Published in: Critical Care 2/2006

Open Access 01-04-2006 | Research

Arterial blood pressure monitoring in overweight critically ill patients: invasive or noninvasive?

Authors: Ali Araghi, Joseph J Bander, Jorge A Guzman

Published in: Critical Care | Issue 2/2006

Login to get access

Abstract

Introduction

Blood pressure measurements frequently guide management in critical care. Direct readings, commonly from a major artery, are considered to be the gold standard. Because arterial cannulation is associated with risks, alternative noninvasive blood pressure (NIBP) measurements are routinely used. However, the accuracy of NIBP determinations in overweight patients in the outpatient setting is variable, and little is known about critically ill patients. This prospective, observational study was performed to compare direct intra-arterial blood pressure (IABP) with NIBP measurements obtained using auscultatory and oscillometric methods in overweight patients admitted to our medical intensive care unit.

Method

Adult critically ill patients with a body mass index (BMI) of 25 kg/m2 or greater and a functional arterial line (assessed using the rapid flush test) were enrolled in the study. IABP measurements were compared with those obtained noninvasively. A calibrated aneroid manometer (auscultatory technique) with arm cuffs compatible with arm sizes and a NIBP monitor (oscillometric technique) were used for NIBP measurements. Agreement between methods was assessed using Bland-Altman analysis.

Results

Fifty-four patients (23 males) with a mean (± standard error) age of 57 ± 3 years were studied. The mean BMI was 34.0 ± 1.4 kg/m2. Mean arm circumference was 32 ± 0.6 cm. IABP readings were obtained from the radial artery in all patients. Only eight patients were receiving vasoactive medications. Mean overall biases for the auscultatory and oscillometric techniques were 4.1 ± 1.9 and -8.0 ± 1.7 mmHg, respectively (P < 0.0001), with wide limits of agreement. The overestimation of blood pressure using the auscultatory technique was more important in patients with a BMI of 30 kg/m2 or greater. In hypertensive patients both NIBP methods underestimated blood pressure as determined using direct IABP measurement.

Conclusion

Oscillometric blood pressure measurements underestimated IABP readings regardless of patient BMI. Auscultatory measurements were also inaccurate, tending to underestimate systolic blood pressure and overestimate mean arterial and diastolic blood pressure. NIBP can be inaccurate among overweight critically ill patients and lead to erroneous interpretations of blood pressure.
Appendix
Available only for authorised users
Literature
1.
go back to reference El-Solh A: Clinical approach to the critically ill, morbidly obese patient. Am J Respir Crit Care 2004, 169: 557-561. 10.1164/rccm.200309-1256CCCrossRef El-Solh A: Clinical approach to the critically ill, morbidly obese patient. Am J Respir Crit Care 2004, 169: 557-561. 10.1164/rccm.200309-1256CCCrossRef
3.
go back to reference Parati G, Boli G, Manci G: Blood pressure measurement in research and in clinical practice: recent evidence. Curr Opin Nephrol Hypertens 2004, 13: 343-357.CrossRefPubMed Parati G, Boli G, Manci G: Blood pressure measurement in research and in clinical practice: recent evidence. Curr Opin Nephrol Hypertens 2004, 13: 343-357.CrossRefPubMed
4.
go back to reference Tholl U, Forstner K, Anlauf M: Measuring blood pressure: pitfalls and recommendations. Nephrol Dial Transplant 2004, 19: 766-770. 10.1093/ndt/gfg602CrossRefPubMed Tholl U, Forstner K, Anlauf M: Measuring blood pressure: pitfalls and recommendations. Nephrol Dial Transplant 2004, 19: 766-770. 10.1093/ndt/gfg602CrossRefPubMed
5.
go back to reference Jones DW, Appel LJ, Sheps SG, Roccella EJ, Lenfant C: Measuring blood pressure accurately. New and persistent changes. JAMA 2003, 289: 1027-1030. 10.1001/jama.289.8.1027CrossRefPubMed Jones DW, Appel LJ, Sheps SG, Roccella EJ, Lenfant C: Measuring blood pressure accurately. New and persistent changes. JAMA 2003, 289: 1027-1030. 10.1001/jama.289.8.1027CrossRefPubMed
6.
go back to reference Bur A, Hirschl MM, Herkner H, Oschatz E, Kofler J, Woisetschläger C, Laggner AN: Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients. Crit Care Med 2000, 28: 371-376. 10.1097/00003246-200002000-00014CrossRefPubMed Bur A, Hirschl MM, Herkner H, Oschatz E, Kofler J, Woisetschläger C, Laggner AN: Accuracy of oscillometric blood pressure measurement according to the relation between cuff size and upper-arm circumference in critically ill patients. Crit Care Med 2000, 28: 371-376. 10.1097/00003246-200002000-00014CrossRefPubMed
7.
go back to reference Stolt M, Sjönell G, Aström H, Rössner S, Hansson L: Improved accuracy of indirect blood pressure measurement in patients with obese arms. Am J Hypertens 1993, 6: 66-71.PubMed Stolt M, Sjönell G, Aström H, Rössner S, Hansson L: Improved accuracy of indirect blood pressure measurement in patients with obese arms. Am J Hypertens 1993, 6: 66-71.PubMed
8.
go back to reference Nielsen PE, Larsen B, Holstein P, Poulsen HL: Accuracy of auscultatory blood pressure measurements in hypertensive and obese subjects. Hypertension 1983, 5: 122-127.CrossRefPubMed Nielsen PE, Larsen B, Holstein P, Poulsen HL: Accuracy of auscultatory blood pressure measurements in hypertensive and obese subjects. Hypertension 1983, 5: 122-127.CrossRefPubMed
9.
go back to reference Gardner RM: Direct blood pressure measurement. Dynamic response requirements. Anesthesiology 1981, 54: 227-236.CrossRefPubMed Gardner RM: Direct blood pressure measurement. Dynamic response requirements. Anesthesiology 1981, 54: 227-236.CrossRefPubMed
10.
go back to reference Pickering TG, Hall JE, Appel L, Falkner BE, Graves J, Hill MN, Jones DW, Kurtz T, Sheps SG, Roccella EJ: Recommendations for blood pressure measurements in humans and experimental animals: part 1: blood pressure management in humans: a statement for professionals from the Subcomittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Circulation 2005, 111: 697-716. 10.1161/01.CIR.0000154900.76284.F6CrossRefPubMed Pickering TG, Hall JE, Appel L, Falkner BE, Graves J, Hill MN, Jones DW, Kurtz T, Sheps SG, Roccella EJ: Recommendations for blood pressure measurements in humans and experimental animals: part 1: blood pressure management in humans: a statement for professionals from the Subcomittee of Professional and Public Education of the American Heart Association Council on High Blood Pressure Research. Circulation 2005, 111: 697-716. 10.1161/01.CIR.0000154900.76284.F6CrossRefPubMed
11.
go back to reference Pickering TG: Principles and techniques of blood pressure measurement. Cardiol Clin 2002, 20: 207-223. 10.1016/S0733-8651(01)00009-1CrossRefPubMed Pickering TG: Principles and techniques of blood pressure measurement. Cardiol Clin 2002, 20: 207-223. 10.1016/S0733-8651(01)00009-1CrossRefPubMed
12.
go back to reference Perloff D, Grim C, Flack J, Frohlich ED, Hill M, McDonald M, Morgenstern BZ: Human blood pressure determination by sphygmomanometry. Circulation 1993, 88: 2460-2470.CrossRefPubMed Perloff D, Grim C, Flack J, Frohlich ED, Hill M, McDonald M, Morgenstern BZ: Human blood pressure determination by sphygmomanometry. Circulation 1993, 88: 2460-2470.CrossRefPubMed
13.
go back to reference Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1: 307-310.CrossRefPubMed Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1: 307-310.CrossRefPubMed
14.
go back to reference Bur A, Herkner H, Vlcek M, Woisetschläger C, Laggner AN, Hirschl MM: Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients. Crit Care Med 2003, 31: 793-799. 10.1097/01.CCM.0000053650.12025.1ACrossRefPubMed Bur A, Herkner H, Vlcek M, Woisetschläger C, Laggner AN, Hirschl MM: Factors influencing the accuracy of oscillometric blood pressure measurement in critically ill patients. Crit Care Med 2003, 31: 793-799. 10.1097/01.CCM.0000053650.12025.1ACrossRefPubMed
15.
go back to reference Heard SO, Lisbon A, Toth I, Ramasubramanian R: An evaluation of a new continuous blood pressure monitoring system in critically ill patients. J Clin Anesth 2000, 12: 509-518. 10.1016/S0952-8180(00)00201-4CrossRefPubMed Heard SO, Lisbon A, Toth I, Ramasubramanian R: An evaluation of a new continuous blood pressure monitoring system in critically ill patients. J Clin Anesth 2000, 12: 509-518. 10.1016/S0952-8180(00)00201-4CrossRefPubMed
16.
go back to reference Pickering TG: What will replace the mercury sphygmomanometer? Blood Press Monit 2003, 8: 23-25. 10.1097/00126097-200302000-00005CrossRefPubMed Pickering TG: What will replace the mercury sphygmomanometer? Blood Press Monit 2003, 8: 23-25. 10.1097/00126097-200302000-00005CrossRefPubMed
17.
go back to reference Finnie KJ, Watts DG, Amstrong PW: Biases in the measurement of arterial pressure. Crit Care Med 1984, 12: 965-968.CrossRefPubMed Finnie KJ, Watts DG, Amstrong PW: Biases in the measurement of arterial pressure. Crit Care Med 1984, 12: 965-968.CrossRefPubMed
Metadata
Title
Arterial blood pressure monitoring in overweight critically ill patients: invasive or noninvasive?
Authors
Ali Araghi
Joseph J Bander
Jorge A Guzman
Publication date
01-04-2006
Publisher
BioMed Central
Published in
Critical Care / Issue 2/2006
Electronic ISSN: 1364-8535
DOI
https://doi.org/10.1186/cc4896

Other articles of this Issue 2/2006

Critical Care 2/2006 Go to the issue