Comparison of MRI, 64-slice MDCT and DSCT in assessing functional cardiac parameters of a moving heart phantom

Authors: J. M. Groen, P. A. van der Vleuten, M. J. W. Greuter, F. Zijlstra, M. Oudkerk

Published in: European Radiology | Issue 3/2009

Login to get access

Abstract

To compare magnetic resonance imaging (MRI), 64-slice multi-detector computed tomography (MDCT) and dual-source computed tomography (DSCT) in assessing global function parameters using a moving heart phantom. A moving heart phantom with known volumes (215–258 ml) moving at 50–100 beats per minute was examined by three different imaging modalities using clinically implemented scanning protocols. End-diastolic and end-systolic volumes were calculated by two experienced observers using dedicated post-processing tools. Ejection fraction (EF) and cardiac output (CO) were calculated and mutually compared using Bland-Altman plots. MRI underestimated the ejection EF by 16.1% with a Bland-Altman interval (B-A) of [-4.35 (-2.48) -0.60]. Sixty-four-slice MDCT overestimated the EF by 2.6% with a relatively wide B-A interval of [-3.40 (0.40) 4.20]. DSCT deviated the least from the known phantom volumes, underestimating the volumes by 0.8% with a B-A interval of [-1.17 (-0.13) 0.91]. CO analysis showed similar results. Furthermore, a good correlation was found between DSCT and MRI for EF and CO results. MRI systematically underestimates functional cardiac parameters, ejection fraction and cardiac output of a moving heart phantom. Sixty-four-slice MDCT underestimates or overestimates these functional parameters depending on the heart rate because of limited spatial resolution. DSCT deviates the least from these functional parameters compared to MRI, EBT and 64-slice MDCT.
Literature
1.
go back to reference Emond M, Mock MB, Davis KB et al (1994) Long-term survival of medically treated patients in the Coronary Artery Surgery Study (CASS) Registry. Circulation 90:2645–2657PubMed Emond M, Mock MB, Davis KB et al (1994) Long-term survival of medically treated patients in the Coronary Artery Surgery Study (CASS) Registry. Circulation 90:2645–2657PubMed
2.
go back to reference Ghali JK, Liao Y, Simmons B et al (1992) The prognostic role of left ventricular hypertrophy in patients with or without coronary artery disease. Ann Intern Med 117:831–836PubMed Ghali JK, Liao Y, Simmons B et al (1992) The prognostic role of left ventricular hypertrophy in patients with or without coronary artery disease. Ann Intern Med 117:831–836PubMed
3.
go back to reference Levy D, Garrison RJ, Savage DD et al (1990) Prognostic implications of echocardiographically determined left ventricular mass in the Framingham Heart Study. N Engl J Med 322:1561–1566PubMed Levy D, Garrison RJ, Savage DD et al (1990) Prognostic implications of echocardiographically determined left ventricular mass in the Framingham Heart Study. N Engl J Med 322:1561–1566PubMed
4.
go back to reference Moise A, Bourassa MG, Theroux P et al (1985) Prognostic significance of progression of coronary artery disease. Am J Cardiol 55:941–946CrossRefPubMed Moise A, Bourassa MG, Theroux P et al (1985) Prognostic significance of progression of coronary artery disease. Am J Cardiol 55:941–946CrossRefPubMed
5.
go back to reference Thiele H, Paetsch I, Schnackenburg B et al (2002) Improved accuracy of quantitative assessment of left ventricular volume and ejection fraction by geometric models with steady-state free precession. J Cardiovasc Magn Reson 4:327–339CrossRefPubMed Thiele H, Paetsch I, Schnackenburg B et al (2002) Improved accuracy of quantitative assessment of left ventricular volume and ejection fraction by geometric models with steady-state free precession. J Cardiovasc Magn Reson 4:327–339CrossRefPubMed
6.
go back to reference Sechtem U, Pflugfelder PW, Gould RG et al (1987) Measurement of right and left ventricular volumes in healthy individuals with cine MR imaging. Radiology 163:697–702PubMed Sechtem U, Pflugfelder PW, Gould RG et al (1987) Measurement of right and left ventricular volumes in healthy individuals with cine MR imaging. Radiology 163:697–702PubMed
7.
go back to reference Rominger MB, Bachmann GF, Pabst W et al (1999) Right ventricular volumes and ejection fraction with fast cine MR imaging in breath-hold technique: applicability, normal values from 52 volunteers, and evaluation of 325 adult cardiac patients. J Magn Reson Imaging 10:908–918CrossRefPubMed Rominger MB, Bachmann GF, Pabst W et al (1999) Right ventricular volumes and ejection fraction with fast cine MR imaging in breath-hold technique: applicability, normal values from 52 volunteers, and evaluation of 325 adult cardiac patients. J Magn Reson Imaging 10:908–918CrossRefPubMed
8.
go back to reference Rerkpattanapipat P, Hundley WG (2007) Dobutamine stress magnetic resonance imaging. Echocardiography 24:309–315CrossRefPubMed Rerkpattanapipat P, Hundley WG (2007) Dobutamine stress magnetic resonance imaging. Echocardiography 24:309–315CrossRefPubMed
9.
go back to reference Futamatsu H, Wilke N, Klassen C et al (2007) Usefulness of cardiac magnetic resonance imaging for coronary artery disease detection. Minerva Cardioangiol 55:105–114PubMed Futamatsu H, Wilke N, Klassen C et al (2007) Usefulness of cardiac magnetic resonance imaging for coronary artery disease detection. Minerva Cardioangiol 55:105–114PubMed
10.
go back to reference Bellenger NG, Grothues F, Smith GC et al (2000) Quantification of right and left ventricular function by cardiovascular magnetic resonance. Herz 25:392–399CrossRefPubMed Bellenger NG, Grothues F, Smith GC et al (2000) Quantification of right and left ventricular function by cardiovascular magnetic resonance. Herz 25:392–399CrossRefPubMed
11.
go back to reference Tornqvist E, Mansson A, Larsson EM et al (2006) It's like being in another world–patients’ lived experience of magnetic resonance imaging. J Clin Nurs 15:954–961CrossRefPubMed Tornqvist E, Mansson A, Larsson EM et al (2006) It's like being in another world–patients’ lived experience of magnetic resonance imaging. J Clin Nurs 15:954–961CrossRefPubMed
12.
go back to reference Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268CrossRefPubMed Flohr TG, McCollough CH, Bruder H et al (2006) First performance evaluation of a dual-source CT (DSCT) system. Eur Radiol 16:256–268CrossRefPubMed
13.
go back to reference Greuter MJ, Dorgelo J, Tukker WG et al (2005) Study on motion artifacts in coronary arteries with an anthropomorphic moving heart phantom on an ECG-gated multidetector computed tomography unit. Eur Radiol 15:995–1007CrossRefPubMed Greuter MJ, Dorgelo J, Tukker WG et al (2005) Study on motion artifacts in coronary arteries with an anthropomorphic moving heart phantom on an ECG-gated multidetector computed tomography unit. Eur Radiol 15:995–1007CrossRefPubMed
14.
go back to reference Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1:307–310PubMed
15.
go back to reference Mao S, Takasu J, Child J et al (2003) Comparison of LV mass and volume measurements derived from electron beam tomography using cine imaging and angiographic imaging. Int J Cardiovasc Imaging 19:439–445CrossRefPubMed Mao S, Takasu J, Child J et al (2003) Comparison of LV mass and volume measurements derived from electron beam tomography using cine imaging and angiographic imaging. Int J Cardiovasc Imaging 19:439–445CrossRefPubMed
16.
go back to reference Mahnken AH, Bruder H, Suess C et al (2007) Dual-source computed tomography for assessing cardiac function: a phantom study. Invest Radiol 42:491–498CrossRefPubMed Mahnken AH, Bruder H, Suess C et al (2007) Dual-source computed tomography for assessing cardiac function: a phantom study. Invest Radiol 42:491–498CrossRefPubMed
17.
go back to reference Mahnken AH, Hohl C, Suess C et al (2006) Influence of heart rate and temporal resolution on left-ventricular volumes in cardiac multislice spiral computed tomography: a phantom study. Invest Radiol 41:429–435CrossRefPubMed Mahnken AH, Hohl C, Suess C et al (2006) Influence of heart rate and temporal resolution on left-ventricular volumes in cardiac multislice spiral computed tomography: a phantom study. Invest Radiol 41:429–435CrossRefPubMed
18.
go back to reference Juergens KU, Maintz D, Grude M et al (2005) Multi-detector row computed tomography of the heart: does a multi-segment reconstruction algorithm improve left ventricular volume measurements? Eur Radiol 15:111–117CrossRefPubMed Juergens KU, Maintz D, Grude M et al (2005) Multi-detector row computed tomography of the heart: does a multi-segment reconstruction algorithm improve left ventricular volume measurements? Eur Radiol 15:111–117CrossRefPubMed
19.
go back to reference Bastarrika G, Arraiza M, De Cecco CN et al (2008) Quantification of left ventricular function and mass in heart transplant recipients using dual-source CT and MRI: initial clinical experience. Eur Radiol 18(9):1784–1790, Epub 2008 May 20CrossRefPubMed Bastarrika G, Arraiza M, De Cecco CN et al (2008) Quantification of left ventricular function and mass in heart transplant recipients using dual-source CT and MRI: initial clinical experience. Eur Radiol 18(9):1784–1790, Epub 2008 May 20CrossRefPubMed
20.
go back to reference Busch S, Johnson TR, Wintersperger BJ et al (2008) Quantitative assessment of left ventricular function with dual-source CT in comparison to cardiac magnetic resonance imaging: initial findings. Eur Radiol 18:570–575CrossRefPubMed Busch S, Johnson TR, Wintersperger BJ et al (2008) Quantitative assessment of left ventricular function with dual-source CT in comparison to cardiac magnetic resonance imaging: initial findings. Eur Radiol 18:570–575CrossRefPubMed
21.
go back to reference Brodoefel H, Kramer U, Reimann A et al (2007) Dual-source CT with improved temporal resolution in assessment of left ventricular function: a pilot study. AJR Am J Roentgenol 189:1064–1070CrossRefPubMed Brodoefel H, Kramer U, Reimann A et al (2007) Dual-source CT with improved temporal resolution in assessment of left ventricular function: a pilot study. AJR Am J Roentgenol 189:1064–1070CrossRefPubMed
22.
go back to reference Steen H, Nasir K, Flynn E et al (2007) Is magnetic resonance imaging the ‘reference standard’ for cardiac functional assessment? Factors influencing measurement of left ventricular mass and volumes. Clin Res Cardiol 96:743–751CrossRefPubMed Steen H, Nasir K, Flynn E et al (2007) Is magnetic resonance imaging the ‘reference standard’ for cardiac functional assessment? Factors influencing measurement of left ventricular mass and volumes. Clin Res Cardiol 96:743–751CrossRefPubMed
23.
go back to reference Ren X, Ristow B, Na B et al (2007) Prevalence and prognosis of asymptomatic left ventricular diastolic dysfunction in ambulatory patients with coronary heart disease. Am J Cardiol 99:1643–1647CrossRefPubMed Ren X, Ristow B, Na B et al (2007) Prevalence and prognosis of asymptomatic left ventricular diastolic dysfunction in ambulatory patients with coronary heart disease. Am J Cardiol 99:1643–1647CrossRefPubMed
24.
go back to reference Maurer MS, Burkhoff D, Fried LP et al (2007) Ventricular structure and function in hypertensive participants with heart failure and a normal ejection fraction: the cardiovascular health study. J Am Coll Cardiol 49:972–981CrossRefPubMed Maurer MS, Burkhoff D, Fried LP et al (2007) Ventricular structure and function in hypertensive participants with heart failure and a normal ejection fraction: the cardiovascular health study. J Am Coll Cardiol 49:972–981CrossRefPubMed
25.
go back to reference Juergens KU, Grude M, Maintz D et al (2004) Multi-detector row CT of left ventricular function with dedicated analysis software versus MR imaging: initial experience. Radiology 230:403–410CrossRefPubMed Juergens KU, Grude M, Maintz D et al (2004) Multi-detector row CT of left ventricular function with dedicated analysis software versus MR imaging: initial experience. Radiology 230:403–410CrossRefPubMed
26.
go back to reference Schlosser T, Pagonidis K, Herborn CU et al (2005) Assessment of left ventricular parameters using 16-MDCT and new software for endocardial and epicardial border delineation. AJR Am J Roentgenol 184:765–773PubMed Schlosser T, Pagonidis K, Herborn CU et al (2005) Assessment of left ventricular parameters using 16-MDCT and new software for endocardial and epicardial border delineation. AJR Am J Roentgenol 184:765–773PubMed
27.
go back to reference Suzuki S, Furui S, Kaminaga T et al (2006) Accuracy and efficiency of left ventricular ejection fraction analysis, using multidetector row computed tomography: effect of image reconstruction window within cardiac phase, slice thickness, and interval of short-axis sections. Circ J 70:289–296CrossRefPubMed Suzuki S, Furui S, Kaminaga T et al (2006) Accuracy and efficiency of left ventricular ejection fraction analysis, using multidetector row computed tomography: effect of image reconstruction window within cardiac phase, slice thickness, and interval of short-axis sections. Circ J 70:289–296CrossRefPubMed
28.
go back to reference Puesken M, Fischbach R, Wenker M et al (2008) Global left-ventricular function assessment using dual-source multidetector CT: effect of improved temporal resolution on ventricular volume measurement. Eur Radiol 18(10):2087–2094, Epub 2008 May 1CrossRefPubMed Puesken M, Fischbach R, Wenker M et al (2008) Global left-ventricular function assessment using dual-source multidetector CT: effect of improved temporal resolution on ventricular volume measurement. Eur Radiol 18(10):2087–2094, Epub 2008 May 1CrossRefPubMed
Metadata
Title
Comparison of MRI, 64-slice MDCT and DSCT in assessing functional cardiac parameters of a moving heart phantom
Authors
J. M. Groen
P. A. van der Vleuten
M. J. W. Greuter
F. Zijlstra
M. Oudkerk
Publication date
01-03-2009
Publisher
Springer-Verlag
Published in
European Radiology / Issue 3/2009
Print ISSN: 0938-7994
Electronic ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-008-1197-1

Other articles of this Issue 3/2009

European Radiology 3/2009 Go to the issue