Published in: Clinical Orthopaedics and Related Research® 2/2013

01-02-2013 | Symposium: Papers Presented at the Annual Meetings of The Hip Society

The 2012 Otto Aufranc Award: The Interpretation of Metal Ion Levels in Unilateral and Bilateral Hip Resurfacing

Authors: Catherine Van Der Straeten, MD, George Grammatopoulos, MD, Harinderjit S. Gill, BEng, DPhil, Alessandro Calistri, MD, Patricia Campbell, PhD, Koen A. De Smet, MD

Published in: Clinical Orthopaedics and Related Research® | Issue 2/2013

Login to get access

Abstract

Background

The interpretation of metal ion concentrations and their role in clinical management of patients with metal-on-metal implants is still controversial.

Questions/Purposes

We questioned whether patients undergoing hip resurfacing with no clinical problems could be differentiated from those with clinical (pain, loss of function) and/or radiographic (component malpositioning, migration, bone loss), problems based on metal ion levels, and if there was a threshold metal level that predicted the need for clinical intervention. Furthermore, we asked if patient and implant factors differed between these functional groups.

Methods

We retrospectively identified 453 unilateral and 139 bilateral patients with ion measurements at minimum followup of 12 months (mean, 4.3 years; range, 1–12.9 years). Patients were designated as well functioning or poorly functioning based on strict criteria. The acceptable upper levels within the well-functioning group were determined from the 75th percentile plus 1.5× interquartile range. The sensitivity and specificity of these levels to predict clinical problems were calculated.

Results

Well-functioning group ions were lower than the poorly functioning group ion levels. The acceptable upper levels were: chromium (Cr) 4.6 μg/L, cobalt (Co) 4.0 μg/L unilateral and Cr 7.4 μg/L, Co 5.0 μg/L bilateral. The specificity of these levels in predicting poor function was high (95%) and sensitivity was low (25%). There were more males in the well-functioning group and more females and smaller femoral components in the poorly functioning group.

Conclusions

Metal levels higher than these proposed safe upper limits can predict problems with metal-on-metal resurfacings and are important parameters in the management of at-risk patients.

Level of Evidence

Level II, diagnostic study. See Guidelines for Authors for a complete description of levels of evidence.
Literature
2.
go back to reference Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34:859–871.PubMedCrossRef Bergmann G, Deuretzbacher G, Heller M, Graichen F, Rohlmann A, Strauss J, Duda GN. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34:859–871.PubMedCrossRef
3.
go back to reference De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.PubMedCrossRef De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.PubMedCrossRef
4.
go back to reference De Smet K, De Haan R, Calistri A, Campbell PA, Ebramzadeh E, Pattyn C, Gill HS. Metal ion measurement as a diagnostic tool to identify problems with metal-on-metal hip resurfacing. J Bone Joint Surg Am. 2008;90(Suppl 4):202–208.PubMedCrossRef De Smet K, De Haan R, Calistri A, Campbell PA, Ebramzadeh E, Pattyn C, Gill HS. Metal ion measurement as a diagnostic tool to identify problems with metal-on-metal hip resurfacing. J Bone Joint Surg Am. 2008;90(Suppl 4):202–208.PubMedCrossRef
5.
go back to reference De Smet KA, Van Der Straeten C, Van Orsouw M, Doubi R, Backers K, Grammatopoulos G. Revisions of metal-on-metal hip resurfacing: lessons learned and improved outcome. Orthop Clin N Am. 2011;42:259–269.CrossRef De Smet KA, Van Der Straeten C, Van Orsouw M, Doubi R, Backers K, Grammatopoulos G. Revisions of metal-on-metal hip resurfacing: lessons learned and improved outcome. Orthop Clin N Am. 2011;42:259–269.CrossRef
7.
go back to reference Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP. The John Charnley Award: Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res. 2010;468:318–325.PubMedCrossRef Garbuz DS, Tanzer M, Greidanus NV, Masri BA, Duncan CP. The John Charnley Award: Metal-on-metal hip resurfacing versus large-diameter head metal-on-metal total hip arthroplasty: a randomized clinical trial. Clin Orthop Relat Res. 2010;468:318–325.PubMedCrossRef
8.
go back to reference Grammatopoulos G, Grammatopoulos H, Pandit Y-M, Kwon R, Gundle P, McLardy-Smith D, Beard DW, Murray HS. Gill. Hip resurfacings revised for inflammatory pseudotumour have a poor outcome. J Bone Joint Surg Br. 2009;91:1019–1024.CrossRef Grammatopoulos G, Grammatopoulos H, Pandit Y-M, Kwon R, Gundle P, McLardy-Smith D, Beard DW, Murray HS. Gill. Hip resurfacings revised for inflammatory pseudotumour have a poor outcome. J Bone Joint Surg Br. 2009;91:1019–1024.CrossRef
9.
go back to reference Griffin WL, Nanson CJ, Springer BD, Fehring TK. Reduced articular surface of one-piece cups: a cause of runaway wear and early failure. Clin Orthop Relat Res. 2010;468:2328–2332.PubMedCrossRef Griffin WL, Nanson CJ, Springer BD, Fehring TK. Reduced articular surface of one-piece cups: a cause of runaway wear and early failure. Clin Orthop Relat Res. 2010;468:2328–2332.PubMedCrossRef
10.
go back to reference Haddad FS, Thakrar RR, Hart AJ, Skinner JA, Nargol AV, Nolan JF, Gill HS, Murray DW, Blom AW, Case CP. Metal-on-metal bearings: the evidence so far. J Bone Joint Surg Br. 2011;93:572–579.PubMed Haddad FS, Thakrar RR, Hart AJ, Skinner JA, Nargol AV, Nolan JF, Gill HS, Murray DW, Blom AW, Case CP. Metal-on-metal bearings: the evidence so far. J Bone Joint Surg Br. 2011;93:572–579.PubMed
11.
go back to reference Heisel C, Streich N, Krachler M, Jakubowitz E, Kretzer JP. Characterization of the running-in period in total hip resurfacing arthroplasty: an in vivo and in vitro metal ion analysis. J Bone Joint Surg Am. 2008;90:125–133.PubMedCrossRef Heisel C, Streich N, Krachler M, Jakubowitz E, Kretzer JP. Characterization of the running-in period in total hip resurfacing arthroplasty: an in vivo and in vitro metal ion analysis. J Bone Joint Surg Am. 2008;90:125–133.PubMedCrossRef
12.
go back to reference Hodge VJ, Austin J. A survey of outlier detection methodologies. Artificial Intelligence Review. 2004;22:85–126.CrossRef Hodge VJ, Austin J. A survey of outlier detection methodologies. Artificial Intelligence Review. 2004;22:85–126.CrossRef
13.
go back to reference Kwon YM, Ostlere SJ, McLardy-Smith P, Athanasou NA, Gill HS, Murray DW. ‘Asymptomatic’ pseudotumors after metal-on-metal hip resurfacing arthroplasty: prevalence and metal ion study. J Arthroplasty. 2011;26:511–518.PubMedCrossRef Kwon YM, Ostlere SJ, McLardy-Smith P, Athanasou NA, Gill HS, Murray DW. ‘Asymptomatic’ pseudotumors after metal-on-metal hip resurfacing arthroplasty: prevalence and metal ion study. J Arthroplasty. 2011;26:511–518.PubMedCrossRef
14.
go back to reference Kwon YM, Xia Z, Glyn-Jones S, Beard D, Gill HS, Murray DW. Dose-dependent cytotoxicity of clinically relevant cobalt nanoparticles and ions on macrophages in vitro. Biomed Mater. 2009;4:025018.PubMedCrossRef Kwon YM, Xia Z, Glyn-Jones S, Beard D, Gill HS, Murray DW. Dose-dependent cytotoxicity of clinically relevant cobalt nanoparticles and ions on macrophages in vitro. Biomed Mater. 2009;4:025018.PubMedCrossRef
15.
go back to reference Langton DJ, Joyce TJ, Jameson SS, Lord J, Van Orsouw M, Holland JP, Nargol AV, De Smet KA. Adverse reaction to metal debris following hip resurfacing: the influence of component type, orientation and volumetric wear. J Bone Joint Surg Br. 2011;93:164–171.PubMedCrossRef Langton DJ, Joyce TJ, Jameson SS, Lord J, Van Orsouw M, Holland JP, Nargol AV, De Smet KA. Adverse reaction to metal debris following hip resurfacing: the influence of component type, orientation and volumetric wear. J Bone Joint Surg Br. 2011;93:164–171.PubMedCrossRef
16.
go back to reference Langton DJ, Joyce TJ, Mangat N, Lord J, Van Orsouw M, De Smet K, Nargol AV. Reducing metal ion release following hip resurfacing arthroplasty. Orthop Clin North Am. 2011;42:169–180.PubMedCrossRef Langton DJ, Joyce TJ, Mangat N, Lord J, Van Orsouw M, De Smet K, Nargol AV. Reducing metal ion release following hip resurfacing arthroplasty. Orthop Clin North Am. 2011;42:169–180.PubMedCrossRef
17.
go back to reference Langton DJ, Sprowson AP, Joyce TJ, Reed M, Carluke I, Partington P, Nargol AV. Blood metal ion concentrations after hip resurfacing arthroplasty: a comparative study of articular surface replacement and Birmingham Hip Resurfacing arthroplasties. J Bone Joint Surg Br. 2009;91:1287–1295.PubMedCrossRef Langton DJ, Sprowson AP, Joyce TJ, Reed M, Carluke I, Partington P, Nargol AV. Blood metal ion concentrations after hip resurfacing arthroplasty: a comparative study of articular surface replacement and Birmingham Hip Resurfacing arthroplasties. J Bone Joint Surg Br. 2009;91:1287–1295.PubMedCrossRef
18.
go back to reference Langton DJ, Sprowson AP, Mahadeva D, Bhatnagar S, Holland JP, Nargol AV. Cup anteversion post hip resurfacing arthroplasty: validation of EBRA and presentation of a simple clinical grading system. J Arthroplasty. 2010;25:607–613.PubMedCrossRef Langton DJ, Sprowson AP, Mahadeva D, Bhatnagar S, Holland JP, Nargol AV. Cup anteversion post hip resurfacing arthroplasty: validation of EBRA and presentation of a simple clinical grading system. J Arthroplasty. 2010;25:607–613.PubMedCrossRef
19.
go back to reference Mac Donald SJ, Brodner W, Jacobs JJ. A consensus paper on metal ions in metal-on-metal hip arthroplasties. J Arthroplasty. 2004;19(Suppl 3):12–16.CrossRef Mac Donald SJ, Brodner W, Jacobs JJ. A consensus paper on metal ions in metal-on-metal hip arthroplasties. J Arthroplasty. 2004;19(Suppl 3):12–16.CrossRef
20.
go back to reference McBryde CW, Theivendran K, Thomas A, Treacy R, Pynsent PB. The influence of head size and sex on the outcome of Birmingham Hip Resurfacing. J Bone Joint Surg Am. 2010;92:105–112.PubMedCrossRef McBryde CW, Theivendran K, Thomas A, Treacy R, Pynsent PB. The influence of head size and sex on the outcome of Birmingham Hip Resurfacing. J Bone Joint Surg Am. 2010;92:105–112.PubMedCrossRef
23.
go back to reference Nunley RM, Della Valle CJ, Barrack RL. Is patient selection important for hip resurfacing? Clin Orthop Relat Res. 2009;467:56–65.PubMedCrossRef Nunley RM, Della Valle CJ, Barrack RL. Is patient selection important for hip resurfacing? Clin Orthop Relat Res. 2009;467:56–65.PubMedCrossRef
24.
go back to reference Smolders JMH, Bisseling P, Hol A, Van Der Straeten C, Schreurs BW, van Susante JLC. Metal ion interpretation in resurfacing versus conventional hip arthroplasty and in whole blood versus serum. How should we interpret metal ion data? Hip Int. 2011;21:587–595.PubMedCrossRef Smolders JMH, Bisseling P, Hol A, Van Der Straeten C, Schreurs BW, van Susante JLC. Metal ion interpretation in resurfacing versus conventional hip arthroplasty and in whole blood versus serum. How should we interpret metal ion data? Hip Int. 2011;21:587–595.PubMedCrossRef
25.
go back to reference Tower SS. Arthroprosthetic cobaltism: neurological and cardiac manifestations in two patients with metal-on-metal arthroplasty: a case report. J Bone Joint Surg Am. 2010;92:2847–2851.PubMedCrossRef Tower SS. Arthroprosthetic cobaltism: neurological and cardiac manifestations in two patients with metal-on-metal arthroplasty: a case report. J Bone Joint Surg Am. 2010;92:2847–2851.PubMedCrossRef
26.
go back to reference Zweig MH, Campbell G. Receiver-operating characteristic (ROC) plots: a fundamental evaluation tool in clinical medicine. Clin Chem. 1993;39:561–577.PubMed Zweig MH, Campbell G. Receiver-operating characteristic (ROC) plots: a fundamental evaluation tool in clinical medicine. Clin Chem. 1993;39:561–577.PubMed
Metadata
Title
The 2012 Otto Aufranc Award: The Interpretation of Metal Ion Levels in Unilateral and Bilateral Hip Resurfacing
Authors
Catherine Van Der Straeten, MD
George Grammatopoulos, MD
Harinderjit S. Gill, BEng, DPhil
Alessandro Calistri, MD
Patricia Campbell, PhD
Koen A. De Smet, MD
Publication date
01-02-2013
Publisher
Springer-Verlag
Published in
Clinical Orthopaedics and Related Research® / Issue 2/2013
Print ISSN: 0009-921X
Electronic ISSN: 1528-1132
DOI
https://doi.org/10.1007/s11999-012-2526-x

Other articles of this Issue 2/2013

Clinical Orthopaedics and Related Research® 2/2013 Go to the issue

Symposium: Papers Presented at the Annual Meetings of The Hip Society

Should We Worry About Periacetabular Interference Gaps in Hip Resurfacing?

Symposium: Papers Presented at the Annual Meetings of The Hip Society

The Shape of the Proximal Femur Influences Acetabular Wear Patterns Over Time

Symposium: Papers Presented at the Annual Meetings of The Hip Society

Patients Report Improvement in Quality of Life and Satisfaction After Hip Resurfacing Arthroplasty