Published in: Archives of Orthopaedic and Trauma Surgery 3/2012

01-03-2012 | Arthroscopy and Sports Medicine

Osteolysis after rotator cuff repair with bioabsorbable anchors

Authors: H. Pilge, J. Spang, T. Rose, H. Wolter, K. Woertler, A. B. Imhoff

Published in: Archives of Orthopaedic and Trauma Surgery | Issue 3/2012

Login to get access

Abstract

Purpose

The current study was designed to evaluate the rate of osteolysis and the clinical and functional outcomes in patients who underwent mini-open rotator cuff repairs using first generation bioabsorbable suture anchors. Our hypothesis was that patients with osteolysis noted on post-repair MRI would have an accompanying decrease in functional and clinical patient outcomes when compared with patients who did not have osteolysis evident on post-repair MRI.

Type of study

Clinical retrospective study.

Methods

Between September 2000 and May 2004, 76 patients were repaired using first generation Bio-Corkscrew suture anchors (Arthrex, Naples, FL). 30 patients were available for complete follow-up evaluation. The mean follow-up time was 36 months (range 24–58). Patients were assessed with the Constant–Murley Score, radiographs and magnetic resonance imaging (MRI). Post-operative strength was measured using an Isokinetic Cybex-Dynamometer. 14 patients had osteolysis evident on MRI (Group A) while 16 patients did not have osteolysis on MRI (Group B).

Results

In both groups, a total of 70 bio-anchors were used. 22 anchors showed osteolytic changes on MRI (Group A). Concerning ROM, there was no statistical difference in both groups. The Constant-Score was statistically not different in both groups A (92.4) and B (83.7). On MRI-scans, there were repair failures in both groups (A, 2 reruptures; B, 3 reruptures).

Conclusions

Although, we found a high rate of osteolysis after rotator cuff repair with bioabsorbable anchors, these results did not change the clinical outcomes after rotator cuff repair. Recurrent tears were not significantly different in both groups and are comparable to rerupture rates in prior studies. More studies are needed to verify the effect of osteolysis over the long term.
Literature
1.
go back to reference Ammon JT, Nyland J, Chang HC, Burden R, Caborn DN (2007) Evaluation of BioCorkscrew and Bioknotless RC suture anchor rotator cuff repair fixation: an in vitro biomechanical study. Knee Surg Sports Traumatol Arthrosc 15(11):1375–1381PubMedCrossRef Ammon JT, Nyland J, Chang HC, Burden R, Caborn DN (2007) Evaluation of BioCorkscrew and Bioknotless RC suture anchor rotator cuff repair fixation: an in vitro biomechanical study. Knee Surg Sports Traumatol Arthrosc 15(11):1375–1381PubMedCrossRef
2.
go back to reference Barber FA, Feder SM, Burkhart SS, Ahrens J (1997) The relationship of suture anchor failure and bone density to proximal humerus location: a cadaveric study. Arthroscopy 13(3):340–345PubMedCrossRef Barber FA, Feder SM, Burkhart SS, Ahrens J (1997) The relationship of suture anchor failure and bone density to proximal humerus location: a cadaveric study. Arthroscopy 13(3):340–345PubMedCrossRef
3.
go back to reference Barber FA, Herbert MA, Coons DA, Boothby MH (2006) Sutures and suture anchors—update 2006. Arthroscopy 22(10):1063PubMedCrossRef Barber FA, Herbert MA, Coons DA, Boothby MH (2006) Sutures and suture anchors—update 2006. Arthroscopy 22(10):1063PubMedCrossRef
4.
go back to reference Bishop J, Klepps S, Lo IK, Bird J, Gladstone JN, Flatow EL (2006) Cuff integrity after arthroscopic versus open rotator cuff repair: a prospective study. J Should Elb Surg 15:290–299CrossRef Bishop J, Klepps S, Lo IK, Bird J, Gladstone JN, Flatow EL (2006) Cuff integrity after arthroscopic versus open rotator cuff repair: a prospective study. J Should Elb Surg 15:290–299CrossRef
5.
go back to reference Bostman O, Pihlajamaki H (2000) Adverse tissue reactions to bioabsorbable fixation devices. Clin Orthop Relat Res 371:216–227PubMedCrossRef Bostman O, Pihlajamaki H (2000) Adverse tissue reactions to bioabsorbable fixation devices. Clin Orthop Relat Res 371:216–227PubMedCrossRef
6.
go back to reference Bravman JT, Guttmann D, Rokito AS, Kummer FJ, Jazrawi LM (2006) A biodegradable button to augment suture attachment in rotator cuff repair. Bull Hosp Jt Dis 63(3–4):126–128PubMed Bravman JT, Guttmann D, Rokito AS, Kummer FJ, Jazrawi LM (2006) A biodegradable button to augment suture attachment in rotator cuff repair. Bull Hosp Jt Dis 63(3–4):126–128PubMed
7.
go back to reference Burkart A, Imhoff AB (2000) Foreign-body reaction to the bioabsorbable suretac device. Arthroscopy 16:91–95PubMedCrossRef Burkart A, Imhoff AB (2000) Foreign-body reaction to the bioabsorbable suretac device. Arthroscopy 16:91–95PubMedCrossRef
8.
go back to reference Burkhart SS (2005) Comment on: case report by Drs. Glueck, Wilson, and Johnson entitled “Extensive osteolysis after rotator cuff repair with a bioabsorbable suture anchor”(May 2005, pages 742–744). Am J Sports Med 33(11):1768PubMedCrossRef Burkhart SS (2005) Comment on: case report by Drs. Glueck, Wilson, and Johnson entitled “Extensive osteolysis after rotator cuff repair with a bioabsorbable suture anchor”(May 2005, pages 742–744). Am J Sports Med 33(11):1768PubMedCrossRef
9.
go back to reference Constant CR, Murley AMG (1987) A clinical method of functional assessment of the shoulder. Clin Orthop Relat Res 214:160–164PubMed Constant CR, Murley AMG (1987) A clinical method of functional assessment of the shoulder. Clin Orthop Relat Res 214:160–164PubMed
10.
go back to reference Fuchs B, Gilbart MK, Hodler J, Gerber C (2006) Clinical and structural results of an open repair of an isolated one-tendon tear of the rotator cuff. J Bone Jt Surg Am 88(2):309–316CrossRef Fuchs B, Gilbart MK, Hodler J, Gerber C (2006) Clinical and structural results of an open repair of an isolated one-tendon tear of the rotator cuff. J Bone Jt Surg Am 88(2):309–316CrossRef
11.
go back to reference Gazielly DF, Gleyze P, Montagnon C (1994) Functional and anatomical results after rotator cuff repair. Clin Orthop 304:43–53PubMed Gazielly DF, Gleyze P, Montagnon C (1994) Functional and anatomical results after rotator cuff repair. Clin Orthop 304:43–53PubMed
12.
go back to reference Glueck D, Wilson TC, Johnson DL (2005) Extensive osteolysis after rotator cuff repair with a bioabsorbable suture anchor: a case report. Am J Sports Med 33(5):742–744PubMedCrossRef Glueck D, Wilson TC, Johnson DL (2005) Extensive osteolysis after rotator cuff repair with a bioabsorbable suture anchor: a case report. Am J Sports Med 33(5):742–744PubMedCrossRef
13.
go back to reference Goutallier D, Postel JM, Bernageau J, Lavau L, Voisin MC (1994) Fatty muscle degeneration in cuff ruptures. Pre- and postoperative evaluation by CT scan. Clin Orthop Relat Res 304:78–83PubMed Goutallier D, Postel JM, Bernageau J, Lavau L, Voisin MC (1994) Fatty muscle degeneration in cuff ruptures. Pre- and postoperative evaluation by CT scan. Clin Orthop Relat Res 304:78–83PubMed
14.
go back to reference Jost B, Pfirrmann CW, Gerber C (2000) Clinical outcome after structural failure of rotator cuff repairs. J Bone Jt Surg Am 82(3):304–314 Jost B, Pfirrmann CW, Gerber C (2000) Clinical outcome after structural failure of rotator cuff repairs. J Bone Jt Surg Am 82(3):304–314
15.
go back to reference Jost B, Zumstein M, Pfirrmann CW, Gerber C (2006) Long-term outcome after structural failure of rotator cuff repairs. J Bone Jt Surg Am 88(3):472–479CrossRef Jost B, Zumstein M, Pfirrmann CW, Gerber C (2006) Long-term outcome after structural failure of rotator cuff repairs. J Bone Jt Surg Am 88(3):472–479CrossRef
16.
go back to reference Klinger HM, Steckel H, Spahn G, Buchhorn GH, Baums MH (2007) Biomechanical comparison of double-loaded suture anchors using arthroscopic Mason-Allen stitches versus traditional transosseous suture technique and modified Mason-Allen stitches for rotator cuff repair. Clin Biomech (Bristol, Avon) 22:106–111CrossRef Klinger HM, Steckel H, Spahn G, Buchhorn GH, Baums MH (2007) Biomechanical comparison of double-loaded suture anchors using arthroscopic Mason-Allen stitches versus traditional transosseous suture technique and modified Mason-Allen stitches for rotator cuff repair. Clin Biomech (Bristol, Avon) 22:106–111CrossRef
17.
go back to reference Lorbach O, Wilmes P, Brogard P, Seil R (2008) Complications related to implants in arthroscopic shoulder surgery. Orthopade 37:1073–1079PubMedCrossRef Lorbach O, Wilmes P, Brogard P, Seil R (2008) Complications related to implants in arthroscopic shoulder surgery. Orthopade 37:1073–1079PubMedCrossRef
18.
go back to reference Moser RC, McManus AJ, Riley SL, Thomas KA (2005) Strength retention of 70:30 poly(l-lactide-co-D, l-lactide) following real-time aging. J Biomed Mater Res B Appl Biomater 75(1):56–63PubMed Moser RC, McManus AJ, Riley SL, Thomas KA (2005) Strength retention of 70:30 poly(l-lactide-co-D, l-lactide) following real-time aging. J Biomed Mater Res B Appl Biomater 75(1):56–63PubMed
19.
go back to reference Müller M, Kääb MJ, Holzach P (2002) Osteolysis after open shoulder stabilization using a new bio-resorbable bone anchor: a prospective, non-randomized clinical trial. Injury 33(Suppl 2):B30–B36PubMedCrossRef Müller M, Kääb MJ, Holzach P (2002) Osteolysis after open shoulder stabilization using a new bio-resorbable bone anchor: a prospective, non-randomized clinical trial. Injury 33(Suppl 2):B30–B36PubMedCrossRef
21.
go back to reference Patte D (1990) Classification of rotator cuff lesions. Clin Orthop 254:55–59PubMed Patte D (1990) Classification of rotator cuff lesions. Clin Orthop 254:55–59PubMed
22.
go back to reference Pihlajamaki H, Bostman O, Tynninen O, Laitinen O (2006) Long-term tissue response to bioabsorbable poly-l-lactide and metallic screws: an experimental study. Bone 39(4):932–937 epub 2006 JunPubMedCrossRef Pihlajamaki H, Bostman O, Tynninen O, Laitinen O (2006) Long-term tissue response to bioabsorbable poly-l-lactide and metallic screws: an experimental study. Bone 39(4):932–937 epub 2006 JunPubMedCrossRef
23.
go back to reference Rehm KE, Helling HJ, Claes L (1995) Report of the working group on biodegradable implants. Akt Traumatol 1995:25 Rehm KE, Helling HJ, Claes L (1995) Report of the working group on biodegradable implants. Akt Traumatol 1995:25
24.
go back to reference Schneeberger AG, Roll Av, Kalberer F, Jacob HA, Gerber C (2002) Mechanical strength of arthroscopic rotator cuff repair techniques: an in vitro study. J Bone Jt Surg Am 84-A(12):2152–2160 Schneeberger AG, Roll Av, Kalberer F, Jacob HA, Gerber C (2002) Mechanical strength of arthroscopic rotator cuff repair techniques: an in vitro study. J Bone Jt Surg Am 84-A(12):2152–2160
25.
go back to reference Shall LM, Cawley PW (1994) Soft tissue reconstruction in the shoulder. Comparison of suture anchors, absorbable staples, and absorbable tacks. Am J Sports Med 22(5):715–718PubMedCrossRef Shall LM, Cawley PW (1994) Soft tissue reconstruction in the shoulder. Comparison of suture anchors, absorbable staples, and absorbable tacks. Am J Sports Med 22(5):715–718PubMedCrossRef
26.
go back to reference Snyder SJ (1993) Evaluation and treatment of the rotator cuff. Orthop Clin North Am 24:173–192PubMed Snyder SJ (1993) Evaluation and treatment of the rotator cuff. Orthop Clin North Am 24:173–192PubMed
27.
go back to reference Spoliti M (2007) Glenoid osteolysis after arthroscopic labrum repair with a bioabsorbable suture anchor. Acta Orthop Belg 73:107–110PubMed Spoliti M (2007) Glenoid osteolysis after arthroscopic labrum repair with a bioabsorbable suture anchor. Acta Orthop Belg 73:107–110PubMed
28.
go back to reference Thomazeau H, Boukobza E, Morcet N, Chaperon J, Langlais F (1997) Prediction of rotator cuff repair results by magnetic resonance imaging. Clin Orthop 344:275–283PubMed Thomazeau H, Boukobza E, Morcet N, Chaperon J, Langlais F (1997) Prediction of rotator cuff repair results by magnetic resonance imaging. Clin Orthop 344:275–283PubMed
Metadata
Title
Osteolysis after rotator cuff repair with bioabsorbable anchors
Authors
H. Pilge
J. Spang
T. Rose
H. Wolter
K. Woertler
A. B. Imhoff
Publication date
01-03-2012
Publisher
Springer-Verlag
Published in
Archives of Orthopaedic and Trauma Surgery / Issue 3/2012
Print ISSN: 0936-8051
Electronic ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-011-1369-3

Other articles of this Issue 3/2012

Archives of Orthopaedic and Trauma Surgery 3/2012 Go to the issue

Arthroscopy and Sports Medicine

Isolated AL bundle reconstruction of the PCL