Published in: Journal of Interventional Cardiac Electrophysiology 1/2021

Open Access 01-06-2021 | Antiarrhythmic Drugs

Long-term safety and effectiveness of paroxysmal atrial fibrillation ablation using a porous tip contact force-sensing catheter from the SMART SF trial

Authors: Andrea Natale, George Monir, Anshul M. Patel, Robert S. Fishel, Francis E. Marchlinski, M. Craig Delaughter, Charles A. Athill, Daniel P. Melby, Mario D. Gonzalez, Ramesh Hariharan, Brett Gidney, Tiffany Tan, Larry A. Chinitz

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 1/2021

Login to get access

Abstract

Purpose

The prospective, multicenter SMART SF trial demonstrated the acute safety and effectiveness of the 56-hole porous tip irrigated contact force (CF) catheter for drug-refractory paroxysmal atrial fibrillation (PAF) ablation with a low primary adverse event rate (2.5%), leading to FDA approval of the catheter. Here, we are reporting the long-term effectiveness and safety results that have not yet been reported.

Methods

Ablations were performed using the 56-hole porous tip irrigated CF catheter guided by the 3D mapping system stability module. The primary effectiveness endpoint was freedom from atrial tachyarrhythmia (including atrial fibrillation, atrial tachycardia, and/or atrial flutter), based on electrocardiographic data at 12 months. Atrial tachyarrhythmia recurrence occurring 3 months post procedure, acute procedural failures such as lack of entrance block confirmation of all PVs, and undergoing repeat procedure for atrial fibrillation in the evaluation period (91 to 365 days post the initial ablation procedure) were considered to be effectiveness failures.

Results

Seventy-eight patients (age 64.8 ± 9.7 years; male 52.6%; Caucasian 96.2%) participated in the 12-month effectiveness evaluation. Mean follow-up time was 373.5 ± 45.4 days. The Kaplan-Meier estimate of freedom from 12-month atrial tachyarrhythmia was 74.9%. Two procedure-related pericardial effusion events were reported at 92 and 180 days post procedure. There were no pulmonary vein stenosis complications or deaths reported through the 12-month follow-up period.

Conclusions

The SMART SF 12-month follow-up evaluation corroborates the early safety and effectiveness success previously reported for PAF ablation with STSF.
Literature
3.
go back to reference Benjamin EJ, Wolf PA, D’Agostino RB, Silbershatz H, Kannel WB, Levy D. Impact of atrial fibrillation on the risk of death: the Framingham Heart Study. Circulation. 1998;98(10):946–52.CrossRef Benjamin EJ, Wolf PA, D’Agostino RB, Silbershatz H, Kannel WB, Levy D. Impact of atrial fibrillation on the risk of death: the Framingham Heart Study. Circulation. 1998;98(10):946–52.CrossRef
4.
go back to reference Wolf PA, Dawber TR, Thomas HE Jr, Kannel WB. Epidemiologic assessment of chronic atrial fibrillation and risk of stroke: the Framingham study. Neurology. 1978;28(10):973–7.CrossRef Wolf PA, Dawber TR, Thomas HE Jr, Kannel WB. Epidemiologic assessment of chronic atrial fibrillation and risk of stroke: the Framingham study. Neurology. 1978;28(10):973–7.CrossRef
6.
go back to reference Calkins H, Kuck KH, Cappato R, Brugada J, Camm AJ, Chen SA, et al. 2012 HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Europace. 2012;14(4):528–606. https://doi.org/10.1093/europace/eus027.CrossRefPubMed Calkins H, Kuck KH, Cappato R, Brugada J, Camm AJ, Chen SA, et al. 2012 HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design. Europace. 2012;14(4):528–606. https://​doi.​org/​10.​1093/​europace/​eus027.CrossRefPubMed
9.
go back to reference Bertaglia E, Fassini G, Anselmino M, Stabile G, Grandinetti G, De Simone A, et al. Comparison of ThermoCool(R) Surround Flow catheter versus ThermoCool(R) catheter in achieving persistent electrical isolation of pulmonary veins: a pilot study. J Cardiovasc Electrophysiol. 2013;24(3):269–73. https://doi.org/10.1111/jce.12031.CrossRefPubMed Bertaglia E, Fassini G, Anselmino M, Stabile G, Grandinetti G, De Simone A, et al. Comparison of ThermoCool(R) Surround Flow catheter versus ThermoCool(R) catheter in achieving persistent electrical isolation of pulmonary veins: a pilot study. J Cardiovasc Electrophysiol. 2013;24(3):269–73. https://​doi.​org/​10.​1111/​jce.​12031.CrossRefPubMed
11.
go back to reference Chinitz LA, Melby DP, Marchlinski FE, Delaughter C, Fishel RS, Monir G, et al. Safety and efficiency of porous-tip contact-force catheter for drug-refractory symptomatic paroxysmal atrial fibrillation ablation: results from the SMART SF trial. Europace. 2017;20(3):392–400. https://doi.org/10.1093/europace/eux264.CrossRef Chinitz LA, Melby DP, Marchlinski FE, Delaughter C, Fishel RS, Monir G, et al. Safety and efficiency of porous-tip contact-force catheter for drug-refractory symptomatic paroxysmal atrial fibrillation ablation: results from the SMART SF trial. Europace. 2017;20(3):392–400. https://​doi.​org/​10.​1093/​europace/​eux264.CrossRef
12.
13.
go back to reference Stabile G, Di Donna P, Schillaci V, Di Monaco A, Iuliano A, Caponi D, et al. Safety and efficacy of pulmonary vein isolation using a surround flow catheter with contact force measurement capabilities: a multicenter registry. J Cardiovasc Electrophysiol. 2017;28(7):762–7. https://doi.org/10.1111/jce.13227.CrossRefPubMed Stabile G, Di Donna P, Schillaci V, Di Monaco A, Iuliano A, Caponi D, et al. Safety and efficacy of pulmonary vein isolation using a surround flow catheter with contact force measurement capabilities: a multicenter registry. J Cardiovasc Electrophysiol. 2017;28(7):762–7. https://​doi.​org/​10.​1111/​jce.​13227.CrossRefPubMed
Metadata
Title
Long-term safety and effectiveness of paroxysmal atrial fibrillation ablation using a porous tip contact force-sensing catheter from the SMART SF trial
Authors
Andrea Natale
George Monir
Anshul M. Patel
Robert S. Fishel
Francis E. Marchlinski
M. Craig Delaughter
Charles A. Athill
Daniel P. Melby
Mario D. Gonzalez
Ramesh Hariharan
Brett Gidney
Tiffany Tan
Larry A. Chinitz
Publication date
01-06-2021
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 1/2021
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-020-00780-4

Other articles of this Issue 1/2021

Journal of Interventional Cardiac Electrophysiology 1/2021 Go to the issue