Published in: Knee Surgery, Sports Traumatology, Arthroscopy 10/2013

01-10-2013 | Knee

Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique

Authors: P. P. Koch, D. Müller, M. Pisan, S. F. Fucentese

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 10/2013

Login to get access

Abstract

Purpose

Patient-specific instrumentation (PSI) technology for the implantation of total knee arthroplasty (TKA) has a rising interest in the orthopaedic community. Data of PSI are controversially discussed. The hypothesis of this paper is that the radiological accuracy of CT-based PSI is similar to the one of navigated TKA published in the literature.

Methods

Since 2010, all 301 consecutively performed PSI TKAs (GMK MyKnee©) were included in this study. The radiological assessment consisted in a preoperative and postoperative standard X-ray and long-standing X-ray. Changes from the planned to the definitively implanted component size were documented. Postoperative analysis included limb alignment and position of femoral and tibial components (for varus/valgus and flexion or tibial slope).

Results

The postoperative average hip–knee–ankle angle was 180.1° ± 2.0°. In the frontal plane a total of 12.4 % of outliers >3°, for the tibial components 4.1 % of outliers >3° and for the femoral components 4.8 % of outliers >3° were measured. A total of 12.3 % of outliers for posterior tibial slope and 9 % of outliers >3° for the femoral flexion were noted. 10.8 % of the 602 planned size components were adapted intraoperatively.

Conclusion

Although it is still unknown which limb axis is the correct one for the best clinical result, a technology providing the aimed axis in a most precise way should be chosen. Comparing the outcome of the current study with the data from the literature, there does not seem to be any difference compared to computer-assisted surgery.

Level of evidence

IV.
Literature
1.
go back to reference Ast MP, Nam D, Haas SB (2012) Patient-specific instrumentation for total knee arthroplasty: a review. Orthop Clin N Am 43:e17–e22CrossRef Ast MP, Nam D, Haas SB (2012) Patient-specific instrumentation for total knee arthroplasty: a review. Orthop Clin N Am 43:e17–e22CrossRef
2.
go back to reference Barrack RL (2011) The results of TKA: what the registries don’t tell us. Orthopedics 34:e485–e487PubMed Barrack RL (2011) The results of TKA: what the registries don’t tell us. Orthopedics 34:e485–e487PubMed
3.
go back to reference Barrack RL, Ruh EL, Williams BM, Ford AD, Foreman K, Nunley RM (2012) Patient specific cutting blocks are currently of no proven value. J Bone Joint Surg Br 94-B(11 Suppl):95–99 Barrack RL, Ruh EL, Williams BM, Ford AD, Foreman K, Nunley RM (2012) Patient specific cutting blocks are currently of no proven value. J Bone Joint Surg Br 94-B(11 Suppl):95–99
4.
go back to reference Bellemans J, Colyn W, Vandenneucker H, Victor J (2012) The Chitranjan Ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus. Clin Orthop Relat Res 470:45–53PubMedCrossRef Bellemans J, Colyn W, Vandenneucker H, Victor J (2012) The Chitranjan Ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus. Clin Orthop Relat Res 470:45–53PubMedCrossRef
5.
go back to reference Bonner TJ, Eardley WG, Patterson P, Gregg PJ (2011) The effect of post-operative mechanical axis alignment on the survival of primary total knee replacements after a follow-up of 15 years. J Bone Joint Surg Br 93-B:1217–1222CrossRef Bonner TJ, Eardley WG, Patterson P, Gregg PJ (2011) The effect of post-operative mechanical axis alignment on the survival of primary total knee replacements after a follow-up of 15 years. J Bone Joint Surg Br 93-B:1217–1222CrossRef
6.
go back to reference Brin YS, Livshetz I, Antoniou J, Greenberg-Dotan S, Zukor DJ (2010) Precise landmarking in computer assisted total knee arthroplasty is critical to final alignment. J Orthop Res 28:1355–1359PubMedCrossRef Brin YS, Livshetz I, Antoniou J, Greenberg-Dotan S, Zukor DJ (2010) Precise landmarking in computer assisted total knee arthroplasty is critical to final alignment. J Orthop Res 28:1355–1359PubMedCrossRef
7.
go back to reference Brin YS, Nikolaou VS, Joseph L, Zukor DJ, Antoniou J (2011) Imageless computer assisted versus conventional total knee replacement. A Bayesian meta-analysis of 23 comparative studies. Int Orthop 35:331–339PubMedCrossRef Brin YS, Nikolaou VS, Joseph L, Zukor DJ, Antoniou J (2011) Imageless computer assisted versus conventional total knee replacement. A Bayesian meta-analysis of 23 comparative studies. Int Orthop 35:331–339PubMedCrossRef
8.
go back to reference Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20:1307–1322PubMedCrossRef Cheng T, Zhao S, Peng X, Zhang X (2012) Does computer-assisted surgery improve postoperative leg alignment and implant positioning following total knee arthroplasty? A meta-analysis of randomized controlled trials? Knee Surg Sports Traumatol Arthrosc 20:1307–1322PubMedCrossRef
9.
go back to reference Choong PF, Dowsey MM, Stoney JD (2009) Does accurate anatomical alignment result in better function and quality of life? Comparing conventional and computer-assisted total knee arthroplasty. J Arthroplasty 24:560–569PubMedCrossRef Choong PF, Dowsey MM, Stoney JD (2009) Does accurate anatomical alignment result in better function and quality of life? Comparing conventional and computer-assisted total knee arthroplasty. J Arthroplasty 24:560–569PubMedCrossRef
10.
go back to reference Collier MB, Engh CA Jr, McAuley JP, Engh GA (2007) Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty. J Bone Joint Surg Am 89-A:1306–1314CrossRef Collier MB, Engh CA Jr, McAuley JP, Engh GA (2007) Factors associated with the loss of thickness of polyethylene tibial bearings after knee arthroplasty. J Bone Joint Surg Am 89-A:1306–1314CrossRef
11.
go back to reference D’Lima DD, Hermida JC, Chen PC, Colwell CW Jr (2001) Polyethylene wear and variations in knee kinematics. Clin Orthop Relat Res 392:124–130PubMedCrossRef D’Lima DD, Hermida JC, Chen PC, Colwell CW Jr (2001) Polyethylene wear and variations in knee kinematics. Clin Orthop Relat Res 392:124–130PubMedCrossRef
12.
go back to reference Daniilidis K, Tibesku CO (2013) Frontal plane alignment after total knee arthroplasty using patient-specific instruments. Int Orthop 37:45–50PubMedCrossRef Daniilidis K, Tibesku CO (2013) Frontal plane alignment after total knee arthroplasty using patient-specific instruments. Int Orthop 37:45–50PubMedCrossRef
13.
go back to reference Hetaimish BM, Khan MM, Simunovic N, Al-Harbi HH, Bhandari M, Zalzal PK (2012) Meta-analysis of navigation vs conventional total knee arthroplasty. J Arthroplasty 27:1177–1182PubMedCrossRef Hetaimish BM, Khan MM, Simunovic N, Al-Harbi HH, Bhandari M, Zalzal PK (2012) Meta-analysis of navigation vs conventional total knee arthroplasty. J Arthroplasty 27:1177–1182PubMedCrossRef
15.
go back to reference Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K (2007) Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty 22:1097–1106PubMedCrossRef Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K (2007) Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty 22:1097–1106PubMedCrossRef
16.
go back to reference Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C (2007) A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three-dimensional evaluation of implant alignment and rotation. J Bone Joint Surg Am 89-A:236–243CrossRef Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C (2007) A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three-dimensional evaluation of implant alignment and rotation. J Bone Joint Surg Am 89-A:236–243CrossRef
17.
go back to reference Ng VY, DeClaire JH, Berend KR, Gulick BC, Lombardi AV Jr (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107PubMedCrossRef Ng VY, DeClaire JH, Berend KR, Gulick BC, Lombardi AV Jr (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107PubMedCrossRef
18.
go back to reference Noble JW Jr, Moore CA, Liu N (2012) The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplasty 27:153–155PubMedCrossRef Noble JW Jr, Moore CA, Liu N (2012) The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplasty 27:153–155PubMedCrossRef
19.
go back to reference Nunley RM, Ellison BS, Ruh EL, Williams BM, Foreman K, Ford AD, Barrack RL (2012) Are patient-specific cutting blocks cost-effective for total knee arthroplasty? Clin Orthop Relat Res 470:889–894PubMedCrossRef Nunley RM, Ellison BS, Ruh EL, Williams BM, Foreman K, Ford AD, Barrack RL (2012) Are patient-specific cutting blocks cost-effective for total knee arthroplasty? Clin Orthop Relat Res 470:889–894PubMedCrossRef
20.
go back to reference Nunley RM, Ellison BS, Zhu J, Ruh EL, Howell SM, Barrack RL (2012) Do patient-specific guides improve coronal alignment in total knee arthroplasty? Clin Orthop Relat Res 470:895–902PubMedCrossRef Nunley RM, Ellison BS, Zhu J, Ruh EL, Howell SM, Barrack RL (2012) Do patient-specific guides improve coronal alignment in total knee arthroplasty? Clin Orthop Relat Res 470:895–902PubMedCrossRef
21.
go back to reference Parratte S, Pagnano MW, Trousdale RT, Berry DJ (2010) Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern, cemented total knee replacements. J Bone Joint Surg Am 92-A:2143–2149CrossRef Parratte S, Pagnano MW, Trousdale RT, Berry DJ (2010) Effect of postoperative mechanical axis alignment on the fifteen-year survival of modern, cemented total knee replacements. J Bone Joint Surg Am 92-A:2143–2149CrossRef
22.
go back to reference Stronach BM, Pelt CE, Erickson J, Peters CL (2013) Patient-specific total knee arthroplasty required frequent surgeon-directed changes. Clin Orthop Relat Res 471:169–174PubMedCrossRef Stronach BM, Pelt CE, Erickson J, Peters CL (2013) Patient-specific total knee arthroplasty required frequent surgeon-directed changes. Clin Orthop Relat Res 471:169–174PubMedCrossRef
23.
go back to reference Watters TS, Mather RC 3rd, Browne JA, Berend KR, Lombardi AV Jr, Bolognesi MP (2011) Analysis of procedure-related costs and proposed benefits of using patient-specific approach in total knee arthroplasty. J Surg Orthop Adv 20:112–116PubMed Watters TS, Mather RC 3rd, Browne JA, Berend KR, Lombardi AV Jr, Bolognesi MP (2011) Analysis of procedure-related costs and proposed benefits of using patient-specific approach in total knee arthroplasty. J Surg Orthop Adv 20:112–116PubMed
Metadata
Title
Radiographic accuracy in TKA with a CT-based patient-specific cutting block technique
Authors
P. P. Koch
D. Müller
M. Pisan
S. F. Fucentese
Publication date
01-10-2013
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 10/2013
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-013-2625-6

Other articles of this Issue 10/2013

Knee Surgery, Sports Traumatology, Arthroscopy 10/2013 Go to the issue