01-08-2018 | Original Research

Non-invasive blood pressure monitoring with an oscillometric brachial cuff: impact of arrhythmia

Authors: Karim Lakhal, Maëlle Martin, Stephan Ehrmann, Sofian Faiz, Bertrand Rozec, Thierry Boulain

Published in: Journal of Clinical Monitoring and Computing | Issue 4/2018

Login to get access

Abstract

Arrhythmia-induced beat-to-beat variability of blood pressure (BP) is deemed to hinder the reliability of non-invasive oscillometric measurements (NIBP) but few data support this belief. We assessed the impact of arrhythmia on a NIBP device never tested for this purpose. We compared, in intensive care unit patients with and without arrhythmia, the agreement between three pairs of NIBP (Infinity™ Delta monitor, Dräger medical systems) and invasive readings. For systolic, diastolic and mean BP, the mean bias between NIBP and invasive measurements was not higher, in 89 patients with arrhythmia, than that observed in 127 patients with regular rhythm (p = 0.93 for mean BP). Averaging three measurements overcame the higher within-subject variability of NIBP measurements during arrhythmia, and yielded similar agreement between the two techniques in patients with arrhythmia and with regular rhythm. The international organization for standardization criteria (mean bias < 5 and SD < 8 mmHg) were satisfied neither during arrhythmia nor during regular rhythm: for mean BP, mean bias of − 8.0 ± 6.5 and − 7.5 ± 6.1 mmHg, respectively. The detection of hypotension (systolic invasive BP < 90 mmHg or mean invasive BP < 65 mmHg) or hypertension (systolic invasive BP > 140 mmHg) by NIBP was similar during arrhythmia and regular rhythm [areas under the receiver operating characteristic curves (AUCROC) of 0.88–0.92, p > 0.13]. The detection of a 10% increase in mean invasive BP after cardiovascular intervention was also associated with similar AUCROCs between the two groups. Provided that triplicates are averaged, the agreement between NIBP measured with the tested device and invasive measurements was not worse during arrhythmia than during regular rhythm.
Literature
3.
go back to reference National Institute for Health and Clinical Excellence: Guidance. Hypertension: the clinical management of primary hypertension in adults: update of clinical guidelines 18 and 34. London: National Institute for Health and Clinical Excellence; 2011. National Institute for Health and Clinical Excellence: Guidance. Hypertension: the clinical management of primary hypertension in adults: update of clinical guidelines 18 and 34. London: National Institute for Health and Clinical Excellence; 2011.
4.
go back to reference Pickering TG, Hall JE, Appel LJ, et al. Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on high blood pressure research. Circulation. 2005;111:697–716. doi:10.1161/01.CIR.0000154900.76284.F6.CrossRefPubMed Pickering TG, Hall JE, Appel LJ, et al. Recommendations for blood pressure measurement in humans and experimental animals: part 1: blood pressure measurement in humans: a statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on high blood pressure research. Circulation. 2005;111:697–716. doi:10.​1161/​01.​CIR.​0000154900.​76284.​F6.CrossRefPubMed
7.
go back to reference International Organization for Standardization (ISO) (2013) Non-invasive sphygmomanometers—Part 2: clinical investigation of automated measurement type. ISO 81060-2:2013. International Organization for Standardization (ISO) (2013) Non-invasive sphygmomanometers—Part 2: clinical investigation of automated measurement type. ISO 81060-2:2013.
10.
go back to reference Lakhal K, Ehrmann S, Martin M, et al. Blood pressure monitoring during arrhythmia: agreement between automated brachial cuff and intra-arterial measurements. Brit J Anaesth. 2015;115:540–9. doi:10.1093/bja/aev304.CrossRefPubMed Lakhal K, Ehrmann S, Martin M, et al. Blood pressure monitoring during arrhythmia: agreement between automated brachial cuff and intra-arterial measurements. Brit J Anaesth. 2015;115:540–9. doi:10.​1093/​bja/​aev304.CrossRefPubMed
11.
go back to reference Lakhal K, Martin M, Faiz S, et al. The CNAP finger cuff for noninvasive beat-to-beat monitoring of arterial blood pressure: an evaluation in intensive care unit patients and a comparison with 2 intermittent devices. Anesth Analg. 2016;123:1126–35. doi:10.1213/ANE.0000000000001324.CrossRefPubMed Lakhal K, Martin M, Faiz S, et al. The CNAP finger cuff for noninvasive beat-to-beat monitoring of arterial blood pressure: an evaluation in intensive care unit patients and a comparison with 2 intermittent devices. Anesth Analg. 2016;123:1126–35. doi:10.​1213/​ANE.​0000000000001324​.CrossRefPubMed
12.
go back to reference Lakhal K, Macq C, Ehrmann S, Boulain T, Capdevila X. Noninvasive monitoring of blood pressure in the critically ill: reliability according to the cuff site (arm, thigh, or ankle). Crit Care Med. 2012;40(4):1207–13.CrossRef Lakhal K, Macq C, Ehrmann S, Boulain T, Capdevila X. Noninvasive monitoring of blood pressure in the critically ill: reliability according to the cuff site (arm, thigh, or ankle). Crit Care Med. 2012;40(4):1207–13.CrossRef
13.
go back to reference Lakhal K, Martin M, Faiz S, et al. The CNAP finger cuff for noninvasive beat-to-beat monitoring of arterial blood pressure: an evaluation in intensive care unit patients and a comparison with 2 intermittent devices. Anesth Analg. 2016;123(5):1126–35.CrossRef Lakhal K, Martin M, Faiz S, et al. The CNAP finger cuff for noninvasive beat-to-beat monitoring of arterial blood pressure: an evaluation in intensive care unit patients and a comparison with 2 intermittent devices. Anesth Analg. 2016;123(5):1126–35.CrossRef
14.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.CrossRef
15.
go back to reference Cecconi M, De Backer D, Antonelli M, et al. Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of intensive care medicine. Intensive Care Med. 2014;40:1795–815. doi:10.1007/s00134-014-3525-z.CrossRef Cecconi M, De Backer D, Antonelli M, et al. Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of intensive care medicine. Intensive Care Med. 2014;40:1795–815. doi:10.​1007/​s00134-014-3525-z.CrossRef
16.
go back to reference Rhodes A, Evans LE, Alhazzani W, et al. Surviving sepsis campaign: international guidelines for management of sepsis and septic shock: 2016. Intensive Care Med. 2017;43(3):304–77.CrossRef Rhodes A, Evans LE, Alhazzani W, et al. Surviving sepsis campaign: international guidelines for management of sepsis and septic shock: 2016. Intensive Care Med. 2017;43(3):304–77.CrossRef
17.
go back to reference Mancia G, Fagard R, Narkiewicz K, et al. 2013 ESH/ESC guidelines for the management of arterial hypertension: the task force for the management of arterial hypertension of the European Society of hypertension (ESH) and of the European Society of cardiology (ESC). Eur Heart J. 2013;34:2159–219. doi:10.3109/08037051.2014.868629.CrossRef Mancia G, Fagard R, Narkiewicz K, et al. 2013 ESH/ESC guidelines for the management of arterial hypertension: the task force for the management of arterial hypertension of the European Society of hypertension (ESH) and of the European Society of cardiology (ESC). Eur Heart J. 2013;34:2159–219. doi:10.​3109/​08037051.​2014.​868629.CrossRef
18.
go back to reference Hanley JA, McNeil BJ. The meaning and use of the area under a receiver operating characteristic (ROC) curve. Radiology. 1982;143:29–36.CrossRef Hanley JA, McNeil BJ. The meaning and use of the area under a receiver operating characteristic (ROC) curve. Radiology. 1982;143:29–36.CrossRef
20.
go back to reference Fortin J, Lerche K, Flot-Zinger D, O’Brien T. Is the standard supplied by the association for the advancement of medical instrumentation the measure of all things for noninvasive continuous hemodynamic devices? Anesthesiology. 2015;122:208–9. doi:10.1097/ALN.0000000000000485.CrossRefPubMed Fortin J, Lerche K, Flot-Zinger D, O’Brien T. Is the standard supplied by the association for the advancement of medical instrumentation the measure of all things for noninvasive continuous hemodynamic devices? Anesthesiology. 2015;122:208–9. doi:10.​1097/​ALN.​0000000000000485​.CrossRefPubMed
22.
go back to reference Annane D, Sebille V, Duboc D, et al. (2008) Incidence and prognosis of sustained arrhythmias in critically ill patients. Am J Resp Crit Care Med 178: 20 – 5. doi:10.1164/rccm.200701-031OC. Annane D, Sebille V, Duboc D, et al. (2008) Incidence and prognosis of sustained arrhythmias in critically ill patients. Am J Resp Crit Care Med 178: 20 – 5. doi:10.​1164/​rccm.​200701-031OC.
24.
go back to reference Sykes D, Dewar R, Mohanaruban K, et al. Measuring blood pressure in the elderly: does atrial fibrillation increase observer variability? BMJ. 1990;300:162–3.CrossRef Sykes D, Dewar R, Mohanaruban K, et al. Measuring blood pressure in the elderly: does atrial fibrillation increase observer variability? BMJ. 1990;300:162–3.CrossRef
26.
go back to reference Antonelli M, Levy M, Andrews PJ, et al. Hemodynamic monitoring in shock and implications for management. International Consensus Conference, Paris, France, 27–28 April 2006. Intensive Care Med. 2007;33:575–90. doi:10.1007/s00134-007-0531-4.CrossRefPubMed Antonelli M, Levy M, Andrews PJ, et al. Hemodynamic monitoring in shock and implications for management. International Consensus Conference, Paris, France, 27–28 April 2006. Intensive Care Med. 2007;33:575–90. doi:10.​1007/​s00134-007-0531-4.CrossRefPubMed
27.
go back to reference Araghi A, Bander JJ, Guzman JA. Arterial blood pressure monitoring in overweight critically ill patients: invasive or noninvasive? Crit Care (London England) 2006. 2006;10:R64. doi:10.1186/cc4896.CrossRef Araghi A, Bander JJ, Guzman JA. Arterial blood pressure monitoring in overweight critically ill patients: invasive or noninvasive? Crit Care (London England) 2006. 2006;10:R64. doi:10.​1186/​cc4896.CrossRef
Metadata
Title
Non-invasive blood pressure monitoring with an oscillometric brachial cuff: impact of arrhythmia
Authors
Karim Lakhal
Maëlle Martin
Stephan Ehrmann
Sofian Faiz
Bertrand Rozec
Thierry Boulain
Publication date
01-08-2018
Publisher
Springer Netherlands
Published in
Journal of Clinical Monitoring and Computing / Issue 4/2018
Print ISSN: 1387-1307
Electronic ISSN: 1573-2614
DOI
https://doi.org/10.1007/s10877-017-0067-2

Other articles of this Issue 4/2018

Journal of Clinical Monitoring and Computing 4/2018 Go to the issue