01-05-2013 | Urogenital
MRI-guided core biopsy of the prostate in the supine position—introduction of a simplified technique using large-bore magnet systems
Published in: European Radiology | Issue 5/2013
Login to get accessAbstract
Objectives
To introduce a simplified technique for MRI-guided core biopsies (MRGB) of the prostate in the supine position using large-bore magnet systems.
Methods
Fifty men with a history of negative transrectal ultrasound-guided biopsies underwent MRGB in either a 1.5-T (13/50) or 3.0-T (37/50) wide-bore MRI unit. MRGBs were conducted with the patients in a supine position using a dedicated MR-compatible biopsy device.
Results
We developed a dedicated positioning device for the supine position. Using this device, the biopsies were performed successfully in all patients. Apart from minor rectal bleeding, only one patient developed a major side effect (urosepsis). Histology revealed prostate cancer in 25/50 (50 %) patients.
Conclusions
The new technique appears feasible. Its major advantage is the more comfortable and patient-friendly supine position during the biopsy without the need to modify the MRI system’s patient table.
Key Points
• A novel positioning device for MRI-guided prostate biopsies has been developed.
• Biopsies can be performed in the patient-friendly supine position.
• The positioning device can be utilised without modifying the MRI’s patient table.