Published in: European Radiology 12/2016

01-12-2016 | Computed Tomography

Visualisation of passive middle ear implants by cone beam and multi-detector computed tomography: a comparative in vitro study

Authors: T. D. Nguyen, S. Kösling, R. Mlynski, S. K. Plontke

Published in: European Radiology | Issue 12/2016

Login to get access

Abstract

Objectives

Modern passive middle ear titanium prostheses are filigree structures, resulting in poorer depiction on CT compared to prostheses used in the past. We compared the visibility of newer prostheses on cone beam CT (CBCT) to multi-detector CT (MDCT) with standard and lower dose in vitro, and analysed image noise and metal artefacts.

Methods

Six different titanium middle ear prostheses (three partial and one total ossicular replacement prostheses, two stapes prostheses) were implanted twice in formalin-fixed head specimens—first correctly and then with displacement. Imaging was performed using standard CBCT and MDCT as well as MDCT with lower dose (36 single imaging investigations). Images were analysed with knowledge of the used types of middle ear prostheses, but blinded with respect to the positioning in the specific case.

Results

On all images the type of prostheses and their positions could be clearly recognized. Their identifiability including their details was rated as statistically significantly higher for all CBCT investigations than for MDCT. MDCT with lower dose showed the worst results. No statistical differences were found in image noise and metal artefacts.

Conclusions

If available, CBCT should be preferred over MDCT in diagnostic evaluation of passive middle ear prostheses.

Key Points

Middle ear prostheses became more filigree, leading to poorer visibility on CT.
High spatial resolution and paraxial reconstructions are necessary requirements for imaging evaluation.
CBCT and MDCT can identify type and positioning of titanium prostheses.
Metal artefacts play a minor part in filigree titanium prostheses.
Regarding visualisation of prostheses details, cone beam CT aids the evaluation.
Literature
1.
go back to reference Kösling S, Bootz F (2001) CT and MR imaging after middle ear surgery. Eur J Radiol 40(2):113–118CrossRefPubMed Kösling S, Bootz F (2001) CT and MR imaging after middle ear surgery. Eur J Radiol 40(2):113–118CrossRefPubMed
2.
go back to reference Beutner D, Hüttenbrink KB (2009) Passive and active middle ear implants. Laryngorhinootologie 88(Suppl 1):32–47CrossRef Beutner D, Hüttenbrink KB (2009) Passive and active middle ear implants. Laryngorhinootologie 88(Suppl 1):32–47CrossRef
3.
go back to reference Jargalkhuu E, Erdenechuluun B, Mlynski R, Hagen R (2009) Middle ear surgery in Mongolia-first results with middle ear prostheses. Laryngorhinootologie 88(6):387–391CrossRefPubMed Jargalkhuu E, Erdenechuluun B, Mlynski R, Hagen R (2009) Middle ear surgery in Mongolia-first results with middle ear prostheses. Laryngorhinootologie 88(6):387–391CrossRefPubMed
4.
go back to reference Shah KD, Bradoo RA, Joshi AA, Sapkale DD (2013) The efficiency of titanium middle ear prosthesis in ossicular chain reconstruction: our experience. Indian J Otolaryngol Head Neck Surg 65(4):298–301CrossRefPubMed Shah KD, Bradoo RA, Joshi AA, Sapkale DD (2013) The efficiency of titanium middle ear prosthesis in ossicular chain reconstruction: our experience. Indian J Otolaryngol Head Neck Surg 65(4):298–301CrossRefPubMed
5.
go back to reference Quesnel S, Teissier N, Viala P, Couloigner V, Van Den Abbeele T (2010) Long term results of ossiculoplasties with partial and total titanium Vario Kurz prostheses in children. Int J Pedriatr Otorhinolaryngol 74(11):1226–1229CrossRef Quesnel S, Teissier N, Viala P, Couloigner V, Van Den Abbeele T (2010) Long term results of ossiculoplasties with partial and total titanium Vario Kurz prostheses in children. Int J Pedriatr Otorhinolaryngol 74(11):1226–1229CrossRef
6.
go back to reference Hahn Y, Bojrab DI (2013) Outcomes following ossicular chain reconstruction with composite prostheses: hydroxyapatite-polyethylene vs. hydroxyapatite-titanium. Ear Nose Throat J 92(6):250, 252, 254, 256, 258, 260 Hahn Y, Bojrab DI (2013) Outcomes following ossicular chain reconstruction with composite prostheses: hydroxyapatite-polyethylene vs. hydroxyapatite-titanium. Ear Nose Throat J 92(6):250, 252, 254, 256, 258, 260
7.
go back to reference Mardassi A, Deveze A, Sanjuan M et al (2011) Titanium ossicular chain replacement protheses: prognostic factors and preliminary functional results. Eur Ann Otorhinolaryngol Head Neck Dis 128(2):53–58CrossRefPubMed Mardassi A, Deveze A, Sanjuan M et al (2011) Titanium ossicular chain replacement protheses: prognostic factors and preliminary functional results. Eur Ann Otorhinolaryngol Head Neck Dis 128(2):53–58CrossRefPubMed
9.
go back to reference Offergeld C, Kromeier J, Aschendorff A et al (2007) Rotational tomography of the normal and reconstructed middle ear in temporal bones: an experimental study. Eur Arch Otorhinolaryngol 264(4):345–351CrossRefPubMed Offergeld C, Kromeier J, Aschendorff A et al (2007) Rotational tomography of the normal and reconstructed middle ear in temporal bones: an experimental study. Eur Arch Otorhinolaryngol 264(4):345–351CrossRefPubMed
10.
go back to reference Offergeld C, Kromeier J, Merchant SN et al (2010) Experimental investigation of rotational tomography in reconstructed middle ears with clinical implications. Hear Res 263(1-2):191–197CrossRefPubMed Offergeld C, Kromeier J, Merchant SN et al (2010) Experimental investigation of rotational tomography in reconstructed middle ears with clinical implications. Hear Res 263(1-2):191–197CrossRefPubMed
11.
go back to reference Bozzato A, Struffert T, Hertel V, Iro H, Hornung J (2010) Analysis of the accuracy of high-resolution computed tomography techniques for the measurement of stapes prostheses. Eur Radiol 20(3):566–571CrossRefPubMed Bozzato A, Struffert T, Hertel V, Iro H, Hornung J (2010) Analysis of the accuracy of high-resolution computed tomography techniques for the measurement of stapes prostheses. Eur Radiol 20(3):566–571CrossRefPubMed
12.
go back to reference Zaoui K, Kromeier J, Neudert M et al (2014) Clinical investigation of flat panel Ct following middle ear reconstruction: a study of 107 patients. Eur Radiol 24(3):587–594CrossRefPubMed Zaoui K, Kromeier J, Neudert M et al (2014) Clinical investigation of flat panel Ct following middle ear reconstruction: a study of 107 patients. Eur Radiol 24(3):587–594CrossRefPubMed
13.
go back to reference De Cock J, Zanca F, Canning J, Pauwels R, Hermans R (2015) A comparative study for image quality and radiation dose of a cone beam computed tomography scanner and a multislice computed tomography scanner for paranasal sinus imaging. Eur Radiol 25(7):1891–1900CrossRefPubMed De Cock J, Zanca F, Canning J, Pauwels R, Hermans R (2015) A comparative study for image quality and radiation dose of a cone beam computed tomography scanner and a multislice computed tomography scanner for paranasal sinus imaging. Eur Radiol 25(7):1891–1900CrossRefPubMed
14.
go back to reference Hofmann E, Schmidt M, Sedlmair M, Banckwitz R, Hirschfelder U, Lell M (2014) Comparative study of image quality and radiation dose of cone beam and low-dose multislice computed tomography—an in-vitro investigation. Clin Oral Invest 18(1):301–311CrossRef Hofmann E, Schmidt M, Sedlmair M, Banckwitz R, Hirschfelder U, Lell M (2014) Comparative study of image quality and radiation dose of cone beam and low-dose multislice computed tomography—an in-vitro investigation. Clin Oral Invest 18(1):301–311CrossRef
15.
go back to reference Carrafiello G, Dizonno M, Colli V et al (2010) Comparative study of jaws with multislice computed tomography and cone-beam computed tomography. Radiol Med 115(4):600–611CrossRefPubMed Carrafiello G, Dizonno M, Colli V et al (2010) Comparative study of jaws with multislice computed tomography and cone-beam computed tomography. Radiol Med 115(4):600–611CrossRefPubMed
16.
go back to reference Bremke M, Lüers JC, Stenner M et al (2013) Radiologic examinations in human temporal bone specimens using digital volume tomography and high-resolution computed tomography after implantation of middle ear prosthesis and cochlear implant electrode array. Otol Neurotol 34(7):1321–1328CrossRefPubMed Bremke M, Lüers JC, Stenner M et al (2013) Radiologic examinations in human temporal bone specimens using digital volume tomography and high-resolution computed tomography after implantation of middle ear prosthesis and cochlear implant electrode array. Otol Neurotol 34(7):1321–1328CrossRefPubMed
17.
go back to reference Teymoortash A, Hamzei S, Murthum T, Eivazi B, Kureck I, Werner JA (2011) Temporal bone imaging using digital volume tomography and computed tomography: a comparative cadaveric radiological study. Surg Radiol Anat 33(2):123–128CrossRefPubMed Teymoortash A, Hamzei S, Murthum T, Eivazi B, Kureck I, Werner JA (2011) Temporal bone imaging using digital volume tomography and computed tomography: a comparative cadaveric radiological study. Surg Radiol Anat 33(2):123–128CrossRefPubMed
18.
19.
go back to reference Pein MK, Brandt S, Plontke SK, Kösling S (2014) Visualisation of subtle temporal bone structures. Comparison of cone beam CT and MDCT. Radiologe 54(3):271–278CrossRefPubMed Pein MK, Brandt S, Plontke SK, Kösling S (2014) Visualisation of subtle temporal bone structures. Comparison of cone beam CT and MDCT. Radiologe 54(3):271–278CrossRefPubMed
20.
go back to reference Lane JI, Lindell EP, Witte RJ, DeLone DR, Driscoll CL (2006) Middle ear and inner ear: improved depiction with multiplanar reconstruction of volumetric CT data. Radiographics 26(1):115–124CrossRefPubMed Lane JI, Lindell EP, Witte RJ, DeLone DR, Driscoll CL (2006) Middle ear and inner ear: improved depiction with multiplanar reconstruction of volumetric CT data. Radiographics 26(1):115–124CrossRefPubMed
21.
go back to reference Dastidar P, Pertti R, Karhuketo T (1997) Axial HRCT, two-dimensional and maximum intensity projection reconstructions in temporal bone lesions. Acta Otolaryngol Suppl 529:43–46CrossRefPubMed Dastidar P, Pertti R, Karhuketo T (1997) Axial HRCT, two-dimensional and maximum intensity projection reconstructions in temporal bone lesions. Acta Otolaryngol Suppl 529:43–46CrossRefPubMed
22.
go back to reference Mehanna AM, Baki FA, Eid M, Negm M (2014) Comparison of different computed tomography post-processing modalities in assessment of various middle ear disorders. Eur Arch Otorhinolaryngol. doi:10.1007/s00405-014-2920-y PubMed Mehanna AM, Baki FA, Eid M, Negm M (2014) Comparison of different computed tomography post-processing modalities in assessment of various middle ear disorders. Eur Arch Otorhinolaryngol. doi:10.​1007/​s00405-014-2920-y PubMed
23.
go back to reference Knörgen M, Brandt S, Kösling S (2012) Comparison of quality on digital X-ray devices with 3D-capacity for ENT-clinical objectives in imaging of temporal bone and paranasal sinuses. Röfo 184(12):1153–1160PubMed Knörgen M, Brandt S, Kösling S (2012) Comparison of quality on digital X-ray devices with 3D-capacity for ENT-clinical objectives in imaging of temporal bone and paranasal sinuses. Röfo 184(12):1153–1160PubMed
24.
go back to reference Peltonen LI, Aarnisalo AA, Kortesniemi MK, Suomalainen A, Jero J, Robinson S (2007) Limited cone-beam computed tomography imaging of the middle ear: a comparison with multislice helical computed tomography. Acta Radiol 48(2):207–212CrossRefPubMed Peltonen LI, Aarnisalo AA, Kortesniemi MK, Suomalainen A, Jero J, Robinson S (2007) Limited cone-beam computed tomography imaging of the middle ear: a comparison with multislice helical computed tomography. Acta Radiol 48(2):207–212CrossRefPubMed
25.
go back to reference Kyriakou Y, Kolditz D, Langner O, Krause J, Kalender W (2011) Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality. Röfo 183(2):144–153PubMed Kyriakou Y, Kolditz D, Langner O, Krause J, Kalender W (2011) Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality. Röfo 183(2):144–153PubMed
26.
go back to reference Mlynski R, Nguyen TD, Plontke SK, Kösling S (2014) Presentation of floating mass transducer and vibroplasty couplers on CT and cone beam CT. Eur Arch Otorhinolaryngol 271(4):665–672CrossRefPubMed Mlynski R, Nguyen TD, Plontke SK, Kösling S (2014) Presentation of floating mass transducer and vibroplasty couplers on CT and cone beam CT. Eur Arch Otorhinolaryngol 271(4):665–672CrossRefPubMed
27.
go back to reference Granström G, Gröndahl HG (2011) Imaging of osseointegrated implants in the temporal bone by Accuitomo 3-dimensional cone beam computed tomography. Otol Neurotol 32(2):199–203CrossRefPubMed Granström G, Gröndahl HG (2011) Imaging of osseointegrated implants in the temporal bone by Accuitomo 3-dimensional cone beam computed tomography. Otol Neurotol 32(2):199–203CrossRefPubMed
28.
go back to reference Liang X, Jacobs R, Hassam B et al (2010) A comparative evaluation of cone beam computed tomography (CBCT) and multi-slice CT (MSCT). Part I. On subjective image quality. Eur J Radiol 75(2):265–269CrossRefPubMed Liang X, Jacobs R, Hassam B et al (2010) A comparative evaluation of cone beam computed tomography (CBCT) and multi-slice CT (MSCT). Part I. On subjective image quality. Eur J Radiol 75(2):265–269CrossRefPubMed
Metadata
Title
Visualisation of passive middle ear implants by cone beam and multi-detector computed tomography: a comparative in vitro study
Authors
T. D. Nguyen
S. Kösling
R. Mlynski
S. K. Plontke
Publication date
01-12-2016
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 12/2016
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-016-4312-8

Other articles of this Issue 12/2016

European Radiology 12/2016 Go to the issue