01-12-2013 | Original Paper

Effect of kVp on image quality and accuracy in coronary CT angiography according to patient body size: a phantom study

Authors: Sang Min Lee, Whal Lee, Jin Wook Chung, Eun-Ah Park, Jae Hyung Park

Published in: The International Journal of Cardiovascular Imaging | Special Issue 2/2013

Login to get access

Abstract

The aim is to investigate the effect of tube voltage and chest wall thickness on image quality, stenosis measurement, and radiation dose in coronary CT angiography (CCTA) in a phantom study. A phantom with tubes in a box at its center and concentric cylindrical plastic chambers of three layers at its periphery was constructed. The concentric cylinders were filled with oil or left empty to simulate different degrees of obesity. Retrospective CT scanning was performed at different kVps and mAs. Image noise, contrast to noise ratio (CNR), stenosis measurement, and radiation dose were obtained. A CNR higher than 10 was considered to be acceptable for clinical practice. Mean image noise was 51.7 at 80 kVp, 31.6 at 100 kVp, and 24.7 at 120 kVp (P < 0.001). A CNR greater than 10 could be achieved with all the images using 80 kVp as well as using 100 or 120 kVp. However, CNRs at 100 and 120 kVp were significantly higher than the CNR at 80 kVp (P < 0.001). There were no significant differences between 100 and 120 kVp. All stenosis measurements were overestimated. Accuracy of stenosis measurement was significantly correlated with CNR (P < 0.05), but not with kVps. Mean doses were 2.07 mSv at 80 kVp, 3.37 mSv at 100 kVp, and 5.17 mSv at 120 kVp (P < 0.001). CNR per radiation dose was highest at 80 kVp, regardless of chest wall thickness. For CCTA, using 80 kVp with high mAs is the best choice, regardless of chest wall thickness, for minimal radiation dose and sufficient image quality.
Literature
1.
go back to reference Haberl R, Tittus J, Bohme E et al (2005) Multislice spiral computed tomographic angiography of coronary arteries in patients with suspected coronary artery disease: an effective filter before catheter angiography? Am Heart J 149(6):1112–1119PubMedCrossRef Haberl R, Tittus J, Bohme E et al (2005) Multislice spiral computed tomographic angiography of coronary arteries in patients with suspected coronary artery disease: an effective filter before catheter angiography? Am Heart J 149(6):1112–1119PubMedCrossRef
2.
go back to reference Janne d’Othee B, Siebert U, Cury R et al (2008) A systematic review on diagnostic accuracy of CT-based detection of significant coronary artery disease. Eur J Radiol 65(3):449–461PubMedCrossRef Janne d’Othee B, Siebert U, Cury R et al (2008) A systematic review on diagnostic accuracy of CT-based detection of significant coronary artery disease. Eur J Radiol 65(3):449–461PubMedCrossRef
3.
go back to reference Francone M, Napoli A, Carbone I et al (2007) Noninvasive imaging of the coronary arteries using a 64-row multidetector CT scanner: initial clinical experience and radiation dose concerns. Radiol Med 112(1):31–46PubMedCrossRef Francone M, Napoli A, Carbone I et al (2007) Noninvasive imaging of the coronary arteries using a 64-row multidetector CT scanner: initial clinical experience and radiation dose concerns. Radiol Med 112(1):31–46PubMedCrossRef
4.
go back to reference Leber AW, Knez A, Becker A et al (2004) Accuracy of multidetector spiral computed tomography in identifying and differentiating the composition of coronary atherosclerotic plaques: a comparative study with intracoronary ultrasound. J Am Coll Cardiol 43(7):1241–1247PubMedCrossRef Leber AW, Knez A, Becker A et al (2004) Accuracy of multidetector spiral computed tomography in identifying and differentiating the composition of coronary atherosclerotic plaques: a comparative study with intracoronary ultrasound. J Am Coll Cardiol 43(7):1241–1247PubMedCrossRef
5.
go back to reference Raff GL, Gallagher MJ, O’Neill WW et al (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46(3):552–557PubMedCrossRef Raff GL, Gallagher MJ, O’Neill WW et al (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46(3):552–557PubMedCrossRef
6.
go back to reference Mollet NR, Cademartiri F, van Mieghem CA et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112(15):2318–2323PubMedCrossRef Mollet NR, Cademartiri F, van Mieghem CA et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112(15):2318–2323PubMedCrossRef
7.
go back to reference Stolzmann P, Scheffel H, Schertler T et al (2008) Radiation dose estimates in dual-source computed tomography coronary angiography. Eur Radiol 18(3):592–599PubMedCrossRef Stolzmann P, Scheffel H, Schertler T et al (2008) Radiation dose estimates in dual-source computed tomography coronary angiography. Eur Radiol 18(3):592–599PubMedCrossRef
8.
go back to reference Achenbach S, Ropers U, Kuettner A et al (2008) Randomized comparison of 64-slice single- and dual-source computed tomography coronary angiography for the detection of coronary artery disease. JACC Cardiovasc Imaging 1(2):177–186PubMedCrossRef Achenbach S, Ropers U, Kuettner A et al (2008) Randomized comparison of 64-slice single- and dual-source computed tomography coronary angiography for the detection of coronary artery disease. JACC Cardiovasc Imaging 1(2):177–186PubMedCrossRef
9.
10.
go back to reference Coles DR, Smail MA, Negus IS et al (2006) Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol 47(9):1840–1845PubMedCrossRef Coles DR, Smail MA, Negus IS et al (2006) Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol 47(9):1840–1845PubMedCrossRef
11.
go back to reference Betsou S, Efstathopoulos EP, Katritsis D et al (1998) Patient radiation doses during cardiac catheterization procedures. Br J Radiol 71(846):634–639PubMed Betsou S, Efstathopoulos EP, Katritsis D et al (1998) Patient radiation doses during cardiac catheterization procedures. Br J Radiol 71(846):634–639PubMed
12.
go back to reference Broadhead DA, Chapple CL, Faulkner K et al (1997) The impact of cardiology on the collective effective dose in the North of England. Br J Radiol 70(833):492–497PubMed Broadhead DA, Chapple CL, Faulkner K et al (1997) The impact of cardiology on the collective effective dose in the North of England. Br J Radiol 70(833):492–497PubMed
13.
go back to reference Jakobs TF, Becker CR, Ohnesorge B et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12(5):1081–1086PubMedCrossRef Jakobs TF, Becker CR, Ohnesorge B et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12(5):1081–1086PubMedCrossRef
14.
go back to reference Abada HT, Larchez C, Daoud B et al (2006) MDCT of the coronary arteries: feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose. AJR Am J Roentgenol 186(6 Suppl 2):S387–S390PubMedCrossRef Abada HT, Larchez C, Daoud B et al (2006) MDCT of the coronary arteries: feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose. AJR Am J Roentgenol 186(6 Suppl 2):S387–S390PubMedCrossRef
15.
go back to reference Hausleiter J, Meyer T, Hadamitzky M et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113(10):1305–1310PubMedCrossRef Hausleiter J, Meyer T, Hadamitzky M et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113(10):1305–1310PubMedCrossRef
16.
go back to reference Hsieh J, Londt J, Vass M et al (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33(11):4236–4248PubMedCrossRef Hsieh J, Londt J, Vass M et al (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33(11):4236–4248PubMedCrossRef
17.
go back to reference Kalva SP, Sahani DV, Hahn PF et al (2006) Using the K-edge to improve contrast conspicuity and to lower radiation dose with a 16-MDCT: a phantom and human study. J Comput Assist Tomogr 30(3):391–397PubMedCrossRef Kalva SP, Sahani DV, Hahn PF et al (2006) Using the K-edge to improve contrast conspicuity and to lower radiation dose with a 16-MDCT: a phantom and human study. J Comput Assist Tomogr 30(3):391–397PubMedCrossRef
18.
go back to reference Stolzmann P, Leschka S, Scheffel H et al (2008) Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology 249(1):71–80PubMedCrossRef Stolzmann P, Leschka S, Scheffel H et al (2008) Dual-source CT in step-and-shoot mode: noninvasive coronary angiography with low radiation dose. Radiology 249(1):71–80PubMedCrossRef
19.
go back to reference Sigal-Cinqualbre AB, Hennequin R, Abada HT et al (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231(1):169–174PubMedCrossRef Sigal-Cinqualbre AB, Hennequin R, Abada HT et al (2004) Low-kilovoltage multi-detector row chest CT in adults: feasibility and effect on image quality and iodine dose. Radiology 231(1):169–174PubMedCrossRef
20.
go back to reference Baek JH, Lee W, Chang KH et al (2009) Optimization of the scan protocol for the reduction of diaphragmatic motion artifacts depicted on CT angiography: a phantom study simulating pediatric patients with free breathing. Korean J Radiol 10(3):260–268PubMedCentralPubMedCrossRef Baek JH, Lee W, Chang KH et al (2009) Optimization of the scan protocol for the reduction of diaphragmatic motion artifacts depicted on CT angiography: a phantom study simulating pediatric patients with free breathing. Korean J Radiol 10(3):260–268PubMedCentralPubMedCrossRef
21.
go back to reference Suzuki S, Furui S, Kaminaga T et al (2004) Measurement of vascular diameter in vitro by automated software for CT angiography: effects of inner diameter, density of contrast medium, and convolution kernel. AJR Am J Roentgenol 182(5):1313–1317PubMedCrossRef Suzuki S, Furui S, Kaminaga T et al (2004) Measurement of vascular diameter in vitro by automated software for CT angiography: effects of inner diameter, density of contrast medium, and convolution kernel. AJR Am J Roentgenol 182(5):1313–1317PubMedCrossRef
22.
go back to reference Suzuki S, Furui S, Kaminaga T (2005) Accuracy of automated CT angiography measurement of vascular diameter in phantoms: effect of size of display field of view, density of contrast medium, and wall thickness. AJR Am J Roentgenol 184(6):1940–1944PubMedCrossRef Suzuki S, Furui S, Kaminaga T (2005) Accuracy of automated CT angiography measurement of vascular diameter in phantoms: effect of size of display field of view, density of contrast medium, and wall thickness. AJR Am J Roentgenol 184(6):1940–1944PubMedCrossRef
23.
go back to reference WHO expert consultation (2004) Appropriate body-mass index for Asian populations and its implications for policy and intervention strategies. Lancet 363(9403):157–163 WHO expert consultation (2004) Appropriate body-mass index for Asian populations and its implications for policy and intervention strategies. Lancet 363(9403):157–163
24.
go back to reference Jee SH, Sull JW, Park J et al (2006) Body-mass index and mortality in Korean men and women. N Engl J Med 355(8):779–787PubMedCrossRef Jee SH, Sull JW, Park J et al (2006) Body-mass index and mortality in Korean men and women. N Engl J Med 355(8):779–787PubMedCrossRef
25.
go back to reference Morin RL (1988) Monte Carlo simulation in the radiological sciences. CRC Press, Boca Raton, Fl Morin RL (1988) Monte Carlo simulation in the radiological sciences. CRC Press, Boca Raton, Fl
26.
go back to reference Nishino TK, Wu X, Johnson RF Jr (2005) Thickness of molybdenum filter and squared contrast-to-noise ratio per dose for digital mammography. AJR Am J Roentgenol 185(4):960–963PubMedCrossRef Nishino TK, Wu X, Johnson RF Jr (2005) Thickness of molybdenum filter and squared contrast-to-noise ratio per dose for digital mammography. AJR Am J Roentgenol 185(4):960–963PubMedCrossRef
27.
go back to reference Siegel MJ, Schmidt B, Bradley D et al (2004) Radiation dose and image quality in pediatric CT: effect of technical factors and phantom size and shape. Radiology 233(2):515–522PubMedCrossRef Siegel MJ, Schmidt B, Bradley D et al (2004) Radiation dose and image quality in pediatric CT: effect of technical factors and phantom size and shape. Radiology 233(2):515–522PubMedCrossRef
28.
go back to reference Park EA, Lee W, Kang JH et al (2009) The image quality and radiation dose of 100-kVp versus 120-kVp ECG-gated 16-slice CT coronary angiography. Korean J Radiol 10(3):235–243PubMedCentralPubMedCrossRef Park EA, Lee W, Kang JH et al (2009) The image quality and radiation dose of 100-kVp versus 120-kVp ECG-gated 16-slice CT coronary angiography. Korean J Radiol 10(3):235–243PubMedCentralPubMedCrossRef
29.
go back to reference Britten AJ, Crotty M, Kiremidjian H et al (2004) The addition of computer simulated noise to investigate radiation dose and image quality in images with spatial correlation of statistical noise: an example application to X-ray CT of the brain. Br J Radiol 77(916):323–328PubMedCrossRef Britten AJ, Crotty M, Kiremidjian H et al (2004) The addition of computer simulated noise to investigate radiation dose and image quality in images with spatial correlation of statistical noise: an example application to X-ray CT of the brain. Br J Radiol 77(916):323–328PubMedCrossRef
Metadata
Title
Effect of kVp on image quality and accuracy in coronary CT angiography according to patient body size: a phantom study
Authors
Sang Min Lee
Whal Lee
Jin Wook Chung
Eun-Ah Park
Jae Hyung Park
Publication date
01-12-2013
Publisher
Springer Netherlands
Published in
The International Journal of Cardiovascular Imaging / Issue Special Issue 2/2013
Print ISSN: 1569-5794
Electronic ISSN: 1875-8312
DOI
https://doi.org/10.1007/s10554-013-0298-3

Other articles of this Special Issue 2/2013

The International Journal of Cardiovascular Imaging 2/2013 Go to the issue