Published in: European Radiology 7/2019

Open Access 01-07-2019 | Computed Tomography | Computed Tomography

Objective and subjective comparison of virtual monoenergetic vs. polychromatic images in patients with pancreatic ductal adenocarcinoma

Authors: Lucian Beer, Michael Toepker, Ahmed Ba-Ssalamah, Christian Schestak, Anja Dutschke, Martin Schindl, Alexander Wressnegger, Helmut Ringl, Paul Apfaltrer

Published in: European Radiology | Issue 7/2019

Login to get access

Abstract

Objectives

The aim of this study was to assess the objective and subjective image characteristics of monoenergetic images (MEI[+]), using a noise-optimized algorithm at different kiloelectron volts (keV) compared to polyenergetic images (PEI), in patients with pancreatic ductal adenocarcinoma (PDAC).

Methods

This retrospective, institutional review board-approved study included 45 patients (18 male, 27 female; mean age 66 years; range, 42–96 years) with PDAC who had undergone a dual-energy CT (DECT) of the abdomen for staging. One standard polyenergetic image (PEI) and five MEI(+) images in 10-keV intervals, ranging from 40 to 80 keV, were reconstructed. Line-density profile analysis, as well as the contrast-to-noise ratio (CNR) of the tumor, the signal-to-noise ratio (SNR) of the regular pancreas parenchyma and the tumor, and the CNR of the three main peripancreatic vessels, was calculated. For subjective quality assessment, two readers independently assessed the images using a 5-point Likert scale. Reader reliability was evaluated using an intraclass correlation coefficient.

Results

Line-density profile analysis revealed the largest gradient in attenuation between PDAC and regular tissue in MEI(+) at 40 keV. Low-keV MEI(+)reconstructions at 40 and 50 keV increased CNR and SNR compared to PEI (40 keV: CNR 46.8 vs. 7.5; SNRPankreas 32.5 vs. 15.7; SNRLesion 13.5 vs. 8.6; p < 0.001). MEI(+) at 40 keV and 50 keV were consistently preferred by the observers (p < 0.05), showing a high intra-observer 0.937 (0.92–0.95) and inter-observer 0.911 (0.89–0.93) agreement.

Conclusion

MEI(+) reconstructions at 40 keV and 50 keV provide better objective and subjective image quality compared to conventional PEI of DECT in patients with PDAC.

Key Points

• Low-keV MEI(+) reconstructions at 40 and 50 keV increase tumor-to-pancreas contrast compared to PEI.
• Low-keV MEI(+) reconstructions improve objective and subjective image quality parameters compared to PEI.
• Dual-energy post-processing might be a valuable tool in the diagnostic workup of patients with PDAC.
Appendix
Available only for authorised users
Literature
1.
go back to reference Arvold ND, Niemierko A, Mamon HJ, Fernandez-del Castillo C, Hong TS (2011) Pancreatic cancer tumor size on CT scan versus pathologic specimen: implications for radiation treatment planning. Int J Radiat Oncol Biol Phys 80:1383–1390CrossRefPubMed Arvold ND, Niemierko A, Mamon HJ, Fernandez-del Castillo C, Hong TS (2011) Pancreatic cancer tumor size on CT scan versus pathologic specimen: implications for radiation treatment planning. Int J Radiat Oncol Biol Phys 80:1383–1390CrossRefPubMed
2.
go back to reference Somers I, Bipat S (2017) Contrast-enhanced CT in determining resectability in patients with pancreatic carcinoma: a meta-analysis of the positive predictive values of CT. Eur Radiol 27:3408–3435CrossRefPubMedPubMedCentral Somers I, Bipat S (2017) Contrast-enhanced CT in determining resectability in patients with pancreatic carcinoma: a meta-analysis of the positive predictive values of CT. Eur Radiol 27:3408–3435CrossRefPubMedPubMedCentral
3.
go back to reference Koelblinger C, Ba-Ssalamah A, Goetzinger P et al (2011) Gadobenate dimeglumine–enhanced 3.0-T MR imaging versus multiphasic 64–detector row CT: prospective evaluation in patients suspected of having pancreatic cancer. Radiology 259:757–766CrossRefPubMed Koelblinger C, Ba-Ssalamah A, Goetzinger P et al (2011) Gadobenate dimeglumine–enhanced 3.0-T MR imaging versus multiphasic 64–detector row CT: prospective evaluation in patients suspected of having pancreatic cancer. Radiology 259:757–766CrossRefPubMed
4.
go back to reference Jeon SK, Lee JM, Joo I et al (2018) Magnetic resonance with diffusion-weighted imaging improves assessment of focal liver lesions in patients with potentially resectable pancreatic cancer on CT. Eur Radiol 28:3484–3493CrossRefPubMed Jeon SK, Lee JM, Joo I et al (2018) Magnetic resonance with diffusion-weighted imaging improves assessment of focal liver lesions in patients with potentially resectable pancreatic cancer on CT. Eur Radiol 28:3484–3493CrossRefPubMed
5.
go back to reference De Cecco CN, Caruso D, Schoepf UJ et al (2018) A noise-optimized virtual monoenergetic reconstruction algorithm improves the diagnostic accuracy of late hepatic arterial phase dual-energy CT for the detection of hypervascular liver lesions. Eur Radiol 28:3393–3404CrossRefPubMed De Cecco CN, Caruso D, Schoepf UJ et al (2018) A noise-optimized virtual monoenergetic reconstruction algorithm improves the diagnostic accuracy of late hepatic arterial phase dual-energy CT for the detection of hypervascular liver lesions. Eur Radiol 28:3393–3404CrossRefPubMed
6.
go back to reference Euler A, Obmann MM, Szucs-Farkas Z et al (2018) Comparison of image quality and radiation dose between split-filter dual-energy images and single-energy images in single-source abdominal CT. Eur Radiol 28:3405–3412CrossRefPubMed Euler A, Obmann MM, Szucs-Farkas Z et al (2018) Comparison of image quality and radiation dose between split-filter dual-energy images and single-energy images in single-source abdominal CT. Eur Radiol 28:3405–3412CrossRefPubMed
7.
go back to reference Kordbacheh H, Baliyan V, Singh P, Eisner BH, Sahani DV, Kambadakone AR (2018) Rapid kVp switching dual-energy CT in the assessment of urolithiasis in patients with large body habitus: preliminary observations on image quality and stone characterization. Abdom Radiol (NY). https://doi.org/10.1007/s00261-018-1808-5 Kordbacheh H, Baliyan V, Singh P, Eisner BH, Sahani DV, Kambadakone AR (2018) Rapid kVp switching dual-energy CT in the assessment of urolithiasis in patients with large body habitus: preliminary observations on image quality and stone characterization. Abdom Radiol (NY). https://​doi.​org/​10.​1007/​s00261-018-1808-5
10.
go back to reference Frellesen C, Fessler F, Hardie AD et al (2015) Dual-energy CT of the pancreas: improved carcinoma-to-pancreas contrast with a noise-optimized monoenergetic reconstruction algorithm. Eur J Radiol 84:2052–2058CrossRefPubMed Frellesen C, Fessler F, Hardie AD et al (2015) Dual-energy CT of the pancreas: improved carcinoma-to-pancreas contrast with a noise-optimized monoenergetic reconstruction algorithm. Eur J Radiol 84:2052–2058CrossRefPubMed
11.
go back to reference Sudarski S, Apfaltrer P, Nance JW et al (2013) Optimization of keV-settings in abdominal and lower extremity dual-source dual-energy CT angiography determined with virtual monoenergetic imaging. Eur J Radiol 82:e574–e581CrossRefPubMed Sudarski S, Apfaltrer P, Nance JW et al (2013) Optimization of keV-settings in abdominal and lower extremity dual-source dual-energy CT angiography determined with virtual monoenergetic imaging. Eur J Radiol 82:e574–e581CrossRefPubMed
12.
go back to reference Toepker M, Czerny C, Ringl H et al (2014) Can dual-energy CT improve the assessment of tumor margins in oral cancer? Oral Oncol 50:221–227CrossRefPubMed Toepker M, Czerny C, Ringl H et al (2014) Can dual-energy CT improve the assessment of tumor margins in oral cancer? Oral Oncol 50:221–227CrossRefPubMed
13.
go back to reference Grant KL, Flohr TG, Krauss B, Sedlmair M, Thomas C, Schmidt B (2014) Assessment of an advanced image-based technique to calculate virtual monoenergetic computed tomographic images from a dual-energy examination to improve contrast-to-noise ratio in examinations using iodinated contrast media. Invest Radiol 49:586–592CrossRefPubMed Grant KL, Flohr TG, Krauss B, Sedlmair M, Thomas C, Schmidt B (2014) Assessment of an advanced image-based technique to calculate virtual monoenergetic computed tomographic images from a dual-energy examination to improve contrast-to-noise ratio in examinations using iodinated contrast media. Invest Radiol 49:586–592CrossRefPubMed
14.
go back to reference Albrecht MH, Scholtz J-E, Hüsers K et al (2016) Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast. Eur Radiol 26:1863–1870CrossRefPubMed Albrecht MH, Scholtz J-E, Hüsers K et al (2016) Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast. Eur Radiol 26:1863–1870CrossRefPubMed
15.
go back to reference Zhao L, Li F, Zhang Z et al (2018) Assessment of an advanced virtual monoenergetic reconstruction technique in cerebral and cervical angiography with third-generation dual-source CT: feasibility of using low-concentration contrast medium. Eur Radiol 28:4379–4388CrossRefPubMed Zhao L, Li F, Zhang Z et al (2018) Assessment of an advanced virtual monoenergetic reconstruction technique in cerebral and cervical angiography with third-generation dual-source CT: feasibility of using low-concentration contrast medium. Eur Radiol 28:4379–4388CrossRefPubMed
16.
go back to reference Rassouli N, Chalian H, Rajiah P, Dhanantwari A, Landeras L (2017) Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging: a comparison study to conventional polychromatic 120-kVp images. Abdom Radiol (NY) 42:2579–2586CrossRef Rassouli N, Chalian H, Rajiah P, Dhanantwari A, Landeras L (2017) Assessment of 70-keV virtual monoenergetic spectral images in abdominal CT imaging: a comparison study to conventional polychromatic 120-kVp images. Abdom Radiol (NY) 42:2579–2586CrossRef
17.
go back to reference Bellini D, Gupta S, Ramirez-Giraldo JC et al (2017) Use of a noise optimized monoenergetic algorithm for patient-size independent selection of an optimal energy level during dual-energy CT of the pancreas. J Comput Assist Tomogr 41:39–47CrossRefPubMed Bellini D, Gupta S, Ramirez-Giraldo JC et al (2017) Use of a noise optimized monoenergetic algorithm for patient-size independent selection of an optimal energy level during dual-energy CT of the pancreas. J Comput Assist Tomogr 41:39–47CrossRefPubMed
18.
go back to reference Mangold S, Cannaó PM, Schoepf UJ et al (2016) Impact of an advanced image-based monoenergetic reconstruction algorithm on coronary stent visualization using third generation dual-source dual-energy CT: a phantom study. Eur Radiol 26:1871–1878CrossRefPubMed Mangold S, Cannaó PM, Schoepf UJ et al (2016) Impact of an advanced image-based monoenergetic reconstruction algorithm on coronary stent visualization using third generation dual-source dual-energy CT: a phantom study. Eur Radiol 26:1871–1878CrossRefPubMed
19.
go back to reference Muenzel D, Lo GC, Yu HS et al (2017) Material density iodine images in dual-energy CT: detection and characterization of hypervascular liver lesions compared to magnetic resonance imaging. Eur J Radiol 95:300–306CrossRefPubMed Muenzel D, Lo GC, Yu HS et al (2017) Material density iodine images in dual-energy CT: detection and characterization of hypervascular liver lesions compared to magnetic resonance imaging. Eur J Radiol 95:300–306CrossRefPubMed
20.
go back to reference Gordic S, Puippe GD, Krauss B et al (2016) Correlation between dual-energy and perfusion CT in patients with hepatocellular carcinoma. Radiology 280:78–87CrossRefPubMed Gordic S, Puippe GD, Krauss B et al (2016) Correlation between dual-energy and perfusion CT in patients with hepatocellular carcinoma. Radiology 280:78–87CrossRefPubMed
21.
go back to reference McNamara MM, Little MD, Alexander LF, Van Carroll L, Beasley TM, Morgan DE (2015) Multireader evaluation of lesion conspicuity in small pancreatic adenocarcinomas: complimentary value of iodine material density and low keV simulated monoenergetic images using multiphasic rapid kVp-switching dual energy CT. Abdom Imaging 40:1230–1240CrossRefPubMed McNamara MM, Little MD, Alexander LF, Van Carroll L, Beasley TM, Morgan DE (2015) Multireader evaluation of lesion conspicuity in small pancreatic adenocarcinomas: complimentary value of iodine material density and low keV simulated monoenergetic images using multiphasic rapid kVp-switching dual energy CT. Abdom Imaging 40:1230–1240CrossRefPubMed
22.
go back to reference Parakh A, Patino M, Muenzel D, Kambadakone A, Sahani DV (2018) Role of rapid kV-switching dual-energy CT in assessment of post-surgical local recurrence of pancreatic adenocarcinoma. Abdom Radiol (NY) 43:497–504CrossRef Parakh A, Patino M, Muenzel D, Kambadakone A, Sahani DV (2018) Role of rapid kV-switching dual-energy CT in assessment of post-surgical local recurrence of pancreatic adenocarcinoma. Abdom Radiol (NY) 43:497–504CrossRef
23.
24.
go back to reference Behrendt FF, Schmidt B, Plumhans C et al (2009) Image fusion in dual energy computed tomography: effect on contrast enhancement, signal-to-noise ratio and image quality in computed tomography angiography. Invest Radiol 44:1–6CrossRefPubMed Behrendt FF, Schmidt B, Plumhans C et al (2009) Image fusion in dual energy computed tomography: effect on contrast enhancement, signal-to-noise ratio and image quality in computed tomography angiography. Invest Radiol 44:1–6CrossRefPubMed
Metadata
Title
Objective and subjective comparison of virtual monoenergetic vs. polychromatic images in patients with pancreatic ductal adenocarcinoma
Authors
Lucian Beer
Michael Toepker
Ahmed Ba-Ssalamah
Christian Schestak
Anja Dutschke
Martin Schindl
Alexander Wressnegger
Helmut Ringl
Paul Apfaltrer
Publication date
01-07-2019
Publisher
Springer Berlin Heidelberg
Published in
European Radiology / Issue 7/2019
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-019-06116-9

Other articles of this Issue 7/2019

European Radiology 7/2019 Go to the issue