Skip to main content
Top
Published in: Neuroradiology 9/2014

01-09-2014 | Head and Neck Radiology

Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose

Authors: Matthias S. May, Manuel R. Kramer, Achim Eller, Wolfgang Wuest, Michael Scharf, Michael Brand, Marc Saake, Bernhard Schmidt, Michael Uder, Michael M. Lell

Published in: Neuroradiology | Issue 9/2014

Login to get access

Abstract

Introduction

Low tube voltage allows for computed tomography (CT) imaging with increased iodine contrast at reduced radiation dose. We sought to evaluate the image quality and potential dose reduction using a combination of attenuation based tube current modulation (TCM) and automated tube voltage adaptation (TVA) between 100 and 120 kV in CT of the head and neck.

Methods

One hundred thirty consecutive patients with indication for head and neck CT were examined with a 128-slice system capable of TCM and TVA. Reference protocol was set at 120 kV. Tube voltage was reduced to 100 kV whenever proposed by automated analysis of the localizer. An additional small scan aligned to the jaw was performed at a fixed 120 kV setting. Image quality was assessed by two radiologists on a standardized Likert-scale and measurements of signal- (SNR) and contrast-to-noise ratio (CNR). Radiation dose was assessed as CTDIvol.

Results

Diagnostic image quality was excellent in both groups and did not differ significantly (p = 0.34). Image noise in the 100 kV data was increased and SNR decreased (17.8/9.6) in the jugular veins and the sternocleidomastoid muscle when compared to 120 kV (SNR 24.4/10.3), but not in fatty tissue and air. However, CNR did not differ statistically significant between 100 (23.5/14.4/9.4) and 120 kV data (24.2/15.3/8.6) while radiation dose was decreased by 7–8 %.

Conclusions

TVA between 100 and 120 kV in combination with TCM led to a radiation dose reduction compared to TCM alone, while keeping CNR constant though maintaining diagnostic image quality.
Literature
1.
go back to reference Agency HP, Centre for Radiation CaEH, Division RP (2008) European Commission Radiation Protection N° 154, European guidance on estimating population doses from medical X-Ray procedures, Annex 1 – DD Report 1 Review of recent National Surveys of Population Exposure from Medical X-rays in Europe. vol 154. Chilton, Didcot, Oxfordshire Agency HP, Centre for Radiation CaEH, Division RP (2008) European Commission Radiation Protection N° 154, European guidance on estimating population doses from medical X-Ray procedures, Annex 1 – DD Report 1 Review of recent National Surveys of Population Exposure from Medical X-rays in Europe. vol 154. Chilton, Didcot, Oxfordshire
2.
go back to reference Brenner DJ, Hall EJ (2007) Computed tomography—an increasing source of radiation exposure. N Engl J Med 357(22):2277–2284PubMedCrossRef Brenner DJ, Hall EJ (2007) Computed tomography—an increasing source of radiation exposure. N Engl J Med 357(22):2277–2284PubMedCrossRef
3.
go back to reference Brenner DJ, Doll R, Goodhead DT, Hall EJ, Land CE, Little JB, Lubin JH, Preston DL, Preston RJ, Puskin JS, Ron E, Sachs RK, Samet JM, Setlow RB, Zaider M (2003) Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. Proc Natl Acad Sci U S A 100(24):13761–13766PubMedCentralPubMedCrossRef Brenner DJ, Doll R, Goodhead DT, Hall EJ, Land CE, Little JB, Lubin JH, Preston DL, Preston RJ, Puskin JS, Ron E, Sachs RK, Samet JM, Setlow RB, Zaider M (2003) Cancer risks attributable to low doses of ionizing radiation: assessing what we really know. Proc Natl Acad Sci U S A 100(24):13761–13766PubMedCentralPubMedCrossRef
4.
go back to reference Berrington de Gonzalez A, Mahesh M, Kim KP, Bhargavan M, Lewis R, Mettler F, Land C (2009) Projected cancer risks from computed tomographic scans performed in the United States in 2007. Arch Intern Med 169(22):2071–2077PubMedCrossRef Berrington de Gonzalez A, Mahesh M, Kim KP, Bhargavan M, Lewis R, Mettler F, Land C (2009) Projected cancer risks from computed tomographic scans performed in the United States in 2007. Arch Intern Med 169(22):2071–2077PubMedCrossRef
5.
go back to reference Kalender WA, Wolf H, Suess C, Gies M, Greess H, Bautz WA (1999) Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers. Eur Radiol 9(2):323–328PubMedCrossRef Kalender WA, Wolf H, Suess C, Gies M, Greess H, Bautz WA (1999) Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers. Eur Radiol 9(2):323–328PubMedCrossRef
6.
go back to reference Kalender WA, Buchenau S, Deak P, Kellermeier M, Langner O, van Straten M, Vollmar S, Wilharm S (2008) Technical approaches to the optimisation of CT. Phys Med 24(2):71–79, an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical PhysicsPubMedCrossRef Kalender WA, Buchenau S, Deak P, Kellermeier M, Langner O, van Straten M, Vollmar S, Wilharm S (2008) Technical approaches to the optimisation of CT. Phys Med 24(2):71–79, an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical PhysicsPubMedCrossRef
7.
go back to reference Leschka S, Stolzmann P, Schmid FT, Scheffel H, Stinn B, Marincek B, Alkadhi H, Wildermuth S (2008) Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose. Eur Radiol 18(9):1809–1817PubMedCrossRef Leschka S, Stolzmann P, Schmid FT, Scheffel H, Stinn B, Marincek B, Alkadhi H, Wildermuth S (2008) Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose. Eur Radiol 18(9):1809–1817PubMedCrossRef
8.
go back to reference Szucs-Farkas Z, Kurmann L, Strautz T, Patak MA, Vock P, Schindera ST (2008) Patient exposure and image quality of low-dose pulmonary computed tomography angiography: comparison of 100- and 80-kVp protocols. Invest Radiol 43(12):871–876PubMedCrossRef Szucs-Farkas Z, Kurmann L, Strautz T, Patak MA, Vock P, Schindera ST (2008) Patient exposure and image quality of low-dose pulmonary computed tomography angiography: comparison of 100- and 80-kVp protocols. Invest Radiol 43(12):871–876PubMedCrossRef
9.
go back to reference Sahani DV, Kalva SP, Hahn PF, Saini S (2007) 16-MDCT angiography in living kidney donors at various tube potentials: impact on image quality and radiation dose. AJR Am J Roentgenol 188(1):115–120PubMedCrossRef Sahani DV, Kalva SP, Hahn PF, Saini S (2007) 16-MDCT angiography in living kidney donors at various tube potentials: impact on image quality and radiation dose. AJR Am J Roentgenol 188(1):115–120PubMedCrossRef
10.
go back to reference Schindera ST, Graca P, Patak MA, Abderhalden S, von Allmen G, Vock P, Szucs-Farkas Z (2009) Thoracoabdominal-aortoiliac multidetector-row CT angiography at 80 and 100 kVp: assessment of image quality and radiation dose. Invest Radiol 44(10):650–655PubMedCrossRef Schindera ST, Graca P, Patak MA, Abderhalden S, von Allmen G, Vock P, Szucs-Farkas Z (2009) Thoracoabdominal-aortoiliac multidetector-row CT angiography at 80 and 100 kVp: assessment of image quality and radiation dose. Invest Radiol 44(10):650–655PubMedCrossRef
11.
go back to reference Schindera ST, Nelson RC, Mukundan S Jr, Paulson EK, Jaffe TA, Miller CM, DeLong DM, Kawaji K, Yoshizumi TT, Samei E (2008) Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection—phantom study. Radiology 246(1):125–132PubMedCrossRef Schindera ST, Nelson RC, Mukundan S Jr, Paulson EK, Jaffe TA, Miller CM, DeLong DM, Kawaji K, Yoshizumi TT, Samei E (2008) Hypervascular liver tumors: low tube voltage, high tube current multi-detector row CT for enhanced detection—phantom study. Radiology 246(1):125–132PubMedCrossRef
12.
go back to reference Marin D, Nelson RC, Samei E, Paulson EK, Ho LM, Boll DT, DeLong DM, Yoshizumi TT, Schindera ST (2009) Hypervascular liver tumors: low tube voltage, high tube current multidetector CT during late hepatic arterial phase for detection—initial clinical experience. Radiology 251(3):771–779PubMedCrossRef Marin D, Nelson RC, Samei E, Paulson EK, Ho LM, Boll DT, DeLong DM, Yoshizumi TT, Schindera ST (2009) Hypervascular liver tumors: low tube voltage, high tube current multidetector CT during late hepatic arterial phase for detection—initial clinical experience. Radiology 251(3):771–779PubMedCrossRef
13.
go back to reference Kyriakou Y, Kachelriess M, Knaup M, Krause JU, Kalender WA (2006) Impact of the z-flying focal spot on resolution and artifact behavior for a 64-slice spiral CT scanner. Eur Radiol 16(6):1206–1215PubMedCrossRef Kyriakou Y, Kachelriess M, Knaup M, Krause JU, Kalender WA (2006) Impact of the z-flying focal spot on resolution and artifact behavior for a 64-slice spiral CT scanner. Eur Radiol 16(6):1206–1215PubMedCrossRef
14.
go back to reference Yu L, Li H, Fletcher JG, McCollough CH (2010) Automatic selection of tube potential for radiation dose reduction in CT: a general strategy. Med Phys 37(1):234–243PubMedCrossRef Yu L, Li H, Fletcher JG, McCollough CH (2010) Automatic selection of tube potential for radiation dose reduction in CT: a general strategy. Med Phys 37(1):234–243PubMedCrossRef
15.
go back to reference Eller A, May MS, Scharf M, Schmid A, Kuefner M, Uder M, Lell MM (2012) Attenuation-based automatic kilovolt selection in abdominal computed tomography: effects on radiation exposure and image quality. Invest Radiol 47(10):559–565PubMedCrossRef Eller A, May MS, Scharf M, Schmid A, Kuefner M, Uder M, Lell MM (2012) Attenuation-based automatic kilovolt selection in abdominal computed tomography: effects on radiation exposure and image quality. Invest Radiol 47(10):559–565PubMedCrossRef
16.
go back to reference Eller A, Wuest W, Scharf M, Brand M, Achenbach S, Uder M, Lell MM (2013) Attenuation-based automatic kilovolt (kV)-selection in computed tomography of the chest: effects on radiation exposure and image quality. Eur J Radiol 82(12):2386–2391PubMedCrossRef Eller A, Wuest W, Scharf M, Brand M, Achenbach S, Uder M, Lell MM (2013) Attenuation-based automatic kilovolt (kV)-selection in computed tomography of the chest: effects on radiation exposure and image quality. Eur J Radiol 82(12):2386–2391PubMedCrossRef
17.
go back to reference Gnannt R, Winklehner A, Goetti R, Schmidt B, Kollias S, Alkadhi H (2012) Low kilovoltage CT of the neck with 70 kVp: comparison with a standard protocol. AJNR 33(6):1014–1019PubMedCrossRef Gnannt R, Winklehner A, Goetti R, Schmidt B, Kollias S, Alkadhi H (2012) Low kilovoltage CT of the neck with 70 kVp: comparison with a standard protocol. AJNR 33(6):1014–1019PubMedCrossRef
18.
go back to reference Winklehner A, Goetti R, Baumueller S, Karlo C, Schmidt B, Raupach R, Flohr T, Frauenfelder T, Alkadhi H (2011) Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness. Invest Radiol 46(12):767–773PubMedCrossRef Winklehner A, Goetti R, Baumueller S, Karlo C, Schmidt B, Raupach R, Flohr T, Frauenfelder T, Alkadhi H (2011) Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: improved dose effectiveness. Invest Radiol 46(12):767–773PubMedCrossRef
19.
go back to reference Eller A, Wuest W, Kramer M, May M, Schmid A, Uder M, Lell MM (2014) Carotid CTA: radiation exposure and image quality with the use of attenuation-based, automated kilovolt selection. AJNR 35(2):237–241PubMedCrossRef Eller A, Wuest W, Kramer M, May M, Schmid A, Uder M, Lell MM (2014) Carotid CTA: radiation exposure and image quality with the use of attenuation-based, automated kilovolt selection. AJNR 35(2):237–241PubMedCrossRef
20.
go back to reference Korn A, Fenchel M, Bender B, Danz S, Thomas C, Ketelsen D, Claussen CD, Moonis G, Krauss B, Heuschmid M, Ernemann U, Brodoefel H (2013) High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation dose. Neuroradiology 55(4):423–430PubMedCrossRef Korn A, Fenchel M, Bender B, Danz S, Thomas C, Ketelsen D, Claussen CD, Moonis G, Krauss B, Heuschmid M, Ernemann U, Brodoefel H (2013) High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation dose. Neuroradiology 55(4):423–430PubMedCrossRef
21.
go back to reference Komlosi P, Zhang Y, Leiva-Salinas C, Ornan D, Patrie JT, Xin W, Grady D, Wintermark M (2014) Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies. Neuroradiology 56(3):187–193PubMedCrossRef Komlosi P, Zhang Y, Leiva-Salinas C, Ornan D, Patrie JT, Xin W, Grady D, Wintermark M (2014) Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies. Neuroradiology 56(3):187–193PubMedCrossRef
22.
go back to reference Zhang WL, Li M, Zhang B, Geng HY, Liang YQ, Xu K, Li SB (2013) CT angiography of the head-and-neck vessels acquired with low tube voltage, low iodine, and iterative image reconstruction: clinical evaluation of radiation dose and image quality. PLoS One 8(12):e81486PubMedCentralPubMedCrossRef Zhang WL, Li M, Zhang B, Geng HY, Liang YQ, Xu K, Li SB (2013) CT angiography of the head-and-neck vessels acquired with low tube voltage, low iodine, and iterative image reconstruction: clinical evaluation of radiation dose and image quality. PLoS One 8(12):e81486PubMedCentralPubMedCrossRef
Metadata
Title
Automated tube voltage adaptation in head and neck computed tomography between 120 and 100 kV: effects on image quality and radiation dose
Authors
Matthias S. May
Manuel R. Kramer
Achim Eller
Wolfgang Wuest
Michael Scharf
Michael Brand
Marc Saake
Bernhard Schmidt
Michael Uder
Michael M. Lell
Publication date
01-09-2014
Publisher
Springer Berlin Heidelberg
Published in
Neuroradiology / Issue 9/2014
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-014-1393-4

Other articles of this Issue 9/2014

Neuroradiology 9/2014 Go to the issue