Head motion evaluation and correction for PET scans with 18F-FDG in the Japanese Alzheimer’s disease neuroimaging initiative (J-ADNI) multi-center study

Authors: Yasuhiko Ikari, Tomoyuki Nishio, Yoko Makishi, Yukari Miya, Kengo Ito, Robert A. Koeppe, Michio Senda

Published in: Annals of Nuclear Medicine | Issue 7/2012

Login to get access

Abstract

Objective

Head motion during 30-min (six 5-min frames) brain PET scans starting 30 min post-injection of FDG was evaluated together with the effect of post hoc motion correction between frames in J-ADNI multicenter study carried out in 24 PET centers on a total of 172 subjects consisting of 81 normal subjects, 55 mild cognitive impairment (MCI) and 36 mild Alzheimer’s disease (AD) patients.

Methods

Based on the magnitude of the between-frame co-registration parameters, the scans were classified into six levels (A–F) of motion degree. The effect of motion and its correction was evaluated using between-frame variation of the regional FDG uptake values on ROIs placed over cerebral cortical areas.

Result

Although AD patients tended to present larger motion (motion level E or F in 22 % of the subjects) than MCI (3 %) and normal (4 %) subjects, unignorable motion was observed in a small number of subjects in the latter groups as well. The between-frame coefficient of variation (SD/mean) was 0.5 % in the frontal, 0.6 % in the parietal and 1.8 % in the posterior cingulate ROI for the scans of motion level 1. The respective values were 1.5, 1.4, and 3.6 % for the scans of motion level F, but reduced by the motion correction to 0.5, 0.4 and 0.8 %, respectively. The motion correction changed the ROI value for the posterior cingulate cortex by 11.6 % in the case of severest motion.

Conclusion

Substantial head motion occurs in a fraction of subjects in a multicenter setup which includes PET centers lacking sufficient experience in imaging demented patients. A simple frame-by-frame co-registration technique that can be applied to any PET camera model is effective in correcting for motion and improving quantitative capability.
Literature
1.
go back to reference Thomas B, Lutz T, Ingo N, Uwe P. On the use of positioning aids to reduce misregistration in the head and neck in whole-body PET/CT studies. J Nucl Med. 2005;46(4):596–602. Thomas B, Lutz T, Ingo N, Uwe P. On the use of positioning aids to reduce misregistration in the head and neck in whole-body PET/CT studies. J Nucl Med. 2005;46(4):596–602.
2.
go back to reference Andrew JM, Kris T, Mitul AM, Federico T, Sanida M, Paul MG. Correction of head movement on PET studies: comparison of methods. J Nucl Med. 2006;47(12):1936–44. Andrew JM, Kris T, Mitul AM, Federico T, Sanida M, Paul MG. Correction of head movement on PET studies: comparison of methods. J Nucl Med. 2006;47(12):1936–44.
3.
go back to reference Jurgen EMM, Mark L, Floris HPV, Adriaan AL, Ronald B. Off-line motion correction methods for multi-frame PET data. Eur J Nucl Med Mol Imaging. 2009;36:2002–13.CrossRef Jurgen EMM, Mark L, Floris HPV, Adriaan AL, Ronald B. Off-line motion correction methods for multi-frame PET data. Eur J Nucl Med Mol Imaging. 2009;36:2002–13.CrossRef
4.
go back to reference Minoshima S, Koeppe RA, Fessler JA, Mintun MA, Berger KL, Taylor SF, et al. Integrated and automated data analysis method for neuronal activation studies using [O-15] water PET. In: Proceedings of PET 93 Akita: quantification of brain function, tracer kinetics and image analysis in brain PET. Elsevier: Amsterdam; 1993. p. 409–15. Minoshima S, Koeppe RA, Fessler JA, Mintun MA, Berger KL, Taylor SF, et al. Integrated and automated data analysis method for neuronal activation studies using [O-15] water PET. In: Proceedings of PET 93 Akita: quantification of brain function, tracer kinetics and image analysis in brain PET. Elsevier: Amsterdam; 1993. p. 409–15.
5.
go back to reference Maes F, Collignon A, Vandermeulen D, Marchal G, Sutens P. Multimodality image registration by maximization of mutual information. IEEE Trans Med Imaging. 1997;16(2):187–98.PubMedCrossRef Maes F, Collignon A, Vandermeulen D, Marchal G, Sutens P. Multimodality image registration by maximization of mutual information. IEEE Trans Med Imaging. 1997;16(2):187–98.PubMedCrossRef
6.
go back to reference Joseph AM, Paul JL, Kraft RobertA, Jonathan HB. An automated method for neuroanatomic and cytoarchitectonic atlas-based interrogation of fMRI data sets. Neuroimage. 2003;19(3):1233–9.CrossRef Joseph AM, Paul JL, Kraft RobertA, Jonathan HB. An automated method for neuroanatomic and cytoarchitectonic atlas-based interrogation of fMRI data sets. Neuroimage. 2003;19(3):1233–9.CrossRef
7.
go back to reference Joseph AM, Paul JL, Kraft RobertA, Jonathan HB. Precentral gyrus discrepancy in electronic versions of the Talairach atlas. Neuroimage. 2004;21(1):450–5.CrossRef Joseph AM, Paul JL, Kraft RobertA, Jonathan HB. Precentral gyrus discrepancy in electronic versions of the Talairach atlas. Neuroimage. 2004;21(1):450–5.CrossRef
8.
go back to reference Tzourio-Mazoyer N, Landeau B, Papathanassiou D, Crivello F, Etard O, Delcroix N, et al. Automated anatomical labelling of activations in SPM using a macroscopic anatomical parcellation of the MNI MRI single subject brain. NeuroImage. 2002;15:273–89.PubMedCrossRef Tzourio-Mazoyer N, Landeau B, Papathanassiou D, Crivello F, Etard O, Delcroix N, et al. Automated anatomical labelling of activations in SPM using a macroscopic anatomical parcellation of the MNI MRI single subject brain. NeuroImage. 2002;15:273–89.PubMedCrossRef
9.
go back to reference Minoshima S, Frey KA, Koeppe RA, Foster NL, Kuhl DE. A diagnostic approach in Alzheimer’s disease using three-dimensional stereotactic surface projections of [18F]FDG. J Nucl Med. 1995;36:1238–48.PubMed Minoshima S, Frey KA, Koeppe RA, Foster NL, Kuhl DE. A diagnostic approach in Alzheimer’s disease using three-dimensional stereotactic surface projections of [18F]FDG. J Nucl Med. 1995;36:1238–48.PubMed
10.
go back to reference R Development Core Team. R: a language and environment for statistical computing. Vienna: R Foundation for Statistical Computing; 2006. R Development Core Team. R: a language and environment for statistical computing. Vienna: R Foundation for Statistical Computing; 2006.
11.
go back to reference Holm S. A simple sequentially rejective multiple test procedure. Scand J Stat. 1979;6:65–70. Holm S. A simple sequentially rejective multiple test procedure. Scand J Stat. 1979;6:65–70.
12.
go back to reference Kendall MG. A new measure of rank correlation. Biometrika. 1938;30(1/2):81–93.CrossRef Kendall MG. A new measure of rank correlation. Biometrika. 1938;30(1/2):81–93.CrossRef
Metadata
Title
Head motion evaluation and correction for PET scans with 18F-FDG in the Japanese Alzheimer’s disease neuroimaging initiative (J-ADNI) multi-center study
Authors
Yasuhiko Ikari
Tomoyuki Nishio
Yoko Makishi
Yukari Miya
Kengo Ito
Robert A. Koeppe
Michio Senda
Publication date
01-08-2012
Publisher
Springer Japan
Published in
Annals of Nuclear Medicine / Issue 7/2012
Print ISSN: 0914-7187
Electronic ISSN: 1864-6433
DOI
https://doi.org/10.1007/s12149-012-0605-4

Other articles of this Issue 7/2012

Annals of Nuclear Medicine 7/2012 Go to the issue