repeat dose administration or full neuromuscular block recovery. Data from NMT cannot be used interchangeably with MMG. Our study revealed that NMT Tri-axial acceleromyography seems to offer no advantage over the MMG gold standard or the classic Mono-axial TOF-Watch SX monitor.
Literature
1.
go back to reference Fuchs-Buder T, Claudius C, Skovgaard LT, et al. Good Clinical research Practice (GCRP) in pharmocodynamic studies of neuromuscular blocking agents II: the Stockholm revision. Acta Anaesthesiol Scand. 2007;51:789–808.CrossRefPubMed Fuchs-Buder T, Claudius C, Skovgaard LT, et al. Good Clinical research Practice (GCRP) in pharmocodynamic studies of neuromuscular blocking agents II: the Stockholm revision. Acta Anaesthesiol Scand. 2007;51:789–808.CrossRefPubMed
2.
go back to reference Loan PB, Paxton LD, Mirakhur RK, Connolly FM, McCoy EP. The TOF-Guard neuromuscular transmission monitor. A comparison with the Myograph 2000. Anaesthesia. 1995;50:699–702.CrossRefPubMed Loan PB, Paxton LD, Mirakhur RK, Connolly FM, McCoy EP. The TOF-Guard neuromuscular transmission monitor. A comparison with the Myograph 2000. Anaesthesia. 1995;50:699–702.CrossRefPubMed
3.
go back to reference Claudius C, Viby-Mogensen J. Acceleromyography for use in scientific and clinical practice. A systematic review of the evidence. Anesthesiology. 2008;108:1117–40.CrossRefPubMed Claudius C, Viby-Mogensen J. Acceleromyography for use in scientific and clinical practice. A systematic review of the evidence. Anesthesiology. 2008;108:1117–40.CrossRefPubMed
4.
go back to reference Bossuyt PM, Reitsma JB, Bruns DE, et al. The STARD statement for reporting studies of diagnostic accuracy: explanation and elaboration. Clin Chem. 2003;49:7–18.CrossRefPubMed Bossuyt PM, Reitsma JB, Bruns DE, et al. The STARD statement for reporting studies of diagnostic accuracy: explanation and elaboration. Clin Chem. 2003;49:7–18.CrossRefPubMed
5.
go back to reference Rowaan CJ, Vandenbrom RHG, Wierda JMKH. The Relaxometer: a complete and comprehensive computer-controlled neuromuscular transmission measurement system developed for clinical research on muscle relaxants. J Clin Monit. 1993;9:38–44.CrossRefPubMed Rowaan CJ, Vandenbrom RHG, Wierda JMKH. The Relaxometer: a complete and comprehensive computer-controlled neuromuscular transmission measurement system developed for clinical research on muscle relaxants. J Clin Monit. 1993;9:38–44.CrossRefPubMed
6.
go back to reference Dahaba AA, Prax N, Gaube W, et al. Haemodynamic and catecholamine stress responses to Laryngeal Tube-Suction Airway and the Proseal Laryngeal Mask Airway. Anaesthesia. 2006;61:330–4.CrossRefPubMed Dahaba AA, Prax N, Gaube W, et al. Haemodynamic and catecholamine stress responses to Laryngeal Tube-Suction Airway and the Proseal Laryngeal Mask Airway. Anaesthesia. 2006;61:330–4.CrossRefPubMed
7.
go back to reference Kopman AF, Kumar S, Klewicka MM, Neuman GG. The staircase phenomenon. Implications for monitoring of neuromuscular transmission. Anesthesiology. 2001;95:403–7.CrossRefPubMed Kopman AF, Kumar S, Klewicka MM, Neuman GG. The staircase phenomenon. Implications for monitoring of neuromuscular transmission. Anesthesiology. 2001;95:403–7.CrossRefPubMed
9.
go back to reference Agoston S. Onset time and evaluation of intubating conditions: rocuronium in perspective. Eur J Anaesth. 1995;12(Suppl. 11):31–7. Agoston S. Onset time and evaluation of intubating conditions: rocuronium in perspective. Eur J Anaesth. 1995;12(Suppl. 11):31–7.
10.
go back to reference Dahaba AA, Klobucar F, Rehak PH, List WF. The Neuromuscular Transmission Module (M-NMT) Vs. the Relaxometer Mechanomyograph for neuromuscular block monitoring. Anesth Analg. 2002;94:591–6.CrossRefPubMed Dahaba AA, Klobucar F, Rehak PH, List WF. The Neuromuscular Transmission Module (M-NMT) Vs. the Relaxometer Mechanomyograph for neuromuscular block monitoring. Anesth Analg. 2002;94:591–6.CrossRefPubMed
11.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.CrossRefPubMed Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10.CrossRefPubMed
12.
go back to reference Myles PS, Cui J. Using the Bland–Altman method to measure agreement with repeated measures. Br J Anaesth. 2007;99:309–11.CrossRefPubMed Myles PS, Cui J. Using the Bland–Altman method to measure agreement with repeated measures. Br J Anaesth. 2007;99:309–11.CrossRefPubMed
13.
go back to reference Hemmerling TM, Schmidt J, Hanusa C, Wolf T, Schmitt H. Simultaneous determination of neuromuscular block at the larynx, diaphragm, adductor pollicis, orbicularis oculi and corrugator supercilii muscles. Br J Anaesth. 2000;85:856–60.CrossRefPubMed Hemmerling TM, Schmidt J, Hanusa C, Wolf T, Schmitt H. Simultaneous determination of neuromuscular block at the larynx, diaphragm, adductor pollicis, orbicularis oculi and corrugator supercilii muscles. Br J Anaesth. 2000;85:856–60.CrossRefPubMed
14.
go back to reference Viby-Mogensen J, Jensen E, Werner MU, Nielsen HK. Measurement of acceleration: a new method of monitoring neuromuscular function. Acta Anaesthesiol Scand. 1988;32:45–8.CrossRefPubMed Viby-Mogensen J, Jensen E, Werner MU, Nielsen HK. Measurement of acceleration: a new method of monitoring neuromuscular function. Acta Anaesthesiol Scand. 1988;32:45–8.CrossRefPubMed
15.
go back to reference Harper NJN, Martlew R, Strang T, Wallace M. Monitoring neuromuscular block by acceleromyography: comparison of the Mini-Accelograph with the Myograph 2000. Br J Anaesth. 1994;72:411–4.CrossRefPubMed Harper NJN, Martlew R, Strang T, Wallace M. Monitoring neuromuscular block by acceleromyography: comparison of the Mini-Accelograph with the Myograph 2000. Br J Anaesth. 1994;72:411–4.CrossRefPubMed
16.
go back to reference Dahaba AA, Rehak PH, List WF. Assessment of Accelerography with the TOF-GUARD. A comparison with electromyography. Eur J Anaesth. 1997;14:623–9.CrossRef Dahaba AA, Rehak PH, List WF. Assessment of Accelerography with the TOF-GUARD. A comparison with electromyography. Eur J Anaesth. 1997;14:623–9.CrossRef
17.
go back to reference Dubois PE, Gourdin M, Russell K, Jamart J. Installation of the hand influences acceleromyography measurement. A comparison with mechanomyography during neuromuscular recovery. Acta Anaesthesiol Belg. 2005;56:163–6.PubMed Dubois PE, Gourdin M, Russell K, Jamart J. Installation of the hand influences acceleromyography measurement. A comparison with mechanomyography during neuromuscular recovery. Acta Anaesthesiol Belg. 2005;56:163–6.PubMed
18.
go back to reference Dahaba AA, Klobucar F, Rehak PH, List WF. Comparison of a new piezoelectric train-of-four neuromuscular monitor, the ParaGraph, and the Relaxometer mechanomyograph. Br J Anaesth. 1999;82:780–2.CrossRefPubMed Dahaba AA, Klobucar F, Rehak PH, List WF. Comparison of a new piezoelectric train-of-four neuromuscular monitor, the ParaGraph, and the Relaxometer mechanomyograph. Br J Anaesth. 1999;82:780–2.CrossRefPubMed
19.
go back to reference Kern SE, Johnson JO, Westenskow DR, Orr JA. An effectiveness study of a new piezoelectric sensor for train-of-four measurement. Anesth Analg. 1994;78:978–82.CrossRefPubMed Kern SE, Johnson JO, Westenskow DR, Orr JA. An effectiveness study of a new piezoelectric sensor for train-of-four measurement. Anesth Analg. 1994;78:978–82.CrossRefPubMed
20.
go back to reference Suzuki T, Fukano N, Kitajima O, Saeki S, Ogawa S. Normalization of acceleromyographic train-of-four ratio by baseline value for detecting residual neuromuscular block. Br J Anaesth. 2006;96:44–7.CrossRefPubMed Suzuki T, Fukano N, Kitajima O, Saeki S, Ogawa S. Normalization of acceleromyographic train-of-four ratio by baseline value for detecting residual neuromuscular block. Br J Anaesth. 2006;96:44–7.CrossRefPubMed
21.
go back to reference Hemmerling TM, Le N. Brief review: Neuromuscular monitoring: an update for the clinician. Can J Anaesth. 2007;54:58–72.CrossRefPubMed Hemmerling TM, Le N. Brief review: Neuromuscular monitoring: an update for the clinician. Can J Anaesth. 2007;54:58–72.CrossRefPubMed
22.
go back to reference Kopman AF, Kopman DJ. An analysis of the TOF-watch algorithm for modifying the displayed train-of-four ratio. Acta Anaesthesiol Scand. 2006;50:1313–4.CrossRefPubMed Kopman AF, Kopman DJ. An analysis of the TOF-watch algorithm for modifying the displayed train-of-four ratio. Acta Anaesthesiol Scand. 2006;50:1313–4.CrossRefPubMed
Metadata
Title
Mindray 3-directional NMT Module (a new generation “Tri-axial” neuromuscular monitor) versus the Relaxometer mechanomyograph and versus the TOF-Watch SX acceleromyograph
Authors
Ashraf A. Dahaba
Ismet Suljevic
Zhao Yang Xiao
Kun Wang
Publication date
01-10-2019
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 5/2019
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-018-0231-3

Other articles of this Issue 5/2019

Journal of Clinical Monitoring and Computing 5/2019 Go to the issue