Acetabular revision arthroplasty using trabecular titanium implants

Authors: Boris Steno, Milan Kokavec, Libor Necas

Published in: International Orthopaedics | Issue 3/2015

Login to get access

Abstract

Purpose

The aim of this study was to evaluate early results of acetabular revisions of total hip replacement using fully cementless trabecular titanium (TT) acetabular modular implants (Delta Trabecular Titanium, Limacorporate, Udine, Italy).

Methods

Between March 2009 and May 2012 TT was used in 81 revisions. The mean age at the time of revision was 68 years (32–84 years). There were nine patients revised for type 1, 11 for type 2A, 27 for type 2B, six for type 2C, 15 for type 3A and 13 for type 3B acetabular defects according to the Paprosky classification. Frozen morselised bone allografts were used in 53 cases and bulk structural allografts in three cases. Clinical evaluations were made using a modified functional Merle d’Aubigné-Postel score. The mean follow-up period was 38.14 months (24–62 months).

Results

The mean pre-operative Merle d’Aubigné-Postel functional score was 4.7 and 9.8 at the time of last follow-up. There was one revision due to instability of the acetabular component. A cage system—Delta Revision TT—was successfully used in this case. Three cases with Paprosky type 3B defect showed cranial migration of the acetabular component by 6 mm, but stabilised after six months. No dislocations associated with acetabular surgery have occurred in the cohort. There have been no dissociations of the modular component. A fatigue fracture of the hemispherical module occurred in the revised case. No other hardware mechanical failures have been recorded.

Conclusions

TT cups, hemispherical modules and augments facilitate reliable and reproducible biological fixation in acetabular revision surgery with excellent results. Extended follow-up is necessary to evaluate the long-term performance of TT modular implants.
Literature
1.
go back to reference Necas L, Katina S, Uhlarova J, Colton CL (2013) Survival analysis of total hip and knee arthroplasty in Slovakia 2003–2011. Acta Chir Orthop Traumatol Cech 80(Suppl):9–67 Necas L, Katina S, Uhlarova J, Colton CL (2013) Survival analysis of total hip and knee arthroplasty in Slovakia 2003–2011. Acta Chir Orthop Traumatol Cech 80(Suppl):9–67
2.
go back to reference Jain S, Grogan RJ, Giannoudis PV (2014) Options for managing severe acetabular bone loss in revision hip arthroplasty. A systematic review. Hip Int 24:109–122CrossRefPubMed Jain S, Grogan RJ, Giannoudis PV (2014) Options for managing severe acetabular bone loss in revision hip arthroplasty. A systematic review. Hip Int 24:109–122CrossRefPubMed
3.
go back to reference Gill TJ, Sledge JB, Müller ME (1998) The Bürch-Schneider anti-protrusio cage in revision total hip arthroplasty: indications, principles and long-term results. J Bone Joint Surg Br 80:946–953CrossRefPubMed Gill TJ, Sledge JB, Müller ME (1998) The Bürch-Schneider anti-protrusio cage in revision total hip arthroplasty: indications, principles and long-term results. J Bone Joint Surg Br 80:946–953CrossRefPubMed
4.
go back to reference Herrera A, Martínez AA, Cuenca J, Canales V (2006) Management of types III and IV acetabular deficiencies with the longitudinal oblong revision cup. J Arthroplasty 21:857–864CrossRefPubMed Herrera A, Martínez AA, Cuenca J, Canales V (2006) Management of types III and IV acetabular deficiencies with the longitudinal oblong revision cup. J Arthroplasty 21:857–864CrossRefPubMed
5.
go back to reference Unger AS, Lewis RJ, Gruen T (2005) Evaluation of porous tantalum uncemented acetabular cup in revision total hip arthroplasty: clinical and radiological results of 60 hips. J Arthroplasty 20:1002–1009CrossRefPubMed Unger AS, Lewis RJ, Gruen T (2005) Evaluation of porous tantalum uncemented acetabular cup in revision total hip arthroplasty: clinical and radiological results of 60 hips. J Arthroplasty 20:1002–1009CrossRefPubMed
6.
go back to reference Paprosky WG, Sporer SS, Murphy BP (2007) Addressing severe bone deficiency: what a cage will not do. J Arthroplasty 22(4 Suppl 1):111–115CrossRefPubMed Paprosky WG, Sporer SS, Murphy BP (2007) Addressing severe bone deficiency: what a cage will not do. J Arthroplasty 22(4 Suppl 1):111–115CrossRefPubMed
7.
go back to reference Abolghasemian M, Tangsataporn S, Sternheim A, Backstein D, Safir O, Gross AE (2013) Combined trabecular metal acetabular shell and augment for acetabular revision with substantial bone loss: a mid-term review. Bone Joint J 95-B:166–172CrossRefPubMed Abolghasemian M, Tangsataporn S, Sternheim A, Backstein D, Safir O, Gross AE (2013) Combined trabecular metal acetabular shell and augment for acetabular revision with substantial bone loss: a mid-term review. Bone Joint J 95-B:166–172CrossRefPubMed
8.
go back to reference Marin E, Fusi S, Pressaco M, Paussa L, Fedrizzi L (2010) Characterization of cellular solids in TiAl4V for orthopaedic implant applications: trabecular titanium. J Mech Behav Biomed Mater 3:373–381CrossRefPubMed Marin E, Fusi S, Pressaco M, Paussa L, Fedrizzi L (2010) Characterization of cellular solids in TiAl4V for orthopaedic implant applications: trabecular titanium. J Mech Behav Biomed Mater 3:373–381CrossRefPubMed
9.
go back to reference Benazzo F, Botta L, Scaffino MF et al (2014) Trabecular titanium can induce in vitro osteogenic differentiation of human adipose derived stem cells without osteogenic factors. J Biomed Mater Res A 102:2061–2071. doi:10.1002/jbm.a.34875 CrossRefPubMed Benazzo F, Botta L, Scaffino MF et al (2014) Trabecular titanium can induce in vitro osteogenic differentiation of human adipose derived stem cells without osteogenic factors. J Biomed Mater Res A 102:2061–2071. doi:10.​1002/​jbm.​a.​34875 CrossRefPubMed
10.
go back to reference Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplasty 9:33–44CrossRefPubMed Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplasty 9:33–44CrossRefPubMed
11.
go back to reference Massin P, Schmidt L, Engh CA (1989) Evaluation of cementless acetabular component migration. An experimental study. J Arthroplasty 4:245–251CrossRefPubMed Massin P, Schmidt L, Engh CA (1989) Evaluation of cementless acetabular component migration. An experimental study. J Arthroplasty 4:245–251CrossRefPubMed
12.
go back to reference D’Aubigne RM, Postel M (1954) Functional results of hip arthroplasty with acrylic prosthesis. J Bone Joint Surg Am 36:451–475PubMed D’Aubigne RM, Postel M (1954) Functional results of hip arthroplasty with acrylic prosthesis. J Bone Joint Surg Am 36:451–475PubMed
13.
go back to reference Charnley J (1979) Low friction arthroplasty. Theory and practice. Springer, New YorkCrossRef Charnley J (1979) Low friction arthroplasty. Theory and practice. Springer, New YorkCrossRef
14.
go back to reference Paprosky WG, Magnus RE (1991) Classification and treatment of the failed acetabulum: a systematic approach. Contemp Orthop 22:121–130 Paprosky WG, Magnus RE (1991) Classification and treatment of the failed acetabulum: a systematic approach. Contemp Orthop 22:121–130
15.
go back to reference Paprosky WG, Bradford MS, Younger TI (1994) Acetabular reconstruction with massive allograft and cementless prosthesis. Chir Organi Mov 79:379–386PubMed Paprosky WG, Bradford MS, Younger TI (1994) Acetabular reconstruction with massive allograft and cementless prosthesis. Chir Organi Mov 79:379–386PubMed
16.
go back to reference Pulido L, Rachala SR, Cabanela ME (2011) Cementless acetabular revision: past, present and future. Revision total hip arthroplasty: the acetabular side using cementless implants. Int Orthop 35:289–298CrossRefPubMedCentralPubMed Pulido L, Rachala SR, Cabanela ME (2011) Cementless acetabular revision: past, present and future. Revision total hip arthroplasty: the acetabular side using cementless implants. Int Orthop 35:289–298CrossRefPubMedCentralPubMed
17.
go back to reference Hernigou P, Pariat J, Queinnec S et al (2014) Supercharging irradiated allografts with mesenchymal stem cells improves acetabular bone grafting in revision arthroplasty. Int Orthop. doi:10.1007/s00264-014-2285-2 Hernigou P, Pariat J, Queinnec S et al (2014) Supercharging irradiated allografts with mesenchymal stem cells improves acetabular bone grafting in revision arthroplasty. Int Orthop. doi:10.​1007/​s00264-014-2285-2
18.
go back to reference Št’astný E, Trč T, Handl M et al (2014) Mid-term results of revision total hip arthroplasty using the oval-shaped uncemented Trč-Cingr cup. Int Orthop 38:935–940CrossRef Št’astný E, Trč T, Handl M et al (2014) Mid-term results of revision total hip arthroplasty using the oval-shaped uncemented Trč-Cingr cup. Int Orthop 38:935–940CrossRef
19.
go back to reference Dennis DA (2003) Management of massive acetabular defects in revision total hip arthroplasty. J Arthroplasty 18(Suppl):121–125CrossRefPubMed Dennis DA (2003) Management of massive acetabular defects in revision total hip arthroplasty. J Arthroplasty 18(Suppl):121–125CrossRefPubMed
20.
go back to reference Sakai T, Ohzono K, Nishii T et al (2013) Modular acetabular reconstructive cup in acetabular revision total hip arthroplasty at a minimum ten year follow-up. Int Orthop 37:605–610CrossRefPubMedCentralPubMed Sakai T, Ohzono K, Nishii T et al (2013) Modular acetabular reconstructive cup in acetabular revision total hip arthroplasty at a minimum ten year follow-up. Int Orthop 37:605–610CrossRefPubMedCentralPubMed
21.
go back to reference Davies JH, Laflamme GY, Delisle J, Fernandes J (2011) Trabecular metal used for major bone loss in acetabular hip revision. J Arthroplasty 26:1245–1250CrossRefPubMed Davies JH, Laflamme GY, Delisle J, Fernandes J (2011) Trabecular metal used for major bone loss in acetabular hip revision. J Arthroplasty 26:1245–1250CrossRefPubMed
22.
go back to reference Flecher X, Sporer S, Paprosky W (2008) Management of severe bone loss in acetabular revision using a trabecular metal shell. J Arthroplasty 23:949–955CrossRefPubMed Flecher X, Sporer S, Paprosky W (2008) Management of severe bone loss in acetabular revision using a trabecular metal shell. J Arthroplasty 23:949–955CrossRefPubMed
23.
go back to reference Macheras GA, Papageloupoulos PJ, Kateros K, Kostakos AT, Baltas D, Karachalios TS (2006) Radiological evaluation of the metal-bone interface of a porous tantalum monoblock acetabular component. J Bone Joint Surg Br 88:304–309CrossRefPubMed Macheras GA, Papageloupoulos PJ, Kateros K, Kostakos AT, Baltas D, Karachalios TS (2006) Radiological evaluation of the metal-bone interface of a porous tantalum monoblock acetabular component. J Bone Joint Surg Br 88:304–309CrossRefPubMed
24.
go back to reference Jaffari SM, Bender B, Coyle C et al (2010) Do tantalum and titanium cups show similar results in revision hip arthroplasty? Clin Orthop Relat Res 468:459–465CrossRef Jaffari SM, Bender B, Coyle C et al (2010) Do tantalum and titanium cups show similar results in revision hip arthroplasty? Clin Orthop Relat Res 468:459–465CrossRef
25.
go back to reference Beckmann NA, Weiss S, Klotz MC, Gondan Mm Jaeger S, Bitsch RG (2014) Loosening after acetabular revision: comparison of trabecular metal and reinforcement rings. A systematic review. J Arthroplasty 29:229–235CrossRefPubMed Beckmann NA, Weiss S, Klotz MC, Gondan Mm Jaeger S, Bitsch RG (2014) Loosening after acetabular revision: comparison of trabecular metal and reinforcement rings. A systematic review. J Arthroplasty 29:229–235CrossRefPubMed
26.
go back to reference Meneghini RM, Meyer C, Buckley CA, Hanssen AD, Lewallen DG (2010) Mechanical stability of novel highly porous metal acetabular components in revision total hip arthroplasty. J Arthroplasty 25:337–341CrossRefPubMed Meneghini RM, Meyer C, Buckley CA, Hanssen AD, Lewallen DG (2010) Mechanical stability of novel highly porous metal acetabular components in revision total hip arthroplasty. J Arthroplasty 25:337–341CrossRefPubMed
27.
go back to reference Skyttä ET, Eiskelinen A, Paavolainen PO, Remes VM (2011) Early results of 827 trabecular metal revision shells in acetabular revision. J Arthroplasty 26:342–345CrossRefPubMed Skyttä ET, Eiskelinen A, Paavolainen PO, Remes VM (2011) Early results of 827 trabecular metal revision shells in acetabular revision. J Arthroplasty 26:342–345CrossRefPubMed
28.
go back to reference Shultz TR, Blaha JD, Gruen TA, Norman TL (2006) Cortical bone viscoelasticity and fixation strength of press-fit femoral stems: finite element model. J Biomech Eng 128:7–12CrossRefPubMed Shultz TR, Blaha JD, Gruen TA, Norman TL (2006) Cortical bone viscoelasticity and fixation strength of press-fit femoral stems: finite element model. J Biomech Eng 128:7–12CrossRefPubMed
29.
go back to reference Simon JP, Bellemans J (2009) Clinical and radiological evaluation of modular trabecular metal acetabular cups. Short-term results in 64 hips. Acta Orthop Belg 75:623–630PubMed Simon JP, Bellemans J (2009) Clinical and radiological evaluation of modular trabecular metal acetabular cups. Short-term results in 64 hips. Acta Orthop Belg 75:623–630PubMed
30.
go back to reference Choi H, Anderson A, Foster S et al (2013) Acetabular cup positioning in revision total hip arthroplasty with Paprosky type III defects: Martell radiographic analysis. Int Orthop 37:1905–1910CrossRefPubMedCentralPubMed Choi H, Anderson A, Foster S et al (2013) Acetabular cup positioning in revision total hip arthroplasty with Paprosky type III defects: Martell radiographic analysis. Int Orthop 37:1905–1910CrossRefPubMedCentralPubMed
31.
go back to reference Sternheim A, Backstein D, Kuzyk PRT, Goshua G, Berkowich Y, Safir O, Gross AE (2012) Porous metal revision shells for management of contained acetabular bone defects at a mean follow-up of six years: a comparison between up to 50% bleeding host bone contact and more than 50% contact. J Bone Joint Surg Br 94:158–162CrossRefPubMed Sternheim A, Backstein D, Kuzyk PRT, Goshua G, Berkowich Y, Safir O, Gross AE (2012) Porous metal revision shells for management of contained acetabular bone defects at a mean follow-up of six years: a comparison between up to 50% bleeding host bone contact and more than 50% contact. J Bone Joint Surg Br 94:158–162CrossRefPubMed
Metadata
Title
Acetabular revision arthroplasty using trabecular titanium implants
Authors
Boris Steno
Milan Kokavec
Libor Necas
Publication date
01-03-2015
Publisher
Springer Berlin Heidelberg
Published in
International Orthopaedics / Issue 3/2015
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-014-2509-5

Other articles of this Issue 3/2015

International Orthopaedics 3/2015 Go to the issue