Published in: Abdominal Radiology 11/2020

01-11-2020 | Hepatobiliary

Evaluation of simultaneous-multislice diffusion-weighted imaging of liver at 3.0 T with different breathing schemes

Authors: Yigang Pei, Siming Xie, Wenzheng Li, Xianjing Peng, Qin Qin, Qian Ye, Mengsi Li, Jiaxi Hu, Jiale Hou, Guijing Li, Shuo Hu

Published in: Abdominal Radiology | Issue 11/2020

Login to get access

Abstract

Purpose

To obtain the optimal simultaneous-multislice (SMS)—accelerated diffusion-weighted imaging (DWI) of the liver at 3.0 T MRI by systematically estimating the repeatability of apparent diffusion coefficient (ADC), signal-to-noise ratio (SNR) and image quality of different breathing schemes in comparison to standard DWI (STD) and other SMS sequences.

Methods

In this institutional review board-approved prospective study, hepatic DWIs (b = 50, 300, 600 s/mm2) were performed in 23 volunteers on 3.0 T MRI using SMS and STD with breath-hold (BH-SMS, BH-STD), free-breathing (FB-SMS, FB-STD) and respiratory-triggered (RT-SMS, RT-STD). Reduction of scan time with SMS-acceleration was calculated. ADC and SNR were measured in nine anatomic locations and image quality was assessed on all SMS and STD sequences. An optimal SMS-DWI was decided by systematically comparing the ADC repeatability, SNR and image quality among above DWIs.

Results

SMS-DWI reduced scan time significantly by comparison with corresponding STD-DWI (27 vs. 42 s for BH, 54 vs. 78 s for FB and 42 vs. 97 s for RT). In all DWIs, BH-SMS had the greatest intraobserver agreement (intraclass correlation coefficient (ICC): 0.920–0.944) and good interobserver agreement (ICC: 0.831–0.886) for ADC measurements, and had the best ADC repeatability (mean ADC absolute differences: 0.046–0.058 × 10−3mm2/s, limits of agreement (LOA): 0.010–0.013 × 10−3mm2/s) in nine locations. BH-SMS had the highest SNR in three representative sections except for RT-STD. There were no significant differences in image quality between BH-SMS and other DWI sequences (median BH-SMS: 4.75, other DWI: 4.5–5.0; P > 0.0.5).

Conclusion

BH-SMS provides considerable scan time reduction with good image quality, sufficient SNR and highest ADC repeatability on 3.0 T MRI, which is thus recommended as the optimal hepatic DWI sequence for those subjects with adequate breath-holding capability.
Literature
3.
8.
20.
go back to reference Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. PSYCHOL BULL. 1979;86:420-28.CrossRef Shrout PE, Fleiss JL. Intraclass correlations: uses in assessing rater reliability. PSYCHOL BULL. 1979;86:420-28.CrossRef
22.
go back to reference Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. LANCET. 1986;1:307-10.CrossRef Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. LANCET. 1986;1:307-10.CrossRef
23.
go back to reference Bland JM, Altman DG. Multiple significance tests: the Bonferroni method. BMJ. 1995;310:170.CrossRef Bland JM, Altman DG. Multiple significance tests: the Bonferroni method. BMJ. 1995;310:170.CrossRef
Metadata
Title
Evaluation of simultaneous-multislice diffusion-weighted imaging of liver at 3.0 T with different breathing schemes
Authors
Yigang Pei
Siming Xie
Wenzheng Li
Xianjing Peng
Qin Qin
Qian Ye
Mengsi Li
Jiaxi Hu
Jiale Hou
Guijing Li
Shuo Hu
Publication date
01-11-2020
Publisher
Springer US
Published in
Abdominal Radiology / Issue 11/2020
Print ISSN: 2366-004X
Electronic ISSN: 2366-0058
DOI
https://doi.org/10.1007/s00261-020-02538-y

Other articles of this Issue 11/2020

Abdominal Radiology 11/2020 Go to the issue