Published in: European Journal of Nuclear Medicine and Molecular Imaging 12/2009

Open Access 01-12-2009 | Original Article

Off-line motion correction methods for multi-frame PET data

Authors: Jurgen E. M. Mourik, Mark Lubberink, Floris H. P. van Velden, Adriaan A. Lammertsma, Ronald Boellaard

Published in: European Journal of Nuclear Medicine and Molecular Imaging | Issue 12/2009

Login to get access

Abstract

Purpose

Patient motion during PET acquisition may affect measured time-activity curves, thereby reducing accuracy of tracer kinetic analyses. The aim of the present study was to evaluate different off-line frame-by-frame methods to correct patient motion, which is of particular interest when no optical motion tracking system is available or when older data sets have to be reanalysed.

Methods

Four different motion correction methods were evaluated. In the first method attenuation-corrected frames were realigned with the summed image of the first 3 min. The second method was identical, except that non-attenuation-corrected images were used. In the third and fourth methods non-attenuation-corrected images were realigned with standard and cupped transmission images, respectively. Two simulation studies were performed, based on [11C]flumazenil and (R)-[11C]PK11195 data sets, respectively. For both simulation studies different types (rotational, translational) and degrees of motion were added. Simulated PET scans were corrected for motion using all correction methods. The optimal method derived from these simulation studies was used to evaluate two (one with and one without visible movement) clinical data sets of [11C]flumazenil, (R)-[11C]PK11195 and [11C]PIB. For these clinical data sets, the volume of distribution (VT) was derived using Logan analysis and values were compared before and after motion correction.

Results

For both [11C]flumazenil and (R)-[11C]PK11195 simulation studies, optimal results were obtained when realignment was based on non-attenuation-corrected images. For the clinical data sets motion disappeared visually after motion correction. Regional differences of up to 433% in VT before and after motion correction were found for scans with visible movement. On the other hand, when no visual motion was present in the original data set, overall differences in VT before and after motion correction were <1.5 ± 1.3%.

Conclusion

Frame-by-frame motion correction using non-attenuation-corrected images improves the accuracy of tracer kinetic analysis compared to non-motion-corrected data.
Appendix
Available only for authorised users
Literature
2.
go back to reference Green MV, Seidel J, Stein SD, Tedder TE, Kempner KM, Kertzman C, et al. Head movement in normal subjects during simulated PET brain imaging with and without head restraint. J Nucl Med 1994;35:1538–46.PubMed Green MV, Seidel J, Stein SD, Tedder TE, Kempner KM, Kertzman C, et al. Head movement in normal subjects during simulated PET brain imaging with and without head restraint. J Nucl Med 1994;35:1538–46.PubMed
3.
go back to reference Beyer T, Tellmann L, Nickel I, Pietrzyk U. On the use of positioning aids to reduce misregistration in the head and neck in whole-body PET/CT studies. J Nucl Med 2005;46:596–602.PubMed Beyer T, Tellmann L, Nickel I, Pietrzyk U. On the use of positioning aids to reduce misregistration in the head and neck in whole-body PET/CT studies. J Nucl Med 2005;46:596–602.PubMed
4.
go back to reference Rahmim A. Advanced motion correction methods in PET. Iran J Nucl Med 2005;13:1–17 Rahmim A. Advanced motion correction methods in PET. Iran J Nucl Med 2005;13:1–17
6.
8.
go back to reference Andersson JL. A rapid and accurate method to realign PET scans utilizing image edge information. J Nucl Med 1995;36:657–69. Andersson JL. A rapid and accurate method to realign PET scans utilizing image edge information. J Nucl Med 1995;36:657–69.
9.
go back to reference Perruchot F, Reilhac A, Grova C, Evans AC, Dagher A. Motion correction of multi-frame PET data. IEEE Nucl Sci Symp Conf Rec 2004;5:3186–90.CrossRef Perruchot F, Reilhac A, Grova C, Evans AC, Dagher A. Motion correction of multi-frame PET data. IEEE Nucl Sci Symp Conf Rec 2004;5:3186–90.CrossRef
11.
go back to reference Montgomery AJ, Thielemans K, Mehta MA, Turkheimer F, Mustafovic S, Grasby PM. Correction of head movement on PET studies: comparison of methods. J Nucl Med 2006;47:1936–44.PubMed Montgomery AJ, Thielemans K, Mehta MA, Turkheimer F, Mustafovic S, Grasby PM. Correction of head movement on PET studies: comparison of methods. J Nucl Med 2006;47:1936–44.PubMed
16.
go back to reference Zamburlini M, Camborde M, de la Fuente-Fernandez R, Stoessl AJ, Ruth TJ, Sossi V. Impact of the spatial normalization template and realignment procedure on the SPM analysis of [11C]raclopride PET studies. IEEE Trans Nucl Sci 2004;51:205–11. doi:10.1109/TNS.2003.823030.CrossRef Zamburlini M, Camborde M, de la Fuente-Fernandez R, Stoessl AJ, Ruth TJ, Sossi V. Impact of the spatial normalization template and realignment procedure on the SPM analysis of [11C]raclopride PET studies. IEEE Trans Nucl Sci 2004;51:205–11. doi:10.​1109/​TNS.​2003.​823030.CrossRef
19.
go back to reference Boellaard R, van Lingen A, van Balen SC, Hoving BG, Lammertsma AA. Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET. Eur J Nucl Med 2001;28:81–9. doi:10.1007/s002590000405.CrossRefPubMed Boellaard R, van Lingen A, van Balen SC, Hoving BG, Lammertsma AA. Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET. Eur J Nucl Med 2001;28:81–9. doi:10.​1007/​s002590000405.CrossRefPubMed
Metadata
Title
Off-line motion correction methods for multi-frame PET data
Authors
Jurgen E. M. Mourik
Mark Lubberink
Floris H. P. van Velden
Adriaan A. Lammertsma
Ronald Boellaard
Publication date
01-12-2009
Publisher
Springer-Verlag
Published in
European Journal of Nuclear Medicine and Molecular Imaging / Issue 12/2009
Print ISSN: 1619-7070
Electronic ISSN: 1619-7089
DOI
https://doi.org/10.1007/s00259-009-1193-y

Other articles of this Issue 12/2009

European Journal of Nuclear Medicine and Molecular Imaging 12/2009 Go to the issue