Remote magnetic navigation system provides a superior catheter stability in acquisition of His bundle electrogram

Authors: Syed Muhammad Ali Tahir, G. Muqtada Chaudhry, Muhammad Ali Syed, Thomas Marchese, Gregory Kotler, Charles I. Haffajee, Michael V. Orlov

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 3/2008

Login to get access

Abstract

Introduction

Catheter stability is a key prerequisite for a successful EP procedure. Remote magnetic navigation system (RMNS) was recently introduced for the manipulation of EP catheters. It may provide an improved catheter stability but this has not been tested prospectively. We performed a prospective cross-over study which compares the amplitude and stability of the His-electrogram obtained by catheters using RMNS vs conventional manual catheter placement.

Methods

His-electrograms were continuously recorded for 3 min with a conventional His mapping catheter and magnetically tipped catheter capable of alignment in the direction of an externally controlled magnetic field (0.08–0.1 Tesla) in the same patient. First 100 beats of each recording were used for data analysis. The amplitudes of each set of His-electrograms were measured and compared in microvolts. Coefficient of variation of the measured His-signal amplitude was calculated for each beat. Miller’s test for the equality of the coefficients of variation and Levene’s test of homogeneity of variance were used for statistical analysis.

Results

Fourteen patients (6 males; aged 48 ± 25 years) were included in the study. His-electrograms were more stable and homogenous with RMNS than manual method in 10/14 patients. Pooled variance of all RMNS recordings was significantly less than pooled variance of manual recordings (p = 0.01).

Conclusion

His-electrogram recording using RMNS is more stable and homogeneous than traditional manual method in most patients. These data indicate a superior catheter stability with RMNS.
Literature
2.
go back to reference Faddis, M. N., Chen, J., Osborn, J., Talcott, M., Cain, M. E., & Lindsay, B. D. (2003). Magnetic guidance system for cardiac electrophysiology: A prospective trial of safety and efficacy in humans. Journal of the American College of Cardiology, 42, 1952–1958. DOI 10.1016/j.jacc.2003.07.023.PubMedCrossRef Faddis, M. N., Chen, J., Osborn, J., Talcott, M., Cain, M. E., & Lindsay, B. D. (2003). Magnetic guidance system for cardiac electrophysiology: A prospective trial of safety and efficacy in humans. Journal of the American College of Cardiology, 42, 1952–1958. DOI 10.​1016/​j.​jacc.​2003.​07.​023.PubMedCrossRef
4.
go back to reference Miller, E. G. M. J. K. (1977). Testing equality of two coefficients of variation, American Statistical Association. Proceedings of the Business and Economic Statistics Section Part 1, 278–283. Miller, E. G. M. J. K. (1977). Testing equality of two coefficients of variation, American Statistical Association. Proceedings of the Business and Economic Statistics Section Part 1, 278–283.
5.
go back to reference Levene, H. (1960). In I. Olkin, et al. (Eds.), Contributions to probability and statistics: Essays in honor of Harold Hotelling (pp. 278–292). CA, USA: Stanford University Press. Levene, H. (1960). In I. Olkin, et al. (Eds.), Contributions to probability and statistics: Essays in honor of Harold Hotelling (pp. 278–292). CA, USA: Stanford University Press.
6.
8.
go back to reference Kerzner, R., Sanchez, J. M., Osborn, J. L., et al. (2006). Radiofrequency ablation of atrioventricular nodal reentrant tachycardia using a novel magnetic guidance system compared with a conventional approach. Heart Rhythm, 3, 261–267. DIO 10.1016/j.hrthm.2005.11.019.CrossRef Kerzner, R., Sanchez, J. M., Osborn, J. L., et al. (2006). Radiofrequency ablation of atrioventricular nodal reentrant tachycardia using a novel magnetic guidance system compared with a conventional approach. Heart Rhythm, 3, 261–267. DIO 10.​1016/​j.​hrthm.​2005.​11.​019.CrossRef
9.
go back to reference Lindsay, B. D. T. G. F., Reddy Vivek, Y., Ruchir, S., Beckman Karen, J., Stanislav, W., Katherine, W., et al. (2006). Magnetic catheter navigation is safe and effective for ablation of supraventricular tachycardia: Results of the multi-center ATTRAC study. Heart Rhythm, 3(Supplement 1), S125–S126.CrossRef Lindsay, B. D. T. G. F., Reddy Vivek, Y., Ruchir, S., Beckman Karen, J., Stanislav, W., Katherine, W., et al. (2006). Magnetic catheter navigation is safe and effective for ablation of supraventricular tachycardia: Results of the multi-center ATTRAC study. Heart Rhythm, 3(Supplement 1), S125–S126.CrossRef
Metadata
Title
Remote magnetic navigation system provides a superior catheter stability in acquisition of His bundle electrogram
Authors
Syed Muhammad Ali Tahir
G. Muqtada Chaudhry
Muhammad Ali Syed
Thomas Marchese
Gregory Kotler
Charles I. Haffajee
Michael V. Orlov
Publication date
01-04-2008
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 3/2008
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-008-9239-9

Other articles of this Issue 3/2008

Journal of Interventional Cardiac Electrophysiology 3/2008 Go to the issue