Abstract

Purpose

Accurate placement of spinal pedicle screws (PS) is mandatory for good primary segmental stabilization allowing consequent osseous fusion, requiring judgmental experience developed during a long training process. Computer navigation offers permanent visual control during screw manipulation and has been shown to significantly lower the risk of pedicle perforation. This study aims to evaluate whether safety, accuracy, and judgmental skills in screw placement, comparable to an experienced surgeon, can be developed during training using computer navigation.

Methods

Lumbosacral PS were placed in 18 patients in a prospective setting, in one segment side with conventional fluoroscopy by a senior spine-surgeon, and computer navigated on the other side by a trainee without prior experience in the technique. At the beginning and at the end of the study, PS were placed freehand in solid foam models by the trainee. PS placement time, intraoperative placement revisions, PS placement accuracy on postoperative CT scans, and postoperative complications were assessed.

Results

Significant improvement of trainee’s PS placement accuracy (Sclafani score 8.2–8.83; p = 0.006) and time (13.3–6.8 min per screw; p = 0.005) to a similar level as the experienced surgeon state (5.2–4.1 min per screw; p = 0.39) was explored; similar improvement was explored in the foam models. The number of intraoperative placement revisions kept on a low level for surgeon (3.3–0.0%) and trainee (5.1–2.6%) during the whole study, no postoperative complications occurred.

Conclusion

Navigated PS insertion allows safe teaching from the early beginning of surgical training, due to steady intraoperative control on PS placement. Adequacy of PS placement is similar to screws placed by an experienced surgeon. Progress in judgmental skills in screw placement can be gained rapidly by the trainee, which can also be transferred to non-computer navigated PS placement.
Literature
2.
go back to reference Gertzbein S, Robbins S (1990) Accuracy of pedicular screw placement in vivo. Spine 15:11–14CrossRef Gertzbein S, Robbins S (1990) Accuracy of pedicular screw placement in vivo. Spine 15:11–14CrossRef
3.
go back to reference Shin BJ, James AR, Njoku IU, Härtl R (2012) Pedicle screw navigation: a systematic review and meta-analysis of perforation risk for computer-navigated versus freehand insertion. A review. J Neurosurg Spine 17:113–122CrossRef Shin BJ, James AR, Njoku IU, Härtl R (2012) Pedicle screw navigation: a systematic review and meta-analysis of perforation risk for computer-navigated versus freehand insertion. A review. J Neurosurg Spine 17:113–122CrossRef
5.
go back to reference Mason A, Paulsen R, Babuska JM, Rajpal S, Burneikiene S, Nelson EL, Villavicencio AT (2014) The accuracy of pedicle screw placement using intraoperative image guidance systems. J Neurosurg Spine 20(2):196–203CrossRef Mason A, Paulsen R, Babuska JM, Rajpal S, Burneikiene S, Nelson EL, Villavicencio AT (2014) The accuracy of pedicle screw placement using intraoperative image guidance systems. J Neurosurg Spine 20(2):196–203CrossRef
6.
go back to reference Härtl R, Lam KS, Wang J, Korge A, Kandziora F, Audigé L (2013) Worldwide survey on the use of navigation in spine surgery. World Neurosurg 79(1):162–172CrossRef Härtl R, Lam KS, Wang J, Korge A, Kandziora F, Audigé L (2013) Worldwide survey on the use of navigation in spine surgery. World Neurosurg 79(1):162–172CrossRef
7.
go back to reference Boon Tow BP, Yue WM, Srivastava A, Lai JM, Guo CM, Wearn Peng BC, Chen JL, Yew AK, Seng C, Tan SB (2015) Does navigation improve accuracy of placement of pedicle screws in single-level lumbar degenerative spondylolisthesis? A comparison between free-hand and three-dimensional O-arm navigation techniques. J Spinal Disord Tech 28(8):E472–E477CrossRef Boon Tow BP, Yue WM, Srivastava A, Lai JM, Guo CM, Wearn Peng BC, Chen JL, Yew AK, Seng C, Tan SB (2015) Does navigation improve accuracy of placement of pedicle screws in single-level lumbar degenerative spondylolisthesis? A comparison between free-hand and three-dimensional O-arm navigation techniques. J Spinal Disord Tech 28(8):E472–E477CrossRef
8.
go back to reference Kraus M, Weiskopf J, Dreyhaupt J, Krischak G, Gebhard F (2015) Computer-aided surgery does not increase the accuracy of dorsal pedicle screw placement in the thoracic and lumbar spine: a retrospective analysis of 2,003 pedicle screws in a level I trauma center. Global Spine J 5(2):93–101CrossRef Kraus M, Weiskopf J, Dreyhaupt J, Krischak G, Gebhard F (2015) Computer-aided surgery does not increase the accuracy of dorsal pedicle screw placement in the thoracic and lumbar spine: a retrospective analysis of 2,003 pedicle screws in a level I trauma center. Global Spine J 5(2):93–101CrossRef
10.
go back to reference Gautschi OP, Schatlo B, Schaller K, Tessitore E (2011) Clinically relevant complications related to pedicle screw placement in thoracolumbar surgery and their management: a literature review of 35,630 pedicle screws. Neurosurg Focus 31(4):E8CrossRef Gautschi OP, Schatlo B, Schaller K, Tessitore E (2011) Clinically relevant complications related to pedicle screw placement in thoracolumbar surgery and their management: a literature review of 35,630 pedicle screws. Neurosurg Focus 31(4):E8CrossRef
11.
go back to reference Puvanesarajah V, Liauw JA, Lo SF, Lina IA, Witham TF (2014) Techniques and accuracy of thoracolumbar pedicle screw placement. World J Orthop 5(2):112–123CrossRef Puvanesarajah V, Liauw JA, Lo SF, Lina IA, Witham TF (2014) Techniques and accuracy of thoracolumbar pedicle screw placement. World J Orthop 5(2):112–123CrossRef
13.
go back to reference Nelson EM, Monazzam SM, Kim KD, Seibert JA, Klineberg EO (2014) Intraoperative fluoroscopy, portable X-ray, and CT: patient and operating room personnel radiation exposure in spinal surgery. Spine J 14(12):2985–2991CrossRef Nelson EM, Monazzam SM, Kim KD, Seibert JA, Klineberg EO (2014) Intraoperative fluoroscopy, portable X-ray, and CT: patient and operating room personnel radiation exposure in spinal surgery. Spine J 14(12):2985–2991CrossRef
14.
go back to reference Sclafani JA, Regev GJ, Webb J, Garfin SR, Kim CW (2011) Use of a quantitative pedicle screw accuracy system to assess new technology: Initial studies on O-arm navigation and its effect on the learning curve of percutaneous pedicle screw insertion. SAS J 5(3):57–62CrossRef Sclafani JA, Regev GJ, Webb J, Garfin SR, Kim CW (2011) Use of a quantitative pedicle screw accuracy system to assess new technology: Initial studies on O-arm navigation and its effect on the learning curve of percutaneous pedicle screw insertion. SAS J 5(3):57–62CrossRef
15.
go back to reference Al-Khouja L, Shweikeh F, Pashman R, Johnson JP, Kim TT, Drazin D (2015) Economics of image guidance and navigation in spine surgery. Surg Neurol Int 6(Suppl 10):S323–S326PubMedPubMedCentral Al-Khouja L, Shweikeh F, Pashman R, Johnson JP, Kim TT, Drazin D (2015) Economics of image guidance and navigation in spine surgery. Surg Neurol Int 6(Suppl 10):S323–S326PubMedPubMedCentral
16.
go back to reference Schulze CJ, Munzinger E, Weber U (1998) Clinical relevance of accuracy of pedicle screw placement. A computed tomographic-supported analysis. Spine 23:2215–2220CrossRef Schulze CJ, Munzinger E, Weber U (1998) Clinical relevance of accuracy of pedicle screw placement. A computed tomographic-supported analysis. Spine 23:2215–2220CrossRef
17.
go back to reference Parker SL, McGirt MJ, Farber SH, Amin AG, Rick AM, Suk I, Bydon A, Sciubba DM, Wolinsky JP, Gokaslan ZL, Witham TF (2011) Accuracy of free-hand pedicle screws in the thoracic and lumbar spine: analysis of 6816 consecutive screws. Neurosurgery 68(1):170–178CrossRef Parker SL, McGirt MJ, Farber SH, Amin AG, Rick AM, Suk I, Bydon A, Sciubba DM, Wolinsky JP, Gokaslan ZL, Witham TF (2011) Accuracy of free-hand pedicle screws in the thoracic and lumbar spine: analysis of 6816 consecutive screws. Neurosurgery 68(1):170–178CrossRef
18.
go back to reference Bergeson RK, Schwend RM, DeLucia T, Silva SR, Smith JE, Avilucea FR (2008) How accurately do novice surgeons place thoracic pedicle screws with the free hand technique? Spine (Phila Pa 1976) 33(15):E501–E507CrossRef Bergeson RK, Schwend RM, DeLucia T, Silva SR, Smith JE, Avilucea FR (2008) How accurately do novice surgeons place thoracic pedicle screws with the free hand technique? Spine (Phila Pa 1976) 33(15):E501–E507CrossRef
Metadata
Title
Navigation versus experience: providing training in accurate lumbar pedicle screw positioning
Authors
L. Leitner
G. Bratschitsch
Patrick Sadoghi
G. Adelsmayr
P. Puchwein
A. Leithner
R. Radl
Publication date
01-12-2019
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 12/2019
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-019-03206-7

Other articles of this Issue 12/2019

Archives of Orthopaedic and Trauma Surgery 12/2019 Go to the issue