Published in: Clinical Oral Investigations 2/2014

01-03-2014 | Original Article

Marginal and internal fit of four-unit zirconia fixed dental prostheses based on digital and conventional impression techniques

Authors: Júnio S. Almeida e Silva, Kurt Erdelt, Daniel Edelhoff, Élito Araújo, Michael Stimmelmayr, Luiz Clovis Cardoso Vieira, Jan-Frederik Güth

Published in: Clinical Oral Investigations | Issue 2/2014

Login to get access

Abstract

Objectives

The aim of this in vitro study was to evaluate the marginal and internal fit of CAD/CAM-generated four-unit zirconia fixed dental prostheses made with digital and conventional impressions.

Materials and method

A titanium master model was used. For group conventional impression (CI), 12 polyether impressions of the master model with ImpregumTM were made. For group digital impression (DI), 12 digital impressions of the master model using LavaTM C.O.S. system were made. The replica technique was applied. The Mann–Whitney U statistical test was applied to detect statistical differences between the groups, in terms of marginal and internal fit. Face-by-face comparisons between groups were also carried out.

Results

Groups DI and CI presented mean marginal fit of 63.96 and 65.33 μm, respectively, and showed no statistically significant difference. Groups DI and CI presented significantly different internal fit with mean values of 58.46 and 65.94 μm, respectively. Group DI showed statistically significantly lower values for marginal and internal fit on premolar mesial face, and on molar distal and palatal faces.

Conclusions

Frameworks fabricated from digital and conventional impressions showed clinically acceptable marginal fit. Frameworks fabricated from digital impression demonstrated better internal fit than ones fabricated from conventional impression. Reviewing each retainer face, digital impression showed better marginal and internal fit at the premolar mesial and molar distal faces.

Clinical relevance

The results of this in vitro study show that digital impressions made with the LavaTM C.O.S. system and its digital workflow are suitable for fabricating four-unit zirconia frameworks, with regard to marginal and internal fit requirements.
Literature
1.
go back to reference Christensen GJ (2008) Will digital impressions eliminate the current problems with conventional impressions? J Am Dent Assoc 139(6):761–763PubMedCrossRef Christensen GJ (2008) Will digital impressions eliminate the current problems with conventional impressions? J Am Dent Assoc 139(6):761–763PubMedCrossRef
2.
go back to reference Cancy JM, Scandrett FR, Ettinger RL (1983) Long-term dimensional stability of three current elastomers. J Oral Rehabil 10(4):325–333CrossRef Cancy JM, Scandrett FR, Ettinger RL (1983) Long-term dimensional stability of three current elastomers. J Oral Rehabil 10(4):325–333CrossRef
3.
go back to reference Endo T, Finger WJ (2006) Dimensional accuracy of a new polyether impression material. Quintessence Int 37(1):47–51PubMed Endo T, Finger WJ (2006) Dimensional accuracy of a new polyether impression material. Quintessence Int 37(1):47–51PubMed
4.
go back to reference Shetty P, Rodrigues S (2006) Accuracy of elastomeric impression materials on repeated pours. J Indian Prosthodont Soc 6(2):68–71CrossRef Shetty P, Rodrigues S (2006) Accuracy of elastomeric impression materials on repeated pours. J Indian Prosthodont Soc 6(2):68–71CrossRef
5.
go back to reference Hondrum SO (2001) Changes in properties of non-aqueous elastomeric impression materials after storage of components. J Prosthet Dent 85(1):73–81PubMedCrossRef Hondrum SO (2001) Changes in properties of non-aqueous elastomeric impression materials after storage of components. J Prosthet Dent 85(1):73–81PubMedCrossRef
6.
go back to reference Thongthammachat S, Moore BK, Barco MT 2nd, Hovijitra S, Brown DT, Andres CJ (2002) Dimensional accuracy of dental casts: influence of tray material, impression material, and time. J Prosthodont 11(2):98–108PubMedCrossRef Thongthammachat S, Moore BK, Barco MT 2nd, Hovijitra S, Brown DT, Andres CJ (2002) Dimensional accuracy of dental casts: influence of tray material, impression material, and time. J Prosthodont 11(2):98–108PubMedCrossRef
7.
go back to reference State of the industry (2000) Lab Management Today; 2000; 9–15 State of the industry (2000) Lab Management Today; 2000; 9–15
8.
go back to reference Christensen GJ (2005) The state of fixed prosthodontic impressions. Room for improvement. J Am Dent Assoc 136:343–346PubMedCrossRef Christensen GJ (2005) The state of fixed prosthodontic impressions. Room for improvement. J Am Dent Assoc 136:343–346PubMedCrossRef
9.
10.
11.
go back to reference Touchstone A, Nieting T, Ulmer N (2010) Digital transition: the collaboration between dentists and laboratory technicians on CAD/CAM restorations. J Am Dent Assoc 141(6 suppl):15S–19SPubMedCrossRef Touchstone A, Nieting T, Ulmer N (2010) Digital transition: the collaboration between dentists and laboratory technicians on CAD/CAM restorations. J Am Dent Assoc 141(6 suppl):15S–19SPubMedCrossRef
12.
go back to reference Mehl A, Hickel R (1999) Current state of development and perspectives of machine based production methods for dental restorations. Int J Comput Dent 2(1):9–35PubMed Mehl A, Hickel R (1999) Current state of development and perspectives of machine based production methods for dental restorations. Int J Comput Dent 2(1):9–35PubMed
13.
go back to reference Beuer F, Schweiger J, Edelhoff D (2008) Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. Br Dent J 204(9):505–511PubMedCrossRef Beuer F, Schweiger J, Edelhoff D (2008) Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. Br Dent J 204(9):505–511PubMedCrossRef
14.
go back to reference Al-Bakri IA, Hussey D, Al-Omari WM (2007) The dimensional accuracy of four impression techniques with the use of addition silicone impression materials. J Clin Dent 18(2):29–33PubMed Al-Bakri IA, Hussey D, Al-Omari WM (2007) The dimensional accuracy of four impression techniques with the use of addition silicone impression materials. J Clin Dent 18(2):29–33PubMed
15.
go back to reference Beuer F, Naumann M, Gernet W (2009) Precision of fit: zirconia three-unit fixed dental prostheses. Clin Oral Invest 13:343–349CrossRef Beuer F, Naumann M, Gernet W (2009) Precision of fit: zirconia three-unit fixed dental prostheses. Clin Oral Invest 13:343–349CrossRef
16.
go back to reference Jablonski S (1993) Illustrated dictionary of dentistry, 1982. J Dent 21:265–273CrossRef Jablonski S (1993) Illustrated dictionary of dentistry, 1982. J Dent 21:265–273CrossRef
17.
go back to reference Fransson B, Oilo G, Gjeitanger R (1985) The fit of metal-ceramic crowns, a clinical study. Dent Mater 1:197–199PubMedCrossRef Fransson B, Oilo G, Gjeitanger R (1985) The fit of metal-ceramic crowns, a clinical study. Dent Mater 1:197–199PubMedCrossRef
18.
go back to reference Karlsson S (1993) The fit of Procera titanium crowns. An in vitro and clinical study. Acta Odontol Scand 51:129–134PubMedCrossRef Karlsson S (1993) The fit of Procera titanium crowns. An in vitro and clinical study. Acta Odontol Scand 51:129–134PubMedCrossRef
19.
go back to reference McLean JW, von Fraunhofer JA (1971) The estimation of cement film thickness by an in vivo technique. Br Dent J 131:107–111PubMedCrossRef McLean JW, von Fraunhofer JA (1971) The estimation of cement film thickness by an in vivo technique. Br Dent J 131:107–111PubMedCrossRef
20.
go back to reference Belser UC, MacEntee MI, Richter WA (1985) Fit of three porcelain-fused-to-metal marginal designs in vivo: a scanning electron microscope study. J Prosthet Dent 53:24–29PubMedCrossRef Belser UC, MacEntee MI, Richter WA (1985) Fit of three porcelain-fused-to-metal marginal designs in vivo: a scanning electron microscope study. J Prosthet Dent 53:24–29PubMedCrossRef
21.
go back to reference Sulaiman F, Chai J, Jameson LM, Wozniak WT (1997) A comparison of the marginal fit of In-Ceram, IPS Empress and Procera crowns. Int J Prosthodont 10:478–484PubMed Sulaiman F, Chai J, Jameson LM, Wozniak WT (1997) A comparison of the marginal fit of In-Ceram, IPS Empress and Procera crowns. Int J Prosthodont 10:478–484PubMed
22.
go back to reference Beschinidt SM, Strub JR (1999) Evaluation of the marginal accuracy of different all-ceramic crown systems after simulation in the artificial mouth. J Oral Rehab 26:582–593CrossRef Beschinidt SM, Strub JR (1999) Evaluation of the marginal accuracy of different all-ceramic crown systems after simulation in the artificial mouth. J Oral Rehab 26:582–593CrossRef
23.
go back to reference Janenko C, Smales RJ (1979) Anterior crowns and gingival health. Australi Dent J 24:225–230CrossRef Janenko C, Smales RJ (1979) Anterior crowns and gingival health. Australi Dent J 24:225–230CrossRef
24.
go back to reference Bader JD, Rozier RG, McFall WT Jr, Ramsey DL (1991) Effect of crowns margins on periodontal conditions in regularly attending patients. J Prosthet Dent 65:75–79PubMedCrossRef Bader JD, Rozier RG, McFall WT Jr, Ramsey DL (1991) Effect of crowns margins on periodontal conditions in regularly attending patients. J Prosthet Dent 65:75–79PubMedCrossRef
25.
go back to reference Kokubo Y, Ohkubo C, Tsumita M, Miyashita A, Vult Von Steyern O, Fukushima S (2005) Clinical marginal and internal gaps of Procera All-Ceram crowns. J Oral Rehab 32:526–530CrossRef Kokubo Y, Ohkubo C, Tsumita M, Miyashita A, Vult Von Steyern O, Fukushima S (2005) Clinical marginal and internal gaps of Procera All-Ceram crowns. J Oral Rehab 32:526–530CrossRef
26.
go back to reference Reich S, Wichmann M, Nkenke E, Proeschel P (2005) Clinical fit of all-ceramic three-unit fixed partial dentures, generated with three different CAD/CAM systems. Eur J Oral Sci 113:174–179PubMedCrossRef Reich S, Wichmann M, Nkenke E, Proeschel P (2005) Clinical fit of all-ceramic three-unit fixed partial dentures, generated with three different CAD/CAM systems. Eur J Oral Sci 113:174–179PubMedCrossRef
27.
go back to reference Reich S, Kappe K, Teschner H, Schimitt J (2008) Clinical fit of four-unit zirconia posterior fixed dental prostheses. Eur J Oral Sci 116:579–584PubMedCrossRef Reich S, Kappe K, Teschner H, Schimitt J (2008) Clinical fit of four-unit zirconia posterior fixed dental prostheses. Eur J Oral Sci 116:579–584PubMedCrossRef
28.
go back to reference Kohorst P, Brinkmann H, Li J, Borchers L, Stiesch M (2009) Marginal accuracy of four-unit zirconia fixed dental prostheses fabricated using different computer-aided manufacturing systems. Eur J Oral Sci 117:319–325PubMedCrossRef Kohorst P, Brinkmann H, Li J, Borchers L, Stiesch M (2009) Marginal accuracy of four-unit zirconia fixed dental prostheses fabricated using different computer-aided manufacturing systems. Eur J Oral Sci 117:319–325PubMedCrossRef
29.
go back to reference Rekow D, Thompson VP (2005) Near-surface damage—a persistent problem in crowns obtained by computer-aided design and manufacturing. Proc Inst Mech Eng 219:233–243CrossRef Rekow D, Thompson VP (2005) Near-surface damage—a persistent problem in crowns obtained by computer-aided design and manufacturing. Proc Inst Mech Eng 219:233–243CrossRef
31.
go back to reference Syrek A, Reich G, Ranftl D, Klein C, Cerny B, Brodesser J (2010) Clinical evaluation of all-ceramic crowns fabricated from digital impressions based on the principle of active wavefront sampling. J Dent 38:553–559PubMedCrossRef Syrek A, Reich G, Ranftl D, Klein C, Cerny B, Brodesser J (2010) Clinical evaluation of all-ceramic crowns fabricated from digital impressions based on the principle of active wavefront sampling. J Dent 38:553–559PubMedCrossRef
32.
go back to reference Kunii J, Hotta Y, Tamaki Y, Ozawa A, Kobayashi Y, Fujishima A, Myasaki T, Fujiwara T (2007) Effect of shrinkage on the marginal and internal fit of CAD/CAM-fabricated zirconia frameworks. Dent Mater J 26(6):820–826PubMedCrossRef Kunii J, Hotta Y, Tamaki Y, Ozawa A, Kobayashi Y, Fujishima A, Myasaki T, Fujiwara T (2007) Effect of shrinkage on the marginal and internal fit of CAD/CAM-fabricated zirconia frameworks. Dent Mater J 26(6):820–826PubMedCrossRef
33.
go back to reference Mehl A, Ender A, Mörman W, Attin T (2009) Accuracy testing of a new intraoral 3D camera. Int J Comput Dent 12(1):11–28PubMed Mehl A, Ender A, Mörman W, Attin T (2009) Accuracy testing of a new intraoral 3D camera. Int J Comput Dent 12(1):11–28PubMed
34.
go back to reference Ender A, Mehl A (2011) Full arch scans: conventional versus digital impressions—an in vitro study. Int J Comput Dent 14(1):11–21PubMed Ender A, Mehl A (2011) Full arch scans: conventional versus digital impressions—an in vitro study. Int J Comput Dent 14(1):11–21PubMed
35.
go back to reference Quaas S, Rudolph H, Luthardt RG (2007) Direct mechanical data acquisition of dental impressions for the manufacturing of CAD/CAM restorations. J Dent 35(12):903–908PubMedCrossRef Quaas S, Rudolph H, Luthardt RG (2007) Direct mechanical data acquisition of dental impressions for the manufacturing of CAD/CAM restorations. J Dent 35(12):903–908PubMedCrossRef
36.
go back to reference Boening KW, Wolf BH, Schmidt AE, Kästner K, Walter MH (2000) Clinical fit of Procera all-ceram crowns. J Prosthet Dent 84:419–124PubMedCrossRef Boening KW, Wolf BH, Schmidt AE, Kästner K, Walter MH (2000) Clinical fit of Procera all-ceram crowns. J Prosthet Dent 84:419–124PubMedCrossRef
37.
go back to reference Molin M, Karlsson S (1993) The fit of gold inlays and three ceramic inlay systems. A clinical and in vitro study. Acta Odontol Scand 51:201–206PubMedCrossRef Molin M, Karlsson S (1993) The fit of gold inlays and three ceramic inlay systems. A clinical and in vitro study. Acta Odontol Scand 51:201–206PubMedCrossRef
38.
go back to reference Holmes JR, Bayne SC, Holland GA, Sulik WD (1989) Considerations in measurement of marginal fit. J Prosthet Dent 62:405–408PubMedCrossRef Holmes JR, Bayne SC, Holland GA, Sulik WD (1989) Considerations in measurement of marginal fit. J Prosthet Dent 62:405–408PubMedCrossRef
39.
go back to reference Tuntiprawon M, Wilson PR (1995) The effect of cement thickness on the fracture strength of all-ceramic crowns. Aust Dent J 40:17–21PubMedCrossRef Tuntiprawon M, Wilson PR (1995) The effect of cement thickness on the fracture strength of all-ceramic crowns. Aust Dent J 40:17–21PubMedCrossRef
40.
go back to reference Rekow ED, Harsono M, Janal M, Thompsin VP, Zhang G (2006) Factorial analysis of variables influencing stress in all-ceramic crowns. Dent Mater 22:125–132PubMedCrossRef Rekow ED, Harsono M, Janal M, Thompsin VP, Zhang G (2006) Factorial analysis of variables influencing stress in all-ceramic crowns. Dent Mater 22:125–132PubMedCrossRef
41.
go back to reference Abduo J, Lyons K, Swain M (2010) Fit of zirconia fixed partial denture: a systematic review. J Oral Rehab 47:866–876CrossRef Abduo J, Lyons K, Swain M (2010) Fit of zirconia fixed partial denture: a systematic review. J Oral Rehab 47:866–876CrossRef
42.
go back to reference Scotti R, Cardelli P, Baldissara P, Monaco C (2011) Clinical fitting of CAD/CAM zirconia single crowns generated from digital intraoral impressions based on active wavefront sampling. J Dent. doi:10.1016/j.jdent.2011.10.005 Scotti R, Cardelli P, Baldissara P, Monaco C (2011) Clinical fitting of CAD/CAM zirconia single crowns generated from digital intraoral impressions based on active wavefront sampling. J Dent. doi:10.​1016/​j.​jdent.​2011.​10.​005
43.
go back to reference Suttor D, Bunke K, Hoescheller S, Hauptmann H, Hertlein G (2001) LAVA—the system for all-ceramic ZrO2 crown and bridge frameworks. Int Comput Dent 4:195–206 Suttor D, Bunke K, Hoescheller S, Hauptmann H, Hertlein G (2001) LAVA—the system for all-ceramic ZrO2 crown and bridge frameworks. Int Comput Dent 4:195–206
44.
go back to reference Moldovan O, Luthardt RG, Corcodel N, Rudolph H (2011) Three-dimensional fit of CAD/CAM-made zirconia copings. Dent Mater 27(12):1273–1278PubMedCrossRef Moldovan O, Luthardt RG, Corcodel N, Rudolph H (2011) Three-dimensional fit of CAD/CAM-made zirconia copings. Dent Mater 27(12):1273–1278PubMedCrossRef
Metadata
Title
Marginal and internal fit of four-unit zirconia fixed dental prostheses based on digital and conventional impression techniques
Authors
Júnio S. Almeida e Silva
Kurt Erdelt
Daniel Edelhoff
Élito Araújo
Michael Stimmelmayr
Luiz Clovis Cardoso Vieira
Jan-Frederik Güth
Publication date
01-03-2014
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 2/2014
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-013-0987-2

Other articles of this Issue 2/2014

Clinical Oral Investigations 2/2014 Go to the issue