Skip to main content
Top
Published in: Clinical Oral Investigations 8/2015

01-11-2015 | Original Article

Accuracy of single-tooth restorations based on intraoral digital and conventional impressions in patients

Authors: Moritz Boeddinghaus, Eva Sabina Breloer, Peter Rehmann, Bernd Wöstmann

Published in: Clinical Oral Investigations | Issue 8/2015

Login to get access

Abstract

Objectives

The purpose of this clinical study was to compare the marginal fit of dental crowns based on three different intraoral digital and one conventional impression methods.

Methods

Forty-nine teeth of altogether 24 patients were prepared to be treated with full-coverage restorations. Digital impressions were made using three intraoral scanners: Sirona CEREC AC Omnicam (OCam), Heraeus Cara TRIOS and 3M Lava True Definition (TDef). Furthermore, a gypsum model based on a conventional impression (EXA’lence, GC, Tokyo, Japan) was scanned with a standard laboratory scanner (3Shape D700). Based on the dataset obtained, four zirconia copings per tooth were produced. The marginal fit of the copings in the patient’s mouth was assessed employing a replica technique.

Results

Overall, seven measurement copings did not fit and, therefore, could not be assessed. The marginal gap was 88 μm (68–136 μm) [median/interquartile range] for the TDef, 112 μm (94–149 μm) for the Cara TRIOS, 113 μm (81–157 μm) for the laboratory scanner and 149 μm (114–218 μm) for the OCam. There was a statistically significant difference between the OCam and the other groups (p < 0.05).

Conclusion

Within the limitations of this study, it can be concluded that zirconia copings based on intraoral scans and a laboratory scans of a conventional model are comparable to one another with regard to their marginal fit.

Clinical relevance

Regarding the results of this study, the digital intraoral impression can be considered as an alternative to a conventional impression with a consecutive digital workflow when the finish line is clearly visible and it is possible to keep it dry.
Literature
2.
go back to reference Tidehag P, Ottosson K, Sjogren G (2013) Accuracy of ceramic restorations made using an in-office optical scanning technique: an in vitro study. Oper Dent 39(3):308–316. doi:10.2341/12-309-L CrossRefPubMed Tidehag P, Ottosson K, Sjogren G (2013) Accuracy of ceramic restorations made using an in-office optical scanning technique: an in vitro study. Oper Dent 39(3):308–316. doi:10.​2341/​12-309-L CrossRefPubMed
3.
go back to reference Syrek A, Reich G, Ranftl D, Klein C, Cerny B, Brodesser J (2010) Clinical evaluation of all-ceramic crowns fabricated from intraoral digital impressions based on the principle of active wavefront sampling. J Dent 38(7):553–559CrossRefPubMed Syrek A, Reich G, Ranftl D, Klein C, Cerny B, Brodesser J (2010) Clinical evaluation of all-ceramic crowns fabricated from intraoral digital impressions based on the principle of active wavefront sampling. J Dent 38(7):553–559CrossRefPubMed
4.
go back to reference Wismeijer D, Mans R, van Genuchten M, Reijers HA (2013) Patients' preferences when comparing analogue implant impressions using a polyether impression material versus digital impressions (intraoral scan) of dental implants. Clin Oral Implants Res. doi:10.1111/clr.12234 Wismeijer D, Mans R, van Genuchten M, Reijers HA (2013) Patients' preferences when comparing analogue implant impressions using a polyether impression material versus digital impressions (intraoral scan) of dental implants. Clin Oral Implants Res. doi:10.​1111/​clr.​12234
5.
go back to reference Yuzbasioglu E, Kurt H, Turunc R, Bilir H (2014) Comparison of digital and conventional impression techniques: evaluation of patients' perception, treatment comfort, effectiveness and clinical outcomes. BMC Oral Health 14:10PubMedCentralCrossRefPubMed Yuzbasioglu E, Kurt H, Turunc R, Bilir H (2014) Comparison of digital and conventional impression techniques: evaluation of patients' perception, treatment comfort, effectiveness and clinical outcomes. BMC Oral Health 14:10PubMedCentralCrossRefPubMed
6.
go back to reference Flügge TV, Schlager S, Nelson K, Nahles S, Metzger MC (2013) Precision of intraoral digital dental impressions with iTero and extraoral digitization with the iTero and a model scanner. Am J Orthod Dentofacial Orthop 144(3):471–478CrossRefPubMed Flügge TV, Schlager S, Nelson K, Nahles S, Metzger MC (2013) Precision of intraoral digital dental impressions with iTero and extraoral digitization with the iTero and a model scanner. Am J Orthod Dentofacial Orthop 144(3):471–478CrossRefPubMed
7.
go back to reference Schaefer O, Decker M, Wittstock F, Kuepper H, Guentsch A (2014) Impact of digital impression techniques on the adaption of ceramic partial crowns in vitro. J Dent Schaefer O, Decker M, Wittstock F, Kuepper H, Guentsch A (2014) Impact of digital impression techniques on the adaption of ceramic partial crowns in vitro. J Dent
8.
go back to reference Rosenstiel SF, Land MF, Fujimoto J (1995) Principles of tooth preparation. In: Duncan L (ed) Contemporary fixed prosthodontics, vol 2. Mosby-Year Book, Inc, St. Louis, pp pp 135–pp 167 Rosenstiel SF, Land MF, Fujimoto J (1995) Principles of tooth preparation. In: Duncan L (ed) Contemporary fixed prosthodontics, vol 2. Mosby-Year Book, Inc, St. Louis, pp pp 135–pp 167
11.
go back to reference Böning KW, Wolf BH, Schmidt AE, Kastner K, Walter MH (2000) Clinical fit of procera all ceram crowns. J Prosthet Dent 84(4):419–424CrossRef Böning KW, Wolf BH, Schmidt AE, Kastner K, Walter MH (2000) Clinical fit of procera all ceram crowns. J Prosthet Dent 84(4):419–424CrossRef
12.
go back to reference Laurent M, Scheer P, Dejou J, Laborde G (2008) Clinical evaluation of the marginal fit of cast crowns—validation of the silicone replica method. J Oral Rehabil 35(2):116–122CrossRefPubMed Laurent M, Scheer P, Dejou J, Laborde G (2008) Clinical evaluation of the marginal fit of cast crowns—validation of the silicone replica method. J Oral Rehabil 35(2):116–122CrossRefPubMed
14.
go back to reference Fransson B, Oilo G, Gjeitanger R (1985) The fit of metal-ceramic crowns, a clinical study. Dent Mater 1(5):197–199CrossRefPubMed Fransson B, Oilo G, Gjeitanger R (1985) The fit of metal-ceramic crowns, a clinical study. Dent Mater 1(5):197–199CrossRefPubMed
15.
go back to reference Renne W, McGill ST, Forshee KV, DeFee MR, Mennito AS (2012) Predicting marginal fit of CAD/CAM crowns based on the presence or absence of common preparation errors. J Prosth Dent 108(5):310–315CrossRef Renne W, McGill ST, Forshee KV, DeFee MR, Mennito AS (2012) Predicting marginal fit of CAD/CAM crowns based on the presence or absence of common preparation errors. J Prosth Dent 108(5):310–315CrossRef
16.
go back to reference Karlsson S (1993) The fit of procera titanium crowns. An in vitro and clinical study. Acta Odontol Scand 51(3):129–134CrossRefPubMed Karlsson S (1993) The fit of procera titanium crowns. An in vitro and clinical study. Acta Odontol Scand 51(3):129–134CrossRefPubMed
17.
go back to reference Holmes JR, Bayne SC, Holland GA, Sulik WD (1989) Considerations in measurement of marginal fit. J Prosthet Dent 62(4):405–408CrossRefPubMed Holmes JR, Bayne SC, Holland GA, Sulik WD (1989) Considerations in measurement of marginal fit. J Prosthet Dent 62(4):405–408CrossRefPubMed
18.
go back to reference Brawek PK, Wolfart S, Endres L, Kirsten A, Reich S (2013) The clinical accuracy of single crowns exclusively fabricated by digital workflow—the comparison of two systems. Clinical oral investigations 17(9):2119–2125. doi:10.1007/s00784-013-0923-5 CrossRefPubMed Brawek PK, Wolfart S, Endres L, Kirsten A, Reich S (2013) The clinical accuracy of single crowns exclusively fabricated by digital workflow—the comparison of two systems. Clinical oral investigations 17(9):2119–2125. doi:10.​1007/​s00784-013-0923-5 CrossRefPubMed
19.
go back to reference Scotti R, Cardelli P, Baldissara P, Monaco C (2011) Clinical fitting of CAD/CAM zirconia single crowns generated from digital intraoral impressions based on active wavefront sampling. J Dent Scotti R, Cardelli P, Baldissara P, Monaco C (2011) Clinical fitting of CAD/CAM zirconia single crowns generated from digital intraoral impressions based on active wavefront sampling. J Dent
20.
go back to reference Bader JD, Rozier RG, McFall WT Jr, Ramsey DL (1991) Effect of crown margins on periodontal conditions in regularly attending patients. J Prosthet Dent 65(1):75–79CrossRefPubMed Bader JD, Rozier RG, McFall WT Jr, Ramsey DL (1991) Effect of crown margins on periodontal conditions in regularly attending patients. J Prosthet Dent 65(1):75–79CrossRefPubMed
21.
go back to reference Valderhaug J, Birkeland JM (1976) Periodontal conditions in patients 5 years following insertion of fixed prostheses. Pocket depth and loss of attachment. J Oral Rehabil 3(3):237–243CrossRefPubMed Valderhaug J, Birkeland JM (1976) Periodontal conditions in patients 5 years following insertion of fixed prostheses. Pocket depth and loss of attachment. J Oral Rehabil 3(3):237–243CrossRefPubMed
22.
go back to reference Silness J (1970) Periodontal conditions in patients treated with dental bridges. 3. The relationship between the location of the crown margin and the periodontal condition. J Periodont Res 5(3):225–229CrossRefPubMed Silness J (1970) Periodontal conditions in patients treated with dental bridges. 3. The relationship between the location of the crown margin and the periodontal condition. J Periodont Res 5(3):225–229CrossRefPubMed
23.
go back to reference Patzelt SB, Emmanouilidi A, Stampf S, Strub JR, Att W Accuracy of full-arch scans using intraoral scanners. Clin Oral Investig 18 (6):1687-1694 Patzelt SB, Emmanouilidi A, Stampf S, Strub JR, Att W Accuracy of full-arch scans using intraoral scanners. Clin Oral Investig 18 (6):1687-1694
24.
go back to reference Kim SY, Kim MJ, Han JS, Yeo IS, Lim YJ, Kwon HB (2013) Accuracy of dies captured by an intraoral digital impression system using parallel confocal imaging. Int J Prosthodont 26(2):161–163CrossRefPubMed Kim SY, Kim MJ, Han JS, Yeo IS, Lim YJ, Kwon HB (2013) Accuracy of dies captured by an intraoral digital impression system using parallel confocal imaging. Int J Prosthodont 26(2):161–163CrossRefPubMed
25.
26.
go back to reference van der Meer WJ, Andriessen FS, Wismeijer D, Ren Y (2012) Application of intra-oral dental scanners in the digital workflow of implantology. PLoS One 7(8):0043312CrossRef van der Meer WJ, Andriessen FS, Wismeijer D, Ren Y (2012) Application of intra-oral dental scanners in the digital workflow of implantology. PLoS One 7(8):0043312CrossRef
27.
go back to reference Seelbach P, Brueckel C, Wostmann B (2013) Accuracy of digital and conventional impression techniques and workflow. Clin Oral Investig 17(7):1759–1764CrossRefPubMed Seelbach P, Brueckel C, Wostmann B (2013) Accuracy of digital and conventional impression techniques and workflow. Clin Oral Investig 17(7):1759–1764CrossRefPubMed
28.
go back to reference Keul C, Stawarczyk B, Erdelt KJ, Beuer F, Edelhoff D, Guth JF (2014) Fit of 4-unit FDPs made of zirconia and CoCr-alloy after chairside and labside digitalization—a laboratory study. Dent Mater 30(4):400–407CrossRefPubMed Keul C, Stawarczyk B, Erdelt KJ, Beuer F, Edelhoff D, Guth JF (2014) Fit of 4-unit FDPs made of zirconia and CoCr-alloy after chairside and labside digitalization—a laboratory study. Dent Mater 30(4):400–407CrossRefPubMed
29.
go back to reference Euán R, Figueras-Álvarez O, Cabratosa-Termes J, Oliver-Parra R (2014) Marginal adaptation of zirconium dioxide copings: influence of the CAD/CAM system and the finish line design. J Prosthet Dent 112(2):155–62. doi:10.1016/j.prosdent.2013.10.012 Euán R, Figueras-Álvarez O, Cabratosa-Termes J, Oliver-Parra R (2014) Marginal adaptation of zirconium dioxide copings: influence of the CAD/CAM system and the finish line design. J Prosthet Dent 112(2):155–62. doi:10.​1016/​j.​prosdent.​2013.​10.​012
30.
go back to reference Almeida ESJS, Erdelt K, Edelhoff D, Araujo E, Stimmelmayr M, Vieira LC, Guth JF (2013) Marginal and internal fit of four-unit zirconia fixed dental prostheses based on digital and conventional impression techniques. Clin Oral Investig 18(2):512–523 Almeida ESJS, Erdelt K, Edelhoff D, Araujo E, Stimmelmayr M, Vieira LC, Guth JF (2013) Marginal and internal fit of four-unit zirconia fixed dental prostheses based on digital and conventional impression techniques. Clin Oral Investig 18(2):512–523
31.
go back to reference Mehl A, Ender A, Mormann W, Attin T (2009) Accuracy testing of a new intraoral 3D camera. Int J Comput Dent 12(1):11–28PubMed Mehl A, Ender A, Mormann W, Attin T (2009) Accuracy testing of a new intraoral 3D camera. Int J Comput Dent 12(1):11–28PubMed
32.
go back to reference Luthardt RG, Loos R, Quaas S (2005) Accuracy of intraoral data acquisition in comparison to the conventional impression. Int J Comput Dent 8(4):283–294PubMed Luthardt RG, Loos R, Quaas S (2005) Accuracy of intraoral data acquisition in comparison to the conventional impression. Int J Comput Dent 8(4):283–294PubMed
33.
go back to reference Ender A, Mehl A (2013) Accuracy of complete-arch dental impressions: a new method of measuring trueness and precision. J Prosthet Dent 109(2):121–128CrossRefPubMed Ender A, Mehl A (2013) Accuracy of complete-arch dental impressions: a new method of measuring trueness and precision. J Prosthet Dent 109(2):121–128CrossRefPubMed
34.
go back to reference Ender A, Mehl A (2011) Full arch scans: conventional versus digital impressions—an in-vitro study. Int J Comput Dent 14(1):11–21PubMed Ender A, Mehl A (2011) Full arch scans: conventional versus digital impressions—an in-vitro study. Int J Comput Dent 14(1):11–21PubMed
35.
go back to reference Wöstmann B (1996) Klinische Bestimmungsvariablen bei der Abformung präparierter Zähne. Stomatologie 93(2):51–57 Wöstmann B (1996) Klinische Bestimmungsvariablen bei der Abformung präparierter Zähne. Stomatologie 93(2):51–57
36.
go back to reference Wöstmann B (1994) Zur Frage der in vivo erreichbaren Abformgenauigkeit. Dtsch Zahnarztl Z 49(9):679–682 Wöstmann B (1994) Zur Frage der in vivo erreichbaren Abformgenauigkeit. Dtsch Zahnarztl Z 49(9):679–682
Metadata
Title
Accuracy of single-tooth restorations based on intraoral digital and conventional impressions in patients
Authors
Moritz Boeddinghaus
Eva Sabina Breloer
Peter Rehmann
Bernd Wöstmann
Publication date
01-11-2015
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 8/2015
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-015-1430-7

Other articles of this Issue 8/2015

Clinical Oral Investigations 8/2015 Go to the issue