Open Access 01-11-2020 | Silicone | Experimental

Feasibility of unconstrained three-material decomposition: imaging an excised human heart using a prototype silicon photon-counting CT detector

Authors: Fredrik Grönberg, Johan Lundberg, Martin Sjölin, Mats Persson, Robert Bujila, Hans Bornefalk, Håkan Almqvist, Staffan Holmin, Mats Danielsson

Published in: European Radiology | Issue 11/2020

Login to get access

Abstract

Rationale and objectives

The purpose of this study was to evaluate the feasibility of unconstrained three-material decomposition in a human tissue specimen containing iodinated contrast agent, using an experimental multi-bin photon-counting silicon detector. It was further to evaluate potential added clinical value compared to a 1st-generation state-of-the-art dual-energy computed tomography system.

Materials and methods

A prototype photon-counting silicon detector in a bench-top setup for x-ray tomographic imaging was calibrated using a multi-material calibration phantom. A heart with calcified plaque was obtained from a deceased patient, and the coronary arteries were injected with an iodinated contrast agent mixed with gelatin. The heart was imaged in the experimental setup and on a 1st-generation state-of-the-art dual-energy computed tomography system. Projection-based three-material decomposition without any constraints was performed with the photon-counting detector data, and the resulting images were compared with those obtained from the dual-energy system.

Results

The photon-counting detector images show better separation of iodine and calcium compared to the dual-energy images. Additional experiments confirmed that unbiased estimates of soft tissue, calcium, and iodine could be achieved without any constraints.

Conclusion

The proposed experimental system could provide added clinical value compared to current dual-energy systems for imaging tasks where mix-up of iodine and calcium is an issue, and the anatomy is sufficiently small to allow iodine to be differentiated from calcium. Considering its previously shown count rate capability, these results show promise for future integration of this detector in a clinical CT scanner.

Key Points

• Spectral photon-counting detectors can solve some of the fundamental problems with conventional single-energy CT.
• Dual-energy methods can be used to differentiate iodine and calcium, but to do so must rely on constraints, since solving for three unknowns with only two measurements is not possible. Photon-counting detectors can improve upon these methods by allowing unconstrained three-material decomposition.
• A prototype photon-counting silicon detector with high count rate capability allows performing unconstrained three-material decomposition and qualitatively shows better differentiation of iodine and calcium than dual-energy CT.
Literature
1.
go back to reference Swank RK (1973) Absorption and noise in x-ray phosphors. J Appl Phys 44(9):4199–4203 Swank RK (1973) Absorption and noise in x-ray phosphors. J Appl Phys 44(9):4199–4203
3.
go back to reference Alvarez RE, Macovski A (1976) Energy-selective reconstructions in x-ray computerised tomography. Phys Med Biol 21(5):733PubMed Alvarez RE, Macovski A (1976) Energy-selective reconstructions in x-ray computerised tomography. Phys Med Biol 21(5):733PubMed
4.
go back to reference Yveborg M, Danielsson M, Bornefalk H (2014) Theoretical comparison of a dual energy system and photon counting silicon detector used for material quantification in spectral CT. IEEE Trans Med Imaging 34(3):796–806PubMed Yveborg M, Danielsson M, Bornefalk H (2014) Theoretical comparison of a dual energy system and photon counting silicon detector used for material quantification in spectral CT. IEEE Trans Med Imaging 34(3):796–806PubMed
5.
go back to reference Roessl E, Brendel B, Engel KJ, Schlomka JP, Thran A, Proksa R (2011) Sensitivity of photon-counting based K-edge imaging in X-ray computed tomography. IEEE Trans Med Imaging 30(9):1678–1690PubMed Roessl E, Brendel B, Engel KJ, Schlomka JP, Thran A, Proksa R (2011) Sensitivity of photon-counting based K-edge imaging in X-ray computed tomography. IEEE Trans Med Imaging 30(9):1678–1690PubMed
6.
go back to reference Si-Mohamed S, Bar-Ness D, Sigovan M et al (2017) Review of an initial experience with an experimental spectral photon-counting computed tomography system. Nucl Inst Methods Phys Res A 873(1):27–35 Si-Mohamed S, Bar-Ness D, Sigovan M et al (2017) Review of an initial experience with an experimental spectral photon-counting computed tomography system. Nucl Inst Methods Phys Res A 873(1):27–35
7.
go back to reference Swap CJ, Nagurney JT (2005) Value and limitations of chest pain history in the evaluation of patients with suspected acute coronary syndromes. JAMA 294(20):2623–2629PubMed Swap CJ, Nagurney JT (2005) Value and limitations of chest pain history in the evaluation of patients with suspected acute coronary syndromes. JAMA 294(20):2623–2629PubMed
8.
go back to reference Fesmire FM, Hughes AD, Fody EP et al (2002) The Erlanger chest pain evaluation protocol: a one-year experience with serial 12-lead ECG monitoring, two-hour delta serum marker measurements, and selective nuclear stress testing to identify and exclude acute coronary syndromes. Ann Emerg Med 40(6):584–594PubMed Fesmire FM, Hughes AD, Fody EP et al (2002) The Erlanger chest pain evaluation protocol: a one-year experience with serial 12-lead ECG monitoring, two-hour delta serum marker measurements, and selective nuclear stress testing to identify and exclude acute coronary syndromes. Ann Emerg Med 40(6):584–594PubMed
9.
go back to reference Greenland P, LaBree L, Azen SP, Doherty TM, Detrano RC (2004) Coronary artery calcium score combined with Framingham score for risk prediction in asymptomatic individuals. JAMA 291(2):210–215PubMed Greenland P, LaBree L, Azen SP, Doherty TM, Detrano RC (2004) Coronary artery calcium score combined with Framingham score for risk prediction in asymptomatic individuals. JAMA 291(2):210–215PubMed
10.
go back to reference Hoffmann U, Bamberg F, Chae CU et al (2009) Coronary computed tomography angiography for early triage of patients with acute chest pain: the ROMICAT (rule out myocardial infarction using computer assisted tomography) trial. J Am Coll Cardiol 53(18):1642–1650PubMedPubMedCentral Hoffmann U, Bamberg F, Chae CU et al (2009) Coronary computed tomography angiography for early triage of patients with acute chest pain: the ROMICAT (rule out myocardial infarction using computer assisted tomography) trial. J Am Coll Cardiol 53(18):1642–1650PubMedPubMedCentral
11.
go back to reference Erbel R, Budoff M (2012) Improvement of cardiovascular risk prediction using coronary imaging: subclinical atherosclerosis: the memory of lifetime risk factor exposure. Eur Heart J 33(10):1201–1213PubMed Erbel R, Budoff M (2012) Improvement of cardiovascular risk prediction using coronary imaging: subclinical atherosclerosis: the memory of lifetime risk factor exposure. Eur Heart J 33(10):1201–1213PubMed
12.
go back to reference Schaar JA, Muller JE, Falk E et al (2004) Terminology for high-risk and vulnerable coronary artery plaques. Eur Heart J 25(12):1077–1082PubMed Schaar JA, Muller JE, Falk E et al (2004) Terminology for high-risk and vulnerable coronary artery plaques. Eur Heart J 25(12):1077–1082PubMed
13.
go back to reference Sanz J, Fayad ZA (2008) Imaging of atherosclerotic cardiovascular disease. Nature 451(7181):953PubMed Sanz J, Fayad ZA (2008) Imaging of atherosclerotic cardiovascular disease. Nature 451(7181):953PubMed
14.
go back to reference Yu Z, Leng S, Jorgensen SM et al (2016) Evaluation of conventional imaging performance in a research whole-body CT system with a photon-counting detector array. Phys Med Biol 61(4):1572PubMedPubMedCentral Yu Z, Leng S, Jorgensen SM et al (2016) Evaluation of conventional imaging performance in a research whole-body CT system with a photon-counting detector array. Phys Med Biol 61(4):1572PubMedPubMedCentral
15.
go back to reference Pourmorteza A, Symons R, Sandfort V et al (2016) Abdominal imaging with contrast-enhanced photon-counting CT: first human experience. Radiology 279(1):239–245PubMedPubMedCentral Pourmorteza A, Symons R, Sandfort V et al (2016) Abdominal imaging with contrast-enhanced photon-counting CT: first human experience. Radiology 279(1):239–245PubMedPubMedCentral
16.
go back to reference Gutjahr R, Halaweish AF, Yu Z et al (2016) Human imaging with photon-counting-based CT at clinical dose levels: contrast-to- noise ratio and cadaver studies. Invest Radiol 51(7):421PubMedPubMedCentral Gutjahr R, Halaweish AF, Yu Z et al (2016) Human imaging with photon-counting-based CT at clinical dose levels: contrast-to- noise ratio and cadaver studies. Invest Radiol 51(7):421PubMedPubMedCentral
17.
go back to reference Symons R, Pourmorteza A, Sandfort V et al (2017) Feasibility of dose-reduced chest CT with photon-counting detectors: initial results in humans. Radiology 285(3):980–989PubMedPubMedCentral Symons R, Pourmorteza A, Sandfort V et al (2017) Feasibility of dose-reduced chest CT with photon-counting detectors: initial results in humans. Radiology 285(3):980–989PubMedPubMedCentral
18.
go back to reference Symons R, Reich DS, Bagheri M et al (2018) Photon-counting computed tomography for vascular imaging of the head and neck: first in vivo human results. Invest Radiol 53(3):135–142PubMedPubMedCentral Symons R, Reich DS, Bagheri M et al (2018) Photon-counting computed tomography for vascular imaging of the head and neck: first in vivo human results. Invest Radiol 53(3):135–142PubMedPubMedCentral
19.
go back to reference Bornefalk H, Danielsson M (2010) Photon-counting spectral computed tomography using silicon strip detectors: a feasibility study. Phys Med Biol 55(7):1999PubMed Bornefalk H, Danielsson M (2010) Photon-counting spectral computed tomography using silicon strip detectors: a feasibility study. Phys Med Biol 55(7):1999PubMed
20.
go back to reference Persson M, Huber B, Karlsson S et al (2014) Energy-resolved CT imaging with a photon-counting silicon-strip detector. Phys Med Biol 59(22):6709PubMed Persson M, Huber B, Karlsson S et al (2014) Energy-resolved CT imaging with a photon-counting silicon-strip detector. Phys Med Biol 59(22):6709PubMed
21.
go back to reference Liu X, Persson M, Bornefalk H et al (2015) Spectral response model for a multibin photon-counting spectral computed tomography detector and its applications. J Med Imaging (Bellingham) 2(3):033502 Liu X, Persson M, Bornefalk H et al (2015) Spectral response model for a multibin photon-counting spectral computed tomography detector and its applications. J Med Imaging (Bellingham) 2(3):033502
22.
go back to reference Wielopolski L, Gardner RP (1976) Prediction of the pulse-height spectral distortion caused by the peak pile-up effect. Nucl Inst Methods 133(2):303–309 Wielopolski L, Gardner RP (1976) Prediction of the pulse-height spectral distortion caused by the peak pile-up effect. Nucl Inst Methods 133(2):303–309
23.
go back to reference Liu X, Grönberg F, Sjölin M, Karlsson S, Danielsson M (2016) Count rate performance of a silicon-strip detector for photon-counting spectral CT. Nucl Instrum Methods Phys Res, Sect A 827:102–106 Liu X, Grönberg F, Sjölin M, Karlsson S, Danielsson M (2016) Count rate performance of a silicon-strip detector for photon-counting spectral CT. Nucl Instrum Methods Phys Res, Sect A 827:102–106
24.
go back to reference Liu X, Bornefalk H, Chen H et al (2014) A silicon-strip detector for photon-counting spectral CT: energy resolution from 40 keV to 120 keV. IEEE Trans Nucl Sci 61(3):1099–1105 Liu X, Bornefalk H, Chen H et al (2014) A silicon-strip detector for photon-counting spectral CT: energy resolution from 40 keV to 120 keV. IEEE Trans Nucl Sci 61(3):1099–1105
25.
go back to reference Persson M, Pelc NJ (2019) Simulation model for evaluating energy-resolving photon-counting CT detectors based on generalized linear-systems frame-work. Medical Imaging 2019: Physics of Medical Imaging 10948:109481V. International Society for Optics and Photonics Persson M, Pelc NJ (2019) Simulation model for evaluating energy-resolving photon-counting CT detectors based on generalized linear-systems frame-work. Medical Imaging 2019: Physics of Medical Imaging 10948:109481V. International Society for Optics and Photonics
26.
go back to reference Schlomka J, Roessl E, Dorscheid R et al (2008) Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography. Phys Med Biol 53(14):4031PubMed Schlomka J, Roessl E, Dorscheid R et al (2008) Experimental feasibility of multi-energy photon-counting K-edge imaging in pre-clinical computed tomography. Phys Med Biol 53(14):4031PubMed
27.
go back to reference Ronaldson JP, Zainon R, Scott NJA et al (2012) Toward quantifying the composition of soft tissues by spectral CT with Medipix3. Med Phys 39(11):6847–6857PubMed Ronaldson JP, Zainon R, Scott NJA et al (2012) Toward quantifying the composition of soft tissues by spectral CT with Medipix3. Med Phys 39(11):6847–6857PubMed
28.
go back to reference Alessio AM, MacDonald LR (2013) Quantitative material characterization from multi-energy photon counting CT. Med Phys 40(3):031108PubMedPubMedCentral Alessio AM, MacDonald LR (2013) Quantitative material characterization from multi-energy photon counting CT. Med Phys 40(3):031108PubMedPubMedCentral
29.
go back to reference Muenzel D, Bar-Ness D, Roessl E et al (2017) Spectral photon-counting CT: initial experience with dual–contrast agent K-edge colonography. Radiology 283(3):723–728PubMed Muenzel D, Bar-Ness D, Roessl E et al (2017) Spectral photon-counting CT: initial experience with dual–contrast agent K-edge colonography. Radiology 283(3):723–728PubMed
30.
go back to reference Si-Mohamed S, Bar-Ness D, Sigovan M et al (2018) Multicolour imaging with spectral photon-counting CT: a phantom study. Eur Radiol Exp 2(1):1–10 Si-Mohamed S, Bar-Ness D, Sigovan M et al (2018) Multicolour imaging with spectral photon-counting CT: a phantom study. Eur Radiol Exp 2(1):1–10
31.
go back to reference Curtis TE, Roeder RK (2019) Quantification of multiple mixed contrast and tissue compositions using photon-counting spectral computed tomography. J Med Imaging (Bellingham) 6(1):013501 Curtis TE, Roeder RK (2019) Quantification of multiple mixed contrast and tissue compositions using photon-counting spectral computed tomography. J Med Imaging (Bellingham) 6(1):013501
32.
go back to reference Boussel L, Coulon P, Thran A et al (2014) Photon counting spectral CT component analysis of coronary artery atherosclerotic plaque samples. Br J Radiol 87(1040):20130798PubMedPubMedCentral Boussel L, Coulon P, Thran A et al (2014) Photon counting spectral CT component analysis of coronary artery atherosclerotic plaque samples. Br J Radiol 87(1040):20130798PubMedPubMedCentral
33.
go back to reference Symons R, Cork TE, Lakshmanan MN et al (2017) Dual-contrast agent photon-counting computed tomography of the heart: initial experience. Int J Cardiovasc Imaging 33(8):1253–1261PubMed Symons R, Cork TE, Lakshmanan MN et al (2017) Dual-contrast agent photon-counting computed tomography of the heart: initial experience. Int J Cardiovasc Imaging 33(8):1253–1261PubMed
34.
go back to reference Symons R, Krauss B, Sahbaee P et al (2017) Photon-counting CT for simultaneous imaging of multiple contrast agents in the abdomen: an in vivo study. Med Phys 44(10):5120–5127PubMedPubMedCentral Symons R, Krauss B, Sahbaee P et al (2017) Photon-counting CT for simultaneous imaging of multiple contrast agents in the abdomen: an in vivo study. Med Phys 44(10):5120–5127PubMedPubMedCentral
35.
go back to reference Xu C, Persson M, Chen H et al (2013) Evaluation of a second-generation ultra-fast energy-resolved ASIC for photon-counting spectral CT. IEEE Trans Nucl Sci 60(1):437–445 Xu C, Persson M, Chen H et al (2013) Evaluation of a second-generation ultra-fast energy-resolved ASIC for photon-counting spectral CT. IEEE Trans Nucl Sci 60(1):437–445
36.
go back to reference Ehn S, Sellerer T, Mechlem K et al (2016) Basis material decomposition in spectral CT using a semi-empirical, polychromatic adaption of the Beer–Lambert model. Phys Med Biol 62:N1–N17PubMed Ehn S, Sellerer T, Mechlem K et al (2016) Basis material decomposition in spectral CT using a semi-empirical, polychromatic adaption of the Beer–Lambert model. Phys Med Biol 62:N1–N17PubMed
37.
go back to reference Brambilla A, Gorecki A, Potop A, Paulus C, Verger L (2017) Basis material decomposition method for material discrimination with a new spectrometric X-ray imaging detector. J Instrum 12(08):P08014 Brambilla A, Gorecki A, Potop A, Paulus C, Verger L (2017) Basis material decomposition method for material discrimination with a new spectrometric X-ray imaging detector. J Instrum 12(08):P08014
38.
go back to reference Berger MJ, Hubbell JH, Seltzer SM et al (2010) XCOM: Photon Cross Section Database (version 1.5). National Institute of Standards and Technology, Gaithersburg, MD. Available via: http://physics.nist.gov/xcom. Accessed 17 Oct 2019 Berger MJ, Hubbell JH, Seltzer SM et al (2010) XCOM: Photon Cross Section Database (version 1.5). National Institute of Standards and Technology, Gaithersburg, MD. Available via: http://​physics.​nist.​gov/​xcom. Accessed 17 Oct 2019
39.
go back to reference Roessl E, Proksa R (2007) K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors. Phys Med Biol 52(15):4679PubMed Roessl E, Proksa R (2007) K-edge imaging in x-ray computed tomography using multi-bin photon counting detectors. Phys Med Biol 52(15):4679PubMed
40.
go back to reference Persson M, Bujila R, Nowik P et al (2016) Upper limits of the photon fluence rate on CT detectors: case study on a commercial scanner. Med Phys 43(7):4398–4411PubMed Persson M, Bujila R, Nowik P et al (2016) Upper limits of the photon fluence rate on CT detectors: case study on a commercial scanner. Med Phys 43(7):4398–4411PubMed
41.
go back to reference Baek J, Pineda AR, Pelc NJ (2013) To bin or not to bin? The effect of CT system limiting resolution on noise and detectability. Phys Med Biol 58(5):1433PubMed Baek J, Pineda AR, Pelc NJ (2013) To bin or not to bin? The effect of CT system limiting resolution on noise and detectability. Phys Med Biol 58(5):1433PubMed
42.
go back to reference Sarwar A, Rieber J, Mooyaart EA et al (2008) Calcified plaque: measurement of area at thin-section flat-panel CT and 64-section multidetector CT and comparison with histopathologic findings. Radiology 249(1):301–306PubMed Sarwar A, Rieber J, Mooyaart EA et al (2008) Calcified plaque: measurement of area at thin-section flat-panel CT and 64-section multidetector CT and comparison with histopathologic findings. Radiology 249(1):301–306PubMed
43.
go back to reference Davies RJ (1990) The relationship between neck circumference, radiographic pharyngeal anatomy, and the obstructive sleep apnoea syndrome. Eur Respir J 3(5):509–514PubMed Davies RJ (1990) The relationship between neck circumference, radiographic pharyngeal anatomy, and the obstructive sleep apnoea syndrome. Eur Respir J 3(5):509–514PubMed
44.
go back to reference Azevedo IG, Holanda NS, Arrais NM, Santos RT, Araujo AG, Pereira SA (2019) Chest circumference in full-term newborns: how can it be predicted? BMC Pediatr 19(1):341PubMedPubMedCentral Azevedo IG, Holanda NS, Arrais NM, Santos RT, Araujo AG, Pereira SA (2019) Chest circumference in full-term newborns: how can it be predicted? BMC Pediatr 19(1):341PubMedPubMedCentral
Metadata
Title
Feasibility of unconstrained three-material decomposition: imaging an excised human heart using a prototype silicon photon-counting CT detector
Authors
Fredrik Grönberg
Johan Lundberg
Martin Sjölin
Mats Persson
Robert Bujila
Hans Bornefalk
Håkan Almqvist
Staffan Holmin
Mats Danielsson
Publication date
01-11-2020
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 11/2020
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-020-07017-y

Other articles of this Issue 11/2020

European Radiology 11/2020 Go to the issue