Published in: Knee Surgery, Sports Traumatology, Arthroscopy 9/2017

01-09-2017 | Knee

Assessment of patient-specific instrumentation precision through bone resection measurements

Authors: F. Zambianchi, A. Colombelli, V. Digennaro, A. Marcovigi, R. Mugnai, F. Fiacchi, D. Sandoni, A. Belluati, F. Catani

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 9/2017

Login to get access

Abstract

Purpose

In the present study, the precision of two patient-specific instrumentation (PSI) systems for total knee arthroplasty (TKA) was evaluated by comparing bony resection thicknesses of the pre-operative PSI planning and intra-operative measurements by a vernier calliper. It was hypothesized that the data provided by pre-operative planning were accurate within ±2 mm of the bone resection thickness measured intra-operatively.

Methods

Forty-one patient-specific TKAs were examined: 25 performed with Visionaire® technology and 16 with OtisMed® system. PSI accuracy was analysed comparing the resected bone thicknesses in the femoral and tibial cuts with pre-operatively planned resections. To determine pre-operative planning precision, the thickness values reported by the PSI planning were subtracted from the values reported intra-operatively by the calliper.

Results

The mean absolute differences between pre-operatively planned resections and corresponding intra-operative thickness measurements ranged from a minimum of 2.6 mm (SD 0.8) to a maximum of 3.6 mm (SD 1.3) in all three anatomical planes in both groups. In every plane, the mean absolute discrepancies between planned resections and measured cuts differed significantly from zero (p < 0.0001). The proportion of differences within ±2 mm between intra-operative measured resections and planned PSI cuts occurred in more than 90 % of the cohort for femoral distal resections. Less precision was reported for the femoral posterior medial cuts (70.7 % within ±2 mm) and the tibial cuts (70.7 % on the medial, 75.6 % on the lateral side). Prosthetic component alignment on the coronal and transverse planes resulted in considerable deviations from the pre-operative planning.

Conclusion

The two examined PSI technologies were accurate in femoral distal cuts, determining acceptable femoral component placement on the coronal plane. Posterior femoral and tibial cuts were less precise. Deviations from the pre-operative resection planning were reported in every plane. Inaccuracy was explained by ambiguous custom-made jigs placement on the bony surface.

Level of evidence

III.
Literature
1.
go back to reference Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KD (2010) Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res 468:57–63CrossRefPubMed Bourne RB, Chesworth BM, Davis AM, Mahomed NN, Charron KD (2010) Patient satisfaction after total knee arthroplasty: who is satisfied and who is not? Clin Orthop Relat Res 468:57–63CrossRefPubMed
2.
go back to reference Catani F, Biasca N, Ensini A, Leardini A, Bianchi L, Digennaro V, Giannini S (2008) Alignment deviations between bone resection and final implant positioning in computer-navigated total knee arthroplasty. J Bone Joint Surg Am 90(4):765–771CrossRefPubMed Catani F, Biasca N, Ensini A, Leardini A, Bianchi L, Digennaro V, Giannini S (2008) Alignment deviations between bone resection and final implant positioning in computer-navigated total knee arthroplasty. J Bone Joint Surg Am 90(4):765–771CrossRefPubMed
3.
go back to reference Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C (2013) A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Joint J 95:354–359CrossRefPubMed Chareancholvanich K, Narkbunnam R, Pornrattanamaneewong C (2013) A prospective randomised controlled study of patient-specific cutting guides compared with conventional instrumentation in total knee replacement. Bone Joint J 95:354–359CrossRefPubMed
4.
go back to reference Choong PF, Dowsey MM, Stoney JD (2009) Does accurate anatomical alignment result in better function and quality of life? Comparing conventional and computer-assisted total knee arthroplasty. J Arthroplasty 24:560–569CrossRefPubMed Choong PF, Dowsey MM, Stoney JD (2009) Does accurate anatomical alignment result in better function and quality of life? Comparing conventional and computer-assisted total knee arthroplasty. J Arthroplasty 24:560–569CrossRefPubMed
5.
go back to reference Conteduca F, Iorio R, Mazza D, Caperna L, Bolle G, Argento G, Ferretti A (2013) Evaluation of the accuracy of a patient-specific instrumentation by navigation. Knee Surg Sports Traumatol Arthrosc 21:2194–2199CrossRefPubMed Conteduca F, Iorio R, Mazza D, Caperna L, Bolle G, Argento G, Ferretti A (2013) Evaluation of the accuracy of a patient-specific instrumentation by navigation. Knee Surg Sports Traumatol Arthrosc 21:2194–2199CrossRefPubMed
6.
go back to reference Ensini A, Timoncini A, Cenni F, Belvedere C, Fusai F, Leardini A, Giannini S (2014) Intra- and post-operative accuracy assessments of two different patient-specific instrumentation systems for total knee replacement. Knee Surg Sports Traumatol Arthrosc 22:621–629CrossRefPubMed Ensini A, Timoncini A, Cenni F, Belvedere C, Fusai F, Leardini A, Giannini S (2014) Intra- and post-operative accuracy assessments of two different patient-specific instrumentation systems for total knee replacement. Knee Surg Sports Traumatol Arthrosc 22:621–629CrossRefPubMed
7.
go back to reference Fang DM, Ritter MA, Davis KE (2009) Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty 24(6 Suppl):39–43CrossRefPubMed Fang DM, Ritter MA, Davis KE (2009) Coronal alignment in total knee arthroplasty: just how important is it? J Arthroplasty 24(6 Suppl):39–43CrossRefPubMed
8.
go back to reference Goyal N, Stulberg SD (2015) Evaluating the precision of preoperative planning in patient specific instrumentation: can a single MRI yield different preoperative plans? J Arthroplasty 30(7):1250–1253CrossRefPubMed Goyal N, Stulberg SD (2015) Evaluating the precision of preoperative planning in patient specific instrumentation: can a single MRI yield different preoperative plans? J Arthroplasty 30(7):1250–1253CrossRefPubMed
9.
go back to reference Hafez MA, Chelule KL, Seedhom BB, Sherman KP (2006) Computer-assisted total knee arthroplasty using patient-specific templating. Clin Orthop Relat Res 444:184–192CrossRefPubMed Hafez MA, Chelule KL, Seedhom BB, Sherman KP (2006) Computer-assisted total knee arthroplasty using patient-specific templating. Clin Orthop Relat Res 444:184–192CrossRefPubMed
10.
go back to reference Heyse TJ, Tibesku CO (2014) Improved femoral component rotation in TKA using patient-specific instrumentation. Knee 21:268–271CrossRefPubMed Heyse TJ, Tibesku CO (2014) Improved femoral component rotation in TKA using patient-specific instrumentation. Knee 21:268–271CrossRefPubMed
11.
go back to reference Hirschmann MT, Knala P, Amsler F, Iranpour F, Friederich NF, Cobb JP (2011) The position and orientation of total knee replacement components: a comparison of conventional radiographs, transverse 2D-CT slices and 3D-CT reconstruction. J Bone Joint Surg Br 93(5):629–633CrossRefPubMed Hirschmann MT, Knala P, Amsler F, Iranpour F, Friederich NF, Cobb JP (2011) The position and orientation of total knee replacement components: a comparison of conventional radiographs, transverse 2D-CT slices and 3D-CT reconstruction. J Bone Joint Surg Br 93(5):629–633CrossRefPubMed
12.
go back to reference Jenny JY, Boeri C, Picard F, Leitner F (2004) Reproducibility of intra-operative measurements of the mechanical axes of the lower limb during total knee replacement with a non-image-based navigation system. Comput Aided Surg 9(4):161–165CrossRefPubMed Jenny JY, Boeri C, Picard F, Leitner F (2004) Reproducibility of intra-operative measurements of the mechanical axes of the lower limb during total knee replacement with a non-image-based navigation system. Comput Aided Surg 9(4):161–165CrossRefPubMed
13.
go back to reference Koch PP, Muller D, Pisan M, Fucentese SF (2013) Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique. Knee Surg Sports Traumatol Arthrosc 21(10):2200–2205CrossRefPubMed Koch PP, Muller D, Pisan M, Fucentese SF (2013) Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique. Knee Surg Sports Traumatol Arthrosc 21(10):2200–2205CrossRefPubMed
14.
go back to reference Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R (2009) Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty 24:570–578CrossRefPubMed Longstaff LM, Sloan K, Stamp N, Scaddan M, Beaver R (2009) Good alignment after total knee arthroplasty leads to faster rehabilitation and better function. J Arthroplasty 24:570–578CrossRefPubMed
15.
go back to reference Lustig S, Scholes CJ, Oussedik SI, Kinzel V, Coolican MRJ, Parker DA (2013) Unsatisfactory accuracy as determined by computer navigation of VISIONAIRE patient-specific instrumentation for total knee arthroplasty. J Arthroplasty 28(3):469–473CrossRefPubMed Lustig S, Scholes CJ, Oussedik SI, Kinzel V, Coolican MRJ, Parker DA (2013) Unsatisfactory accuracy as determined by computer navigation of VISIONAIRE patient-specific instrumentation for total knee arthroplasty. J Arthroplasty 28(3):469–473CrossRefPubMed
16.
go back to reference Mahfouz M, Abdel Fatah EE, Bowers LS, Scuderi G (2012) Three-dimensional morphology of the knee reveals ethnic differences. Clin Orthop Relat Res 470(1):172–185CrossRefPubMed Mahfouz M, Abdel Fatah EE, Bowers LS, Scuderi G (2012) Three-dimensional morphology of the knee reveals ethnic differences. Clin Orthop Relat Res 470(1):172–185CrossRefPubMed
17.
go back to reference Noble JW Jr, Moore CA, Liu N (2012) The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplasty 27(1):153–155CrossRefPubMed Noble JW Jr, Moore CA, Liu N (2012) The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplasty 27(1):153–155CrossRefPubMed
18.
go back to reference Nogler M, Hozack W, Collopy D, Mayr E, Deirmengian G, Sekyra K (2012) Alignment for total knee replacement: a comparison of kinematic axis versus mechanical axis techniques. A cadaver study. Int Orthop 36:2249–2253CrossRefPubMedPubMedCentral Nogler M, Hozack W, Collopy D, Mayr E, Deirmengian G, Sekyra K (2012) Alignment for total knee replacement: a comparison of kinematic axis versus mechanical axis techniques. A cadaver study. Int Orthop 36:2249–2253CrossRefPubMedPubMedCentral
19.
go back to reference Nunley RM, Ellison BS, Zhu J, Ruh Barrack RL (2012) Do patient-specific guides improve coronal alignment in total knee arthroplasty? Clin Orthop Relat Res 470:895–902CrossRefPubMed Nunley RM, Ellison BS, Zhu J, Ruh Barrack RL (2012) Do patient-specific guides improve coronal alignment in total knee arthroplasty? Clin Orthop Relat Res 470:895–902CrossRefPubMed
20.
go back to reference Otani T, Whiteside LA, White SE (1993) Cutting errors in preparation of the femoral components in total knee arthroplasty. J Arthroplasty 8(5):503–510CrossRefPubMed Otani T, Whiteside LA, White SE (1993) Cutting errors in preparation of the femoral components in total knee arthroplasty. J Arthroplasty 8(5):503–510CrossRefPubMed
21.
go back to reference Parratte S, Blanc G, Boussermart T, Ollivier M, Le Corroller T, Argenson JN (2013) Rotation in total knee arthroplasty: no difference between patient-specific and conventional instrumentation. Knee Surg Sports Traumatol Arthrosc 21(10):2213–2219CrossRefPubMed Parratte S, Blanc G, Boussermart T, Ollivier M, Le Corroller T, Argenson JN (2013) Rotation in total knee arthroplasty: no difference between patient-specific and conventional instrumentation. Knee Surg Sports Traumatol Arthrosc 21(10):2213–2219CrossRefPubMed
22.
go back to reference Scholes C, Sahni V, Lustig S, Parker DA, Coolican MRJ (2014) Patient-specific instrumentation for total knee arthroplasty does not match the pre-operative plan as assessed by intra-operative computer-assisted navigation. Knee Surg Sports Traumatol Arthrosc 22:660–665CrossRefPubMed Scholes C, Sahni V, Lustig S, Parker DA, Coolican MRJ (2014) Patient-specific instrumentation for total knee arthroplasty does not match the pre-operative plan as assessed by intra-operative computer-assisted navigation. Knee Surg Sports Traumatol Arthrosc 22:660–665CrossRefPubMed
23.
go back to reference Stoeckl B, Nogler M, Krismer M, Beimel C, de la Barrera JL, Kessler O (2006) Reliability of the transepicondylar axis as an anatomical landmark in total knee arthroplasty. J Arthroplasty 21(6):878–882CrossRefPubMed Stoeckl B, Nogler M, Krismer M, Beimel C, de la Barrera JL, Kessler O (2006) Reliability of the transepicondylar axis as an anatomical landmark in total knee arthroplasty. J Arthroplasty 21(6):878–882CrossRefPubMed
24.
go back to reference Thienpont E, Schwab PE, Fennema P (2014) A systematic review and meta-analysis of patient-specific instrumentation for improving alignment of the components in total knee replacement. Bone Joint J 96:1052–1061CrossRefPubMed Thienpont E, Schwab PE, Fennema P (2014) A systematic review and meta-analysis of patient-specific instrumentation for improving alignment of the components in total knee replacement. Bone Joint J 96:1052–1061CrossRefPubMed
25.
go back to reference Victor J, Dujardin J, Vandenneucker H, Arnout N, Bellemans J (2014) Patient-specific guides do not improve accuracy in total knee arthroplasty. a prospective randomized controlled trial. Clin Orthop Relat Res 472:263–271CrossRefPubMed Victor J, Dujardin J, Vandenneucker H, Arnout N, Bellemans J (2014) Patient-specific guides do not improve accuracy in total knee arthroplasty. a prospective randomized controlled trial. Clin Orthop Relat Res 472:263–271CrossRefPubMed
26.
go back to reference Victor J, Van Doninck D, Labey L, Innocenti B, Parizel PM, Bellemans J (2009) How precise can bony landmarks be determined on a CT scan of the knee? Knee 16(5):358–365CrossRefPubMed Victor J, Van Doninck D, Labey L, Innocenti B, Parizel PM, Bellemans J (2009) How precise can bony landmarks be determined on a CT scan of the knee? Knee 16(5):358–365CrossRefPubMed
27.
go back to reference Yue B, Varadarajan KM, Ai S, Tang T, Rubash HE, Li G (2011) Differences of knee anthropometry between Chinese and white men and women. J Arthroplasty 26(1):124–130CrossRefPubMed Yue B, Varadarajan KM, Ai S, Tang T, Rubash HE, Li G (2011) Differences of knee anthropometry between Chinese and white men and women. J Arthroplasty 26(1):124–130CrossRefPubMed
Metadata
Title
Assessment of patient-specific instrumentation precision through bone resection measurements
Authors
F. Zambianchi
A. Colombelli
V. Digennaro
A. Marcovigi
R. Mugnai
F. Fiacchi
D. Sandoni
A. Belluati
F. Catani
Publication date
01-09-2017
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 9/2017
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-015-3949-1

Other articles of this Issue 9/2017

Knee Surgery, Sports Traumatology, Arthroscopy 9/2017 Go to the issue