Published in: Journal of Nuclear Cardiology 4/2011

01-08-2011 | Original Article

Comparison of high efficiency CZT SPECT MPI to coronary angiography

Authors: W. Lane Duvall, MD, Joseph M. Sweeny, MD, Lori B. Croft, MD, Maya H. Barghash, MD, Nitin K. Kulkarni, MD, Krista A. Guma, CNMT, Milena J. Henzlova, MD

Published in: Journal of Nuclear Cardiology | Issue 4/2011

Login to get access

Abstract

Background

The recently introduced cadmium zinc telluride (CZT) SPECT cameras have the potential to reduce radiation exposure to patients and shorten imaging time. So far, there has been only one small study comparing the results of high efficiency CZT SPECT myocardial perfusion imaging (MPI) to invasive coronary angiography.

Methods

All patients who had either a Tc-99m sestamibi or Tl-201 SPECT MPI study using a CZT camera (GE Discovery NM 530c) over a 1-year period followed by a coronary angiogram within 2 months were included. Only patients with a history of CABG surgery were excluded. Standard stress protocols were employed. Rest images were acquired for 5 min and stress supine and prone images for 3 min each. Both MPI studies and coronary angiograms were interpreted by blinded readers. A standard 17-segment model was employed for MPI interpretation, and coronary angiograms were interpreted for the presence of obstructive epicardial coronary artery disease (CAD) defined as ≥70% luminal narrowing. Correlation was based on the ability to diagnose obstructive epicardial CAD.

Results

Of the 3,111 patients who underwent SPECT imaging using the CZT camera during this time period, 230 patients qualified for the correlation study (mean age 64.2 ± 11.0 years old, 69% male, and 49% had a history of intracoronary stenting). Tc-99m was used in 76% vs Tl-201 in 24% of the studies. Exercise stress was performed in 60% of patients and vasodilator pharmacologic stress in 40%. Sensitivity was 95%, normalcy rate was 97%, and accuracy was 69% for detecting obstructive CAD.

Conclusions

In this so far largest correlation study between coronary angiography and high efficiency CZT SPECT imaging, a high sensitivity and accuracy for detecting obstructive epicardial CAD was found for this new SPECT camera technology.
Literature
1.
go back to reference Garcia EV, Faber TL, Esteves FP. Cardiac dedicated ultrafast SPECT cameras: New designs and clinical implications. J Nucl Med 2011;52:210-7.PubMedCrossRef Garcia EV, Faber TL, Esteves FP. Cardiac dedicated ultrafast SPECT cameras: New designs and clinical implications. J Nucl Med 2011;52:210-7.PubMedCrossRef
2.
go back to reference Slomka PJ, Patton JA, Berman DS, Germano G. Advances in technical aspects of myocardial perfusion SPECT imaging. J Nucl Cardiol 2009;16:255-76.PubMedCrossRef Slomka PJ, Patton JA, Berman DS, Germano G. Advances in technical aspects of myocardial perfusion SPECT imaging. J Nucl Cardiol 2009;16:255-76.PubMedCrossRef
3.
go back to reference Sharir T, Ben-Haim S, Merzon K, Prochorov V, Dickman D, Berman DS. High-speed myocardial perfusion imaging initial clinical comparison with conventional dual detector anger camera imaging. JACC Cardiovasc Imaging 2008;1:156-63.PubMedCrossRef Sharir T, Ben-Haim S, Merzon K, Prochorov V, Dickman D, Berman DS. High-speed myocardial perfusion imaging initial clinical comparison with conventional dual detector anger camera imaging. JACC Cardiovasc Imaging 2008;1:156-63.PubMedCrossRef
4.
go back to reference Sharir T, Slomka PJ, Hayes SW, DiCarli MF, Ziffer JA, Martin WH, et al. Multicenter trial of high-speed versus conventional single-photon emission computed tomography imaging: Quantitative results of myocardial perfusion and left ventricular function. J Am Coll Cardiol 2010;55:1965-74.PubMedCrossRef Sharir T, Slomka PJ, Hayes SW, DiCarli MF, Ziffer JA, Martin WH, et al. Multicenter trial of high-speed versus conventional single-photon emission computed tomography imaging: Quantitative results of myocardial perfusion and left ventricular function. J Am Coll Cardiol 2010;55:1965-74.PubMedCrossRef
5.
go back to reference Esteves FP, Raggi P, Folks RD, Keidar Z, Askew JW, Rispler S, et al. Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: Multicenter comparison with standard dual detector cameras. J Nucl Cardiol 2009;16:927-34.PubMedCrossRef Esteves FP, Raggi P, Folks RD, Keidar Z, Askew JW, Rispler S, et al. Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: Multicenter comparison with standard dual detector cameras. J Nucl Cardiol 2009;16:927-34.PubMedCrossRef
6.
go back to reference Buechel RR, Herzog BA, Husmann L, Burger IA, Pazhenkottil AP, Treyer V, et al. Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: First clinical validation. Eur J Nucl Med Mol Imaging 2010;37:773-8.PubMedCrossRef Buechel RR, Herzog BA, Husmann L, Burger IA, Pazhenkottil AP, Treyer V, et al. Ultrafast nuclear myocardial perfusion imaging on a new gamma camera with semiconductor detector technique: First clinical validation. Eur J Nucl Med Mol Imaging 2010;37:773-8.PubMedCrossRef
7.
go back to reference Herzog BA, Buechel RR, Katz R, Brueckner M, Husmann L, Burger IA, et al. Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: Optimized protocol for scan time reduction. J Nucl Med 2010;51:46-51.PubMedCrossRef Herzog BA, Buechel RR, Katz R, Brueckner M, Husmann L, Burger IA, et al. Nuclear myocardial perfusion imaging with a cadmium-zinc-telluride detector technique: Optimized protocol for scan time reduction. J Nucl Med 2010;51:46-51.PubMedCrossRef
8.
go back to reference Duvall WL, Croft LB, Ginsberg ES, Einstein AJ, Guma KA, George T et al. Reduced isotope dose and imaging time with a high-efficiency CZT SPECT camera. J Nucl Cardiol 2011. doi:10.1007/s12350-011-9379-7. Duvall WL, Croft LB, Ginsberg ES, Einstein AJ, Guma KA, George T et al. Reduced isotope dose and imaging time with a high-efficiency CZT SPECT camera. J Nucl Cardiol 2011. doi:10.​1007/​s12350-011-9379-7.
9.
go back to reference Nakazato R, Tamarappoo BK, Kang X, Wolak A, Kite F, Hayes SW, et al. Quantitative upright-supine high-speed SPECT myocardial perfusion imaging for detection of coronary artery disease: Correlation with invasive coronary angiography. J Nucl Med 2010;51:1724-31.PubMedCrossRef Nakazato R, Tamarappoo BK, Kang X, Wolak A, Kite F, Hayes SW, et al. Quantitative upright-supine high-speed SPECT myocardial perfusion imaging for detection of coronary artery disease: Correlation with invasive coronary angiography. J Nucl Med 2010;51:1724-31.PubMedCrossRef
10.
go back to reference Hansen CL, Goldstein RA, Berman DS, Churchwell KB, Cooke CD, Corbett JR, et al. Myocardial perfusion and function single photon emission computed tomography. J Nucl Cardiol 2006;13:e97-120.PubMedCrossRef Hansen CL, Goldstein RA, Berman DS, Churchwell KB, Cooke CD, Corbett JR, et al. Myocardial perfusion and function single photon emission computed tomography. J Nucl Cardiol 2006;13:e97-120.PubMedCrossRef
11.
go back to reference Henzlova MJ, Cerqueira MD, Hansen CL, Taillefer R, Yao SS. ASNC announcement. J Nucl Cardiol. 2009;16:331.CrossRef Henzlova MJ, Cerqueira MD, Hansen CL, Taillefer R, Yao SS. ASNC announcement. J Nucl Cardiol. 2009;16:331.CrossRef
12.
go back to reference Henzlova MJ, Cerqueira MD, Mahmarian JJ, Yao SS. Stress protocols and tracers. J Nucl Cardiol 2006;13:e80-90.PubMedCrossRef Henzlova MJ, Cerqueira MD, Mahmarian JJ, Yao SS. Stress protocols and tracers. J Nucl Cardiol 2006;13:e80-90.PubMedCrossRef
13.
go back to reference Hendel RC, Berman DS, DiCarli MF, Heidenreich PA, Henkin RE, Pellikka PA, et al. ACCF/ASNC/ACR/AHA/ASE/SCCT/SCMR/SNM 2009 appropriate use criteria for cardiac radionuclide imaging: A report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the American Society of Nuclear Cardiology, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the Society of Cardiovascular Computed Tomography, the Society for Cardiovascular Magnetic Resonance, and the Society of Nuclear Medicine. Circulation 2009;119:e561-87.PubMedCrossRef Hendel RC, Berman DS, DiCarli MF, Heidenreich PA, Henkin RE, Pellikka PA, et al. ACCF/ASNC/ACR/AHA/ASE/SCCT/SCMR/SNM 2009 appropriate use criteria for cardiac radionuclide imaging: A report of the American College of Cardiology Foundation Appropriate Use Criteria Task Force, the American Society of Nuclear Cardiology, the American College of Radiology, the American Heart Association, the American Society of Echocardiography, the Society of Cardiovascular Computed Tomography, the Society for Cardiovascular Magnetic Resonance, and the Society of Nuclear Medicine. Circulation 2009;119:e561-87.PubMedCrossRef
14.
go back to reference Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. J Nucl Cardiol 2002;9:240-5.PubMedCrossRef Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart: A statement for healthcare professionals from the Cardiac Imaging Committee of the Council on Clinical Cardiology of the American Heart Association. J Nucl Cardiol 2002;9:240-5.PubMedCrossRef
15.
go back to reference Duvall WL, Croft LB, Godiwala T, Ginsberg E, George T, Henzlova MJ. Reduced isotope dose with rapid SPECT MPI imaging: Initial experience with a CZT SPECT camera. J Nucl Cardiol 2010;17:1009-14.PubMedCrossRef Duvall WL, Croft LB, Godiwala T, Ginsberg E, George T, Henzlova MJ. Reduced isotope dose with rapid SPECT MPI imaging: Initial experience with a CZT SPECT camera. J Nucl Cardiol 2010;17:1009-14.PubMedCrossRef
16.
go back to reference Udelson JE, Dilsizian V, Bonow RO. Chapter 16—nuclear cardiology. In: Libby P, Bonow RO, Mann DL, Zipes DP, editors. Braunwald’s heart disease: A textbook of cardiovascular medicine. Philadelphia, PA: WB Saunders; 2008. Udelson JE, Dilsizian V, Bonow RO. Chapter 16—nuclear cardiology. In: Libby P, Bonow RO, Mann DL, Zipes DP, editors. Braunwald’s heart disease: A textbook of cardiovascular medicine. Philadelphia, PA: WB Saunders; 2008.
17.
go back to reference Klocke FJ, Baird MG, Lorell BH, Bateman TM, Messer JV, Berman DS, et al. ACC/AHA/ASNC guidelines for the clinical use of cardiac radionuclide imaging–executive summary: A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASNC Committee to Revise the 1995 Guidelines for the Clinical Use of Cardiac Radionuclide Imaging). Circulation 2003;108:1404-18.PubMedCrossRef Klocke FJ, Baird MG, Lorell BH, Bateman TM, Messer JV, Berman DS, et al. ACC/AHA/ASNC guidelines for the clinical use of cardiac radionuclide imaging–executive summary: A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/ASNC Committee to Revise the 1995 Guidelines for the Clinical Use of Cardiac Radionuclide Imaging). Circulation 2003;108:1404-18.PubMedCrossRef
18.
go back to reference Begg CB, Greenes RA. Assessment of diagnostic tests when disease verification is subject to selection bias. Biometrics 1983;39:207-15.PubMedCrossRef Begg CB, Greenes RA. Assessment of diagnostic tests when disease verification is subject to selection bias. Biometrics 1983;39:207-15.PubMedCrossRef
19.
go back to reference Diamond GA. Reverend Bayes’ silent majority. An alternative factor affecting sensitivity and specificity of exercise electrocardiography. Am J Cardiol 1986;57:1175-80.PubMedCrossRef Diamond GA. Reverend Bayes’ silent majority. An alternative factor affecting sensitivity and specificity of exercise electrocardiography. Am J Cardiol 1986;57:1175-80.PubMedCrossRef
20.
go back to reference Diamond GA. Affirmative actions: Can the discriminant accuracy of a test be determined in the face of selection bias? Med Decis Making 1991;11:48-56.PubMedCrossRef Diamond GA. Affirmative actions: Can the discriminant accuracy of a test be determined in the face of selection bias? Med Decis Making 1991;11:48-56.PubMedCrossRef
21.
go back to reference Herzog BA, Husmann L, Buechel RR, Pazhenkottil AP, Burger IA, Valenta I et al. Rapid cardiac hybrid imaging with minimized radiation dose for accurate non-invasive assessment of ischemic coronary artery disease. Int J Cardiol 2010. doi:10.1016/j.ijcard.2010.08.023. Herzog BA, Husmann L, Buechel RR, Pazhenkottil AP, Burger IA, Valenta I et al. Rapid cardiac hybrid imaging with minimized radiation dose for accurate non-invasive assessment of ischemic coronary artery disease. Int J Cardiol 2010. doi:10.​1016/​j.​ijcard.​2010.​08.​023.
Metadata
Title
Comparison of high efficiency CZT SPECT MPI to coronary angiography
Authors
W. Lane Duvall, MD
Joseph M. Sweeny, MD
Lori B. Croft, MD
Maya H. Barghash, MD
Nitin K. Kulkarni, MD
Krista A. Guma, CNMT
Milena J. Henzlova, MD
Publication date
01-08-2011
Publisher
Springer-Verlag
Published in
Journal of Nuclear Cardiology / Issue 4/2011
Print ISSN: 1071-3581
Electronic ISSN: 1532-6551
DOI
https://doi.org/10.1007/s12350-011-9382-z

Other articles of this Issue 4/2011

Journal of Nuclear Cardiology 4/2011 Go to the issue