Single column locking plate fixation is inadequate in two column acetabular fractures. A biomechanical analysis

Authors: Kiarash Khajavi, Arthur T Lee, Derek P Lindsey, Philipp Leucht, Michael J Bellino, Nicholas J Giori

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2010

Login to get access

Abstract

Background

The objective of this study was to determine whether one can achieve stable fixation of a two column (transverse) acetabular fracture by only fixing a single column with a locking plate and unicortical locking screws. We hypothesized that a locking plate applied to the anterior column of a transverse acetabular fracture would create a construct that is more rigid than a non-locking plate, and that this construct would be biomechanically comparable to two column fixation.

Methods

Using urethane foam models of the pelvis, we simulated transverse acetabular fractures and stabilized them with 1) an anterior column plate with bicortical screws, 2) an anterior locking plate with unicortical screws, 3) an anterior plate and posterior column lag screw, and 4) a posterior plate with an anterior column lag screw. These constructs were mechanically loaded on a servohydraulic material testing machine. Construct stiffness and fracture displacement were measured.

Result and Discussion

We found that two column fixation is 54% stiffer than a single column fixation with a conventional plate with bicortical screws. There was no significant difference between fixation with an anterior column locking plate with unicortical screws and an anterior plate with posterior column lag screw. We detected a non-significant trend towards more stiffness for the anterior locking plate compared to the anterior non-locking plate.

Conclusion

In conclusion, a locking plate construct of the anterior column provides less stability than a traditional both column construct with posterior plate and anterior column lag screw. However, the locking construct offers greater strength than a non-locking, bicortical construct, which in addition often requires extensive contouring and its application is oftentimes accompanied by the risk of neurovascular damage.
Appendix
Available only for authorised users
Literature
1.
go back to reference Letournel E: Fractures of the acetabulum. 1993, New York: Springer-Verlag, 2CrossRef Letournel E: Fractures of the acetabulum. 1993, New York: Springer-Verlag, 2CrossRef
2.
go back to reference Shazar N, Brumback RJ, Novak VP, Belkoff SM: Biomechanical evaluation of transverse acetabular fracture fixation. Clin Orthop Relat Res. 1998, 215-222. 352 Shazar N, Brumback RJ, Novak VP, Belkoff SM: Biomechanical evaluation of transverse acetabular fracture fixation. Clin Orthop Relat Res. 1998, 215-222. 352
3.
go back to reference Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ: Biomechanics of locked plates and screws. J Orthop Trauma. 2004, 18 (8): 488-493. 10.1097/00005131-200409000-00003.CrossRefPubMed Egol KA, Kubiak EN, Fulkerson E, Kummer FJ, Koval KJ: Biomechanics of locked plates and screws. J Orthop Trauma. 2004, 18 (8): 488-493. 10.1097/00005131-200409000-00003.CrossRefPubMed
4.
go back to reference Wong MK, Leung F, Chow SP: Treatment of distal femoral fractures in the elderly using a less-invasive plating technique. Int Orthop. 2005, 29 (2): 117-120. 10.1007/s00264-004-0609-3.PubMedCentralCrossRefPubMed Wong MK, Leung F, Chow SP: Treatment of distal femoral fractures in the elderly using a less-invasive plating technique. Int Orthop. 2005, 29 (2): 117-120. 10.1007/s00264-004-0609-3.PubMedCentralCrossRefPubMed
5.
go back to reference Kregor PJ, Stannard JA, Zlowodzki M, Cole PA: Treatment of distal femur fractures using the less invasive stabilization system: surgical experience and early clinical results in 103 fractures. J Orthop Trauma. 2004, 18 (8): 509-520. 10.1097/00005131-200409000-00006.CrossRefPubMed Kregor PJ, Stannard JA, Zlowodzki M, Cole PA: Treatment of distal femur fractures using the less invasive stabilization system: surgical experience and early clinical results in 103 fractures. J Orthop Trauma. 2004, 18 (8): 509-520. 10.1097/00005131-200409000-00006.CrossRefPubMed
6.
go back to reference Farouk O, Krettek C, Miclau T, Schandelmaier P, Tscherne H: Effects of percutaneous and conventional plating techniques on the blood supply to the femur. Archives of Orthopaedic and Trauma Surgery. 1998, 117 (8): 438-441. 10.1007/s004020050288.CrossRefPubMed Farouk O, Krettek C, Miclau T, Schandelmaier P, Tscherne H: Effects of percutaneous and conventional plating techniques on the blood supply to the femur. Archives of Orthopaedic and Trauma Surgery. 1998, 117 (8): 438-441. 10.1007/s004020050288.CrossRefPubMed
7.
go back to reference Farouk O, Krettek C, Miclau T, Schandelmaier P, Guy P, Tscherne H: Minimally invasive plate osteosynthesis: does percutaneous plating disrupt femoral blood supply less than the traditional technique?. J Orthop Trauma. 1999, 13 (6): 401-406. 10.1097/00005131-199908000-00002.CrossRefPubMed Farouk O, Krettek C, Miclau T, Schandelmaier P, Guy P, Tscherne H: Minimally invasive plate osteosynthesis: does percutaneous plating disrupt femoral blood supply less than the traditional technique?. J Orthop Trauma. 1999, 13 (6): 401-406. 10.1097/00005131-199908000-00002.CrossRefPubMed
8.
go back to reference Farouk O, Krettek C, Miclau T, Schandelmaier P, Guy P, Tscherne H: Minimally invasive plate osteosynthesis and vascularity: preliminary results of a cadaver injection study. Injury. 1997, 28 (Suppl 1): A7-12. 10.1016/S0020-1383(97)90110-8.CrossRefPubMed Farouk O, Krettek C, Miclau T, Schandelmaier P, Guy P, Tscherne H: Minimally invasive plate osteosynthesis and vascularity: preliminary results of a cadaver injection study. Injury. 1997, 28 (Suppl 1): A7-12. 10.1016/S0020-1383(97)90110-8.CrossRefPubMed
9.
go back to reference Bergmann G, Graichen F, Rohlmann A: Hip joint loading during walking and running, measured in two patients. J Biomech. 1993, 26 (8): 969-990. 10.1016/0021-9290(93)90058-M.CrossRefPubMed Bergmann G, Graichen F, Rohlmann A: Hip joint loading during walking and running, measured in two patients. J Biomech. 1993, 26 (8): 969-990. 10.1016/0021-9290(93)90058-M.CrossRefPubMed
10.
go back to reference Chang JK, Gill SS, Zura RD, Krause WR, Wang GJ: Comparative strength of three methods of fixation of transverse acetabular fractures. Clin Orthop Relat Res. 2001, 433-441. 10.1097/00003086-200111000-00057. 392 Chang JK, Gill SS, Zura RD, Krause WR, Wang GJ: Comparative strength of three methods of fixation of transverse acetabular fractures. Clin Orthop Relat Res. 2001, 433-441. 10.1097/00003086-200111000-00057. 392
Metadata
Title
Single column locking plate fixation is inadequate in two column acetabular fractures. A biomechanical analysis
Authors
Kiarash Khajavi
Arthur T Lee
Derek P Lindsey
Philipp Leucht
Michael J Bellino
Nicholas J Giori
Publication date
01-12-2010
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2010
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/1749-799X-5-30

Other articles of this Issue 1/2010

Journal of Orthopaedic Surgery and Research 1/2010 Go to the issue