Skip to main content
Top
Published in: Surgical Endoscopy 5/2012

Open Access 01-05-2012

In vivo MRI visualization of mesh shrinkage using surgical implants loaded with superparamagnetic iron oxides

Authors: Nicolas Kuehnert, Nils A. Kraemer, Jens Otto, Hank C. W. Donker, Ioana Slabu, Martin Baumann, Christiane K. Kuhl, Uwe Klinge

Published in: Surgical Endoscopy | Issue 5/2012

Login to get access

Abstract

Background

Prosthetic mesh implants are widely used in hernia surgery. To show long-term mesh-related complications such as shrinkage or adhesions, a precise visualization of meshes and their vicinity in vivo is important. By supplementing mesh fibers with ferro particles, magnetic resonance imaging (MRI) can help to delineate the mesh itself. This study aimed to demonstrate and quantify time-dependent mesh shrinkage in vivo by MRI.

Methods

Polyvinylidenfluoride (PVDF) meshes with incorporated superparamagnetic iron oxides (SPIOs) were implanted as an abdominal wall replacement in 30 rats. On days 1, 7, 14, or 21, MRI was performed using a gradient echo sequence with repetition time (TR)/echo time (TE) of 50/4.6 and a flip angle of 20°. The length, width, and area of the device were measured on axial, coronal, and sagittal images, and geometric deformations were assessed by surgical explantation.

Results

In all cases, the meshes were visualized and their area estimated by measuring the length and width of the mesh. The MRI presented a mean area shrinkage in vivo of 13% on day 7, 23% on day 14, and 23% on day 21. Postmortem measurements differed statistically from MRI, with a mean area shrinkage of 23% on day 7, 28% on day 14, and 30% on day 21. Ex vivo measurements of shrinkage showed in vivo measurements to be overestimated approximately 8%. Delineation of the mesh helped to show folding or adhesions close to the intestine.

Conclusion

Loading of surgical meshes with SPIOs allows their precise visualization during MRI and guarantees an accurate in vivo assessment of their shrinkage. The authors’ observation clearly indicates that shrinkage in vivo is remarkably less than that shown by illustrated explantation measurements. The use of MRI with such meshes could be a reliable technique for checking on proper operation of implanted meshes and showing related complications, obviating the need for exploratory open surgical revision.
Literature
1.
go back to reference McGreevy JM, Goodney PP, Birkmeyer CM, Finlayson SRG, Laycock WS, Birkmeyer JD (2003) A prospective study comparing the complication rates between laparoscopic and open ventral hernia repairs. Surg Endosc 17:1778–1780PubMedCrossRef McGreevy JM, Goodney PP, Birkmeyer CM, Finlayson SRG, Laycock WS, Birkmeyer JD (2003) A prospective study comparing the complication rates between laparoscopic and open ventral hernia repairs. Surg Endosc 17:1778–1780PubMedCrossRef
2.
go back to reference Andersen LPH, Klein M, Goegenur I, Rosenberg J (2009) Long-term recurrence and complication rates after incisional hernia repair with the open onlay technique. BMC Surg 9:6PubMedCrossRef Andersen LPH, Klein M, Goegenur I, Rosenberg J (2009) Long-term recurrence and complication rates after incisional hernia repair with the open onlay technique. BMC Surg 9:6PubMedCrossRef
3.
go back to reference Junge K, Klinge U, Rosch R, Klosterhalfen B, Schumpelick V (2002) Functional and morphologic properties of a modified mesh for inguinal hernia repair. World J Surg 26:1472–1480PubMedCrossRef Junge K, Klinge U, Rosch R, Klosterhalfen B, Schumpelick V (2002) Functional and morphologic properties of a modified mesh for inguinal hernia repair. World J Surg 26:1472–1480PubMedCrossRef
4.
go back to reference Lomanto D, Iyer SG, Shabbir A, Cheah WK (2006) Laparoscopic versus open ventral hernia mesh repair: a prospective study. Surg Endosc 20:1030–1035PubMedCrossRef Lomanto D, Iyer SG, Shabbir A, Cheah WK (2006) Laparoscopic versus open ventral hernia mesh repair: a prospective study. Surg Endosc 20:1030–1035PubMedCrossRef
5.
go back to reference Schoenmaeckers EJP, van der Valk SBA, van den Hout HW, Raymakers JFTJ, Rakic S (2009) Computed tomographic measurements of mesh shrinkage after laparoscopic ventral incisional hernia repair with an expanded polytetrafluoroethylene mesh. Surg Endosc 23:1620–1623PubMedCrossRef Schoenmaeckers EJP, van der Valk SBA, van den Hout HW, Raymakers JFTJ, Rakic S (2009) Computed tomographic measurements of mesh shrinkage after laparoscopic ventral incisional hernia repair with an expanded polytetrafluoroethylene mesh. Surg Endosc 23:1620–1623PubMedCrossRef
6.
go back to reference Fischer T, Ladurner R, Gangkofer A, Mussack T, Reiser M, Lienemann A (2007) Functional cine MRI of the abdomen for the assessment of implanted synthetic mesh in patients after incisional hernia repair: initial results. Eur Radiol 17:3123–3129PubMedCrossRef Fischer T, Ladurner R, Gangkofer A, Mussack T, Reiser M, Lienemann A (2007) Functional cine MRI of the abdomen for the assessment of implanted synthetic mesh in patients after incisional hernia repair: initial results. Eur Radiol 17:3123–3129PubMedCrossRef
7.
go back to reference Zogbi L, Portella AOV, Trindade MRM, Trindade EN (2010) Retraction and fibroplasia in a polypropylene prosthesis: experimental study in rats. Hernia J Hernias Abdom Wall Surg 14:291–298 Zogbi L, Portella AOV, Trindade MRM, Trindade EN (2010) Retraction and fibroplasia in a polypropylene prosthesis: experimental study in rats. Hernia J Hernias Abdom Wall Surg 14:291–298
8.
go back to reference Parra JA, Revuelta S, Gallego T, Bueno J, Berrio JI, Fariñas MC (2004) Prosthetic mesh used for inguinal and ventral hernia repair: normal appearance and complications in ultrasound and CT. Br J Radiol 77:261–265PubMedCrossRef Parra JA, Revuelta S, Gallego T, Bueno J, Berrio JI, Fariñas MC (2004) Prosthetic mesh used for inguinal and ventral hernia repair: normal appearance and complications in ultrasound and CT. Br J Radiol 77:261–265PubMedCrossRef
9.
go back to reference Jenkins ED, Yom V, Melman L, Brunt LM, Eagon JC, Frisella MM, Matthews BD. Prospective evaluation of adhesion characteristics to intraperitoneal mesh and adhesiolysis-related complications during laparoscopic re-exploration after prior ventral hernia repair. Surg Endosc 24:3002–3007 Jenkins ED, Yom V, Melman L, Brunt LM, Eagon JC, Frisella MM, Matthews BD. Prospective evaluation of adhesion characteristics to intraperitoneal mesh and adhesiolysis-related complications during laparoscopic re-exploration after prior ventral hernia repair. Surg Endosc 24:3002–3007
10.
go back to reference Ferrari GC, Miranda A, Di Lernia S, Sansonna F, Magistro C, Maggioni D, Scandroglio I, Costanzi A, Franzetti M, Pugliese R (2008) Laparoscopic repair of incisional hernia: outcomes of 100 consecutive cases comprising 25 wall defects larger than 15 cm. Surg Endosc 22:1173–1179PubMedCrossRef Ferrari GC, Miranda A, Di Lernia S, Sansonna F, Magistro C, Maggioni D, Scandroglio I, Costanzi A, Franzetti M, Pugliese R (2008) Laparoscopic repair of incisional hernia: outcomes of 100 consecutive cases comprising 25 wall defects larger than 15 cm. Surg Endosc 22:1173–1179PubMedCrossRef
11.
go back to reference Conze J, Krones CJ, Schumpelick V, Klinge U (2007) Incisional hernia: challenge of reoperations after mesh repair. Langenbeck’s Arch Surg Deutsche Gesellschaft Chir 392:453–457CrossRef Conze J, Krones CJ, Schumpelick V, Klinge U (2007) Incisional hernia: challenge of reoperations after mesh repair. Langenbeck’s Arch Surg Deutsche Gesellschaft Chir 392:453–457CrossRef
12.
go back to reference Lienemann A, Sprenger D, Steitz HO, Korell M, Reiser M (2000) Detection and mapping of intraabdominal adhesions by using functional cine MR imaging: preliminary results. Radiology 217:421–425PubMed Lienemann A, Sprenger D, Steitz HO, Korell M, Reiser M (2000) Detection and mapping of intraabdominal adhesions by using functional cine MR imaging: preliminary results. Radiology 217:421–425PubMed
13.
go back to reference Zinther NB, Zeuten A, Marinovskij E, Haislund M, Wara P, Friis-Andersen H (2010) Functional cine MRI and transabdominal ultrasonography for the assessment of adhesions to implanted synthetic mesh 5–7 years after laparoscopic ventral hernia repair. Hernia J Hernias Abdom Wall Surg 14:499–504 Zinther NB, Zeuten A, Marinovskij E, Haislund M, Wara P, Friis-Andersen H (2010) Functional cine MRI and transabdominal ultrasonography for the assessment of adhesions to implanted synthetic mesh 5–7 years after laparoscopic ventral hernia repair. Hernia J Hernias Abdom Wall Surg 14:499–504
14.
go back to reference Mussack T, Fischer T, Ladurner R, Gangkofer A, Bensler S, Hallfeldt KK, Reiser M, Lienemann A (2005) Cine magnetic resonance imaging vs high-resolution ultrasonography for detection of adhesions after laparoscopic and open incisional hernia repair: a matched pair pilot analysis. Surg Endosc 19:1538–1543PubMedCrossRef Mussack T, Fischer T, Ladurner R, Gangkofer A, Bensler S, Hallfeldt KK, Reiser M, Lienemann A (2005) Cine magnetic resonance imaging vs high-resolution ultrasonography for detection of adhesions after laparoscopic and open incisional hernia repair: a matched pair pilot analysis. Surg Endosc 19:1538–1543PubMedCrossRef
15.
go back to reference Zinther NB, Zeuten A, Marinovskij E, Haislund Mt, Friis-Andersen H (2010) Detection of abdominal wall adhesions using visceral slide. Surg Endosc 24:3161–3166PubMedCrossRef Zinther NB, Zeuten A, Marinovskij E, Haislund Mt, Friis-Andersen H (2010) Detection of abdominal wall adhesions using visceral slide. Surg Endosc 24:3161–3166PubMedCrossRef
16.
go back to reference Kirchhoff S, Ladurner R, Kirchhoff C, Mussack T, Reiser MF, Lienemann A (2010) Detection of recurrent hernia and intraabdominal adhesions following incisional hernia repair: a functional cine MRI-study. Abdom Imaging 35:224–231PubMedCrossRef Kirchhoff S, Ladurner R, Kirchhoff C, Mussack T, Reiser MF, Lienemann A (2010) Detection of recurrent hernia and intraabdominal adhesions following incisional hernia repair: a functional cine MRI-study. Abdom Imaging 35:224–231PubMedCrossRef
17.
go back to reference Kraemer NA, Donker HCW, Otto J, Hodenius M, Senegas J, Slabu I, Klinge U, Baumann M, Muellen A, Obolenski B, Guenther RW, Krombach GA (2010) A concept for magnetic resonance visualization of surgical textile implants. Invest Radiol 45:477–483CrossRef Kraemer NA, Donker HCW, Otto J, Hodenius M, Senegas J, Slabu I, Klinge U, Baumann M, Muellen A, Obolenski B, Guenther RW, Krombach GA (2010) A concept for magnetic resonance visualization of surgical textile implants. Invest Radiol 45:477–483CrossRef
18.
go back to reference Kuhry E, Schwenk W, Gaupset R, Romild U, Bonjer J (2008) Long-term outcome of laparoscopic surgery for colorectal cancer: a Cochrane systematic review of randomised controlled trials. Cancer Treat Rev 34:498–504PubMedCrossRef Kuhry E, Schwenk W, Gaupset R, Romild U, Bonjer J (2008) Long-term outcome of laparoscopic surgery for colorectal cancer: a Cochrane systematic review of randomised controlled trials. Cancer Treat Rev 34:498–504PubMedCrossRef
19.
20.
go back to reference Klinge U, Junge K, Stumpf M, Ap APO, Klosterhalfen B (2002) Functional and morphological evaluation of a low-weight, monofilament polypropylene mesh for hernia repair. J Biomed Mater Res 63:129–136PubMedCrossRef Klinge U, Junge K, Stumpf M, Ap APO, Klosterhalfen B (2002) Functional and morphological evaluation of a low-weight, monofilament polypropylene mesh for hernia repair. J Biomed Mater Res 63:129–136PubMedCrossRef
21.
go back to reference García Ureña MA, Vega Ruiz V, Díaz Godoy A, Báez Perea JM, Marín Gómez LM, Carnero Hernández FJ, MAngel Velasco García (2007) Differences in polypropylene shrinkage depending on mesh position in an experimental study. Am J Surg 193:538–542PubMedCrossRef García Ureña MA, Vega Ruiz V, Díaz Godoy A, Báez Perea JM, Marín Gómez LM, Carnero Hernández FJ, MAngel Velasco García (2007) Differences in polypropylene shrinkage depending on mesh position in an experimental study. Am J Surg 193:538–542PubMedCrossRef
22.
go back to reference Harrell AG, Novitsky YW, Peindl RD, Cobb WS, Austin CE, Cristiano JA, Norton JH, Kercher KW, Heniford BT (2006) Prospective evaluation of adhesion formation and shrinkage of intraabdominal prosthetics in a rabbit model. Am Surg 72:808–833 discussion 813–814PubMed Harrell AG, Novitsky YW, Peindl RD, Cobb WS, Austin CE, Cristiano JA, Norton JH, Kercher KW, Heniford BT (2006) Prospective evaluation of adhesion formation and shrinkage of intraabdominal prosthetics in a rabbit model. Am Surg 72:808–833 discussion 813–814PubMed
23.
go back to reference Johnson EK, Hoyt CH, Dinsmore RC (2004) Abdominal wall hernia repair: a long-term comparison of Sepramesh and Dualmesh in a rabbit hernia model. Am Surg 70:657–661PubMed Johnson EK, Hoyt CH, Dinsmore RC (2004) Abdominal wall hernia repair: a long-term comparison of Sepramesh and Dualmesh in a rabbit hernia model. Am Surg 70:657–661PubMed
24.
go back to reference Jacob BP, Hogle NJ, Durak E, Kim T, Fowler DL (2007) Tissue ingrowth and bowel adhesion formation in an animal comparative study: polypropylene versus Proceed versus Parietex Composite. Surg Endosc 21:629–633PubMedCrossRef Jacob BP, Hogle NJ, Durak E, Kim T, Fowler DL (2007) Tissue ingrowth and bowel adhesion formation in an animal comparative study: polypropylene versus Proceed versus Parietex Composite. Surg Endosc 21:629–633PubMedCrossRef
25.
go back to reference Burger JWA, Halm JA, Wijsmuller AR, ten Raa S, Jeekel J (2006) Evaluation of new prosthetic meshes for ventral hernia repair. Surg Endosc 20:1320–1325PubMedCrossRef Burger JWA, Halm JA, Wijsmuller AR, ten Raa S, Jeekel J (2006) Evaluation of new prosthetic meshes for ventral hernia repair. Surg Endosc 20:1320–1325PubMedCrossRef
26.
go back to reference Taupitz M, Schmitz S, Hamm B (2003) Superparamagnetic iron oxide particles: current state and future development. RoeFo Fortschr Geb Roentgenstr Nuklearmedizin 175:752–765CrossRef Taupitz M, Schmitz S, Hamm B (2003) Superparamagnetic iron oxide particles: current state and future development. RoeFo Fortschr Geb Roentgenstr Nuklearmedizin 175:752–765CrossRef
27.
go back to reference Kircher MF, Allport JR, Graves EE, Love V, Josephson L, Lichtman AH, Weissleder R (2003) In vivo high resolution three-dimensional imaging of antigen-specific cytotoxic T-lymphocyte trafficking to tumors. Cancer Res 63:6838–6846PubMed Kircher MF, Allport JR, Graves EE, Love V, Josephson L, Lichtman AH, Weissleder R (2003) In vivo high resolution three-dimensional imaging of antigen-specific cytotoxic T-lymphocyte trafficking to tumors. Cancer Res 63:6838–6846PubMed
28.
go back to reference Junge K, Binneboesel M, Rosch R, Jansen M, Kaemmer D, Otto J, Schumpelick V, Klinge U (2009) Adhesion formation of a polyvinylidenfluoride/polypropylene mesh for intraabdominal placement in a rodent animal model. Surg Endosc 23:327–333PubMedCrossRef Junge K, Binneboesel M, Rosch R, Jansen M, Kaemmer D, Otto J, Schumpelick V, Klinge U (2009) Adhesion formation of a polyvinylidenfluoride/polypropylene mesh for intraabdominal placement in a rodent animal model. Surg Endosc 23:327–333PubMedCrossRef
29.
go back to reference Berger D, Bientzle M (2009) Polyvinylidenfluoride: a suitable mesh material for laparoscopic incisional and parastomal hernia repair! A prospective, observational study with 344 patients. Hernia J Hernias Abdom Wall Surg 13:167–172 Berger D, Bientzle M (2009) Polyvinylidenfluoride: a suitable mesh material for laparoscopic incisional and parastomal hernia repair! A prospective, observational study with 344 patients. Hernia J Hernias Abdom Wall Surg 13:167–172
30.
go back to reference Conze J, Junge K, Weiss C, Anurov M, Oettinger A, Klinge U, Schumpelick V (2008) New polymer for intraabdominal meshes: PVDF copolymer. J Biomed Mater Res Part B Appl Biomater 87:321–328PubMedCrossRef Conze J, Junge K, Weiss C, Anurov M, Oettinger A, Klinge U, Schumpelick V (2008) New polymer for intraabdominal meshes: PVDF copolymer. J Biomed Mater Res Part B Appl Biomater 87:321–328PubMedCrossRef
31.
go back to reference Dauendorffer JN, Bastuji-Garin S, Guéro S, Brousse N, Fraitag S (2009) Shrinkage of skin excision specimens: formalin fixation is not the culprit. Br J Dermatol 160:810–814PubMedCrossRef Dauendorffer JN, Bastuji-Garin S, Guéro S, Brousse N, Fraitag S (2009) Shrinkage of skin excision specimens: formalin fixation is not the culprit. Br J Dermatol 160:810–814PubMedCrossRef
32.
go back to reference Blasdale C, Charlton FG, Weatherhead SC, Ormond P, Lawrence CM (2010) Effect of tissue shrinkage on histological tumour-free margin after excision of basal cell carcinoma. Br J Dermatol 162:607–610PubMedCrossRef Blasdale C, Charlton FG, Weatherhead SC, Ormond P, Lawrence CM (2010) Effect of tissue shrinkage on histological tumour-free margin after excision of basal cell carcinoma. Br J Dermatol 162:607–610PubMedCrossRef
33.
go back to reference Wheeler TL, Koohmaraie M (1994) Prerigor and postrigor changes in tenderness of ovine longissimus muscle. J Anim Sci 72:1232–1238PubMed Wheeler TL, Koohmaraie M (1994) Prerigor and postrigor changes in tenderness of ovine longissimus muscle. J Anim Sci 72:1232–1238PubMed
Metadata
Title
In vivo MRI visualization of mesh shrinkage using surgical implants loaded with superparamagnetic iron oxides
Authors
Nicolas Kuehnert
Nils A. Kraemer
Jens Otto
Hank C. W. Donker
Ioana Slabu
Martin Baumann
Christiane K. Kuhl
Uwe Klinge
Publication date
01-05-2012
Publisher
Springer-Verlag
Published in
Surgical Endoscopy / Issue 5/2012
Print ISSN: 0930-2794
Electronic ISSN: 1432-2218
DOI
https://doi.org/10.1007/s00464-011-2057-7

Other articles of this Issue 5/2012

Surgical Endoscopy 5/2012 Go to the issue