01-10-2021 | Acetabular Fracture | Original Article

Antegrade versus retrograde screw fixation of anterior column acetabular fractures: a biomechanical in vitro study

Authors: Georg Osterhoff, Dag Wulsten, Satish Babu, Mark Heyland, Carlotta Pari

Published in: European Journal of Trauma and Emergency Surgery | Issue 5/2021

Login to get access

Abstract

Background

To compare the mechanical strength of antegrade versus retrograde lag screw fixation of anterior column acetabular fractures.

Methods

Standardised anterior column fractures were created in synthetic pelvis models and stabilised by either antegrade (ANTE, n = 4) or retrograde (RETRO, n = 4) anterior column screw fixation. In a validated setup, a cyclic loading protocol was applied with increasing axial force (750 cycles, 250–750 N) followed by load to failure. Construct survival, energy absorbed, construct stiffness, and load to failure were assessed. Descriptive and opto-metric methods were used to describe the mode of failure.

Results

All constructs failed with loads below 1500 N. With regard to energy absorbed until failure, the ANTE group resisted to 3.763 × 105 N*cycles (range 3.760 × 105–3.763 × 105) and the RETRO group to 3.762 × 105 N*cycles (range 3.761 × 105–3.765 × 105; p = 1.0). The load to failure was 1254 N (range 977–1299) in the ANTE group and 1234 N (range 1087–1456) in the RETRO group (p = 1.0). Construct stiffness with 250 N was not different between the two groups (ANTE 192 N/mm vs. RETRO 215 N/mm, p = 0.486). In all samples, the mode of failure was a transiliac fracture with screw breakout due to rotation of the pubic fragment around the axis of the screw with a range of rotational motion [ROM] during cyclic testing of 0.96° in one ANTE sample and 0.82° in one RETRO sample for 750 N, and ROM at failure of 2.53° in one ANTE sample and 2.23° in one RETRO sample. There was some plastic deformation of the screws in all cases but no breakage.

Conclusions

In this in vitro mechanical study, antegrade screw fixation of an anterior column acetabular fracture was not different in construct survival, load to failure, stiffness, and mode of failure when compared to retrograde screw fixation.
Literature
4.
go back to reference Starr AJ, Reinert CM, Jones AL. Percutaneous fixation of the columns of the acetabulum: a new technique. J Orthop Trauma. 1998;12(1):51–8.CrossRefPubMed Starr AJ, Reinert CM, Jones AL. Percutaneous fixation of the columns of the acetabulum: a new technique. J Orthop Trauma. 1998;12(1):51–8.CrossRefPubMed
10.
go back to reference Schäffler A, Döbele S, Stuby F, et al. Die neue anatomische Flügelplatte für osteoporotische Azetabulumfrakturen: biomechanische Testung und erste klinische Erfahrungen (A new anatomical wing plate for osteoporotic acetabular fractures: biomechanical testing and first clinical experience). Z Orthop Unfall. 2014;152(1):26–322. https://doi.org/10.1055/s-0033-1360351.CrossRefPubMed Schäffler A, Döbele S, Stuby F, et al. Die neue anatomische Flügelplatte für osteoporotische Azetabulumfrakturen: biomechanische Testung und erste klinische Erfahrungen (A new anatomical wing plate for osteoporotic acetabular fractures: biomechanical testing and first clinical experience). Z Orthop Unfall. 2014;152(1):26–322. https://​doi.​org/​10.​1055/​s-0033-1360351.CrossRefPubMed
16.
go back to reference Chow S, Shao J, Wang H. Sample size calculations in clinical research, 2nd edn. CRC biostatistics series. London: Chapman & Hall; 2008. Chow S, Shao J, Wang H. Sample size calculations in clinical research, 2nd edn. CRC biostatistics series. London: Chapman & Hall; 2008.
18.
go back to reference Bergmann G, Deuretzbacher G, Heller M, et al. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34(7):859–71.CrossRefPubMed Bergmann G, Deuretzbacher G, Heller M, et al. Hip contact forces and gait patterns from routine activities. J Biomech. 2001;34(7):859–71.CrossRefPubMed
22.
go back to reference Starr AJ, Jones AL, Reinert CM, et al. Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury. 2001;32(Suppl 1):SA45–SA50.CrossRefPubMed Starr AJ, Jones AL, Reinert CM, et al. Preliminary results and complications following limited open reduction and percutaneous screw fixation of displaced fractures of the acetabulum. Injury. 2001;32(Suppl 1):SA45–SA50.CrossRefPubMed
Metadata
Title
Antegrade versus retrograde screw fixation of anterior column acetabular fractures: a biomechanical in vitro study
Authors
Georg Osterhoff
Dag Wulsten
Satish Babu
Mark Heyland
Carlotta Pari
Publication date
01-10-2021
Publisher
Springer Berlin Heidelberg
Published in
European Journal of Trauma and Emergency Surgery / Issue 5/2021
Print ISSN: 1863-9933
Electronic ISSN: 1863-9941
DOI
https://doi.org/10.1007/s00068-019-01255-6

Other articles of this Issue 5/2021

European Journal of Trauma and Emergency Surgery 5/2021 Go to the issue