Published in: European Radiology 5/2013

01-05-2013 | Magnetic Resonance

Assessment of reduced field of view in diffusion tensor imaging of the lumbar nerve roots at 3 T

Authors: Jean-François Budzik, Sébastien Verclytte, Guillaume Lefebvre, Aurélien Monnet, Gerard Forzy, Anne Cotten

Published in: European Radiology | Issue 5/2013

Login to get access

Abstract

Objectives

To assess the value of reduced field of view (rFOV) imaging in diffusion tensor imaging (DTI) and tractography of the lumbar nerve roots at 3 T from the perspective of future clinical trials.

Methods

DTI images of the lumbar nerves were obtained in eight healthy volunteers, with and without the rFOV technique. Non-coplanar excitation and refocusing pulses associated with outer volume suppression (OVS) were used to achieve rFOV imaging. Tractography was performed. A visual evaluation of image quality was made by two observers, both senior musculoskeletal radiologists. Fractional anisotropy (FA) and apparent diffusion coefficient (ADC) were measured in L5 and S1 roots.

Results

rFOV images of the L5 and S1 roots were assessed as being superior to full FOV (fFOV) images. Image quality was rated as good to excellent by both observers. Interobserver agreement was good. No significant difference was found in FA and ADC measurements of the L5 or S1 roots. On the contrary, only poor-quality images could be obtained with fFOV imaging as major artefacts were present.

Conclusion

The rFOV approach was essential to achieve high-quality DTI imaging of lumbar nerve roots on 3-T MRI.

Key Points

Diffusion tensor 3-T MR imaging of lumbar nerve roots creates severe artefacts.
A reduced field of view drastically reduces artefacts, thereby improving image quality.
Good-quality tractography images can even be obtained with rFOV imaging.
rFOV DTI is better than fFOV DTI for clinical studies.
Literature
1.
go back to reference Budzik JF, Balbi V, Le Thuc V, Duhamel A, Assaker R, Cotten A (2011) Diffusion tensor imaging and fibre tracking in cervical spondylotic myelopathy. Eur Radiol 21:426–433PubMedCrossRef Budzik JF, Balbi V, Le Thuc V, Duhamel A, Assaker R, Cotten A (2011) Diffusion tensor imaging and fibre tracking in cervical spondylotic myelopathy. Eur Radiol 21:426–433PubMedCrossRef
2.
go back to reference Budzik JF, Le Thuc V, Demondion X, Morel M, Chechin D, Cotten A (2007) In vivo MR tractography of thigh muscles using diffusion imaging: initial results. Eur Radiol 17:3079–3085PubMedCrossRef Budzik JF, Le Thuc V, Demondion X, Morel M, Chechin D, Cotten A (2007) In vivo MR tractography of thigh muscles using diffusion imaging: initial results. Eur Radiol 17:3079–3085PubMedCrossRef
3.
go back to reference Khalil C, Budzik JF, Kermarrec E, Balbi V, Le Thuc V, Cotten A (2010) Tractography of peripheral nerves and skeletal muscles. Eur J Radiol 76:391–397PubMedCrossRef Khalil C, Budzik JF, Kermarrec E, Balbi V, Le Thuc V, Cotten A (2010) Tractography of peripheral nerves and skeletal muscles. Eur J Radiol 76:391–397PubMedCrossRef
4.
go back to reference Hiltunen J, Kirveskari E, Numminen J, Lindfors N, Goransson H, Hari R (2012) Pre- and post-operative diffusion tensor imaging of the median nerve in carpal tunnel syndrome. Eur Radiol 22:1310–1319PubMedCrossRef Hiltunen J, Kirveskari E, Numminen J, Lindfors N, Goransson H, Hari R (2012) Pre- and post-operative diffusion tensor imaging of the median nerve in carpal tunnel syndrome. Eur Radiol 22:1310–1319PubMedCrossRef
5.
go back to reference Khalil C, Hancart C, Le Thuc V, Chantelot C, Chechin D, Cotten A (2008) Diffusion tensor imaging and tractography of the median nerve in carpal tunnel syndrome: preliminary results. Eur Radiol 18:2283–2291PubMedCrossRef Khalil C, Hancart C, Le Thuc V, Chantelot C, Chechin D, Cotten A (2008) Diffusion tensor imaging and tractography of the median nerve in carpal tunnel syndrome: preliminary results. Eur Radiol 18:2283–2291PubMedCrossRef
6.
go back to reference Balbi V, Budzik JF, Duhamel A, Bera-Louville A, Le Thuc V, Cotten A (2011) Tractography of lumbar nerve roots: initial results. Eur Radiol 21:1153–1159PubMedCrossRef Balbi V, Budzik JF, Duhamel A, Bera-Louville A, Le Thuc V, Cotten A (2011) Tractography of lumbar nerve roots: initial results. Eur Radiol 21:1153–1159PubMedCrossRef
7.
go back to reference Eguchi Y, Ohtori S, Orita S et al (2011) Quantitative evaluation and visualization of lumbar foraminal nerve root entrapment by using diffusion tensor imaging: preliminary results. AJNR Am J Neuroradiol 32:1824–1829PubMedCrossRef Eguchi Y, Ohtori S, Orita S et al (2011) Quantitative evaluation and visualization of lumbar foraminal nerve root entrapment by using diffusion tensor imaging: preliminary results. AJNR Am J Neuroradiol 32:1824–1829PubMedCrossRef
8.
go back to reference Barakat N, Mohamed FB, Hunter LN et al (2012) Diffusion tensor imaging of the normal pediatric spinal cord using an inner field of view echo-planar imaging sequence. AJNR Am J Neuroradiol 33:1127–1133PubMedCrossRef Barakat N, Mohamed FB, Hunter LN et al (2012) Diffusion tensor imaging of the normal pediatric spinal cord using an inner field of view echo-planar imaging sequence. AJNR Am J Neuroradiol 33:1127–1133PubMedCrossRef
9.
go back to reference Karampinos DC, Van AT, Olivero WC, Georgiadis JG, Sutton BP (2009) High-resolution diffusion tensor imaging of the human pons with a reduced field-of-view, multishot, variable-density, spiral acquisition at 3 T. Magn Reson Med 62:1007–1016PubMedCrossRef Karampinos DC, Van AT, Olivero WC, Georgiadis JG, Sutton BP (2009) High-resolution diffusion tensor imaging of the human pons with a reduced field-of-view, multishot, variable-density, spiral acquisition at 3 T. Magn Reson Med 62:1007–1016PubMedCrossRef
10.
go back to reference Saritas EU, Cunningham CH, Lee JH, Han ET, Nishimura DG (2008) DWI of the spinal cord with reduced FOV single-shot EPI. Magn Reson Med 60:468–473PubMedCrossRef Saritas EU, Cunningham CH, Lee JH, Han ET, Nishimura DG (2008) DWI of the spinal cord with reduced FOV single-shot EPI. Magn Reson Med 60:468–473PubMedCrossRef
11.
go back to reference Wilm BJ, Svensson J, Henning A, Pruessmann KP, Boesiger P, Kollias SS (2007) Reduced field-of-view MRI using outer volume suppression for spinal cord diffusion imaging. Magn Reson Med 57:625–630PubMedCrossRef Wilm BJ, Svensson J, Henning A, Pruessmann KP, Boesiger P, Kollias SS (2007) Reduced field-of-view MRI using outer volume suppression for spinal cord diffusion imaging. Magn Reson Med 57:625–630PubMedCrossRef
12.
go back to reference Price RR, Axel L, Morgan T et al (1990) Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1. Med Phys 17:287–295PubMedCrossRef Price RR, Axel L, Morgan T et al (1990) Quality assurance methods and phantoms for magnetic resonance imaging: report of AAPM nuclear magnetic resonance Task Group No. 1. Med Phys 17:287–295PubMedCrossRef
13.
go back to reference Zaharchuk G, Saritas EU, Andre JB et al (2011) Reduced field-of-view diffusion imaging of the human spinal cord: comparison with conventional single-shot echo-planar imaging. AJNR Am J Neuroradiol 32:813–820PubMedCrossRef Zaharchuk G, Saritas EU, Andre JB et al (2011) Reduced field-of-view diffusion imaging of the human spinal cord: comparison with conventional single-shot echo-planar imaging. AJNR Am J Neuroradiol 32:813–820PubMedCrossRef
14.
go back to reference Finsterbusch J (2009) High-resolution diffusion tensor imaging with inner field-of-view EPI. J Magn Reson Imaging 29:987–993PubMedCrossRef Finsterbusch J (2009) High-resolution diffusion tensor imaging with inner field-of-view EPI. J Magn Reson Imaging 29:987–993PubMedCrossRef
15.
go back to reference Santarelli X, Garbin G, Ukmar M, Longo R (2010) Dependence of the fractional anisotropy in cervical spine from the number of diffusion gradients, repeated acquisition and voxel size. Magn Reson Imaging 28:70–76PubMedCrossRef Santarelli X, Garbin G, Ukmar M, Longo R (2010) Dependence of the fractional anisotropy in cervical spine from the number of diffusion gradients, repeated acquisition and voxel size. Magn Reson Imaging 28:70–76PubMedCrossRef
16.
go back to reference Farrell JA, Landman BA, Jones CK et al (2007) Effects of signal-to-noise ratio on the accuracy and reproducibility of diffusion tensor imaging-derived fractional anisotropy, mean diffusivity, and principal eigenvector measurements at 1.5 T. J Magn Reson Imaging 26:756–767PubMedCrossRef Farrell JA, Landman BA, Jones CK et al (2007) Effects of signal-to-noise ratio on the accuracy and reproducibility of diffusion tensor imaging-derived fractional anisotropy, mean diffusivity, and principal eigenvector measurements at 1.5 T. J Magn Reson Imaging 26:756–767PubMedCrossRef
17.
go back to reference Wardlaw JM, Brindle W, Casado AM et al (2012) A systematic review of the utility of 1.5 versus 3 Tesla magnetic resonance brain imaging in clinical practice and research. Eur Radiol. doi:10.1007/s00330-012-2500-8 Wardlaw JM, Brindle W, Casado AM et al (2012) A systematic review of the utility of 1.5 versus 3 Tesla magnetic resonance brain imaging in clinical practice and research. Eur Radiol. doi:10.​1007/​s00330-012-2500-8
18.
go back to reference Wattjes MP, Barkhof F (2012) Diagnostic relevance of high field MRI in clinical neuroradiology: the advantages and challenges of driving a sports car. Eur Radiol. doi:10.1007/s00330-012-2552-9 Wattjes MP, Barkhof F (2012) Diagnostic relevance of high field MRI in clinical neuroradiology: the advantages and challenges of driving a sports car. Eur Radiol. doi:10.​1007/​s00330-012-2552-9
Metadata
Title
Assessment of reduced field of view in diffusion tensor imaging of the lumbar nerve roots at 3 T
Authors
Jean-François Budzik
Sébastien Verclytte
Guillaume Lefebvre
Aurélien Monnet
Gerard Forzy
Anne Cotten
Publication date
01-05-2013
Publisher
Springer-Verlag
Published in
European Radiology / Issue 5/2013
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-012-2710-0

Other articles of this Issue 5/2013

European Radiology 5/2013 Go to the issue