Published in: Clinical Neuroradiology 1/2019

01-03-2019 | Original Article

Low-Dose CCT to Exclude Contraindications to Lumbar Puncture

Benefits and Limitations

Authors: Stella Blasel, Sarah Alex, Hanns Ackermann, Julia Tichy, Joachim Berkefeld, Marlies Wagner

Published in: Clinical Neuroradiology | Issue 1/2019

Login to get access

Abstract

Background

Low-dose cranial computed tomography (LD-CCT) based on iterative reconstruction has been shown to have sufficient image quality to assess cerebrospinal fluid spaces (CSF) and midline structures but not to exclude subtle parenchymal pathologies. Patients without focal neurological deficits often undergo CCT before lumbar puncture (LP) to exclude contraindications to LP including brain herniation or increased CSF pressure. We performed LD-CCT to assess if image quality is appropriate for this indication.

Methods

A total of 58 LD-CCT (220 mA/120 kV) of patients before LP were retrospectively evaluated and compared to 79 normal standard dose cranial computed tomography (SD-CCT) (350 mA/120 kV). Iterative reconstruction used for both dose levels was increased by one factor for LD-CCT. We assessed the signal-to-noise (SNR) and contrast-to-noise ratio (CNR), the dose estimates and scored diagnostic image quality by two raters independently. Significance level was set at p < 0.05.

Results

The inner and outer CSF spaces except the sulci were equally well depicted by the LD-CCT and SD-CCT; however, depiction of the subtle density differences of the brain parenchyma and the sulci was significantly worse in the LD-CCT (p < 0.0001). The SNR in the gray matter (9.35 vs. 10.61, p < 0.05) and white matter (7.23 vs. 8.15, p < 0.001) were significantly lower in LD-CCT than in SD-CCT with significantly lower dose estimates (1.04 vs. 1.69 mSv, respectively p < 0.0001).

Conclusion

The use of LD-CCT with a dose reduction of almost 50% is sufficient to exclude contraindications to LP; however, LD-CCT cannot exclude subtle parenchymal pathologies. Therefore, in patients with suspected parenchymal pathology, SD-CCT is still the method of choice.
Literature
1.
go back to reference Brenner DJ, Hall EJ. Computed tomography—an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–84.CrossRefPubMed Brenner DJ, Hall EJ. Computed tomography—an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–84.CrossRefPubMed
6.
go back to reference Kalra MK, Maher MM, Toth TL, Hamberg LM, Blake MA, Shepard JA, Saini S. Strategies for CT radiation dose optimization. Radiology. 2004;230:619–28.CrossRefPubMed Kalra MK, Maher MM, Toth TL, Hamberg LM, Blake MA, Shepard JA, Saini S. Strategies for CT radiation dose optimization. Radiology. 2004;230:619–28.CrossRefPubMed
7.
go back to reference Kalra MK, Maher MM, Toth TL, Schmidt B, Westerman BL, Morgan HT, Saini S. Techniques and applications of automatic tube current modulation for CT. Radiology. 2004;233:649–57.CrossRefPubMed Kalra MK, Maher MM, Toth TL, Schmidt B, Westerman BL, Morgan HT, Saini S. Techniques and applications of automatic tube current modulation for CT. Radiology. 2004;233:649–57.CrossRefPubMed
8.
go back to reference Hounsfield GN. Computerized transverse axial scanning (tomography). 1. Description of system. Br J Radiol. 1973;46:1016–22.CrossRefPubMed Hounsfield GN. Computerized transverse axial scanning (tomography). 1. Description of system. Br J Radiol. 1973;46:1016–22.CrossRefPubMed
9.
go back to reference Willemink MJ, de Jong PA, Leiner T, de Heer LM, Nievelstein RA, Budde RP, Schilham AM. Iterative reconstruction techniques for computed tomography part 1: technical principles. Eur Radiol. 2013;23:1623–31.CrossRefPubMed Willemink MJ, de Jong PA, Leiner T, de Heer LM, Nievelstein RA, Budde RP, Schilham AM. Iterative reconstruction techniques for computed tomography part 1: technical principles. Eur Radiol. 2013;23:1623–31.CrossRefPubMed
10.
go back to reference Blasel S, Huck L, Lescher S, Ackermann H, Berkefeld J, Wagner M. Low dose CT of the brain in the follow-up of Intracranial hemorrhage. Int J Radiol Imaging Technol. 2016;2(2):015.CrossRef Blasel S, Huck L, Lescher S, Ackermann H, Berkefeld J, Wagner M. Low dose CT of the brain in the follow-up of Intracranial hemorrhage. Int J Radiol Imaging Technol. 2016;2(2):015.CrossRef
11.
go back to reference Bodelle B, Klein E, Naguib NN, Bauer RW, Kerl JM, Al-Butmeh F, Wichmann JL, Ackermann H, Lehnert T, Vogl TJ, Schulz B. Acute intracranial hemorrhage in CT: benefits of sinogram-affirmed iterative reconstruction techniques. AJNR Am J Neuroradiol. 2014;35(3):445–9.CrossRefPubMed Bodelle B, Klein E, Naguib NN, Bauer RW, Kerl JM, Al-Butmeh F, Wichmann JL, Ackermann H, Lehnert T, Vogl TJ, Schulz B. Acute intracranial hemorrhage in CT: benefits of sinogram-affirmed iterative reconstruction techniques. AJNR Am J Neuroradiol. 2014;35(3):445–9.CrossRefPubMed
12.
go back to reference Corcuera-Solano I, Doshi AH, Noor A, Tanenbaum LN. Repeated head CT in the neurosurgical intensive care unit: feasibility of sinogram-affirmed iterative reconstruction-based ultra-low-dose CT for surveillance. AJNR Am J Neuroradiol. 2014;35:1281–7.CrossRefPubMed Corcuera-Solano I, Doshi AH, Noor A, Tanenbaum LN. Repeated head CT in the neurosurgical intensive care unit: feasibility of sinogram-affirmed iterative reconstruction-based ultra-low-dose CT for surveillance. AJNR Am J Neuroradiol. 2014;35:1281–7.CrossRefPubMed
13.
go back to reference Kaul D, Kahn J, Huizing L, Wiener E, Grupp U, Böning G, Ghadjar P, Renz DM, Streitparth F. Reducing Radiation Dose in Adult Head CT using Iterative Reconstruction—A Clinical Study in 177 Patients. Rofo. 2016;188:155–62.CrossRefPubMed Kaul D, Kahn J, Huizing L, Wiener E, Grupp U, Böning G, Ghadjar P, Renz DM, Streitparth F. Reducing Radiation Dose in Adult Head CT using Iterative Reconstruction—A Clinical Study in 177 Patients. Rofo. 2016;188:155–62.CrossRefPubMed
14.
go back to reference von Kummer R, Meyding-Lamadé U, Forsting M, Rosin L, Rieke K, Hacke W, Sartor K. Sensitivity and prognostic value of early CT in occlusion of the middle cerebral artery trunk. AJNR Am J Neuroradiol. 1994;15:9–15. von Kummer R, Meyding-Lamadé U, Forsting M, Rosin L, Rieke K, Hacke W, Sartor K. Sensitivity and prognostic value of early CT in occlusion of the middle cerebral artery trunk. AJNR Am J Neuroradiol. 1994;15:9–15.
15.
go back to reference Diener HC, Ackermann H. Leitlinien für Diagnostik und Therapie in der Neurologie. 5th ed. Stuttgart: Thieme; 2012. Diener HC, Ackermann H. Leitlinien für Diagnostik und Therapie in der Neurologie. 5th ed. Stuttgart: Thieme; 2012.
17.
go back to reference Gröschel K, Gröschel S. How to do: the diagnostic lumbar puncture. Dtsch Med Wochenschr. 2015;140:738–40.CrossRefPubMed Gröschel K, Gröschel S. How to do: the diagnostic lumbar puncture. Dtsch Med Wochenschr. 2015;140:738–40.CrossRefPubMed
18.
go back to reference Jinkins JR, Athale S, Xiong L, Yuh WT, Rothman MI, Nguyen PT. MR of optic papilla protrusion in patients with high intracranial pressure. AJNR Am J Neuroradiol. 1996;17:665–8.PubMed Jinkins JR, Athale S, Xiong L, Yuh WT, Rothman MI, Nguyen PT. MR of optic papilla protrusion in patients with high intracranial pressure. AJNR Am J Neuroradiol. 1996;17:665–8.PubMed
19.
go back to reference Kilic K, Erbas G, Guryildirim M, Arac M, Ilgit E, Coskun B. Lowering the dose in head CT using adaptive statistical iterative reconstruction. AJNR Am J Neuroradiol. 2011;32:1578–82.CrossRefPubMed Kilic K, Erbas G, Guryildirim M, Arac M, Ilgit E, Coskun B. Lowering the dose in head CT using adaptive statistical iterative reconstruction. AJNR Am J Neuroradiol. 2011;32:1578–82.CrossRefPubMed
20.
go back to reference Korn A, Fenchel M, Bender B, Danz S, Hauser TK, Ketelsen D, Flohr T, Claussen CD, Heuschmid M, Ernemann U, Brodoefel H. Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection. AJNR Am J Neuroradiol. 2012;33:218–24.CrossRefPubMed Korn A, Fenchel M, Bender B, Danz S, Hauser TK, Ketelsen D, Flohr T, Claussen CD, Heuschmid M, Ernemann U, Brodoefel H. Iterative reconstruction in head CT: image quality of routine and low-dose protocols in comparison with standard filtered back-projection. AJNR Am J Neuroradiol. 2012;33:218–24.CrossRefPubMed
21.
go back to reference Korn A, Bender B, Fenchel M, Spira D, Schabel C, Thomas C, Flohr T, Claussen CD, Bhadelia R, Ernemann U, Brodoefel H. Sinogram affirmed iterative reconstruction in head CT: Improvement of objective and subjective image quality with concomitant radiation dose reduction. Eur J Radiol. 2013;82:1431–5.CrossRefPubMed Korn A, Bender B, Fenchel M, Spira D, Schabel C, Thomas C, Flohr T, Claussen CD, Bhadelia R, Ernemann U, Brodoefel H. Sinogram affirmed iterative reconstruction in head CT: Improvement of objective and subjective image quality with concomitant radiation dose reduction. Eur J Radiol. 2013;82:1431–5.CrossRefPubMed
22.
go back to reference Rapalino O, Kamalian S, Kamalian S, Payabvash S, Souza LCS, Zhang D, Mukta J, Sahani DV, Lev MH, Pomerantz SR. Cranial CT with adaptive statistical iterative reconstruction: improved image quality with concomitant radiation dose reduction. AJNR Am J Neuroradiol. 2012;33:609–15.CrossRefPubMed Rapalino O, Kamalian S, Kamalian S, Payabvash S, Souza LCS, Zhang D, Mukta J, Sahani DV, Lev MH, Pomerantz SR. Cranial CT with adaptive statistical iterative reconstruction: improved image quality with concomitant radiation dose reduction. AJNR Am J Neuroradiol. 2012;33:609–15.CrossRefPubMed
23.
go back to reference Ren Q, Dewan SK, Li M, Li J, Mao D, Wang Z, Hua Y. Comparison of adaptive statistical iterative and filtered back projection reconstruction techniques in brain CT. Eur J Radiol. 2012;81:2597–601.CrossRefPubMed Ren Q, Dewan SK, Li M, Li J, Mao D, Wang Z, Hua Y. Comparison of adaptive statistical iterative and filtered back projection reconstruction techniques in brain CT. Eur J Radiol. 2012;81:2597–601.CrossRefPubMed
24.
go back to reference Wu TH, Hung SC, Sun JY, Lin CJ, Lin CH, Chiu CF, Liu MJ, Teng MM, Guo WY, Chang CY. How far can the radiation dose be lowered in head CT with iterative reconstruction? Analysis of imaging quality and diagnostic accuracy. Eur Radiol. 2013;23:2612–21.CrossRefPubMed Wu TH, Hung SC, Sun JY, Lin CJ, Lin CH, Chiu CF, Liu MJ, Teng MM, Guo WY, Chang CY. How far can the radiation dose be lowered in head CT with iterative reconstruction? Analysis of imaging quality and diagnostic accuracy. Eur Radiol. 2013;23:2612–21.CrossRefPubMed
25.
go back to reference Bodelle B, Wichmann JL, Scholtz J‑E, Lehnert T, Vogl TJ, Luboldt W, Schulz B. Iterative reconstruction leads to increased subjective and objective image quality in cranial CT in patients with stroke. AJR Am J Roentgenol. 2015;205:618–22.CrossRefPubMed Bodelle B, Wichmann JL, Scholtz J‑E, Lehnert T, Vogl TJ, Luboldt W, Schulz B. Iterative reconstruction leads to increased subjective and objective image quality in cranial CT in patients with stroke. AJR Am J Roentgenol. 2015;205:618–22.CrossRefPubMed
26.
go back to reference Komlosi P, Zhang Y, Leiva-Salinas C, Ornan D, Patrie JT, Xin W, Grady D, Wintermark M. Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies. Neuroradiology. 2014;56:187–93.CrossRefPubMed Komlosi P, Zhang Y, Leiva-Salinas C, Ornan D, Patrie JT, Xin W, Grady D, Wintermark M. Adaptive statistical iterative reconstruction reduces patient radiation dose in neuroradiology CT studies. Neuroradiology. 2014;56:187–93.CrossRefPubMed
27.
go back to reference Ryska P, Kvasnicka T, Jandura J, Klzo L, Grepl J, Zizka J. Reduction of effective dose and organ dose to the eye lens in head MDCT using iterative image reconstruction and automatic tube current modulation. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2014;158(2):265–72.CrossRefPubMed Ryska P, Kvasnicka T, Jandura J, Klzo L, Grepl J, Zizka J. Reduction of effective dose and organ dose to the eye lens in head MDCT using iterative image reconstruction and automatic tube current modulation. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2014;158(2):265–72.CrossRefPubMed
28.
go back to reference Noël PB, Fingerle AA, Renger B, Münzel D, Rummeny EJ, Dobritz M. Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol. 2011;197:1404–9.CrossRefPubMed Noël PB, Fingerle AA, Renger B, Münzel D, Rummeny EJ, Dobritz M. Initial performance characterization of a clinical noise-suppressing reconstruction algorithm for MDCT. AJR Am J Roentgenol. 2011;197:1404–9.CrossRefPubMed
29.
go back to reference Buhk JH, Laqmani A, von Schultzendorff HC, Hammerle D, Sehner S, Adam G, Fiehler J, Nagel HD, Regier M. Intraindividual evaluation of the influence of iterative reconstruction and filter kernel on subjective and objective image quality in computed tomography of the brain. Rofo. 2013;185:741–8.CrossRefPubMed Buhk JH, Laqmani A, von Schultzendorff HC, Hammerle D, Sehner S, Adam G, Fiehler J, Nagel HD, Regier M. Intraindividual evaluation of the influence of iterative reconstruction and filter kernel on subjective and objective image quality in computed tomography of the brain. Rofo. 2013;185:741–8.CrossRefPubMed
31.
go back to reference Pexman JH, Barber PA, Hill MD, Sevick RJ, Demchuk AM, Hudon ME, Hu WY, Buchan AM. Use of the Alberta Stroke Program Early CT Score (ASPECTS) for assessing CT scans in patients with acute stroke. AJNR Am J Neuroradiol. 2001;22:1534–42.PubMed Pexman JH, Barber PA, Hill MD, Sevick RJ, Demchuk AM, Hudon ME, Hu WY, Buchan AM. Use of the Alberta Stroke Program Early CT Score (ASPECTS) for assessing CT scans in patients with acute stroke. AJNR Am J Neuroradiol. 2001;22:1534–42.PubMed
32.
go back to reference Goyal M, Menon BK, van Zwam WH, Dippel DW, Mitchell PJ, Demchuk AM, Dávalos A, Majoie CB, van der Lugt A, de Miquel MA, Donnan GA, Roos YB, Bonafe A, Jahan R, Diener HC, van den Berg LA, Levy EI, Berkhemer OA, Pereira VM, Rempel J, Millán M, Davis SM, Roy D, Thornton J, Román LS, Ribó M, Beumer D, Stouch B, Brown S, Campbell BC, van Oostenbrugge RJ, Saver JL, Hill MD, Jovin TG; HERMES collaborators. Endovascular thrombectomy after large-vessel ischaemic stroke: a meta-analysis of individual patient data from five randomised trials. Lancet. 2016;387(10029):1723–31.CrossRefPubMed Goyal M, Menon BK, van Zwam WH, Dippel DW, Mitchell PJ, Demchuk AM, Dávalos A, Majoie CB, van der Lugt A, de Miquel MA, Donnan GA, Roos YB, Bonafe A, Jahan R, Diener HC, van den Berg LA, Levy EI, Berkhemer OA, Pereira VM, Rempel J, Millán M, Davis SM, Roy D, Thornton J, Román LS, Ribó M, Beumer D, Stouch B, Brown S, Campbell BC, van Oostenbrugge RJ, Saver JL, Hill MD, Jovin TG; HERMES collaborators. Endovascular thrombectomy after large-vessel ischaemic stroke: a meta-analysis of individual patient data from five randomised trials. Lancet. 2016;387(10029):1723–31.CrossRefPubMed
33.
go back to reference Tumani H, Petzold A, Wick M, Kühn HJ, Uhr M, Otto M, Regeniter A, Brettschneider J. Liquordiagnostik bei CT-negativer Subarachnoidalblutung. Nervenarzt. 2010;81:973–9.CrossRefPubMed Tumani H, Petzold A, Wick M, Kühn HJ, Uhr M, Otto M, Regeniter A, Brettschneider J. Liquordiagnostik bei CT-negativer Subarachnoidalblutung. Nervenarzt. 2010;81:973–9.CrossRefPubMed
35.
go back to reference Dubosh NM, Bellolio MF, Rabinstein AA, Edlow JA. Sensitivity of early brain computed tomography to exclude aneurysmal subarachnoid hemorrhage: a systematic review and meta-analysis. Stroke. 2016;47:750–5.CrossRefPubMed Dubosh NM, Bellolio MF, Rabinstein AA, Edlow JA. Sensitivity of early brain computed tomography to exclude aneurysmal subarachnoid hemorrhage: a systematic review and meta-analysis. Stroke. 2016;47:750–5.CrossRefPubMed
36.
go back to reference Ewen K. Die effektive Dosis in der Röntgendiagnostik. Z Med Phys. 2000;10:119–22.CrossRef Ewen K. Die effektive Dosis in der Röntgendiagnostik. Z Med Phys. 2000;10:119–22.CrossRef
Metadata
Title
Low-Dose CCT to Exclude Contraindications to Lumbar Puncture
Benefits and Limitations
Authors
Stella Blasel
Sarah Alex
Hanns Ackermann
Julia Tichy
Joachim Berkefeld
Marlies Wagner
Publication date
01-03-2019
Publisher
Springer Berlin Heidelberg
Published in
Clinical Neuroradiology / Issue 1/2019
Print ISSN: 1869-1439
Electronic ISSN: 1869-1447
DOI
https://doi.org/10.1007/s00062-017-0628-2

Other articles of this Issue 1/2019

Clinical Neuroradiology 1/2019 Go to the issue

Information

Information