Correlation of serum metal ion levels with pathological changes of ARMD in failed metal-on-metal-hip-resurfacing arthroplasties

Authors: George Grammatopoulos, Mitsuru Munemoto, Athanasios Pollalis, Nicholas A. Athanasou

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 8/2017

Login to get access

Abstract

Background

Metal-on-metal-hip-resurfacing arthroplasties (MoMHRAs) have been associated with an increased failure rates due to an adverse-response-to-metal-debris (ARMD) associated with a spectrum of pathological features. Serum levels of cobalt (Co) and chromium (Cr) are used to assess MoMHRAs, with regard to ARMD, but it is not certain whether ion levels correlate with pathological changes in periprosthetic tissues.

Methods

Serum Co and Cr levels were correlated with histological findings in 38 revised MoMHRAs (29 pseudotumour cases and 9 non-pseudotumour cases revised for pain). The extent of necrosis and macrophage infiltrate as well as the aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL) response was assessed semi-quantitatively; the prosthesis linear wear rate (PLWR) was also determined in ten cases.

Results

Cr levels were elevated in 82% and Co levels elevated in 53% of cases; the PLWR correlated with Cr level (rho = 0.8, p = 0.006). Tissue necrosis and macrophage infiltration were noted in all, most of which also exhibited significant ALVAL. Although a discrete correlation was not seen between Co and/or Cr ion levels and the extent of necrosis, degree of macrophage infiltration, or ALVAL score, it was noted that cases with acceptable metal ions levels had high ALVAL score.

Conclusion

Histological features of both innate and adaptive immune response to metal wear are seen in periprosthetic tissues in cases with both elevated and non-elevated metal ion levels. MoMHRA failures with acceptable ion levels exhibited a pronounced ALVAL response. Although metal ion levels are elevated in most cases of MoMHRA failure due to ARMD, the finding of a normal metal ion level does not exclude this diagnosis.
Literature
3.
go back to reference Jacobs JJ, Skipor AK, Campbell PA, Hallab NJ, Urban RM, Amstutz HC (2004) Can metal levels be used to monitor metal-on-metal hip arthroplasties? J Arthroplasty 19(8 Suppl 3):59–65 CrossRefPubMed Jacobs JJ, Skipor AK, Campbell PA, Hallab NJ, Urban RM, Amstutz HC (2004) Can metal levels be used to monitor metal-on-metal hip arthroplasties? J Arthroplasty 19(8 Suppl 3):59–65 CrossRefPubMed
4.
go back to reference Clarke MT, Lee PT, Arora A, Villar RN (2003) Levels of metal ions after small- and large-diameter metal-on-metal hip arthroplasty. J Bone Jt Surg Br 85(6):913–917 Clarke MT, Lee PT, Arora A, Villar RN (2003) Levels of metal ions after small- and large-diameter metal-on-metal hip arthroplasty. J Bone Jt Surg Br 85(6):913–917
5.
7.
go back to reference De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K (2008) Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Jt Surg Br 90(10):1291–1297. doi:10.1302/0301-620X.90B10.20533 CrossRef De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K (2008) Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Jt Surg Br 90(10):1291–1297. doi:10.​1302/​0301-620X.​90B10.​20533 CrossRef
8.
go back to reference Smolders JM, Bisseling P, Hol A, Der Van, Straeten C, Schreurs BW, van Susante JL (2011) Metal ion interpretation in resurfacing versus conventional hip arthroplasty and in whole blood versus serum. How should we interpret metal ion data. Hip Int 21(5):587–595. doi:10.5301/HIP.2011.8643 CrossRefPubMed Smolders JM, Bisseling P, Hol A, Der Van, Straeten C, Schreurs BW, van Susante JL (2011) Metal ion interpretation in resurfacing versus conventional hip arthroplasty and in whole blood versus serum. How should we interpret metal ion data. Hip Int 21(5):587–595. doi:10.​5301/​HIP.​2011.​8643 CrossRefPubMed
9.
go back to reference Hart AJ, Sabah SA, Sampson B, Skinner JA, Powell JJ, Palla L, Pajamaki KJ, Puolakka T, Reito A, Eskelinen A (2014) Surveillance of patients with metal-on-metal hip resurfacing and total hip prostheses: a prospective cohort study to investigate the relationship between blood metal ion levels and implant failure. J Bone Jt Surg Am 96(13):1091–1099. doi:10.2106/JBJS.M.00957 CrossRef Hart AJ, Sabah SA, Sampson B, Skinner JA, Powell JJ, Palla L, Pajamaki KJ, Puolakka T, Reito A, Eskelinen A (2014) Surveillance of patients with metal-on-metal hip resurfacing and total hip prostheses: a prospective cohort study to investigate the relationship between blood metal ion levels and implant failure. J Bone Jt Surg Am 96(13):1091–1099. doi:10.​2106/​JBJS.​M.​00957 CrossRef
12.
go back to reference Hailer NP, Bengtsson M, Lundberg C, Milbrink J (2014) High metal ion levels after use of the ASR device correlate with development of pseudotumors and T cell activation. Clin Orthop Relat Res 472(3):953–961. doi:10.1007/s11999-013-3307-x CrossRefPubMed Hailer NP, Bengtsson M, Lundberg C, Milbrink J (2014) High metal ion levels after use of the ASR device correlate with development of pseudotumors and T cell activation. Clin Orthop Relat Res 472(3):953–961. doi:10.​1007/​s11999-013-3307-x CrossRefPubMed
13.
go back to reference Hjorth MH, Stilling M, Soballe K, Bolvig LH, Thyssen JP, Mechlenburg I, Jakobsen SS (2016) No association between pseudotumors, high serum metal-ion levels and metal hypersensitivity in large-head metal-on-metal total hip arthroplasty at 5–7-year follow-up. Skelet Radiol 45(1):115–125. doi:10.1007/s00256-015-2264-8 CrossRef Hjorth MH, Stilling M, Soballe K, Bolvig LH, Thyssen JP, Mechlenburg I, Jakobsen SS (2016) No association between pseudotumors, high serum metal-ion levels and metal hypersensitivity in large-head metal-on-metal total hip arthroplasty at 5–7-year follow-up. Skelet Radiol 45(1):115–125. doi:10.​1007/​s00256-015-2264-8 CrossRef
15.
go back to reference Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH (2005) Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Jt Surg Am 87(1):28–36. doi:10.2106/JBJS.A.02039pp Willert HG, Buchhorn GH, Fayyazi A, Flury R, Windler M, Koster G, Lohmann CH (2005) Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J Bone Jt Surg Am 87(1):28–36. doi:10.​2106/​JBJS.​A.​02039pp
17.
go back to reference Grammatopoulos G, Pandit H, Kamali A, Maggiani F, Glyn-Jones S, Gill HS, Murray DW, Athanasou N (2013) The correlation of wear with histological features after failed hip resurfacing arthroplasty. J Bone Jt Surg Am 95(12):e81. doi:10.2106/JBJS.L.00775 CrossRef Grammatopoulos G, Pandit H, Kamali A, Maggiani F, Glyn-Jones S, Gill HS, Murray DW, Athanasou N (2013) The correlation of wear with histological features after failed hip resurfacing arthroplasty. J Bone Jt Surg Am 95(12):e81. doi:10.​2106/​JBJS.​L.​00775 CrossRef
18.
19.
go back to reference Van Der Straeten C, Grammatopoulos G, Gill HS, Calistri A, Campbell P, De Smet KA (2013) The 2012 Otto Aufranc Award: the interpretation of metal ion levels in unilateral and bilateral hip resurfacing. Clin Orthop Relat Res 471(2):377–385. doi:10.1007/s11999-012-2526-x CrossRef Van Der Straeten C, Grammatopoulos G, Gill HS, Calistri A, Campbell P, De Smet KA (2013) The 2012 Otto Aufranc Award: the interpretation of metal ion levels in unilateral and bilateral hip resurfacing. Clin Orthop Relat Res 471(2):377–385. doi:10.​1007/​s11999-012-2526-x CrossRef
20.
go back to reference Langton DJ (2014) Are metal ion levels a trigger for surgical intervention in metal-on-metal bearings? Metal on metal bearings: a clinical practicum, Springer, New York, pp 63–82 Langton DJ (2014) Are metal ion levels a trigger for surgical intervention in metal-on-metal bearings? Metal on metal bearings: a clinical practicum, Springer, New York, pp 63–82
22.
go back to reference Lohmann CH, Meyer H, Nuechtern JV, Singh G, Junk-Jantsch S, Schmotzer H, Morlock MM, Pfluger G (2013) Periprosthetic tissue metal content but not serum metal content predicts the type of tissue response in failed small-diameter metal-on-metal total hip arthroplasties. J Bone Jt Surg Am 95(17):1561–1568. doi:10.2106/JBJS.L.01273 CrossRef Lohmann CH, Meyer H, Nuechtern JV, Singh G, Junk-Jantsch S, Schmotzer H, Morlock MM, Pfluger G (2013) Periprosthetic tissue metal content but not serum metal content predicts the type of tissue response in failed small-diameter metal-on-metal total hip arthroplasties. J Bone Jt Surg Am 95(17):1561–1568. doi:10.​2106/​JBJS.​L.​01273 CrossRef
23.
go back to reference De Smet K, De Haan R, Calistri A, Campbell PA, Ebramzadeh E, Pattyn C, Gill HS (2008) Metal ion measurement as a diagnostic tool to identify problems with metal-on-metal hip resurfacing. J Bone Jt Surg Am 90(Suppl 4):202–208. doi:10.2106/JBJS.H.00672 CrossRef De Smet K, De Haan R, Calistri A, Campbell PA, Ebramzadeh E, Pattyn C, Gill HS (2008) Metal ion measurement as a diagnostic tool to identify problems with metal-on-metal hip resurfacing. J Bone Jt Surg Am 90(Suppl 4):202–208. doi:10.​2106/​JBJS.​H.​00672 CrossRef
24.
go back to reference De Pasquale D, Stea S, Squarzoni S, Bordini B, Amabile M, Catalani S, Apostoli P, Toni A (2014) Metal-on-metal hip prostheses: correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream. Int Orthop 38(3):469–475. doi:10.1007/s00264-013-2137-5 CrossRefPubMed De Pasquale D, Stea S, Squarzoni S, Bordini B, Amabile M, Catalani S, Apostoli P, Toni A (2014) Metal-on-metal hip prostheses: correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream. Int Orthop 38(3):469–475. doi:10.​1007/​s00264-013-2137-5 CrossRefPubMed
25.
go back to reference Langton DJ, Sprowson AP, Joyce TJ, Reed M, Carluke I, Partington P, Nargol AV (2009) Blood metal ion concentrations after hip resurfacing arthroplasty: a comparative study of articular surface replacement and Birmingham hip resurfacing arthroplasties. J Bone Jt Surg Br 91(10):1287–1295. doi:10.1302/0301-620X.91B10.22308 CrossRef Langton DJ, Sprowson AP, Joyce TJ, Reed M, Carluke I, Partington P, Nargol AV (2009) Blood metal ion concentrations after hip resurfacing arthroplasty: a comparative study of articular surface replacement and Birmingham hip resurfacing arthroplasties. J Bone Jt Surg Br 91(10):1287–1295. doi:10.​1302/​0301-620X.​91B10.​22308 CrossRef
26.
go back to reference Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Jt Surg Br 92(1):38–46. doi:10.1302/0301-620X.92B1.22770 CrossRef Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV (2010) Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Jt Surg Br 92(1):38–46. doi:10.​1302/​0301-620X.​92B1.​22770 CrossRef
27.
go back to reference Pelt CE, Erickson J, Clarke I, Donaldson T, Layfield L, Peters CL (2013) Histologic, serologic, and tribologic findings in failed metal-on-metal total hip arthroplasty: AAOS exhibit selection. J Bone Jt Surg Am 95(21):e163. doi:10.2106/JBJS.L.01446 CrossRef Pelt CE, Erickson J, Clarke I, Donaldson T, Layfield L, Peters CL (2013) Histologic, serologic, and tribologic findings in failed metal-on-metal total hip arthroplasty: AAOS exhibit selection. J Bone Jt Surg Am 95(21):e163. doi:10.​2106/​JBJS.​L.​01446 CrossRef
29.
go back to reference Daniel J, Ziaee H, Pradhan C, Pynsent PB, McMinn DJ (2010) Renal clearance of cobalt in relation to the use of metal-on-metal bearings in hip arthroplasty. J Bone Jt Surg Am 92(4):840–845. doi:10.2106/JBJS.H.01821 CrossRef Daniel J, Ziaee H, Pradhan C, Pynsent PB, McMinn DJ (2010) Renal clearance of cobalt in relation to the use of metal-on-metal bearings in hip arthroplasty. J Bone Jt Surg Am 92(4):840–845. doi:10.​2106/​JBJS.​H.​01821 CrossRef
31.
go back to reference Maezawa K, Nozawa M, Hirose T, Matsuda K, Yasuma M, Shitoto K, Kurosawa H (2002) Cobalt and chromium concentrations in patients with metal-on-metal and other cementless total hip arthroplasty. Arch Orthop Trauma Surg 122(5):283–287. doi:10.1007/s00402-001-0382-3 CrossRefPubMed Maezawa K, Nozawa M, Hirose T, Matsuda K, Yasuma M, Shitoto K, Kurosawa H (2002) Cobalt and chromium concentrations in patients with metal-on-metal and other cementless total hip arthroplasty. Arch Orthop Trauma Surg 122(5):283–287. doi:10.​1007/​s00402-001-0382-3 CrossRefPubMed
33.
go back to reference Mittal S, Revell M, Barone F, Hardie DL, Matharu GS, Davenport AJ, Martin RA, Grant M, Mosselmans F, Pynsent P, Sumathi VP, Addison O, Revell PA, Buckley CD (2013) Lymphoid aggregates that resemble tertiary lymphoid organs define a specific pathological subset in metal-on-metal hip replacements. PLoS One 8(5):e63470. doi:10.1371/journal.pone.0063470 CrossRefPubMedPubMedCentral Mittal S, Revell M, Barone F, Hardie DL, Matharu GS, Davenport AJ, Martin RA, Grant M, Mosselmans F, Pynsent P, Sumathi VP, Addison O, Revell PA, Buckley CD (2013) Lymphoid aggregates that resemble tertiary lymphoid organs define a specific pathological subset in metal-on-metal hip replacements. PLoS One 8(5):e63470. doi:10.​1371/​journal.​pone.​0063470 CrossRefPubMedPubMedCentral
34.
go back to reference Van Der Straeten C, Van Quickenborne D, De Roest B, Calistri A, Victor J, De Smet K (2013) Metal ion levels from well-functioning Birmingham hip resurfacings decline significantly at ten years. Bone Jt J 95(10):1332–1338. doi:10.1302/0301-620X.95B10.32022 CrossRef Van Der Straeten C, Van Quickenborne D, De Roest B, Calistri A, Victor J, De Smet K (2013) Metal ion levels from well-functioning Birmingham hip resurfacings decline significantly at ten years. Bone Jt J 95(10):1332–1338. doi:10.​1302/​0301-620X.​95B10.​32022 CrossRef
35.
go back to reference (MHRA) MHpRA (2010) Medical device alert: ASR hip replacement implant manufactured by DePuy internation Ltd (MHRA) MHpRA (2010) Medical device alert: ASR hip replacement implant manufactured by DePuy internation Ltd
36.
go back to reference Hannemann F, Hartmann A, Schmitt J, Lutzner J, Seidler A, Campbell P, Delaunay CP, Drexler H, Ettema HB, Garcia-Cimbrelo E, Huberti H, Knahr K, Kunze J, Langton DJ, Lauer W, Learmonth I, Lohmann CH, Morlock M, Wimmer MA, Zagra L, Gunther KP (2013) European multidisciplinary consensus statement on the use and monitoring of metal-on-metal bearings for total hip replacement and hip resurfacing. Orthop Traumatol Surg Res 99(3):263–271. doi:10.1016/j.otsr.2013.01.005 CrossRefPubMed Hannemann F, Hartmann A, Schmitt J, Lutzner J, Seidler A, Campbell P, Delaunay CP, Drexler H, Ettema HB, Garcia-Cimbrelo E, Huberti H, Knahr K, Kunze J, Langton DJ, Lauer W, Learmonth I, Lohmann CH, Morlock M, Wimmer MA, Zagra L, Gunther KP (2013) European multidisciplinary consensus statement on the use and monitoring of metal-on-metal bearings for total hip replacement and hip resurfacing. Orthop Traumatol Surg Res 99(3):263–271. doi:10.​1016/​j.​otsr.​2013.​01.​005 CrossRefPubMed
Metadata
Title
Correlation of serum metal ion levels with pathological changes of ARMD in failed metal-on-metal-hip-resurfacing arthroplasties
Authors
George Grammatopoulos
Mitsuru Munemoto
Athanasios Pollalis
Nicholas A. Athanasou
Publication date
01-08-2017
Publisher
Springer Berlin Heidelberg
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 8/2017
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-017-2723-x

Other articles of this Issue 8/2017

Archives of Orthopaedic and Trauma Surgery 8/2017 Go to the issue