01-03-2008 | Cardiac

Radiation dose estimates in dual-source computed tomography coronary angiography

Authors: Paul Stolzmann, Hans Scheffel, Thomas Schertler, Thomas Frauenfelder, Sebastian Leschka, Lars Husmann, Thomas G. Flohr, Borut Marincek, Philipp A. Kaufmann, Hatem Alkadhi

Published in: European Radiology | Issue 3/2008

Login to get access

Abstract

The purpose of this study was to quantify radiation dose parameters of dual-source CT coronary angiography. Eighty patients underwent contrast-enhanced, retrospectively ECG-gated dual-source CT coronary angiography with heart rate-adapted ECG pulsing using two algorithms: In 40 patients, the tube current was reduced to 20% (Amin1) of the normal tube current (Amax) outside the pulsing window; in 40 patients tube current was reduced to 4% (Amin2) of Amax. Mean CTDIvol in the Amin1 group was 45.1 ± 3.6 mGy; the mean CTDIvol in the Amin2 group was 39.1 ± 3.2 mGy, with CTDIvol in the Amin2 group being significantly reduced when compared to the Amin1 group (P < 0.001). A significant negative correlation was found between CTDIvol and heart rate in group Amin1 (r = −0.82, P < 0.001), whereas no correlation was found between CTDIvol and heart rate in group Amin2 (r = −0.066). Using the conversion coefficient for the chest, dual-source CT coronary angiography resulted in an estimated mean effective dose of 8.8 mSv in the Amin1 group and 7.8 mSv in the Amin2. Radiation exposure of dual-source CT coronary angiography using an ECG-pulsing protocol reducing the tube current to 20% significantly decreases with increasing heart rates, despite using wider pulsing windows at higher heart rates. When using a protocol with reduced tube current of 4%, the radiation dose is significantly lower, irrespective of the heart rate.
Literature
1.
go back to reference Leber AW et al (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef Leber AW et al (2005) Quantification of obstructive and nonobstructive coronary lesions by 64-slice computed tomography: a comparative study with quantitative coronary angiography and intravascular ultrasound. J Am Coll Cardiol 46:147–154PubMedCrossRef
2.
go back to reference Leschka S et al (2005) Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J 26:1482–1487PubMedCrossRef Leschka S et al (2005) Accuracy of MSCT coronary angiography with 64-slice technology: first experience. Eur Heart J 26:1482–1487PubMedCrossRef
3.
go back to reference Mollet NR et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323PubMedCrossRef Mollet NR et al (2005) High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation 112:2318–2323PubMedCrossRef
4.
go back to reference Raff GL, Gallagher MJ, O’Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557PubMedCrossRef Raff GL, Gallagher MJ, O’Neill WW, Goldstein JA (2005) Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol 46:552–557PubMedCrossRef
5.
go back to reference Flohr TG et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268PubMedCrossRef Flohr TG et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268PubMedCrossRef
6.
go back to reference Achenbach S et al (2006) Contrast-enhanced coronary artery visualization by dual-source computed tomography–initial experience. Eur J Radiol 57:331–335PubMedCrossRef Achenbach S et al (2006) Contrast-enhanced coronary artery visualization by dual-source computed tomography–initial experience. Eur J Radiol 57:331–335PubMedCrossRef
7.
go back to reference Johnson TR et al (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415PubMedCrossRef Johnson TR et al (2006) Dual-source CT cardiac imaging: initial experience. Eur Radiol 16:1409–1415PubMedCrossRef
8.
go back to reference Scheffel H et al (2006) Accuracy of dual-source CT coronary angiography: First experience in a high pre-test probability population without heart rate control. Eur Radiol 16:2739–2747PubMedCrossRef Scheffel H et al (2006) Accuracy of dual-source CT coronary angiography: First experience in a high pre-test probability population without heart rate control. Eur Radiol 16:2739–2747PubMedCrossRef
9.
go back to reference Yates SJ, Pike LC, Goldstone KE (2004) Effect of multislice scanners on patient dose from routine CT examinations in East Anglia. Br J Radiol 77:472–478PubMedCrossRef Yates SJ, Pike LC, Goldstone KE (2004) Effect of multislice scanners on patient dose from routine CT examinations in East Anglia. Br J Radiol 77:472–478PubMedCrossRef
10.
go back to reference Gerber TC, Kuzo RS, Morin RL (2005) Techniques and parameters for estimating radiation exposure and dose in cardiac computed tomography. Int J Cardiovasc Imaging 21:165–176PubMedCrossRef Gerber TC, Kuzo RS, Morin RL (2005) Techniques and parameters for estimating radiation exposure and dose in cardiac computed tomography. Int J Cardiovasc Imaging 21:165–176PubMedCrossRef
11.
go back to reference Primak AN, McCollough CH, Bruesewitz MR, Zhang J, Fletcher JG (2006) Relationship between noise, dose, and pitch in cardiac multi-detector row CT. Radiographics 26:1785–1794PubMedCrossRef Primak AN, McCollough CH, Bruesewitz MR, Zhang J, Fletcher JG (2006) Relationship between noise, dose, and pitch in cardiac multi-detector row CT. Radiographics 26:1785–1794PubMedCrossRef
12.
go back to reference Jakobs TF et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12:1081–1086PubMedCrossRef Jakobs TF et al (2002) Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol 12:1081–1086PubMedCrossRef
13.
go back to reference McCollough CH et al (2007) Dose performance of a 64-channel dual-source CT scanner. Radiology 243:775–784PubMedCrossRef McCollough CH et al (2007) Dose performance of a 64-channel dual-source CT scanner. Radiology 243:775–784PubMedCrossRef
14.
go back to reference Leschka S et al (2007) Image Quality and Reconstruction Intervals of Dual-Source CT Coronary Angiography: Recommendations for ECG-Pulsing Windowing. Invest Radiol 42:543–549PubMedCrossRef Leschka S et al (2007) Image Quality and Reconstruction Intervals of Dual-Source CT Coronary Angiography: Recommendations for ECG-Pulsing Windowing. Invest Radiol 42:543–549PubMedCrossRef
15.
go back to reference Menzel H, Schibilla H, Teunen D, eds. (2000) European guidelines on quality criteria for computed tomography. Luxembourg: European Commission Publication No. EUR 16262 EN Menzel H, Schibilla H, Teunen D, eds. (2000) European guidelines on quality criteria for computed tomography. Luxembourg: European Commission Publication No. EUR 16262 EN
16.
go back to reference Hausleiter J et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113:1305–1310PubMedCrossRef Hausleiter J et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113:1305–1310PubMedCrossRef
17.
go back to reference Morin RL (1988) Monte carlo simulation in the radiological sciences. CRC Press, Boca Raton, FL Morin RL (1988) Monte carlo simulation in the radiological sciences. CRC Press, Boca Raton, FL
18.
go back to reference Austen WG et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51:5–40PubMed Austen WG et al (1975) A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association. Circulation 51:5–40PubMed
21.
go back to reference Schardt P et al (2004) New x-ray tube performance in computed tomography by introducing the rotating envelope tube technology. Med Phys 31:2699–2706PubMedCrossRef Schardt P et al (2004) New x-ray tube performance in computed tomography by introducing the rotating envelope tube technology. Med Phys 31:2699–2706PubMedCrossRef
22.
go back to reference Wierzbicki M, Guiraudon GM, Jones DL, Peters T (2007) Dose reduction for cardiac CT using a registration-based approach. Med Phys 34:1884–1895PubMedCrossRef Wierzbicki M, Guiraudon GM, Jones DL, Peters T (2007) Dose reduction for cardiac CT using a registration-based approach. Med Phys 34:1884–1895PubMedCrossRef
23.
go back to reference Kuon E, Robinson DM, Empen K, Dahm JB (2005) [Fluoroscopy time – an overestimated factor for patient radiation exposure in invasive cardiology]. Rofo 177:812–817PubMed Kuon E, Robinson DM, Empen K, Dahm JB (2005) [Fluoroscopy time – an overestimated factor for patient radiation exposure in invasive cardiology]. Rofo 177:812–817PubMed
24.
go back to reference Coles DR et al (2006) Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol 47:1840–1845PubMedCrossRef Coles DR et al (2006) Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol 47:1840–1845PubMedCrossRef
25.
go back to reference Aroua A, Trueb P, Vader JP, Valley JF, Verdun FR (2007) Exposure of the Swiss population by radiodiagnostics: 2003 review. Health Phys 92:442–448PubMedCrossRef Aroua A, Trueb P, Vader JP, Valley JF, Verdun FR (2007) Exposure of the Swiss population by radiodiagnostics: 2003 review. Health Phys 92:442–448PubMedCrossRef
26.
go back to reference Zanzonico P, Rothenberg LN, Strauss HW (2006) Radiation exposure of computed tomography and direct intracoronary angiography: risk has its reward. J Am Coll Cardiol 47:1846–1849PubMedCrossRef Zanzonico P, Rothenberg LN, Strauss HW (2006) Radiation exposure of computed tomography and direct intracoronary angiography: risk has its reward. J Am Coll Cardiol 47:1846–1849PubMedCrossRef
27.
go back to reference Becker C et al (1999) [Assessment of the effective dose for routine protocols in conventional CT, electron beam CT and coronary angiography]. Rofo 170:99–104PubMed Becker C et al (1999) [Assessment of the effective dose for routine protocols in conventional CT, electron beam CT and coronary angiography]. Rofo 170:99–104PubMed
28.
go back to reference Das M et al (2005) Individually adapted examination protocols for reduction of radiation exposure for 16-MDCT chest examinations. AJR Am J Roentgenol 184:1437–1443PubMed Das M et al (2005) Individually adapted examination protocols for reduction of radiation exposure for 16-MDCT chest examinations. AJR Am J Roentgenol 184:1437–1443PubMed
29.
go back to reference Nyman U, Ahl TL, Kristiansson M, Nilsson L, Wettemark S (2005) Patient-circumference-adapted dose regulation in body computed tomography. A practical and flexible formula. Acta Radiol 46:396–406PubMedCrossRef Nyman U, Ahl TL, Kristiansson M, Nilsson L, Wettemark S (2005) Patient-circumference-adapted dose regulation in body computed tomography. A practical and flexible formula. Acta Radiol 46:396–406PubMedCrossRef
Metadata
Title
Radiation dose estimates in dual-source computed tomography coronary angiography
Authors
Paul Stolzmann
Hans Scheffel
Thomas Schertler
Thomas Frauenfelder
Sebastian Leschka
Lars Husmann
Thomas G. Flohr
Borut Marincek
Philipp A. Kaufmann
Hatem Alkadhi
Publication date
01-03-2008
Publisher
Springer-Verlag
Published in
European Radiology / Issue 3/2008
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-007-0786-8

Other articles of this Issue 3/2008

European Radiology 3/2008 Go to the issue