Light transmittance and micro-mechanical properties of bulk fill vs. conventional resin based composites

Authors: Stefan Bucuta, Nicoleta Ilie

Published in: Clinical Oral Investigations | Issue 8/2014

Login to get access

Abstract

Objectives

The aim of this study was to quantify the blue light that passes through different incremental thicknesses of bulk fill in comparison to conventional resin-based composites (RBCs) and to relate it to the induced mechanical properties.

Materials and methods

Seven bulk fill, five nanohybrid and two flowable RBCs were analysed. Specimens (n = 5) of three incremental thicknesses (2, 4 and 6 mm) were cured from the top for 20 s, while at the bottom, a spectrometer monitored in real time the transmitted irradiance. Micro-mechanical properties (Vickers hardness, HV, and indentation modulus, E) were measured at the top and bottom after 24 h of storage in distilled water at 37 °C. Electron microscope images were taken for assessing the filler distribution and size.

Results

Bulk fill RBCs (except SonicFill) were more translucent than conventional RBCs. Low-viscosity bulk fill materials showed the lowest mechanical properties. HV depends highly on the following parameters: material (ηp 2 = 0.952), incremental thickness (0.826), filler volume (0.747), filler weight (0.746) and transmitted irradiance (0.491). The bottom-to-top HV ratio (HVbt) was higher than 80 % in all materials in 2- and 4-mm increments (except for Premise), whereas in 6-mm increments, this is valid only in four bulk fill materials (Venus Bulk Fill, SDR, x-tra fil, Tetric EvoCeram Bulk Fill).

Conclusions

The depth of cure is dependent on the RBC’s translucency. Low-viscosity bulk fill RBCs have lower mechanical properties than all other types of analysed materials. All bulk fill RBCs (except SonicFill) are more translucent for blue light than conventional RBCs.

Significance

Although bulk fill RBCs are generally more translucent, the practitioner has to follow the manufacturer’s recommendations on curing technique and maximum incremental thickness.
Literature
3.
go back to reference Rueggeberg F (1999) Contemporary issues in photocuring. Compend Contin Educ Dent Suppl (Jamesburg, NJ: 1995) (25):S4–15; quiz S73. Rueggeberg F (1999) Contemporary issues in photocuring. Compend Contin Educ Dent Suppl (Jamesburg, NJ: 1995) (25):S4–15; quiz S73.
4.
go back to reference Radhika M, Sajjan GS, Kumara Swami BN, Mittal N (2010) Effect of different placement techniques on marginal microleakage of deep class-II cavities restored with two composite resin formulations. J Conserv Dent 13(1):9–15. doi:10.4103/0972-0707.62633 CrossRef Radhika M, Sajjan GS, Kumara Swami BN, Mittal N (2010) Effect of different placement techniques on marginal microleakage of deep class-II cavities restored with two composite resin formulations. J Conserv Dent 13(1):9–15. doi:10.​4103/​0972-0707.​62633 CrossRef
14.
19.
go back to reference Caldas DB, de Almeida JB, Correr-Sobrinho L, Sinhoreti MA, Consani S (2003) Influence of curing tip distance on resin composite Knoop hardness number, using three different light curing units. Oper Dent 28(3):315–320PubMed Caldas DB, de Almeida JB, Correr-Sobrinho L, Sinhoreti MA, Consani S (2003) Influence of curing tip distance on resin composite Knoop hardness number, using three different light curing units. Oper Dent 28(3):315–320PubMed
20.
go back to reference Yearn JA (1985) Factors affecting cure of visible light activated composites. Int Dent J 35(3):218–225PubMed Yearn JA (1985) Factors affecting cure of visible light activated composites. Int Dent J 35(3):218–225PubMed
24.
go back to reference Ilie N, Hickel R (2007) Quality of curing in relation to hardness, degree of cure and polymerization depth measured on a nano-hybrid composite. Am J Dent 20(4):263–268PubMed Ilie N, Hickel R (2007) Quality of curing in relation to hardness, degree of cure and polymerization depth measured on a nano-hybrid composite. Am J Dent 20(4):263–268PubMed
28.
go back to reference Bouschlicher MR, Rueggeberg FA, Wilson BM (2004) Correlation of bottom-to-top surface microhardness and conversion ratios for a variety of resin composite compositions. Oper Dent 29(6):698–704PubMed Bouschlicher MR, Rueggeberg FA, Wilson BM (2004) Correlation of bottom-to-top surface microhardness and conversion ratios for a variety of resin composite compositions. Oper Dent 29(6):698–704PubMed
29.
go back to reference ISO (2013) Dentistry—polymer-based restorative materials. ISO 4049:2009 ISO (2013) Dentistry—polymer-based restorative materials. ISO 4049:2009
31.
go back to reference ISO (2013) Metallic materials—instrumented indentation test for hardness and materials parameters—part 1: test method. ISO 14577–1:2002 ISO (2013) Metallic materials—instrumented indentation test for hardness and materials parameters—part 1: test method. ISO 14577–1:2002
33.
go back to reference Arikawa H, Kanie T, Fujii K, Takahashi H, Ban S (2007) Effect of filler properties in composite resins on light transmittance characteristics and color. Dent Mater J 26(1):38–44PubMedCrossRef Arikawa H, Kanie T, Fujii K, Takahashi H, Ban S (2007) Effect of filler properties in composite resins on light transmittance characteristics and color. Dent Mater J 26(1):38–44PubMedCrossRef
41.
go back to reference Braem M, Finger W, Van Doren VE, Lambrechts P, Vanherle G (1989) Mechanical properties and filler fraction of dental composites. Dent Mater 5(5):346–348PubMedCrossRef Braem M, Finger W, Van Doren VE, Lambrechts P, Vanherle G (1989) Mechanical properties and filler fraction of dental composites. Dent Mater 5(5):346–348PubMedCrossRef
Metadata
Title
Light transmittance and micro-mechanical properties of bulk fill vs. conventional resin based composites
Authors
Stefan Bucuta
Nicoleta Ilie
Publication date
01-11-2014
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 8/2014
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-013-1177-y

Other articles of this Issue 8/2014

Clinical Oral Investigations 8/2014 Go to the issue