Skip to main content
Top
Published in: Knee Surgery, Sports Traumatology, Arthroscopy 11/2016

01-11-2016 | Knee

Assessing the accuracy of patient-specific guides for total knee arthroplasty

Authors: Jong-Keun Seon, Hyeong-Won Park, Seung-Hyun Yoo, Eun-Kyoo Song

Published in: Knee Surgery, Sports Traumatology, Arthroscopy | Issue 11/2016

Login to get access

Abstract

Purpose

Patient-specific guides have been introduced recently as a means of making accurate bone cuts through custom cutting blocks constructed based on pre-operative three-dimensional imaging. However, the controversy concerning the improved results of patient-specific guides have not been resolved yet; in addition, there have been no studies to investigate the causes of variable with inconsistent results and solutions for the causes.

Methods

Thirty eight patients (38 knees) underwent total knee arthroplasty with patient-specific guides. The mean age of the patients was 68 years (SD ± 6.3), and all patients had a minimum 2-year follow-up. An intra-operative alignment using navigation and the causes of outliers were evaluated.

Results

An average coronal alignment of PSI jigs was 0.5° (SD ± 0.9°) in femur and 0.1° (SD ± 0.8°) in tibia, and the number of outliers was two and three cases, respectively. An average sagittal alignment was 0.6° (SD ± 0.9°) in femur and 5.5° (SD ± 1.1°) in tibia, and the number of outliers was three and five cases, respectively. All outliers resulted from large osteophytes near the contact point of patient-specific guides which disturb sitting of the guide.

Conclusion

It was suggested that patient-specific guides were an effective and safe method to achieve accurate alignments, with no additional intra-operative complication. It is important to note that surgeons need to be precautious using the patient-specific instrumentation in patients with severe varus deformity. In addition, existing osteophytes which disturb sitting of the guides should be carefully evaluated pre-operatively and intra-operatively.

Level of evidence

IV.
Literature
1.
go back to reference Barrack RL, Ruh EL, Williams BM, Ford AD, Foreman K, Nunley RM (2012) Patient specific cutting blocks are currently of no proven value. J Bone Joint Surg Br 94:95–99CrossRefPubMed Barrack RL, Ruh EL, Williams BM, Ford AD, Foreman K, Nunley RM (2012) Patient specific cutting blocks are currently of no proven value. J Bone Joint Surg Br 94:95–99CrossRefPubMed
2.
go back to reference Conteduca F, Iorio R, Mazza D, Caperna L, Bolle G, Argento G, Ferretti A (2013) Evaluation of the accuracy of a patient-specific instrumentation by navigation. Knee Surg Sports Traumatol Arthrosc 21:2194–2199CrossRefPubMed Conteduca F, Iorio R, Mazza D, Caperna L, Bolle G, Argento G, Ferretti A (2013) Evaluation of the accuracy of a patient-specific instrumentation by navigation. Knee Surg Sports Traumatol Arthrosc 21:2194–2199CrossRefPubMed
3.
go back to reference Hernandez-Vaquero D, Suarez-Vazquez A, Iglesias-Fernandez S (2011) Can computer assistance improve the clinical and functional scores in total knee arthroplasty? Clin Orthop Relat Res 469:3436–3442CrossRefPubMedPubMedCentral Hernandez-Vaquero D, Suarez-Vazquez A, Iglesias-Fernandez S (2011) Can computer assistance improve the clinical and functional scores in total knee arthroplasty? Clin Orthop Relat Res 469:3436–3442CrossRefPubMedPubMedCentral
4.
go back to reference Jenny JY, Boeri C, Picard F, Leitner F (2004) Reproducibility of intra-operative measurement of the mechanical axes of the lower limb during total knee replacement with a non-image-based navigation system. Comput Aided Surg 9:161–165CrossRefPubMed Jenny JY, Boeri C, Picard F, Leitner F (2004) Reproducibility of intra-operative measurement of the mechanical axes of the lower limb during total knee replacement with a non-image-based navigation system. Comput Aided Surg 9:161–165CrossRefPubMed
5.
go back to reference Klatt BA, Goyal N, Austin MS, Hozack WJ (2008) Custom-fit total knee arthroplasty (OtisKnee) results in malalignment. J Arthroplast 23:26–29CrossRef Klatt BA, Goyal N, Austin MS, Hozack WJ (2008) Custom-fit total knee arthroplasty (OtisKnee) results in malalignment. J Arthroplast 23:26–29CrossRef
6.
go back to reference Lombardi AV Jr, Berend KR, Adams JB (2008) Patient-specific approach in total knee arthroplasty. Orthopedics 31:927–930CrossRefPubMed Lombardi AV Jr, Berend KR, Adams JB (2008) Patient-specific approach in total knee arthroplasty. Orthopedics 31:927–930CrossRefPubMed
7.
go back to reference Lustig S, Fleury C, Goy D, Neyret P, Donell ST (2011) The accuracy of acquisition of an imageless computer-assisted system and its implication for knee arthroplasty. Knee 18:15–20CrossRefPubMed Lustig S, Fleury C, Goy D, Neyret P, Donell ST (2011) The accuracy of acquisition of an imageless computer-assisted system and its implication for knee arthroplasty. Knee 18:15–20CrossRefPubMed
8.
go back to reference Lustig S, Scholes CJ, Oussedik SI, Kinzel V, Coolican MR, Parker DA (2013) Unsatisfactory accuracy as determined by computer navigation of VISIONAIRE patient-specific instrumentation for total knee arthroplasty. J Arthroplast 28:469–473CrossRef Lustig S, Scholes CJ, Oussedik SI, Kinzel V, Coolican MR, Parker DA (2013) Unsatisfactory accuracy as determined by computer navigation of VISIONAIRE patient-specific instrumentation for total knee arthroplasty. J Arthroplast 28:469–473CrossRef
9.
go back to reference Magnussen RA, Weppe F, Demey G, Servien E, Lustig S (2011) Residual varus alignment does not compromise results of TKAs in patients with preoperative varus. Clin Orthop Relat Res 469:3443–3450CrossRefPubMedPubMedCentral Magnussen RA, Weppe F, Demey G, Servien E, Lustig S (2011) Residual varus alignment does not compromise results of TKAs in patients with preoperative varus. Clin Orthop Relat Res 469:3443–3450CrossRefPubMedPubMedCentral
10.
go back to reference Marx RG, Grimm P, Lillemoe KA, Robertson CM, Ayeni OR, Lyman S, Bogner EA, Pavlov H (2011) Reliability of lower extremity alignment measurement using radiographs and PACS. Knee Surg Sports Traumatol Arthrosc 19:1693–1698CrossRefPubMed Marx RG, Grimm P, Lillemoe KA, Robertson CM, Ayeni OR, Lyman S, Bogner EA, Pavlov H (2011) Reliability of lower extremity alignment measurement using radiographs and PACS. Knee Surg Sports Traumatol Arthrosc 19:1693–1698CrossRefPubMed
11.
go back to reference Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K (2007) Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplast 22:1097–1106CrossRef Mason JB, Fehring TK, Estok R, Banel D, Fahrbach K (2007) Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplast 22:1097–1106CrossRef
12.
go back to reference Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C (2007) A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three-dimensional evaluation of implant alignment and rotation. J Bone Joint Surg Am 89:236–243PubMed Matziolis G, Krocker D, Weiss U, Tohtz S, Perka C (2007) A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three-dimensional evaluation of implant alignment and rotation. J Bone Joint Surg Am 89:236–243PubMed
13.
go back to reference Ng VY, DeClaire JH, Berend KR, Gulick BC, Lombardi AV Jr (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107CrossRefPubMed Ng VY, DeClaire JH, Berend KR, Gulick BC, Lombardi AV Jr (2012) Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 470:99–107CrossRefPubMed
14.
go back to reference Nunley RM, Ellison BS, Zhu J, Ruh EL, Howell SM, Barrack RL (2012) Do patient-specific guides improve coronal alignment in total knee arthroplasty? Clin Orthop Relat Res 470:895–902CrossRefPubMed Nunley RM, Ellison BS, Zhu J, Ruh EL, Howell SM, Barrack RL (2012) Do patient-specific guides improve coronal alignment in total knee arthroplasty? Clin Orthop Relat Res 470:895–902CrossRefPubMed
15.
go back to reference Pitto RP, Graydon AJ, Bradley L, Malak SF, Walker CG, Anderson IA (2006) Accuracy of a computer-assisted navigation system for total knee replacement. J Bone Joint Surg Br 88:601–605CrossRefPubMed Pitto RP, Graydon AJ, Bradley L, Malak SF, Walker CG, Anderson IA (2006) Accuracy of a computer-assisted navigation system for total knee replacement. J Bone Joint Surg Br 88:601–605CrossRefPubMed
16.
go back to reference Scholes C, Sahni V, Lustig S, Parker DA, Coolican MR (2014) Patient-specific instrumentation for total knee arthroplasty does not match the pre-operative plan as assessed by intra-operative computer-assisted navigation. Knee Surg Sports Traumatol Arthrosc 22:660–665CrossRefPubMed Scholes C, Sahni V, Lustig S, Parker DA, Coolican MR (2014) Patient-specific instrumentation for total knee arthroplasty does not match the pre-operative plan as assessed by intra-operative computer-assisted navigation. Knee Surg Sports Traumatol Arthrosc 22:660–665CrossRefPubMed
17.
go back to reference Spencer BA, Mont MA, McGrath MS, Boyd B, Mitrick MF (2009) Initial experience with custom-fit total knee replacement: intra-operative events and long-leg coronal alignment. Int Orthop 33:1571–1575CrossRefPubMed Spencer BA, Mont MA, McGrath MS, Boyd B, Mitrick MF (2009) Initial experience with custom-fit total knee replacement: intra-operative events and long-leg coronal alignment. Int Orthop 33:1571–1575CrossRefPubMed
18.
go back to reference Stronach BM, Pelt CE, Erickson J, Peters CL (2013) Patient-specific total knee arthroplasty required frequent surgeon-directed changes. Clin Orthop Relat Res 471:169–174CrossRefPubMed Stronach BM, Pelt CE, Erickson J, Peters CL (2013) Patient-specific total knee arthroplasty required frequent surgeon-directed changes. Clin Orthop Relat Res 471:169–174CrossRefPubMed
19.
go back to reference Stulberg SD, Yaffe MA, Koo SS (2006) Computer-assisted surgery versus manual total knee arthroplasty: a case-controlled study. J Bone Joint Surg Am 88(4):47–54PubMed Stulberg SD, Yaffe MA, Koo SS (2006) Computer-assisted surgery versus manual total knee arthroplasty: a case-controlled study. J Bone Joint Surg Am 88(4):47–54PubMed
20.
go back to reference Thienpont E, Bellemans J, Delport H, Van Overschelde P, Stuyts B, Brabants K, Victor J (2013) Patient-specific instruments: industry’s innovation with a surgeon’s interest. Knee Surg Sports Traumatol Arthrosc 21:2227–2233CrossRefPubMed Thienpont E, Bellemans J, Delport H, Van Overschelde P, Stuyts B, Brabants K, Victor J (2013) Patient-specific instruments: industry’s innovation with a surgeon’s interest. Knee Surg Sports Traumatol Arthrosc 21:2227–2233CrossRefPubMed
21.
go back to reference Victor J, Dujardin J, Vandenneucker H, Arnout N, Bellemans J (2014) Patient-specific guides do not improve accuracy in total knee arthroplasty: a prospective randomized controlled trial. Clin Orthop Relat Res 472:263–271CrossRefPubMed Victor J, Dujardin J, Vandenneucker H, Arnout N, Bellemans J (2014) Patient-specific guides do not improve accuracy in total knee arthroplasty: a prospective randomized controlled trial. Clin Orthop Relat Res 472:263–271CrossRefPubMed
22.
go back to reference Yaffe MA, Patel A, Mc Coy BW, Luo M, Cayo M, Ghate R, Stulberg SD (2012) Component sizing in total knee arthroplasty: patient-specific guides vs. computer-assisted navigation. Biomed Tech 57:277–282CrossRef Yaffe MA, Patel A, Mc Coy BW, Luo M, Cayo M, Ghate R, Stulberg SD (2012) Component sizing in total knee arthroplasty: patient-specific guides vs. computer-assisted navigation. Biomed Tech 57:277–282CrossRef
Metadata
Title
Assessing the accuracy of patient-specific guides for total knee arthroplasty
Authors
Jong-Keun Seon
Hyeong-Won Park
Seung-Hyun Yoo
Eun-Kyoo Song
Publication date
01-11-2016
Publisher
Springer Berlin Heidelberg
Published in
Knee Surgery, Sports Traumatology, Arthroscopy / Issue 11/2016
Print ISSN: 0942-2056
Electronic ISSN: 1433-7347
DOI
https://doi.org/10.1007/s00167-014-3429-z

Other articles of this Issue 11/2016

Knee Surgery, Sports Traumatology, Arthroscopy 11/2016 Go to the issue