Skip to main content
Top
Published in: Pediatric Cardiology 3/2018

01-03-2018 | Original Article

Transcatheter Closure of Arterial Duct in Infants < 6 kg: Amplatzer Duct Occluder Type I vs Amplatzer Duct Occluder II Additional Sizes

Authors: Giuseppe Santoro, Mario Giordano, Gianpiero Gaio, Maria Teresa Palladino, Giovanbattista Capozzi, Carola Iacono, Maria Giovanna Russo

Published in: Pediatric Cardiology | Issue 3/2018

Login to get access

Abstract

Transcatheter closure of arterial duct (AD) remains challenging in low-weight patients and using Amplatzer Duct Occluder devices is still considered off-label in infants < 6 kg. This study aimed to report a large, single-center experience of percutaneous AD closure in infants < 6 kg as well as to compare the most frequently used devices, Amplatzer Duct Occluder type I (ADO I) and Amplatzer Duct Occluder type II Additional Sizes devices (ADO II-AS) (St. Jude Medical Corp, St. Paul, MN, USA). From March 2000 to March 2017, among the 762 patients submitted to percutaneous closure of AD at our Institution, 33 were infants < 6 kg (age 4.8 ± 2.1 months; weight 5.0 ± 0.9 kg). Fourteen patients (45%) underwent ADO I (Group I) and 19 patients ADO II-AS (Group II) device implantation. AD diameter was 2.6 ± 0.8 (range 1.5–4.0) mm resulting in QP/QS of 2.6 ± 0.0.9 (range 1.1–4.5). Successful device deployment was achieved in all patients without procedural morbidity or mortality. Procedural and fluoroscopy times were not significantly different between the groups. However, total X-ray absorbed dose was significantly higher in Group I (121 ± 69 vs 29 ± 16 mGy/cm2, p < 0.01). Immediate, 24 h, and mid-term (46 ± 37 months, median 39) complete occlusion were recorded in 72.7, 90.9, and 97% of patients, respectively without significant difference between the groups. In conclusion, transcatheter closure of AD with Amplatzer Duct Occluder devices is feasible, safe, and effective also in infants less than 6 kg, without significant difference between the most commonly used devices, namely, ADO I and ADO II-AS.
Literature
1.
go back to reference Baspinar O, Sahin DA, Sulu A et al (2015) Transcatheter closure of patent ductus arteriosus in under 6 kg and premature infants. J Interv Cardiol 28(2):180–189CrossRefPubMed Baspinar O, Sahin DA, Sulu A et al (2015) Transcatheter closure of patent ductus arteriosus in under 6 kg and premature infants. J Interv Cardiol 28(2):180–189CrossRefPubMed
2.
go back to reference Moysich A, Happel CM, Laser KT, Kececioglu D, Haas NA (2015) Implantation of the new Nit-Occlud PDA-R device in children below 10 kilogram. Catheter Cardiovasc Interv 85(7):1203–1208CrossRefPubMed Moysich A, Happel CM, Laser KT, Kececioglu D, Haas NA (2015) Implantation of the new Nit-Occlud PDA-R device in children below 10 kilogram. Catheter Cardiovasc Interv 85(7):1203–1208CrossRefPubMed
3.
go back to reference Narin N, Pamukcu O, Baykan A, Sunkak S, Ulgey A, Uzum K (2016) Percutaneous PDA closure in extremely low birth weight babies. J Interv Cardiol 29(6):654–660CrossRefPubMed Narin N, Pamukcu O, Baykan A, Sunkak S, Ulgey A, Uzum K (2016) Percutaneous PDA closure in extremely low birth weight babies. J Interv Cardiol 29(6):654–660CrossRefPubMed
4.
go back to reference Ali S, El Sisi A (2016) Transcatheter closure of patent ductus arteriosus in children weighing 10 kg or less: initial experience at Sohag University Hospital. J Saudi Heart Assoc 28:95–100CrossRefPubMed Ali S, El Sisi A (2016) Transcatheter closure of patent ductus arteriosus in children weighing 10 kg or less: initial experience at Sohag University Hospital. J Saudi Heart Assoc 28:95–100CrossRefPubMed
5.
go back to reference Zahn EM, Peck D, Phillips A et al (2016) Transcatheter closure of patent ductus arteriosus in extremely premature newborns: early results and midterm follow-up. JACC Cardiovasc Interv 9(23):2429–2437CrossRefPubMed Zahn EM, Peck D, Phillips A et al (2016) Transcatheter closure of patent ductus arteriosus in extremely premature newborns: early results and midterm follow-up. JACC Cardiovasc Interv 9(23):2429–2437CrossRefPubMed
7.
go back to reference Wang-Giuffre EW, Breinholt JP (2017) Novel use of the medtronic micro vascular plug for PDA closure in preterm infants. Catheter Cardiovasc Interv 89(6):1059–1065CrossRefPubMed Wang-Giuffre EW, Breinholt JP (2017) Novel use of the medtronic micro vascular plug for PDA closure in preterm infants. Catheter Cardiovasc Interv 89(6):1059–1065CrossRefPubMed
9.
go back to reference Santoro G, Cappelli Bigazzi M, Palladino MT, Russo MG, Carrozza M, Calabrò R (2004) Comparison of percutaneous closure of large patent ductus arteriosus by multiple coils vs the Amplatzer Duct Occluder device. Am J Cardiol 94:252–255CrossRefPubMed Santoro G, Cappelli Bigazzi M, Palladino MT, Russo MG, Carrozza M, Calabrò R (2004) Comparison of percutaneous closure of large patent ductus arteriosus by multiple coils vs the Amplatzer Duct Occluder device. Am J Cardiol 94:252–255CrossRefPubMed
10.
go back to reference Sungur M, Karakurt C, Ozbarlas N, Baspinar O (2013) Closure of patent ductus arteriosus in children, small infants, and premature babies with Amplatzer Duct Occluder II additional sizes: multicenter study. Catheter Cardiovasc Interv 82(2):245–252CrossRefPubMed Sungur M, Karakurt C, Ozbarlas N, Baspinar O (2013) Closure of patent ductus arteriosus in children, small infants, and premature babies with Amplatzer Duct Occluder II additional sizes: multicenter study. Catheter Cardiovasc Interv 82(2):245–252CrossRefPubMed
11.
go back to reference Mahmoud HT, Santoro G, Gaio G et al (2017) Single-center experience in percutaneous closure of arterial duct with Amplatzer Duct Occluder II additional sizes. Catheter Cardiovasc Interv 89(6):1045–1050CrossRefPubMed Mahmoud HT, Santoro G, Gaio G et al (2017) Single-center experience in percutaneous closure of arterial duct with Amplatzer Duct Occluder II additional sizes. Catheter Cardiovasc Interv 89(6):1045–1050CrossRefPubMed
12.
go back to reference Kang SL, Morgan G, Forsey J, Tometzki A, Martin R (2014) Long-Term clinical experience with Amplatzer Ductal Occluder II for closure of the persistent arterial duct in children. Catheter Cardiovasc Interv 83:1102–1108CrossRefPubMed Kang SL, Morgan G, Forsey J, Tometzki A, Martin R (2014) Long-Term clinical experience with Amplatzer Ductal Occluder II for closure of the persistent arterial duct in children. Catheter Cardiovasc Interv 83:1102–1108CrossRefPubMed
13.
go back to reference Baspinar O, Irdem A, Sivasli E, Sahin DA, Kilinc M (2013) Comparison of the efficacy of different-sized Amplatzer Duct Occluders (I, II, and II AS) in children weighing less than 10 kg. Pediatr Cardiol 34(1):88–94CrossRefPubMed Baspinar O, Irdem A, Sivasli E, Sahin DA, Kilinc M (2013) Comparison of the efficacy of different-sized Amplatzer Duct Occluders (I, II, and II AS) in children weighing less than 10 kg. Pediatr Cardiol 34(1):88–94CrossRefPubMed
15.
go back to reference Parra-Bravo R, Cruz-Ramírez A, Rebolledo-Pineda V et al (2009) Transcatheter closure of patent ductus arteriosus using the Amplatzer Duct Occluder in infants under 1 year of age. Rev Esp Cardiol 62(8):867–874CrossRef Parra-Bravo R, Cruz-Ramírez A, Rebolledo-Pineda V et al (2009) Transcatheter closure of patent ductus arteriosus using the Amplatzer Duct Occluder in infants under 1 year of age. Rev Esp Cardiol 62(8):867–874CrossRef
16.
go back to reference Wang JK, Wu MH, Hwang JJ, Chiang FT, Lin MT, Lue HC (2007) Transcatheter closure of moderate to large patent ductus arteriosus with the Amplatzer Duct Occluder. Catheter Cardiovasc Interv 69(4):572–578CrossRefPubMed Wang JK, Wu MH, Hwang JJ, Chiang FT, Lin MT, Lue HC (2007) Transcatheter closure of moderate to large patent ductus arteriosus with the Amplatzer Duct Occluder. Catheter Cardiovasc Interv 69(4):572–578CrossRefPubMed
17.
go back to reference Butera G, De Rosa G, Chessa M et al (2004) Transcatheter closure of persistent ductus arteriosus with the Amplatzer Duct Occluder in very young symptomatic children. Heart 90(12):1467–1470CrossRefPubMedPubMedCentral Butera G, De Rosa G, Chessa M et al (2004) Transcatheter closure of persistent ductus arteriosus with the Amplatzer Duct Occluder in very young symptomatic children. Heart 90(12):1467–1470CrossRefPubMedPubMedCentral
18.
go back to reference Krichenko A, Benson LN, Burrows P, Moes CA, McLaughlin P, Freedon RM (1989) Angiographic classification of the isolated, persistently patent ductus arteriosus and implications for percutaneous catheter occlusion. Am J Cardiol 63:877–879CrossRefPubMed Krichenko A, Benson LN, Burrows P, Moes CA, McLaughlin P, Freedon RM (1989) Angiographic classification of the isolated, persistently patent ductus arteriosus and implications for percutaneous catheter occlusion. Am J Cardiol 63:877–879CrossRefPubMed
19.
go back to reference Brunetti MA, Ringel R, Owada C et al (2010) Percutaneous closure of patent ductus arteriosus: a multi-institutional registry comparing multiple devices. Catheter Cardiovasc Interv 76(5):696–702CrossRefPubMed Brunetti MA, Ringel R, Owada C et al (2010) Percutaneous closure of patent ductus arteriosus: a multi-institutional registry comparing multiple devices. Catheter Cardiovasc Interv 76(5):696–702CrossRefPubMed
20.
go back to reference Sathanandam S, Justino H, Waller BR 3rd, Radtke W, Qureshi AM (2017) Initial clinical experience with the Medtronic Micro Vascular Plug™ in transcatheter occlusion of PDAs in extremely premature infants. Catheter Cardiovasc Interv 89(6):1051–1058CrossRefPubMed Sathanandam S, Justino H, Waller BR 3rd, Radtke W, Qureshi AM (2017) Initial clinical experience with the Medtronic Micro Vascular Plug™ in transcatheter occlusion of PDAs in extremely premature infants. Catheter Cardiovasc Interv 89(6):1051–1058CrossRefPubMed
21.
go back to reference Liddy S, Oslizlok P, Walsh KP (2013) Comparison of the results of transcatheter closure of patent ductus arteriosus with newer Amplatzer devices. Catheter Cardiovasc Interv 82:253–259CrossRefPubMed Liddy S, Oslizlok P, Walsh KP (2013) Comparison of the results of transcatheter closure of patent ductus arteriosus with newer Amplatzer devices. Catheter Cardiovasc Interv 82:253–259CrossRefPubMed
22.
go back to reference Kenny D, Morgan GJ, Bentham JR et al (2013) Early clinical experience with a modified Amplatzer Ductal Occluder for transcatheter arterial duct occlusion in infants and small children. Catheter Cardiovasc Interv 82(4):534–540PubMed Kenny D, Morgan GJ, Bentham JR et al (2013) Early clinical experience with a modified Amplatzer Ductal Occluder for transcatheter arterial duct occlusion in infants and small children. Catheter Cardiovasc Interv 82(4):534–540PubMed
23.
go back to reference Bruckheimer E, Godfrey M, Dagan T, Levinzon M, Amir G, Birk E (2014) The Amplatzer Duct Occluder II additional sizes device for transcatheter PDA closure: initial experience. Catheter Cardiovasc Interv 83(7): 1097–1101CrossRefPubMed Bruckheimer E, Godfrey M, Dagan T, Levinzon M, Amir G, Birk E (2014) The Amplatzer Duct Occluder II additional sizes device for transcatheter PDA closure: initial experience. Catheter Cardiovasc Interv 83(7): 1097–1101CrossRefPubMed
Metadata
Title
Transcatheter Closure of Arterial Duct in Infants < 6 kg: Amplatzer Duct Occluder Type I vs Amplatzer Duct Occluder II Additional Sizes
Authors
Giuseppe Santoro
Mario Giordano
Gianpiero Gaio
Maria Teresa Palladino
Giovanbattista Capozzi
Carola Iacono
Maria Giovanna Russo
Publication date
01-03-2018
Publisher
Springer US
Published in
Pediatric Cardiology / Issue 3/2018
Print ISSN: 0172-0643
Electronic ISSN: 1432-1971
DOI
https://doi.org/10.1007/s00246-018-1828-5

Other articles of this Issue 3/2018

Pediatric Cardiology 3/2018 Go to the issue