Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique

Authors: M. Cecconi, D. Dawson, R. M. Grounds, A. Rhodes

Published in: Intensive Care Medicine | Issue 3/2009

Login to get access

Abstract

Background

Lithium dilution cardiac output by LiDCO™plus (LiDCO, Cambridge, UK) is a validated methodology for measuring cardiac output. It is used to calibrate a pulse pressure analysis algorithm (PulseCO) for the continuous measurement of subsequent changes in this variable. The variability of measurements, or precision, within patients of lithium dilution cardiac output has not previously been described.

Material and methods

Thirty-five hemodynamically stable patients in intensive care, with no significant variability in heart rate, mean arterial pressure or central venous pressure, were recruited. Fifty-three determinations of cardiac output were made, each using four lithium dilution measurement curves performed consecutively within a maximum period of 10 min. The coefficient of variation of the measurements was determined and used to derive the least significant change in cardiac output that this technique could reliably detect.

Results

For a single measurement, the coefficient of variation was 8%. This equates to the technique being able to detect a change (least significant change) between two measurements of 24%. Averaging two lithium dilution measurements improved the coefficient of variation to 6% with a least significant change of 17%. Using the average of three curves reduced the coefficient of variation to 5% with a least significant change of 14%.

Conclusions

To achieve a good precision with this technique, three lithium dilution measurements should be averaged. This will allow changes in cardiac output of more than 14% to be reliably detected. The understanding of the precision of this technique allows the user to know when a real change has happened to their patient.
Appendix
Available only for authorised users
Literature
1.
go back to reference Rivers E, Nguyen B, Havstad S, Ressler J, Muzzin A, Knoblich B, Peterson E, Tomlanovich M (2001) Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med 345:1368–1377PubMedCrossRef Rivers E, Nguyen B, Havstad S, Ressler J, Muzzin A, Knoblich B, Peterson E, Tomlanovich M (2001) Early goal-directed therapy in the treatment of severe sepsis and septic shock. N Engl J Med 345:1368–1377PubMedCrossRef
2.
go back to reference Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED (2005) Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial [ISRCTN38797445]. Crit Care 9:R687–R693PubMedCrossRef Pearse R, Dawson D, Fawcett J, Rhodes A, Grounds RM, Bennett ED (2005) Early goal-directed therapy after major surgery reduces complications and duration of hospital stay. A randomised, controlled trial [ISRCTN38797445]. Crit Care 9:R687–R693PubMedCrossRef
3.
go back to reference Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D (1970) Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med 283:447–451PubMed Swan HJ, Ganz W, Forrester J, Marcus H, Diamond G, Chonette D (1970) Catheterization of the heart in man with use of a flow-directed balloon-tipped catheter. N Engl J Med 283:447–451PubMed
4.
go back to reference Connors AF Jr, Speroff T, Dawson NV, Thomas C, Harrell FE Jr, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ Jr, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA (1996) The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 276:889–897PubMedCrossRef Connors AF Jr, Speroff T, Dawson NV, Thomas C, Harrell FE Jr, Wagner D, Desbiens N, Goldman L, Wu AW, Califf RM, Fulkerson WJ Jr, Vidaillet H, Broste S, Bellamy P, Lynn J, Knaus WA (1996) The effectiveness of right heart catheterization in the initial care of critically ill patients. SUPPORT Investigators. JAMA 276:889–897PubMedCrossRef
5.
go back to reference Cecconi M, Wilson J, Rhodes A (2006) Pulse pressure analysis. In: Vincent JL (ed) Yearbook of intensive care and emergency medicine. Springer, Berlin, pp 176–184CrossRef Cecconi M, Wilson J, Rhodes A (2006) Pulse pressure analysis. In: Vincent JL (ed) Yearbook of intensive care and emergency medicine. Springer, Berlin, pp 176–184CrossRef
6.
go back to reference Rhodes A, Sunderland R (2004) Arterial pulse pressure analysis: the LiDCOplus system. Update in intensive care and emergency medicine. In: Pinsky MR, Payen D (eds) Functional hemodynamic monitoring. Springer, Heidelberg, pp 183–192 Rhodes A, Sunderland R (2004) Arterial pulse pressure analysis: the LiDCOplus system. Update in intensive care and emergency medicine. In: Pinsky MR, Payen D (eds) Functional hemodynamic monitoring. Springer, Heidelberg, pp 183–192
7.
go back to reference Linton RA, Band DM, Haire KM (1993) A new method of measuring cardiac output in man using lithium dilution. Br J Anaesth 71:262–266PubMedCrossRef Linton RA, Band DM, Haire KM (1993) A new method of measuring cardiac output in man using lithium dilution. Br J Anaesth 71:262–266PubMedCrossRef
8.
go back to reference Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160PubMedCrossRef Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8:135–160PubMedCrossRef
9.
go back to reference Cecconi M, Al-Subaie N, Canete M, Dawson D, Puntis M, Poloniecki J, Grounds R, Rhodes A (2007) Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique. Crit Care 11:P294CrossRef Cecconi M, Al-Subaie N, Canete M, Dawson D, Puntis M, Poloniecki J, Grounds R, Rhodes A (2007) Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique. Crit Care 11:P294CrossRef
10.
go back to reference Lodder MC, Lems WF, Ader HJ, Marthinsen AE, van Coeverden SC, Lips P, Netelenbos JC, Dijkmans BA, Roos JC (2004) Reproducibility of bone mineral density measurement in daily practice. Ann Rheum Dis 63:285–289PubMedCrossRef Lodder MC, Lems WF, Ader HJ, Marthinsen AE, van Coeverden SC, Lips P, Netelenbos JC, Dijkmans BA, Roos JC (2004) Reproducibility of bone mineral density measurement in daily practice. Ann Rheum Dis 63:285–289PubMedCrossRef
11.
go back to reference Jansen JR, Schreuder JJ, Settels JJ, Kornet L, Penn OC, Mulder PG, Versprille A, Wesseling KH (1996) Single injection thermodilution. A flow-corrected method. Anesthesiology 85:481–490PubMedCrossRef Jansen JR, Schreuder JJ, Settels JJ, Kornet L, Penn OC, Mulder PG, Versprille A, Wesseling KH (1996) Single injection thermodilution. A flow-corrected method. Anesthesiology 85:481–490PubMedCrossRef
12.
go back to reference Jansen JR, Schreuder JJ, Punt KD, van den Berg PC, Alfieri O (2001) Mean cardiac output by thermodilution with a single controlled injection. Crit Care Med 29:1868–1873PubMedCrossRef Jansen JR, Schreuder JJ, Punt KD, van den Berg PC, Alfieri O (2001) Mean cardiac output by thermodilution with a single controlled injection. Crit Care Med 29:1868–1873PubMedCrossRef
13.
go back to reference Jansen JR, Schreuder JJ, Bogaard JM, van Rooyen W, Versprille A (1981) Thermodilution technique for measurement of cardiac output during artificial ventilation. J Appl Physiol 51:584–591PubMed Jansen JR, Schreuder JJ, Bogaard JM, van Rooyen W, Versprille A (1981) Thermodilution technique for measurement of cardiac output during artificial ventilation. J Appl Physiol 51:584–591PubMed
14.
go back to reference Jansen JR, Versprille A (1986) Improvement of cardiac output estimation by the thermodilution method during mechanical ventilation. Intensive Care Med 12:71–79PubMedCrossRef Jansen JR, Versprille A (1986) Improvement of cardiac output estimation by the thermodilution method during mechanical ventilation. Intensive Care Med 12:71–79PubMedCrossRef
15.
go back to reference Jansen JR, Schreuder JJ, Settels JJ, Kloek JJ, Versprille A (1990) An adequate strategy for the thermodilution technique in patients during mechanical ventilation. Intensive Care Med 16:422–425PubMedCrossRef Jansen JR, Schreuder JJ, Settels JJ, Kloek JJ, Versprille A (1990) An adequate strategy for the thermodilution technique in patients during mechanical ventilation. Intensive Care Med 16:422–425PubMedCrossRef
16.
go back to reference Jansen JR (1995) The thermodilution method for the clinical assessment of cardiac output. Intensive Care Med 21:691–697PubMedCrossRef Jansen JR (1995) The thermodilution method for the clinical assessment of cardiac output. Intensive Care Med 21:691–697PubMedCrossRef
17.
go back to reference Berthelsen PG, Eldrup N, Nilsson LB, Rasmussen JP (2002) Thermodilution cardiac output. Cold vs room temperature injectate and the importance of measuring the injectate temperature in the right atrium. Acta Anaesthesiol Scand 46:1103–1110PubMedCrossRef Berthelsen PG, Eldrup N, Nilsson LB, Rasmussen JP (2002) Thermodilution cardiac output. Cold vs room temperature injectate and the importance of measuring the injectate temperature in the right atrium. Acta Anaesthesiol Scand 46:1103–1110PubMedCrossRef
18.
go back to reference Nilsson LB, Nilsson JC, Skovgaard LT, Berthelsen PG (2004) Thermodilution cardiac output—are three injections enough? Acta Anaesthesiol Scand 48:1322–1327PubMedCrossRef Nilsson LB, Nilsson JC, Skovgaard LT, Berthelsen PG (2004) Thermodilution cardiac output—are three injections enough? Acta Anaesthesiol Scand 48:1322–1327PubMedCrossRef
19.
go back to reference Ostergaard M, Nilsson LB, Nilsson JC, Rasmussen JP, Berthelsen PG (2005) Precision of bolus thermodilution cardiac output measurements in patients with atrial fibrillation. Acta Anaesthesiol Scand 49:366–372PubMedCrossRef Ostergaard M, Nilsson LB, Nilsson JC, Rasmussen JP, Berthelsen PG (2005) Precision of bolus thermodilution cardiac output measurements in patients with atrial fibrillation. Acta Anaesthesiol Scand 49:366–372PubMedCrossRef
20.
go back to reference Pearse RM, Ikram K, Barry J (2004) Equipment review: an appraisal of the LiDCO plus method of measuring cardiac output. Crit Care 8:190–195PubMedCrossRef Pearse RM, Ikram K, Barry J (2004) Equipment review: an appraisal of the LiDCO plus method of measuring cardiac output. Crit Care 8:190–195PubMedCrossRef
21.
go back to reference Costa MG, Della Rocca G, Chiarandini P, Mattelig S, Pompei L, Barriga MS, Reynolds T, Cecconi M, Pietropaoli P (2008) Continuous and intermittent cardiac output measurement in hyperdynamic conditions: pulmonary artery catheter vs. lithium dilution technique. Intensive Care Med 34:257–263PubMedCrossRef Costa MG, Della Rocca G, Chiarandini P, Mattelig S, Pompei L, Barriga MS, Reynolds T, Cecconi M, Pietropaoli P (2008) Continuous and intermittent cardiac output measurement in hyperdynamic conditions: pulmonary artery catheter vs. lithium dilution technique. Intensive Care Med 34:257–263PubMedCrossRef
22.
go back to reference Jonas MM, Tanser SJ (2002) Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output. Curr Opin Crit Care 8:257–261PubMedCrossRef Jonas MM, Tanser SJ (2002) Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output. Curr Opin Crit Care 8:257–261PubMedCrossRef
23.
go back to reference Jonas MM, Kelly FE, Linton RA, Band DM, O’Brien TK, Linton NW (1999) A comparison of lithium dilution cardiac output measurements made using central and antecubital venous injection of lithium chloride. J Clin Monit Comput 15:525–528PubMedCrossRef Jonas MM, Kelly FE, Linton RA, Band DM, O’Brien TK, Linton NW (1999) A comparison of lithium dilution cardiac output measurements made using central and antecubital venous injection of lithium chloride. J Clin Monit Comput 15:525–528PubMedCrossRef
24.
go back to reference McGee WT, Horswell JL, Calderon J, Janvier G, Van Severen T, Van den Berghe G, Kozikowski L (2007) Validation of a continuous, arterial pressure-based cardiac output measurement: a multicenter, prospective clinical trial. Crit Care 11:R105PubMedCrossRef McGee WT, Horswell JL, Calderon J, Janvier G, Van Severen T, Van den Berghe G, Kozikowski L (2007) Validation of a continuous, arterial pressure-based cardiac output measurement: a multicenter, prospective clinical trial. Crit Care 11:R105PubMedCrossRef
25.
go back to reference Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C (2007) Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med 33:1191–1194PubMedCrossRef Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C (2007) Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med 33:1191–1194PubMedCrossRef
26.
go back to reference Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15:85–91PubMedCrossRef Critchley LA, Critchley JA (1999) A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput 15:85–91PubMedCrossRef
27.
go back to reference Cecconi M, Grounds M, Rhodes A (2007) Methodologies for assessing agreement between two methods of clinical measurement: are we as good as we think we are? Curr Opin Crit Care 13:294–296PubMedCrossRef Cecconi M, Grounds M, Rhodes A (2007) Methodologies for assessing agreement between two methods of clinical measurement: are we as good as we think we are? Curr Opin Crit Care 13:294–296PubMedCrossRef
Metadata
Title
Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique
Authors
M. Cecconi
D. Dawson
R. M. Grounds
A. Rhodes
Publication date
01-03-2009
Publisher
Springer-Verlag
Published in
Intensive Care Medicine / Issue 3/2009
Print ISSN: 0342-4642
Electronic ISSN: 1432-1238
DOI
https://doi.org/10.1007/s00134-008-1292-4

Other articles of this Issue 3/2009

Intensive Care Medicine 3/2009 Go to the issue