Published in: Journal of Interventional Cardiac Electrophysiology 2/2014

01-03-2014

Initial experience of three-minute freeze cycles using the second-generation cryoballoon ablation: acute and short-term procedural outcomes

Authors: Gian-Battista Chierchia, Giacomo Di Giovanni, Juan Sieira-Moret, Carlo de Asmundis, Giulio Conte, Moises Rodriguez-Mañero, Ruben Casado-Arroyo, Giannis Baltogiannis, Gaetano Paparella, Giuseppe Ciconte, Andrea Sarkozy, Pedro Brugada

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 2/2014

Login to get access

Abstract

Background

The second-generation cryoballoon (CB-A) (Arctic Front Advance, Cryocath, Medtronic, MN, USA) might significantly improve procedural outcome with respect to the first-generation balloon. These technological improvements might also question the current recommendation of the need a 4-min freeze to achieve durable pulmonary vein isolation (PVI).

Objective

The main aim of the study was to analyze the procedural efficacy of a 3-min freeze–thaw cycles with the CB-A balloon in the terms of rates of acute PVI and 6-month outcome.

Methods

Patients having undergone CB-A for PAF or early persistent AF, with 3-min freeze–thaw cycles were consecutively included in our analysis. Acute procedural success was measured in terms of the rate of PVI. Short-term follow-up was evaluated by the means of 24-h Holters and clinical examinations at regular intervals.

Results

Fifty-two consecutive patients (35 male (67 %); mean age, 59.8 ± 10.5) were included. Mean procedure and fluoroscopy times were 96 ± 15 and 13.2 ± 8.3 min, respectively. Mean time from groin puncture to catheter extraction was 60.4 ± 20 min. After a mean of 1.5 freeze cycles per vein of 3 min in duration, all 208 (100 %) PVs could be isolated with the CB-A. A total 192 (91 %) veins were isolated during the first freeze. At a mean of 5.7-month follow-up, 82 % of patients were free of AF.

Conclusion

CB-A is effective in producing PVI by using 3-min-duration freeze cycles. After a mean of 1.5 freeze per vein, freedom from AF was achieved in 82 % of patients at 6-month follow-up.
Literature
1.
go back to reference Fürnkranz, A., Bordignon, S., Schmidt, B., et al. (2013). Improved procedural efficacy of pulmonary vein isolation using the novel second-generation cryoballoon. Journal of Cardiovascular Electrophysiology, 24(5), 492–497.PubMedCrossRef Fürnkranz, A., Bordignon, S., Schmidt, B., et al. (2013). Improved procedural efficacy of pulmonary vein isolation using the novel second-generation cryoballoon. Journal of Cardiovascular Electrophysiology, 24(5), 492–497.PubMedCrossRef
2.
go back to reference Andrade, J. G., Dubuc, M., Guerra, P. G., et al. (2013). Pulmonary vein isolation using a second-generation cryoballoon catheter: a randomized comparison of ablation duration and method of deflation. Journal of Cardiovascular Electrophysiology, 24(6), 692–698.PubMedCrossRef Andrade, J. G., Dubuc, M., Guerra, P. G., et al. (2013). Pulmonary vein isolation using a second-generation cryoballoon catheter: a randomized comparison of ablation duration and method of deflation. Journal of Cardiovascular Electrophysiology, 24(6), 692–698.PubMedCrossRef
3.
go back to reference Coulombe, N., Paulin, J., & Su, W. (2013). Improved in vivo performance of second-generation cryoballoon for pulmonary vein isolation. J Cardiovasc Electrophysiol, 24(8), 919–25. doi:10.1111/jce.12157.PubMedCrossRef Coulombe, N., Paulin, J., & Su, W. (2013). Improved in vivo performance of second-generation cryoballoon for pulmonary vein isolation. J Cardiovasc Electrophysiol, 24(8), 919–25. doi:10.​1111/​jce.​12157.PubMedCrossRef
4.
go back to reference Dubuc, M., Roy, D., Thibault, B., et al. (1999). Transvenous catheter ice mapping and cryoablation of the atrioventricular node in dogs. Pacing and Clinical Electrophysiology, 22(10), 1488–1498.PubMedCrossRef Dubuc, M., Roy, D., Thibault, B., et al. (1999). Transvenous catheter ice mapping and cryoablation of the atrioventricular node in dogs. Pacing and Clinical Electrophysiology, 22(10), 1488–1498.PubMedCrossRef
5.
go back to reference Chierchia, G. B., Namdar, M., Sarkozy, A., et al. (2012). Verification of pulmonary vein isolation during single transseptal cryoballoon ablation: a comparison between the classical circular mapping catheter and the inner lumen mapping catheter. Europace, 14(12), 1708–1714.PubMedCrossRef Chierchia, G. B., Namdar, M., Sarkozy, A., et al. (2012). Verification of pulmonary vein isolation during single transseptal cryoballoon ablation: a comparison between the classical circular mapping catheter and the inner lumen mapping catheter. Europace, 14(12), 1708–1714.PubMedCrossRef
6.
go back to reference Neumann, T., Vogt, J., Schumacher, B., Dorszewski, A., Kuniss, M., Neuser, H., et al. (2008). Circumferential pulmonary vein isolation with the cryoballoon technique results from a prospective 3-center study. Journal of the American College of Cardiology, 52, 273–278.PubMedCrossRef Neumann, T., Vogt, J., Schumacher, B., Dorszewski, A., Kuniss, M., Neuser, H., et al. (2008). Circumferential pulmonary vein isolation with the cryoballoon technique results from a prospective 3-center study. Journal of the American College of Cardiology, 52, 273–278.PubMedCrossRef
7.
go back to reference Chun, K. R., Schmidt, B., Metzner, A., Tilz, R., Zerm, T., Köster, I., et al. (2009). The ‘single big cryoballoon’ technique for acute pulmonary vein isolation in patients with paroxysmal atrial fibrillation: a prospective observational single centre study. European Heart Journal, 30(6), 699–709.PubMedCentralPubMedCrossRef Chun, K. R., Schmidt, B., Metzner, A., Tilz, R., Zerm, T., Köster, I., et al. (2009). The ‘single big cryoballoon’ technique for acute pulmonary vein isolation in patients with paroxysmal atrial fibrillation: a prospective observational single centre study. European Heart Journal, 30(6), 699–709.PubMedCentralPubMedCrossRef
8.
go back to reference Metzner, A., Burchard, A., Wohlmuth, P., Rausch, P., Bardyszewski, A., Gienapp, C., Tilz, R. R., Rillig, A., Mathew, S., Deiss, S., Makimoto, H., Ouyang, F., Kuck, K. H., & Wissner, E. (2013). Increased incidence of esophageal thermal lesions using the second-generation 28 mm cryoballoon. Circ Arrhythm Electrophysiol. doi:10.1161/CIRCEP.113.000228. Metzner, A., Burchard, A., Wohlmuth, P., Rausch, P., Bardyszewski, A., Gienapp, C., Tilz, R. R., Rillig, A., Mathew, S., Deiss, S., Makimoto, H., Ouyang, F., Kuck, K. H., & Wissner, E. (2013). Increased incidence of esophageal thermal lesions using the second-generation 28 mm cryoballoon. Circ Arrhythm Electrophysiol. doi:10.​1161/​CIRCEP.​113.​000228.
9.
go back to reference Kühne, M., Knecht, S., Altmann, D., et al. (2013). Validation of a novel spiral mapping catheter for real-time recordings from the pulmonary veins during cryoballoon ablation of atrial fibrillation. Heart Rhythm, 10(2), 241–246.PubMedCrossRef Kühne, M., Knecht, S., Altmann, D., et al. (2013). Validation of a novel spiral mapping catheter for real-time recordings from the pulmonary veins during cryoballoon ablation of atrial fibrillation. Heart Rhythm, 10(2), 241–246.PubMedCrossRef
10.
go back to reference Ghosh, J., Sepahpour, A., Chan, K. H., Singarayar, S., & McGuire, M. A. (2013). Immediate balloon deflation for prevention of persistent phrenic nerve palsy during pulmonary vein isolation by balloon cryoablation. Heart Rhythm, 10(5), 646–652.PubMedCrossRef Ghosh, J., Sepahpour, A., Chan, K. H., Singarayar, S., & McGuire, M. A. (2013). Immediate balloon deflation for prevention of persistent phrenic nerve palsy during pulmonary vein isolation by balloon cryoablation. Heart Rhythm, 10(5), 646–652.PubMedCrossRef
11.
go back to reference Kühne, M., Knecht, S., Altmann, D., et al. (2013). Phrenic nerve palsy during ablation of atrial fibrillation using a 28-mm cryoballoon catheter: predictors and prevention. Journal of Interventional Cardiac Electrophysiology, 36(1), 47–54.PubMedCrossRef Kühne, M., Knecht, S., Altmann, D., et al. (2013). Phrenic nerve palsy during ablation of atrial fibrillation using a 28-mm cryoballoon catheter: predictors and prevention. Journal of Interventional Cardiac Electrophysiology, 36(1), 47–54.PubMedCrossRef
12.
go back to reference Franceschi, F., Dubuc, M., Guerra, P. G., & Khairy, P. (2011). Phrenic nerve monitoring with diaphragmatic electromyography during cryoballoon ablation for atrial fibrillation: the first human application. Heart Rhythm, 8(7), 1068–1071.PubMedCrossRef Franceschi, F., Dubuc, M., Guerra, P. G., & Khairy, P. (2011). Phrenic nerve monitoring with diaphragmatic electromyography during cryoballoon ablation for atrial fibrillation: the first human application. Heart Rhythm, 8(7), 1068–1071.PubMedCrossRef
13.
go back to reference Franceschi, F., Dubuc, M., Guerra, P. G., et al. (2011). Diaphragmatic electromyography during cryoballoon ablation: a novel concept in the prevention of phrenic nerve palsy. Heart Rhythm, 8(6), 885–891.PubMedCrossRef Franceschi, F., Dubuc, M., Guerra, P. G., et al. (2011). Diaphragmatic electromyography during cryoballoon ablation: a novel concept in the prevention of phrenic nerve palsy. Heart Rhythm, 8(6), 885–891.PubMedCrossRef
14.
go back to reference Fürnkranz, A., Bordignon, S., Schmidt, B., Böhmig, M., Böhmer, M. C., Bode, F., et al. (2013). Luminal esophageal temperature predicts esophageal lesions after second-generation cryoballoon pulmonary vein isolation. Heart Rhythm, 10(6), 789–793.PubMedCrossRef Fürnkranz, A., Bordignon, S., Schmidt, B., Böhmig, M., Böhmer, M. C., Bode, F., et al. (2013). Luminal esophageal temperature predicts esophageal lesions after second-generation cryoballoon pulmonary vein isolation. Heart Rhythm, 10(6), 789–793.PubMedCrossRef
15.
go back to reference Neumann, T., Wójcik, M., Berkowitsch, A., et al. (2013). Cryoballoon ablation of paroxysmal atrial fibrillation: 5-year outcome after single procedure and predictors of success. Europace, 15(8), 1143–9. doi:10.1093/europace/eut021.PubMedCrossRef Neumann, T., Wójcik, M., Berkowitsch, A., et al. (2013). Cryoballoon ablation of paroxysmal atrial fibrillation: 5-year outcome after single procedure and predictors of success. Europace, 15(8), 1143–9. doi:10.​1093/​europace/​eut021.PubMedCrossRef
16.
go back to reference Rao, J.Y., Chierchia, G.B., de Asmundis, C., et al. (2013). Cryoballoon ablation as index procedure for paroxysmal atrial fibrillation: long-term results from a single center early experience. J Cardiovasc Med (Hagerstown). 2013 Jun 10. doi:10.2459/JCM.0b013e3283623838. Rao, J.Y., Chierchia, G.B., de Asmundis, C., et al. (2013). Cryoballoon ablation as index procedure for paroxysmal atrial fibrillation: long-term results from a single center early experience. J Cardiovasc Med (Hagerstown). 2013 Jun 10. doi:10.​2459/​JCM.​0b013e3283623838​.
17.
go back to reference Vogt, J., Heintze, J., Gutleben, K. J., Muntean, B., Horstkotte, D., & Nölker, G. (2013). Long-term outcomes after cryoballoon pulmonary vein isolation: results from a prospective study in 605 patients. Journal of the American College of Cardiology, 61(16), 1707–1712.PubMedCrossRef Vogt, J., Heintze, J., Gutleben, K. J., Muntean, B., Horstkotte, D., & Nölker, G. (2013). Long-term outcomes after cryoballoon pulmonary vein isolation: results from a prospective study in 605 patients. Journal of the American College of Cardiology, 61(16), 1707–1712.PubMedCrossRef
18.
go back to reference Klein, G., Oswald, H., Gardiwal, A., et al. (2008). Efficacy of pulmonary vein isolation by cryoballoon ablation in patients with paroxysmal atrial fibrillation. Heart Rhythm, 5(6), 802–806.PubMedCrossRef Klein, G., Oswald, H., Gardiwal, A., et al. (2008). Efficacy of pulmonary vein isolation by cryoballoon ablation in patients with paroxysmal atrial fibrillation. Heart Rhythm, 5(6), 802–806.PubMedCrossRef
19.
go back to reference Chierchia, G. B., de Asmundis, C., Müller-Burri, S. A., et al. (2009). Early recovery of pulmonary vein conduction after cryoballoon ablation for paroxysmal atrial fibrillation: a prospective study. Europace, 11(4), 445–449.PubMedCrossRef Chierchia, G. B., de Asmundis, C., Müller-Burri, S. A., et al. (2009). Early recovery of pulmonary vein conduction after cryoballoon ablation for paroxysmal atrial fibrillation: a prospective study. Europace, 11(4), 445–449.PubMedCrossRef
Metadata
Title
Initial experience of three-minute freeze cycles using the second-generation cryoballoon ablation: acute and short-term procedural outcomes
Authors
Gian-Battista Chierchia
Giacomo Di Giovanni
Juan Sieira-Moret
Carlo de Asmundis
Giulio Conte
Moises Rodriguez-Mañero
Ruben Casado-Arroyo
Giannis Baltogiannis
Gaetano Paparella
Giuseppe Ciconte
Andrea Sarkozy
Pedro Brugada
Publication date
01-03-2014
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 2/2014
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-013-9855-x

Other articles of this Issue 2/2014

Journal of Interventional Cardiac Electrophysiology 2/2014 Go to the issue