Purpose

Reconnection of pulmonary veins (PVs) remains common following radiofrequency catheter ablation for atrial fibrillation (AF). Ablation Index (AI) is a novel ablation quality marker that incorporates stability, contact force (CF), time, and power in a weighted formula. Its use seems to improve lesion durability. This is a prospective, single-arm registry to investigate on the safety and mid-term efficacy of AF ablation guided by the AI.

Methods

One hundred fifty-six consecutive patients (mean age 58 ± 10 years, 49% males, 44% with structural heart disease) referred for paroxysmal (124) or persistent (32) AF underwent antral PV isolation using a surround flow CF-sensing catheter guided by the AI. Radiofrequency was delivered targeting interlesion distance ≤ 6 mm and Ablation Index of 330–350 at posterior wall and 400–450 at anterior wall.

Results

Mean overall procedure time was 95 ± 30 min with a mean fluoroscopy time of 5 ± 6 min. Mean ablation time was 26 ± 10 min, 627/628 targeted PV were isolated. One pericardial effusion and two groin hematomas were reported; none required intervention. During a mean follow-up of 14 ± 6 months, 17 (10.8%) (9% paroxysmal AF vs 22% persistent AF, p = 0.09) patients had an atrial arrhythmia recurrence.

Conclusions

PV ablation guided by AI resulted feasible, achieving a high rate of isolated PVs, with a low complication rate, and allowed a high single-procedure arrhythmia-free survival at 14 months.
Literature
2.
go back to reference Cheema A, Dong J, Dalal D, Marine JE, Henrikson CA, Spragg D, et al. Incidence and time course of early recovery of pulmonary vein conduction after catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2007;18:387–91.CrossRefPubMed Cheema A, Dong J, Dalal D, Marine JE, Henrikson CA, Spragg D, et al. Incidence and time course of early recovery of pulmonary vein conduction after catheter ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2007;18:387–91.CrossRefPubMed
3.
go back to reference Pratola C, Baldo E, Notarstefano P, Toselli T, Ferrari R. Radiofrequency ablation of atrial fibrillation: is the persistence of all intraprocedural targets necessary for long-term maintenance of sinus rhythm? Circulation. 2008;117:136–43.CrossRefPubMed Pratola C, Baldo E, Notarstefano P, Toselli T, Ferrari R. Radiofrequency ablation of atrial fibrillation: is the persistence of all intraprocedural targets necessary for long-term maintenance of sinus rhythm? Circulation. 2008;117:136–43.CrossRefPubMed
6.
go back to reference Nakagawa H, Ikeda A, Constantine G, Govari A, Sharma T, Pitha JV, et al. Controlling lesion size and incidence of steam pop by controlling contact force, radiofrequency power and application time (Force-Power-Time Index) in canine beating heart. [Abstract] Heart Rhythm. 2012;9:S5. Nakagawa H, Ikeda A, Constantine G, Govari A, Sharma T, Pitha JV, et al. Controlling lesion size and incidence of steam pop by controlling contact force, radiofrequency power and application time (Force-Power-Time Index) in canine beating heart. [Abstract] Heart Rhythm. 2012;9:S5.
7.
go back to reference Nakagawa H, Ikeda A, Govari A, Papaioannou T, Costantine G, Bar-Tal M, et al. Prospective study to test the ability to create RF lesion at predicted depth and diameter using a new formula incorporating contact force, radiofrequency power and application time (force-power-time index) in beating heart. [Abstract] Heart Rhythm. 2014;11(Suppl):S548. Nakagawa H, Ikeda A, Govari A, Papaioannou T, Costantine G, Bar-Tal M, et al. Prospective study to test the ability to create RF lesion at predicted depth and diameter using a new formula incorporating contact force, radiofrequency power and application time (force-power-time index) in beating heart. [Abstract] Heart Rhythm. 2014;11(Suppl):S548.
12.
go back to reference Stabile G, Di Donna P, Schillaci V, Di Monaco A, Iuliano A, Caponi D, et al. Safety and efficacy of pulmonary vein isolation using a surround flow catheter with contact force measurement capabilities: a multicenter registry. J Cardiovasc Electrophysiol. 2017;28:762–7. https://doi.org/10.1111/jce.13227.CrossRefPubMed Stabile G, Di Donna P, Schillaci V, Di Monaco A, Iuliano A, Caponi D, et al. Safety and efficacy of pulmonary vein isolation using a surround flow catheter with contact force measurement capabilities: a multicenter registry. J Cardiovasc Electrophysiol. 2017;28:762–7. https://​doi.​org/​10.​1111/​jce.​13227.CrossRefPubMed
14.
go back to reference Bertaglia E, Fassini G, Anselmino M, Stabile G, Grandinetti G, Simone AD, et al. Comparison of ThermoCool(®) Surround Flow catheter versus ThermoCool(®) catheter in achieving persistent electrical isolation of pulmonary veins: a pilot study. J Cardiovasc Electrophysiol. 2013;24:269–73. https://doi.org/10.1111/jce.12031.CrossRefPubMed Bertaglia E, Fassini G, Anselmino M, Stabile G, Grandinetti G, Simone AD, et al. Comparison of ThermoCool(®) Surround Flow catheter versus ThermoCool(®) catheter in achieving persistent electrical isolation of pulmonary veins: a pilot study. J Cardiovasc Electrophysiol. 2013;24:269–73. https://​doi.​org/​10.​1111/​jce.​12031.CrossRefPubMed
18.
go back to reference Marijon E, Fazaa S, Narayanan K, Guy-Moyat B, Bouzeman A, Providencia R, et al. Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: procedural and 1-year results. J Cardiovasc Electrophysiol. 2014;25:130–7. https://doi.org/10.1111/jce.12303.CrossRefPubMed Marijon E, Fazaa S, Narayanan K, Guy-Moyat B, Bouzeman A, Providencia R, et al. Real-time contact force sensing for pulmonary vein isolation in the setting of paroxysmal atrial fibrillation: procedural and 1-year results. J Cardiovasc Electrophysiol. 2014;25:130–7. https://​doi.​org/​10.​1111/​jce.​12303.CrossRefPubMed
20.
go back to reference Sigmund E, Puererfellner H, Derndorfer M, Kollias G, Winter S, Aichinger J, et al. Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement-a case-matched comparison in 198 patients. Pacing Clin Electrophysiol. 2015;38:201–8. https://doi.org/10.1111/pace.12549.CrossRefPubMed Sigmund E, Puererfellner H, Derndorfer M, Kollias G, Winter S, Aichinger J, et al. Optimizing radiofrequency ablation of paroxysmal and persistent atrial fibrillation by direct catheter force measurement-a case-matched comparison in 198 patients. Pacing Clin Electrophysiol. 2015;38:201–8. https://​doi.​org/​10.​1111/​pace.​12549.CrossRefPubMed
22.
go back to reference Anter E, Tschabrunn CM, Contreras-Valdes FM, Buxton AE, Josephson ME. Radiofrequency ablation annotation algorithm reduces the incidence of linear gaps and reconnection after pulmonary vein isolation. Heart Rhythm. 2014;11:783–90.CrossRefPubMed Anter E, Tschabrunn CM, Contreras-Valdes FM, Buxton AE, Josephson ME. Radiofrequency ablation annotation algorithm reduces the incidence of linear gaps and reconnection after pulmonary vein isolation. Heart Rhythm. 2014;11:783–90.CrossRefPubMed
23.
go back to reference Zucchelli G, Sirico G, Rebellato L, Marini M, Stabile G, Del Greco M, et al. Contiguity between ablation lesions and strict catheter stability settings assessed by VISITAG™ module improve clinical outcome of paroxysmal atrial fibrillation ablation. Results from the VISITALY Study. Circ J. 2018;82:974–82.CrossRefPubMed Zucchelli G, Sirico G, Rebellato L, Marini M, Stabile G, Del Greco M, et al. Contiguity between ablation lesions and strict catheter stability settings assessed by VISITAG™ module improve clinical outcome of paroxysmal atrial fibrillation ablation. Results from the VISITALY Study. Circ J. 2018;82:974–82.CrossRefPubMed
Metadata
Title
Safety and efficacy of atrial fibrillation ablation guided by Ablation Index module
Publication date
01-01-2019
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 1/2019
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-018-0420-5

Other articles of this Issue 1/2019

Journal of Interventional Cardiac Electrophysiology 1/2019 Go to the issue